PXI-4132 Safety

This product is intended to be used by qualified personnel who are familiar with shock hazards and the safety precautions required to avoid possible injury. Read and follow all installation and operation information carefully before using the product. The protection provided by this equipment might be impaired if it is used in a manner not described in the documentation.

For safety, always operate the PXI-4132 with suitably rated cables and the backshell provided in the shipping kit. Additional backshell kits can be purchased from NI (part number 781175-01). Operating the instrument without the provided backshell can expose users to high voltage.

Caution Shock hazards exist when voltage levels are greater than 30 V RMS, 42.4 V peak, or 60 V DC. Extreme caution should be used when a shock hazard is present. Always ensure the test system is deenergized before connecting or disconnecting the backshell or cables from the PXI-4132.
Caution Do not connect to MAINS. Do not connect to signals or use for measurements within CAT II, CAT III, or CAT IV.

Figure 1 illustrates the PXI-4132 and backshell connector.