Internal Channels for NI ELVIS II Family
- Updated2024-10-22
- 2 minute(s) read
Internal Channels for NI ELVIS II Family
Internal Channel Name | Description |
---|---|
_aignd_vs_aignd | A differential terminal with the positive and negative terminals both connected to the ground reference for analog input. |
_ao0_vs_aognd | A differential terminal with the positive terminal connected to physical channel ao0 and the negative terminal connected to the ground reference for analog output. |
_ao1_vs_aognd | A differential terminal with the positive terminal connected to physical channel ao1 and the negative terminal connected to the ground reference for analog output. |
_calref_vs_aignd | A differential terminal with the positive terminal connected to the internal calibration reference voltage and the negative terminal connected to the ground reference for analog input. |
_aignd_vs_aisense | A differential terminal with the positive terminal connected to the ground reference for analog input and the negative terminal connected to physical channel AI SENSE. |
_aignd_vs_aisense2 | A differential terminal with the positive terminal connected to the ground reference for analog input and the negative terminal connected to physical channel AI SENSE2. |
_calSrcHi_vs_aignd | A differential terminal with the positive terminal connected to the calibration PWM and the negative terminal connected to the ground reference for analog input. |
_calref_vs_calSrcHi | A differential terminal with the positive terminal connected to the internal calibration reference voltage and the negative terminal connected to the calibration PWM. |
_calSrcHi_vs_calSrcHi | A differential terminal with the positive and negative terminals connected to the calibration PWM. |
_aignd_vs_calSrcHi | A differential terminal with the positive terminal connected to the ground reference for analog input and the negative terminal connected to the calibration PWM. |
_calSrcMid_vs_aignd | A differential terminal with the positive terminal connected to the calibration PWM and the negative terminal connected to the ground reference for analog input. _calSrcMid is the divided down version of _calSrcHi. |
_calSrcLo_vs_aignd | A differential terminal with the positive terminal connected to the calibration PWM and the negative terminal connected to the ground reference for analog input. _calSrcLo is the divided down version of _calSrcHi. |
_ai0_vs_calSrcHi | A differential terminal with the positive terminal connected to physical channel ai0 and the negative terminal connected to the calibration PWM. |
_ai8_vs_calSrcHi | A differential terminal with the positive terminal connected to physical channel ai8 and the negative terminal connected to the calibration PWM. |
_boardTempSensor_vs_aignd | A differential terminal with the positive terminal connected to the onboard temperature sensor and the negative terminal connected to the ground reference for analog input.. |
_ai16:31 | A differential or RSE terminal used to access internally connected AI terminals. |
_vpsPosCurrent | A differential terminal connected to the shunt resistor on the variable power supply. |
_vpsNegCurrent | A differential terminal connected to the shunt resistor on the variable power supply. |
_vpsPos_vs_gnd | A differential terminal connected to a voltage divider on the variable power supply. |
_vpsNeg_vs_gnd | A differential terminal connected to a voltage divider on the variable power supply. |
_dutNeg | Reads the voltage equivalent of Current into the DUT- pin. |
_base | Reads the voltage at the base pin (base voltage for 3-wire analyzer). |
_dutPos | Reads the voltage at the DUT+ pin. |
_fgenImpedance | Internally routes the function generator to the impedance analyzer circuit. |
_ao0Impedance | Internally routes AO 0 to the impedance analyzer for Three-Wire Current-Voltage Analyzer measurements. |