Replacing the PXI-2567 Internal Drive Power Fuse

The front panel LEDs show the status of the +5 V and +12 V fuses. If an LED is on, the corresponding fuse is intact. This section explains how to replace the fuse in the PXI-2567.
Table 7. Replacement Fuses for PXI-2567
Accessory +5 V Internal Drive Power Fuse +12 V Internal Drive Power Fuse
Fuse Rating F 2 A, 125 V F 0.75 A, 125 V
Manufacturer and Part Number Littelfuse, NANO2, 0453.002 Littelfuse, NANO2, 0453.750
  1. Ground yourself with a grounding strap or with a ground to your PXI chassis. Proper grounding prevents damage to your PXI-2567 from electrostatic discharge.
  2. Power off the PXI chassis and remove the PXI-2567 with the blown fuse.
  3. Remove the blown fuse with pliers and replace it. Refer to the following figure for the fuse locations.
    Figure 15. PXI-2567 Internal Drive Power Fuse Locations

    1378

    1. +5 V Fuse
    2. +12 V Fuse