Fuse Replacement
- Updated2023-05-10
- 2 minute(s) read
Fuse Replacement
This section explains how to replace the fuse in the NI 4080/4082.
Caution For protection against fire, replace the fuse only with fuses of the same type and rating. Refer to the following table for fuse types.
DMM | Fuse Rating | Fuse Type |
---|---|---|
NI 4080/4082 | 1 A 400 V | Time-lag |
To replace the fuse, complete the following steps.
- Remove all front panel connections from the device.
- Power off the chassis, and remove the device.
- Locate the fuse hold shown in the image below.
- Insert a screwdriver into the fuse hole.
Caution For protection against fire, replace the fuse only with fuses of the same type and rating. Refer to the following table for fuse types.
- Gently press the fuse with the screwdriver until one of the fuse clamps releases.
- Locate the fuse slot shown in the image below.
- Pry the fuse loose from the fuse slot.
- Verify that the fuse is blown by measuring discontinuity across the fuse using another meter.
- Insert a new fuse into the fuse slot as shown in the image below.
- Gently press the fuse with the screwdriver until both fuse clamps snap the fuse into place.