NI 5105 Trigger Types
- Updated2023-09-20
- 1 minute(s) read
NI 5105 Trigger Types
The NI 5105 supports the following trigger types: edge, hysteresis, immediate, digital, software, and window triggers.
The NI 5105 uses the predecimated digitized data from the input channel ADC when performing an analog trigger on any of the channels. This behavior ensures a level and time trigger accuracy of 1 timebase clock period. Because the trigger detector circuit is after the ADC, there is a 24- to 30-sample delay when you export an analog trigger to another device.