NI 5105 Trigger Delay
- Updated2023-09-20
- 1 minute(s) read
NI 5105 Trigger Delay
Trigger delay, which is specified in seconds, is achieved by adding the appropriate number of posttrigger samples to the record while keeping the allocated onboard memory equal to the record size you request. NI-SCOPE then corrects the trigger time by the delay you specify. To determine the maximum delay for a particular actual sample rate, use the following formula:
Max trigger delay in seconds = [(232 – 1) – requested posttrigger samples] × (1/actual sample rate)
Note
The maximum trigger delay changes when you use an external
sample clock, and when you sample at rates other than full rate while using the internal
sample clock.