USB-6343 Specifications
- Updated2023-02-20
- 11 minute(s) read
USB-6343 Specifications
These specifications apply to the USB-6343 BNC and USB-6343 Spring Terminal.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid at 25 °C unless otherwise noted.
Analog Input
Number of channels | 32 single ended or 16 differential | ||||||||||||||||||
ADC resolution | 16 bits | ||||||||||||||||||
DNL | No missing codes guaranteed | ||||||||||||||||||
INL | Refer to AI Absolute Accuracy. | ||||||||||||||||||
| |||||||||||||||||||
Timing resolution | 10 ns | ||||||||||||||||||
Timing accuracy | 50 ppm of sample rate | ||||||||||||||||||
Input coupling | DC | ||||||||||||||||||
Input range | ±0.2 V, ±1 V, ±5 V, ±10 V | ||||||||||||||||||
Maximum working voltage for analog inputs (signal + common mode) | ±11 V of AI GND | ||||||||||||||||||
CMRR (DC to 60 Hz) | 100 dB | ||||||||||||||||||
| |||||||||||||||||||
Input bias current | ±100 pA | ||||||||||||||||||
| |||||||||||||||||||
Small signal bandwidth (-3 dB) | 1.2 MHz | ||||||||||||||||||
Input FIFO size | 2,047 samples | ||||||||||||||||||
Scan list memory | 4,095 entries | ||||||||||||||||||
Data transers | USB Signal Stream, programmed I/O | ||||||||||||||||||
| |||||||||||||||||||
Input current during overvoltage condition | ±20 mA maximum/AI pin | ||||||||||||||||||
|
AI Absolute Accuracy (Warranted)
Nominal Range Positive Full Scale (V) | Nominal Range Negative Full Scale (V) | Residual Gain Error (ppm of Reading) | Residual Offset Error (ppm of Range) | Offset Tempco (ppm of Range/°C) | Random Noise, σ (μV RMS) | Absolute Accuracy at Full Scale (μV) |
---|---|---|---|---|---|---|
10 | -10 | 65 | 13 | 23 | 270 | 2,190 |
5 | -5 | 72 | 13 | 23 | 135 | 1,130 |
1 | -1 | 78 | 17 | 26 | 28 | 240 |
0.2 | -0.2 | 105 | 27 | 39 | 9 | 60 |
For more information about absolute accuracy at full scale, refer to the AI Absolute Accuracy Example section.
Gain tempco | 7.3 ppm/°C |
Reference tempco | 5 ppm/°C |
INL error | 60 ppm of range |
- TempChangeFromLastExternalCal = 10 °C
- TempChangeFromLastInternalCal = 1 °C
- NumberOfReadings = 10,000
- CoverageFactor = 3 σ
AI Absolute Accuracy Equation
AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainity
- GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
- OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError
-
NoiseUncertainty = for a coverage factor of 3 σ and averaging 10,000 points.
AI Absolute Accuracy Example
For example, on the 10 V range, the absolute accuracy at full scale is as follows:
- GainError: 65 ppm + 7.3 ppm · 1 + 5 ppm · 10 = 122 ppm
- OffsetError: 13 ppm + 23 ppm · 1 + 60 ppm = 96 ppm
- NoiseUncertainty: = 8.1 μV
- AbsoluteAccuracy: 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty = 2,190 μV
Analog Output
Number of channels | 4 | ||||||||||
DAC resolution | 16 bits | ||||||||||
DNL | ±1 LSB | ||||||||||
Monotonicty | 16 bit guaranteed | ||||||||||
| |||||||||||
Timing accuracy | 50 ppm of sample rate | ||||||||||
Timing resolution | 10 ns | ||||||||||
Output range | ±10 V | ||||||||||
Output coupling | DC | ||||||||||
Output impedance | 0.2 Ω | ||||||||||
Output current drive | ±5 mA | ||||||||||
Overdrive protection | ±15 V | ||||||||||
Overdrive current | 15 mA | ||||||||||
Power-on state | ±20 mV | ||||||||||
Power-on/off glitch | 1.5 V for 1.2 s[1] | ||||||||||
Output FIFO size | 8,191 samples shared among channels used | ||||||||||
Data transfers | USB Signal Stream, programmed I/O | ||||||||||
AO waveform modes | Non-periodic waveform, periodic waveform regeneration mode from onboard FIFO, periodic waveform regeneration from host buffer including dynamic update | ||||||||||
Settling time, full-scale step, 15 ppm (1 LSB) | 6 µs | ||||||||||
Slew rate | 15 V/µs | ||||||||||
|
AO Absolute Accuracy (Warranted)
Absolute accuracy at full-scale numbers is valid immediately following self calibration and assumes the device is operating within 10 °C of the last external calibration.
Nominal Range Positive Full Scale | Nominal Range Negative Full Scale | Residual Gain Error (ppm of Reading) | Gain Tempco (ppm/°C) | Reference Tempco (ppm/°C) | Residual Offset Error (ppm of Range) | Offset Tempco (ppm of Range/°C) | INL Error (ppm of Range) | Absolute Accuracy at Full Scale (μV) |
---|---|---|---|---|---|---|---|---|
10 | -10 | 80 | 11.3 | 5 | 53 | 4.8 | 128 | 3,271 |
AO Absolute Accuracy Equation
AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError)
- GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
- OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError
Digital I/O/PFI
Static Characteristics
Number of channels | 48 total, 32 (P0.<0..31>), 16 (PFI <0..7>/P1, PFI <8..15>/P2) |
Ground reference | D GND |
Direction control | Each terminal individually programmable as input or output |
Pull-down resistor | 50 kΩ typical, 20 kΩ minimum |
Input voltage protection | ±20 V on up to two pins |
Waveform Characteristics (Port 0 Only)
Terminals used | Port 0 (P0.<0..31>) |
Port/sample size | Up to 32 bits |
Waveform generation (DO) FIFO | 2,047 samples |
Waveform acquisition (DI) FIFO | 255 samples |
DO or DI Sample Clock frequency | 0 to 1 MHz, system and bus activity dependent |
Data transfers | USB Signal Stream, programmed I/O |
Digital line filter settings | 160 ns, 10.24 μs, 5.12 ms, disable |
PFI/Port 1/Port 2 Functionality
Functionality | Static digital input, static digital output, timing input, timing output |
Timing output sources | Many AI, AO, counter, DI, DO timing signals |
Debounce filter settings | 90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable high and low transitions; selectable per input |
Recommended Operating Conditions
| |||||||
| |||||||
| |||||||
|
Digital I/O Characteristics
Positive-going threshold (VT+) | 2.2 V maximum |
Negative-going threshold (VT-) | 0.8 V minimum |
Delta VT hysteresis (VT+ - VT-) | 0.2 V minimum |
IIL input low current (VIN = 0 V) | -10 μA maximum |
IIH input high current (VIN = 5 V) | 250 μA maximum |
General-Purpose Counters
Number of counter/timers | 4 |
Resolution | 32 bits |
Counter measurements | Edge counting, pulse, pulse width, semi-period, period, two-edge separation |
Position measurements | X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding |
Output applications | Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling |
Internal base clocks | 100 MHz, 20 MHz, 100 kHz |
External base clock frequency | 0 MHz to 25 MHz |
Base clock accuracy | 50 ppm |
Inputs | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock |
Routing options for inputs | Any PFI, many internal signals |
FIFO | 127 samples per counter |
Data transfers | USB Signal Stream, programmed I/O |
Frequency Generator
Number of channels | 1 |
Base clocks | 20 MHz, 10 MHz, 100 kHz |
Divisors | 1 to 16 |
Base clock accuracy | 50 ppm |
Output can be available on any PFI terminal.
Phase-Locked Loop (PLL)
Number of PLLs | 1 |
PFI <0..15> reference clock locking frequency | 10 MHz |
Output of PLL | 100 MHz Timebase; other signals derived from 100 MHz Timebase including 20 MHz and 100 kHz Timebases |
External Digital Triggers
Source | Any PFI |
Polarity | Software-selectable for most signals |
Analog input function | Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase |
Analog output function | Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Counter/timer functions | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock |
Digital waveform generation (DO) function | Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Digital waveform acquisition (DI) function | Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Bus Interface
USB compatibility | USB 2.0 Hi-Speed or full-speed[2] |
USB Signal Stream | 8, can be used for analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1, counter/timer 2, counter/timer 3 |
Power Requirements
Power supply requirements | 11 V DC to 30 V DC, 30 W, 2 positions 3.5 mm pitch pluggable screw terminal with screw locks similar to Phoenix Contact MC 1,5/2-STF-3,5 BK |
Power input mating connector | Phoenix Contact MC 1,5/2-GF-3,5 BK or equivalent |
Current Limits
+5 V terminal (connector 0) | 1 A maximum[3] |
P0/PFI/P1/P2 and +5 V terminals combined | 2 A maximum |
Physical Characteristics
| |||||||
| |||||||
| |||||||
Screw terminal wiring gauge | 0.2047 mm2 to 1.3087 mm2 (16 AWG to 24 AWG) |
Calibration
Recommended warm-up time | 15 minutes |
Calibration interval | 2 years |
Maximum Working Voltage
Maximum working voltage refers to the signal voltage plus the common-mode voltage.
Channel to earth | 11 V, Measurement Category I |
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Environmental
| |||||||
| |||||||
Pollution Degree | 2 | ||||||
Maximum altitude | 2,000 m |
Indoor use only.
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility Standards
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-003: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
- 2011/65/EU; Restriction of Hazardous Substances (RoHS)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
1 Typical behavior. Time period may be longer due to host system USB performance. Time period will be longer during firmware updates.
2 Operating on a full-speed bus results in lower performance, and you might not be able to achieve maximum sampling/update rates.
3 Has self-resetting fuse that opens when current exceeds this specification.
In This Section
- Definitions
- Conditions
- Analog Input
- Analog Output
- Digital I/O/PFI
- General-Purpose Counters
- Frequency Generator
- Phase-Locked Loop (PLL)
- External Digital Triggers
- Bus Interface
- Power Requirements
- Current Limits
- Physical Characteristics
- Calibration
- Maximum Working Voltage
- Environmental
- Safety Compliance Standards
- Electromagnetic Compatibility Standards
- CE Compliance
- Product Certifications and Declarations
- Environmental Management