PXI-4132 Ranges
- Updated2023-02-18
- 2 minute(s) read
PXI-4132 Ranges
Note If a range other than what is
listed in the following tables is selected, the range is coerced to the
next-highest range. For example, requesting the 20
V voltage limit range on channel 0
of the PXI-4132 coerces the voltage limit
range to 100 V.
PXI-4132
Voltage Ranges
For voltage output, the PXI-4132 uses the following ranges.
Channel | Voltage Level Range | Voltage Level |
---|---|---|
0 | 10 V | -10 V to +10 V |
100 V | -100 V to +100 V |
Channel | Current Limit Range | Current Limit |
---|---|---|
0 | 10 µA | +0.2 µA to +10 µA |
100 µA | +2 µA to +100 µA | |
1 mA | +0.02 mA to +1 mA | |
10 mA | +0.2 mA to +10 mA | |
100 mA | +2 mA to +100 mA |
PXI-4132
Current Ranges
For current output, the PXI-4132 uses the following ranges.
Channel | Current Level Range | Current Level |
---|---|---|
0 | 10 µA | +0.2 µA to +10 µA and -0.2 µA to -10 µA |
100 µA | +2 µA to +100 µA and -2 µA to -100 µA | |
1 mA | +0.02 mA to +1 mA and -0.02 mA to -1 mA | |
10 mA | +0.2 mA to +10 mA and -0.2 mA to -10 mA | |
100 mA | +2 mA to +100 mA and -2 mA to -100 mA |
Channel | Voltage Limit Range | Voltage Limit |
---|---|---|
0 | 10 V | 0 V to +10 V |
100 V | 0 V to +100 V |
PXI-4132 Overranging Capability
Enabling overranging for a particular channel of this instrument extends current and voltage output capabilities from 100% to 105% and current capabilities down from 2% to 1% for the output range.
Overranging is applicable to output ranges only and does not apply to measurement ranges. Measurements in any given range may be made up to 105% of the range by default without enabling overranging.