Fuse Replacement
- Updated2023-05-10
- 2 minute(s) read
Fuse Replacement
This section explains how to replace the fuse in the NI 4081.
Caution For protection
against fire, replace the fuse only with fuses of the same type and rating. Refer to
the following table for fuse types.
DMM | Fuse Rating | Fuse Type |
---|---|---|
NI 4081 | 3.5 A 1000 V | Time-Delay User-Replaceable |
To replace the fuse, complete the following steps.
- Remove all front panel connections from the device.
- Power off the chassis, and remove the device.
- Unscrew the shield to locate the fuse slot, as shown in the image below.
- Use a screwdriver to pry the fuse loose from the fuse slot, as shown in the image below. Caution For protection against fire, replace the fuse only with fuses of the same type and rating. Refer to the following table for fuse types.
- Verify that the fuse is blown by measuring discontinuity across the fuse using another meter.
- Insert a new fuse into the fuse slot as shown in the image below.
- Gently press the fuse with the screwdriver until both fuse clamps snap the fuse into place.