What Should I Do if the PXIe-5831 Fails the Self-Test in Max or Hardware Configuration Utility?
- Restart the system.
-
Launch MAX or Hardware Configuration Utility.
- Failed self-test—Perform self-calibration, then perform the self-test again. The PXIe-5831 must be calibrated to pass the self-test.
- Failed self-calibration—Perform self-calibration again.
- Power off the chassis.
- Reinstall the failed module in a different slot.
- Power on the chassis
- Perform the self-test again.