Acquisition Modes
- Updated2024-01-05
- 1 minute(s) read
Acquisition Modes
The NI-568x instrument driver supports three acquisition modes for RF power sensors: continuous, scope, and time slot. All modes are not supported by all devices. The following table summarizes the supported acquisition modes for each device.
Acquisition Mode | Description | Main Use Case | Supported Devices |
---|---|---|---|
Continuous (default) | The RF power sensor starts measuring power immediately after the operation is initiated and continues until you stop the operation. You do not use triggers in continuous mode. | You want to collect data at all times while the device is enabled. | USB-5680/5681/5683/5684 |
Scope | You can use gate and fence to accurately measure the power at any specific time. An internal or external trigger initiates the power sensor measurement with respect to time. | You want to measure waveforms that contain very short duty cycle timing information that otherwise skews average power measurement. | USB-5681/5683/5684 |
Time Slot | The RF power sensor divides the power measurement into "time slots" that you define for the measurement, and it calculates the average power reading for each individual time slot. | You want to perform measurements on time-division multiple access (TDMA) waveforms such as GSM/EDGE. | USB-5681/5683/5684 |