TS-15010 Specifications
- Updated2024-11-11
- 4 minute(s) read
TS-15010 Specifications
Conditions
Specifications are typical under the following conditions unless otherwise noted.
- Ambient temperature of 0 °C to 55 °C. The ambient temperature is defined as the temperature at the enclosure fan inlet (air intake). Refer to the Thermal Considerations topic in the backplanes documentation on ni.com/docs for additional information on ambient temperature characteristics and cooling requirements.
- All voltages relative to COM
Analog Input
Input FIFO size | 127 samples per slot |
Maximum sample rate | Determined by the TestScale I/O module or modules |
Timing accuracy | 50 ppm of sample rate |
Timing resolution | 12.5 ns |
Number of channels supported | Determined by the TestScale I/O module or modules |
Analog Output
|
Number of channels supported, non-hardware-timed task | Determined by the TestScale I/O module or modules |
|
Timing accuracy | 50 ppm of sample rate |
Timing resolution | 12.5 ns |
|
TS-15010 Pinout
The TestScale backplanes have two four-pin power connectors to supply power to the entire TestScale assembly.
- Vsup—Supplies power to the backplane components and to all module slots.
- Vaux—Increases available current to any installed modules with high-power requirements, such as programmable power supplies. Refer to your module specifications for Vaux power consumption information.
Digital Waveform Characteristics
Waveform acquisition (DI) FIFO | 127 samples per slot |
|
|
|
Timing accuracy | 50 ppm |
General-Purpose Counters/Timers
Number of counters/timers | 4 |
Resolution | 32 bits |
Counter measurements | Edge counting, pulse, semi-period, period, two-edge separation, pulse width |
Position measurements | X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding |
Output applications | Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling |
Internal base clocks | 80 MHz, 20 MHz, 100 kHz |
External base clock frequency | 0 MHz to 20 MHz |
Base clock accuracy | 50 ppm |
Output frequency | 0 MHz to 20 MHz |
Inputs | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
Routing options for inputs | Any module PFI, analog trigger, many internal signals |
FIFO | Dedicated 127-sample FIFO |
Frequency Generator
Number of channels | 1 |
Base clocks | 20 MHz, 10 MHz, 100 kHz |
Divisors | 1 to 16 (integers) |
Base clock accuracy | 50 ppm |
Output | Any module PFI terminal |
Module PFI Characteristics
Functionality | Static digital input, static digital output, timing input, and timing output |
Timing output sources | Many analog input, analog output, counter, digital input, and digital output timing signals |
Timing input frequency | 0 MHz to 20 MHz |
Timing output frequency | 0 MHz to 20 MHz |
Digital Triggers
Source | Any module PFI terminal |
Polarity | Software-selectable for most signals |
Analog input function | Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Analog output function | Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
Counter/timer function | Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
Module I/O States
At power-on | Module-dependent. Refer to the documentation for each TestScale I/O module. |
USB Port Interface
| |||||||||
|
Power Requirements
Input voltage range (Vsup and Vaux) | 11 V to 28 V |
Power consumption from Vsup | 13.5 W maximum |
Thermal dissipation | 2.6 W maximum |
Power input connector, Vsup and Vaux | 4 positions, 3.5 mm pitch, dual row connector rated to 11 A maximum, similar to Molex Ultra-Fit Vertical Header 172299-1204 |
Power input mating connector | Molex Ultra-Fit Tangless Receptacle Housing 1722583104 with Molex crimp terminal 1722536011, or equivalent |
Power consumption from USB, 4.10 to 5.25 V | 500 µA maximum |
Cooling
Module cooling system | Forced air circulation (positive pressurization) through single static 18,000 RPM fan |
Module cooling intake | Rear of the system |
Module cooling exhaust | Along both sides of the system |
Physical Characteristics
Dimensions | Visit ni.com/dimensions and search by backplane model number. |
If you need to clean the device, wipe it with a dry towel.
Environmental Characteristics
Operating | 0 °C to 55 °C |
Storage | -40 °C to 71 °C |
Operating | 10% to 90%, noncondensing |
Storage | 5% to 95%, noncondensing |
Pollution degree | 2 |
Maximum altitude | 2,000 m |
Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS |
Operating shock | 30 g, half-sine, 11 ms pulse |