PXI-4132 Ranges

Note If a range other than what is listed in the following tables is selected, the range is coerced to the next-highest range. For example, requesting the 20 V voltage limit range on channel 0 of the PXI-4132 coerces the voltage limit range to 100 V.

PXI-4132 Voltage Ranges

For voltage output, the PXI-4132 uses the following ranges.

Table 4. PXI-4132 Voltage Levels
Channel Voltage Level Range Voltage Level
0 10 V -10 V to +10 V
100 V -100 V to +100 V
Table 5. PXI-4132 Current Limits
Channel Current Limit Range Current Limit
0 10 µA +0.2 µA to +10 µA
100 µA +2 µA to +100 µA
1 mA +0.02 mA to +1 mA
10 mA +0.2 mA to +10 mA
100 mA +2 mA to +100 mA

PXI-4132 Current Ranges

For current output, the PXI-4132 uses the following ranges.

Table 6. PXI-4132 Current Levels
Channel Current Level Range Current Level
0 10 µA +0.2 µA to +10 µA and -0.2 µA to -10 µA
100 µA +2 µA to +100 µA and -2 µA to -100 µA
1 mA +0.02 mA to +1 mA and -0.02 mA to -1 mA
10 mA +0.2 mA to +10 mA and -0.2 mA to -10 mA
100 mA +2 mA to +100 mA and -2 mA to -100 mA
Table 7. PXI-4132 Voltage Limits
Channel Voltage Limit Range Voltage Limit
0 10 V 0 V to +10 V
100 V 0 V to +100 V

PXI-4132 Overranging Capability

Enabling overranging for a particular channel of this instrument extends current and voltage output capabilities from 100% to 105% and current capabilities down from 2% to 1% for the output range.

Overranging is applicable to output ranges only and does not apply to measurement ranges. Measurements in any given range may be made up to 105% of the range by default without enabling overranging.