NI-9234 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid for the range -40 °C to 70 °C unless otherwise noted.

Input Characteristics

Number of channels

4 analog input channels

ADC resolution

24 bits

Type of ADC

Delta-Sigma (with analog prefiltering)

Sampling mode

Simultaneous

Type of TEDS supported

IEEE 1451.4 TEDS Class I

Internal master timebase (fM)

Frequency

13.1072 MHz

Accuracy

±50 ppm maximum

Data rate range (fs)
Using internal master timebase

Minimum

1.652 kS/s

Maximum

51.2 kS/s

Using external master timebase

Minimum

0.391 kS/s

Maximum

52.734 kS/s

Data rates [1]1 The data rate must remain within the appropriate data range. (fs)

(fM ÷ 256)/n, n = 1, 2, ..., 31

Input coupling

AC/DC (software-selectable)

AC cutoff frequency

-3 dB

0.5 Hz

-0.1 dB

4.6 Hz maximum

Figure 1. AC Cutoff Frequency Response

1378

Input range

±5 V

AC voltage full-scale range

Minimum

±5 Vpk

Typical

±5.1 Vpk

Maximum

±5.2 Vpk

Common-mode voltage range (AI- to earth ground)

±2 V maximum

IEPE excitation current (software-selectable on/off)

Minimum

2.0 mA

Typical

2.1 mA

Power-on glitch

90 μA for 10 μs

IEPE compliance voltage

19 V maximum

If you are using an IEPE sensor, use the following equation to make sure your configuration meets the IEPE compliance voltage range.

  • (Vcommon-mode + Vbias ± Vfull-scale) must be 0 to 19

Where

  • V common-mode is the common-mode voltage applied to the NI 9234
  • V bias is the bias voltage of the IEPE sensor
  • V full-scale is the full-scale voltage of the IEPE sensor
Overvoltage protection (with respect to chassis ground)

For a signal source connected to AI+ and AI-

±30 V

For a low-impedance source connected to AI+ and AI-

-6 V to 30 V

Input delay

(40 + 13/512)/fs + 2.6 μs

Table 1. Accuracy
Measurement Conditions Percent of Reading (Gain Error) Percent of Range[2]2 Range = 5.1 Vpk (Offset Error)
Calibrated Maximum (-40 °C to 70 °C) 0.34%, ±0.03 dB ±0.14%, 7.1 mV
Typical (25 °C±5 °C) 0.05%, ±0.005 dB ±0.006%, 0.3 mV
Uncalibrated[3]3 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. Maximum (-40 °C to 70 °C) 1.9%, ±0.16 dB ±0.27%, 13.9 mV
Typical (25 °C±5 °C) 0.48%, ±0.04 dB ±0.04%, 2.3 mV
Gain drift

Typical

0.14 mdB/°C (16 ppm/°C)

Maximum

0.45 mdB/°C (52 ppm/°C)

Offset drift

Typical

19.2 μV/°C

Maximum

118 μV/°C

Channel-to-channel matching

Phase (fin in kHz)

(fin * 0.045° + 0.04 maximum)

Gain

Typical

0.01 dB

Maximum

0.04 dB

Passband

Frequency

0.45 * fs

Flatness (fs = 51.2 kS/s)

40 mdB (pk-to-pk maximum)

Phase nonlinearity (fs = 51.2 kS/s)

±0.45° maximum

Stopband

Frequency

0.55 * fs

Rejection

100 dB

Alias-free bandwidth

0.45 * fs

Oversample rate

64 * fs

Crosstalk (1 kHz)

-110 dB

CMRR (fin1 kHz)

Minimum

40 dB

Typical

47 dB

SFDR (fin = 1 kHz, -60 dBFS)

120 dB

Table 2. Idle Channel Noise and Noise Density
Idle Channel 51.2 kS/s 25.6 kS/s 2.048 kS/s
Noise 97 dBFS 99 dBFS 103 dBFS
50 μVrms 40 μVrms 25 μVrms
Noise density 310 nV/√Hz 350 nV/√Hz 780 nV/√Hz
Input impedance

Differential

305 kΩ

AI- (shield) to chassis ground

50 Ω

Table 3. Total Harmonic Distortion (THD)
Input Amplitude 1 kHz 8 kHz
-1 dBFS -95 dB -87 dB
-20 dBFS -95 dB -80 dB
Intermodulation distortion (-1 dBFS)

DIN 250 Hz/8 kHz 4:1 amplitude ratio

-80 dB

CCIF 11 kHz/12 kHz 1:1 amplitude ratio

-93 dB

MTBF

390,362 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method

Power Requirements

Power consumption from chassis

Active mode

900 mW maximum

Sleep mode

25 μW maximum

Thermal dissipation (at 70 °C)

Active mode

930 mW maximum

Sleep mode

25 μW maximum

Physical Characteristics

Dimensions

Visit ni.com/dimensions and search by module number.

Weight

173 g (6.1 oz)

Safety Voltages

Connect only voltages that are within the following limits:

Channel-to-earth ground

±30 V maximum, Measurement Category I

Isolation

Channel-to-channel

None

Channel-to-earth ground

None

Measurement Category I

Warning Do not connect the product to signals or use for measurements within Measurement Categories II, III, or IV, or for measurements on MAINs circuits or on circuits derived from Overvoltage Category II, III, or IV which may have transient overvoltages above what the product can withstand. The product must not be connected to circuits that have a maximum voltage above the continuous working voltage, relative to earth or to other channels, or this could damage and defeat the insulation. The product can only withstand transients up to the transient overvoltage rating without breakdown or damage to the insulation. An analysis of the working voltages, loop impedances, temporary overvoltages, and transient overvoltages in the system must be conducted prior to making measurements.
Mise en garde Ne pas connecter le produit à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de surtensions de catégorie II, III ou IV pouvant présenter des surtensions transitoires supérieures à ce que le produit peut supporter. Le produit ne doit pas être raccordé à des circuits ayant une tension maximale supérieure à la tension de fonctionnement continu, par rapport à la terre ou à d'autres voies, sous peine d'endommager et de compromettre l'isolation. Le produit peut tomber en panne et son isolation risque d'être endommagée si les tensions transitoires dépassent la surtension transitoire nominale. Une analyse des tensions de fonctionnement, des impédances de boucle, des surtensions temporaires et des surtensions transitoires dans le système doit être effectuée avant de procéder à des mesures.

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Environmental Characteristics

Temperature

Operating

-40 °C to 70 °C

Storage

-40 °C to 85 °C

Humidity

Operating

10% RH to 90% RH, noncondensing

Storage

5% RH to 95% RH, noncondensing

Ingress protection

IP40

Pollution Degree

2

Maximum altitude

5,000 m

Shock and Vibration
Operating vibration

Random

5 g RMS, 10 Hz to 500 Hz

Sinusoidal

5 g, 10 Hz to 500 Hz

Operating shock

30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Calibration

You can obtain the calibration certificate and information about calibration services for the NI-9234 at ni.com/calibration.

Calibration interval

1 year

1 The data rate must remain within the appropriate data range.

2 Range = 5.1 Vpk

3 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.