USB-6343 Specifications

These specifications apply to the USB-6343 BNC and USB-6343 Spring Terminal.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid at 25 °C unless otherwise noted.

Analog Input

Number of channels

32 single ended or 16 differential

ADC resolution

16 bits

DNL

No missing codes guaranteed

INL

Refer to AI Absolute Accuracy.

Sample rate

Single channel maximum

500 kSample/s

Multichannel maximum (aggregate)

500 kSample/s

Minimum

No minimum

Timing resolution

10 ns

Timing accuracy

50 ppm of sample rate

Input coupling

DC

Input range

±0.2 V, ±1 V, ±5 V, ±10 V

Maximum working voltage for analog inputs (signal + common mode)

±11 V of AI GND

CMRR (DC to 60 Hz)

100 dB

Input impedance
Device on

AI+ to AI GND

>10 GΩ in parallel with 100 pF

AI- to AI GND

>10 GΩ in parallel with 100 pF

Device off

AI+ to AI GND

1,200 Ω

AI- to AI GND

1,200 Ω

Input bias current

±100 pA

Crosstalk (at 100 kHz)

Adjacent channels

-75 dB

Non-adjacent channels

-90 dB

Small signal bandwidth (-3 dB)

1.2 MHz

Input FIFO size

2,047 samples

Scan list memory

4,095 entries

Data transers

USB Signal Stream, programmed I/O

Overvoltage protection for all analog input and sense channels

Device on

±25 V for up to two AI pins

Device off

±15 V for up to two AI pins

Input current during overvoltage condition

±20 mA maximum/AI pin

Settling time for multichannel measurements, accuracy, full-scale step, all ranges

±90 ppm of step (±6 LSB)

2 μs convert interval

±30 ppm of step (±2 LSB)

3 μs convert interval

±15 ppm of step (±1 LSB)

5 μs convert interval

Figure 1. Settling Error versus Time for Different Source Impedances

1378

AI Absolute Accuracy (Warranted)

Table 1. AI Absolute Accuracy
Nominal Range Positive Full Scale (V) Nominal Range Negative Full Scale (V) Residual Gain Error (ppm of Reading) Residual Offset Error (ppm of Range) Offset Tempco (ppm of Range/°C) Random Noise, σ (μV RMS) Absolute Accuracy at Full Scale (μV)
10 -10 65 13 23 270 2,190
5 -5 72 13 23 135 1,130
1 -1 78 17 26 28 240
0.2 -0.2 105 27 39 9 60

For more information about absolute accuracy at full scale, refer to the AI Absolute Accuracy Example section.

Gain tempco

7.3 ppm/°C

Reference tempco

5 ppm/°C

INL error

60 ppm of range

Note Absolute Accuracy at Full Scale is determined using the following assumptions:
  • TempChangeFromLastExternalCal = 10 °C
  • TempChangeFromLastInternalCal = 1 °C
  • NumberOfReadings = 10,000
  • CoverageFactor = 3 σ
Note Accuracies listed are valid for up to two years from the device external calibration.

AI Absolute Accuracy Equation

AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainity

  • GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
  • OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError
  • NoiseUncertainty =
    RandomNoise310,000
    for a coverage factor of 3 σ and averaging 10,000 points.

AI Absolute Accuracy Example

For example, on the 10 V range, the absolute accuracy at full scale is as follows:

  • GainError: 65 ppm + 7.3 ppm · 1 + 5 ppm · 10 = 122 ppm
  • OffsetError: 13 ppm + 23 ppm · 1 + 60 ppm = 96 ppm
  • NoiseUncertainty:
    270µV310,000
    = 8.1 μV
  • AbsoluteAccuracy: 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty = 2,190 μV

Analog Output

Number of channels

4

DAC resolution

16 bits

DNL

±1 LSB

Monotonicty

16 bit guaranteed

Maximum update rate (simultaneous)

1 channel

900 kSample/s

2 channels

840 kSample/s per channel

3 channels

775 kSample/s per channel

4 channels

719 kSample/s per channel

Timing accuracy

50 ppm of sample rate

Timing resolution

10 ns

Output range

±10 V

Output coupling

DC

Output impedance

0.2 Ω

Output current drive

±5 mA

Overdrive protection

±15 V

Overdrive current

15 mA

Power-on state

±20 mV

Power-on/off glitch

1.5 V for 1.2 s[1]

Output FIFO size

8,191 samples shared among channels used

Data transfers

USB Signal Stream, programmed I/O

AO waveform modes

Non-periodic waveform, periodic waveform regeneration mode from onboard FIFO, periodic waveform regeneration from host buffer including dynamic update

Settling time, full-scale step, 15 ppm (1 LSB)

6 µs

Slew rate

15 V/µs

Glitch energy

Magnitude

100 mV

Duration

2.6 µs

AO Absolute Accuracy (Warranted)

Absolute accuracy at full-scale numbers is valid immediately following self calibration and assumes the device is operating within 10 °C of the last external calibration.

Table 2. AO Absolute Accuracy
Nominal Range Positive Full Scale Nominal Range Negative Full Scale Residual Gain Error (ppm of Reading) Gain Tempco (ppm/°C) Reference Tempco (ppm/°C) Residual Offset Error (ppm of Range) Offset Tempco (ppm of Range/°C) INL Error (ppm of Range) Absolute Accuracy at Full Scale (μV)
10 -10 80 11.3 5 53 4.8 128 3,271
Note Absolute Accuracy at Full Scale numbers are valid immediately following self calibration and assumes the device is operating within 10 °C of the last external calibration.
Note Accuracies listed are valid for up to two years from the device external calibration.

AO Absolute Accuracy Equation

AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError)

  • GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal)
  • OffsetError = ResidualOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INLError

Digital I/O/PFI

Static Characteristics

Number of channels

48 total, 32 (P0.<0..31>), 16 (PFI <0..7>/P1, PFI <8..15>/P2)

Ground reference

D GND

Direction control

Each terminal individually programmable as input or output

Pull-down resistor

50 kΩ typical, 20 kΩ minimum

Input voltage protection

±20 V on up to two pins

Caution Stresses beyond those listed under the Input voltage protection specification may cause permanent damage to the device.

Waveform Characteristics (Port 0 Only)

Terminals used

Port 0 (P0.<0..31>)

Port/sample size

Up to 32 bits

Waveform generation (DO) FIFO

2,047 samples

Waveform acquisition (DI) FIFO

255 samples

DO or DI Sample Clock frequency

0 to 1 MHz, system and bus activity dependent

Data transfers

USB Signal Stream, programmed I/O

Digital line filter settings

160 ns, 10.24 μs, 5.12 ms, disable

PFI/Port 1/Port 2 Functionality

Functionality

Static digital input, static digital output, timing input, timing output

Timing output sources

Many AI, AO, counter, DI, DO timing signals

Debounce filter settings

90 ns, 5.12 μs, 2.56 ms, custom interval, disable; programmable high and low transitions; selectable per input

Recommended Operating Conditions

Input high voltage (VIH)

Minimum

2.2 V

Maximum

5.25 V

Input low voltage (VIL)

Minimum

0 V

Maximum

0.8 V

Output high current (IOH)

P0.<0..31>

-24 mA maximum

PFI <0..15>/P1/P2

-16 mA maximum

Output low current (IOL)

P0.<0..31>

24 mA maximum

PFI <0..15>/P1/P2

16 mA maximum

Digital I/O Characteristics

Positive-going threshold (VT+)

2.2 V maximum

Negative-going threshold (VT-)

0.8 V minimum

Delta VT hysteresis (VT+ - VT-)

0.2 V minimum

IIL input low current (VIN = 0 V)

-10 μA maximum

IIH input high current (VIN = 5 V)

250 μA maximum

Figure 2. P0.<0..31>: IOH versus VOH

1378

Figure 3. P0.<0..31>: IOL versus VOL

1378

Figure 4. PFI <0..15>/P1/P2: IOH versus VOH

1378

Figure 5. PFI <0..15>/P1/P2: IOL versus VOL

1378

General-Purpose Counters

Number of counter/timers

4

Resolution

32 bits

Counter measurements

Edge counting, pulse, pulse width, semi-period, period, two-edge separation

Position measurements

X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding

Output applications

Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling

Internal base clocks

100 MHz, 20 MHz, 100 kHz

External base clock frequency

0 MHz to 25 MHz

Base clock accuracy

50 ppm

Inputs

Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock

Routing options for inputs

Any PFI, many internal signals

FIFO

127 samples per counter

Data transfers

USB Signal Stream, programmed I/O

Frequency Generator

Number of channels

1

Base clocks

20 MHz, 10 MHz, 100 kHz

Divisors

1 to 16

Base clock accuracy

50 ppm

Output can be available on any PFI terminal.

Phase-Locked Loop (PLL)

Number of PLLs

1

PFI <0..15> reference clock locking frequency

10 MHz

Output of PLL

100 MHz Timebase; other signals derived from 100 MHz Timebase including 20 MHz and 100 kHz Timebases

External Digital Triggers

Source

Any PFI

Polarity

Software-selectable for most signals

Analog input function

Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase

Analog output function

Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase

Counter/timer functions

Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down, Sample Clock

Digital waveform generation (DO) function

Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase

Digital waveform acquisition (DI) function

Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase

Bus Interface

USB compatibility

USB 2.0 Hi-Speed or full-speed[2]

USB Signal Stream

8, can be used for analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1, counter/timer 2, counter/timer 3

Power Requirements

Caution The protection provided by the device can be impaired if the device is used in a manner not described in the X Series User Manual.
Caution The USB-6343 must be powered with an NI-offered AC adapter or a National Electric Code (NEC) Class 2 DC source that meets the power requirements for the device and has appropriate safety certification marks for country of use.

Power supply requirements

11 V DC to 30 V DC, 30 W, 2 positions 3.5 mm pitch pluggable screw terminal with screw locks similar to Phoenix Contact MC 1,5/2-STF-3,5 BK

Power input mating connector

Phoenix Contact MC 1,5/2-GF-3,5 BK or equivalent

Current Limits

Caution Exceeding the current limits may cause unpredictable device behavior.

+5 V terminal (connector 0)

1 A maximum[3]

P0/PFI/P1/P2 and +5 V terminals combined

2 A maximum

Physical Characteristics

Enclosure dimensions (includes connectors)

BNC

20.3 cm × 18.5 cm × 6.8 cm (8.0 in. × 7.3 in. × 2.7 in.)

Screw terminal

26.4 cm × 17.3 cm × 3.6 cm (10.4 in. × 6.8 in. × 1.4 in.)

Weight

Screw Terminal

1.445 kg (3 lb 3 oz)

BNC

1.803 kg (3 lb 15.6 oz)

I/O Connector

Screw Terminal

128 screw terminals

BNC

30 BNCs and 60 screw terminals

Screw terminal wiring gauge

0.2047 mm2 to 1.3087 mm2 (16 AWG to 24 AWG)

Note For more information about the connectors used for DAQ devices, refer to the document, NI DAQ Device Custom Cables, Replacement Connectors, and Screws, by going to ni.com/info and entering the Info Code rdspmb.

Calibration

Recommended warm-up time

15 minutes

Calibration interval

2 years

Maximum Working Voltage

Maximum working voltage refers to the signal voltage plus the common-mode voltage.

Channel to earth

11 V, Measurement Category I

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Caution Do not connect the system to signals or use for measurements within Measurement Categories II, III, or IV.
Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Environmental

Temperature

Operating

0 °C to 45 °C

Storage

-40 °C to 70 °C

Humidity

Operating

10% to 90% RH, noncondensing

Storage

5% to 95% RH, noncondensing

Pollution Degree

2

Maximum altitude

2,000 m

Indoor use only.

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility Standards

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:

  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-003: Class A emissions
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.
Notice For EMC declarations and certifications, and additional information, refer to the Product Certifications and Declarations section.

CE Compliance 1378

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)
  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
  • 2011/65/EU; Restriction of Hazardous Substances (RoHS)

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • 1378 Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 1378 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 Typical behavior. Time period may be longer due to host system USB performance. Time period will be longer during firmware updates.

    2 Operating on a full-speed bus results in lower performance, and you might not be able to achieve maximum sampling/update rates.

    3 Has self-resetting fuse that opens when current exceeds this specification.