NI-VNA

Content Type
Programming Language
Current manual

RF Calibration

Each user calibration type has a variety of implementations. Which method you use depends on the DUT, test system, and measurement requirements. The following four methods are commonly used in network analyzer calibration.

Standard Calibrations

With user calibration, various standards are used to calibrate the number of discrete frequency points configured for the sweep. Refer to the NI-VNA LabVIEW and C/CVI Reference sections for information about how to configure the frequency range and number of frequencies for the sweep.

Changing some NI-VNA settings after running a calibration invalidates the calibration. These settings include the following NI-VNA properties and attributes, which are described in more detail in the NI-VNA Programming section of this help file.

  • IF bandwidth
  • Source power
  • Averaging type
  • Frequency sweep characteristics
  • Measurement enabled properties/attributes
  • Calibration type
  • Port connector gender and type
  • Most AutoCal module properties/attributes

The one-port calibration is a short/open/load (SOL) calibration that removes source match, directivity, and frequency response errors. The one-path two-port calibration is a short/open/load/through (SOLT) calibration. Full, two-port calibrations can be completed with a through-reflect-line (TRL) calibration or a short/open/load/through (SOLT) calibration. Both TRL and SOLT calibration remove transmission response errors and transmission source match errors in addition to reflection error terms. You can also perform a reflection response calibration or a transmission response calibration, which corrects the frequency response of the respective path. Reflection response calibration requires an open connected to PORT 1, and transmission response calibration requires a through standard connected between PORT 1 and PORT 2.

Note Note  Reduced measurement accuracy may occur if you use a one-path two-port calibration instead of a full two-port calibration, because the one-path two-port calibration does not account for the load match of port two.

The following figure provides a state diagram for the NI-VNA calibration process.

For information about how to complete a calibration of your NI-VNA device using the NI-VNA SFP, refer to Performing an Automatic Calibration (AutoCal) Using the NI-VNA SFP, Performing a Manual Calibration Using the NI-VNA SFP, or Performing a User Calibration Using the NI-VNA API.

Calibration and Correction

When the device has a valid calibration, you can choose whether to "correct" the measurements by understanding the relationship between enabling correction and the calibration. This relationship is shown in the following table.

Correction Enabled Calibration State
Uncalibrated Calibrated Calibrating
FALSE NI-VNA does not have a valid set of calibration data for the current device configuration. Measurement data is not corrected using calibration data. NI-VNA does have a valid set of calibration data for the current device configuration; however, measurement data is not corrected using calibration data. N/A
TRUE NI-VNA does not have a valid set of calibration data for the current device configuration; however, measurement data is modified using (potentially invalid) calibration data. NI-VNA does have a valid set of calibration data for the current device configuration. Measurement data is corrected using calibration data. N/A
Was this information helpful?

Log in to get a better experience

Help us improve your future ni.com experience.

What are you trying to do?

Submit