PXIe-5693 Specifications
- Updated2023-02-20
- 12 minute(s) read
PXIe-5693 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Typical-95 specifications describe the performance met by 95% (≈2σ) of models with a 95% confidence.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- 30 minute warm-up time before operation
- Calibration cycle maintained
- Chassis fan speed set to High. NI recommends using slot blockers and EMC filler panels in empty module slots to minimize temperature drift.
- NI-RFSA 2.6 or later is used
Frequency
| |||||||||
Preselector Filters
| Preselector Filter Band | Center Frequency Tune Range[1] (MHz) | Passband Frequency Range[2] (MHz) | Lower Stopband Frequency Range (MHz) | Upper Stopband Frequency Range (MHz) | Stopband Rejection (dB) |
|---|---|---|---|---|---|
| 1 | 20 to 34 | 19 to 35 | <14 | >42 | >20 |
| 2 | >34 to 60 | 33 to 61 | <27 | >70 | >20 |
| 3 | >60 to 100 | 59 to 110 | <49 | >128 | >20 |
| 4 | >100 to 160 | 90 to 170 | <75 | >185 | >20 |
| 5 | >160 to 225 | 140 to 245 | <115 | >285 | >20 |
| 6 | >225 to 350 | 205 to 370 | <170 | >420 | >20 |
| 7 | >350 to 555 | 330 to 575 | <280 | >660 | >20 |
| 8 | >555 to 950 | 530 to 975 | <450 | >1,120 | >20 |
| 9 | >950 to 1,560 | 910 to 1,600 | <775 | >1,920 | >20 |
| 10 | >1,560 to 2,000 | 1,520 to 2,040 | <1,350 | >2,320 | >20 |
| 11 | >2,000 to 2,500 | 1,960 to 2,540 | <1,700 | >2,860 | >20 |
| 12 | >2,500 to 3,000 | 2,460 to 3,040 | <2,140 | >3,460 | >20 |
| 13 | >3,000 to 3,800 | 2,960 to 3,840 | <2,650 | >4,350 | >20 |
| 14 | >3,800 to 4,600 | 3,560 to 4,640 | <3,350 | >5,050 | >20 |
| 15 | >4,600 to 5,800 | 4,560 to 5,840 | <3,850 | >6,550 | >20 |
| 16 | >5,800 to 7,000 | 5,760 to 7,040 | <4,900 | >8,250 | >20 |
Notch Filters
Amplitude
Gain
| Tune Frequency Range (MHz) | 23 °C ± 5 °C (dB) | 0 °C to 55 °C (dB) | ||
|---|---|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | Preamplifier Enabled | Preamplifier Disabled | |
| 20 to 950 | >23.5 | >1.3 | >22.6 | >0.7 |
| >950 to 2,000 | >26.9 | >-0.2 | >25.9 | >-0.9 |
| >2,000 to 3,000 | >15.9 | >-2.0 | >14.7 | >-2.8 |
| >3,000 to 5,800 | >15.6 | >0.1 | >14.5 | >-0.8 |
| >5,800 to 7,000 | >14.4 | >-1.3 | >13.2 | >-2.3 |
| Notch filters | — | >0.5 | — | >-0.3 |
| External filter | — | >1.14 | — | >0.5 |
Gain Range
| Tune Frequency Range (MHz) | 23 °C ± 5 °C (dB) | 0 °C to 55 °C (dB) | ||
|---|---|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | Preamplifier Enabled | Preamplifier Disabled | |
| Preselector path 20 to 950 | >16.8 | >21.6 | >15.1 | >20.2 |
| Preselector path >950 to 2,000 | >20.6 | >20.9 | >18.8 | >19.4 |
| Preselector path >2,000 to 3,000 | >13.5 | >21.3 | >11.3 | >19.6 |
| Preselector path >3,000 to 5,800 | >21.1 | >21.6 | >18.7 | >19.5 |
| Preselector path >5,800 to 7,000 | >21.3 | >21.2 | >18.8 | >19.1 |
| Notch filters | — | >22.4 | — | >21.1 |
| External filter | — | >21.8 | — | >20.3 |
Gain Accuracy
| Frequency | 23 °C ± 5 °C (dB) | 0 °C to 55 °C (dB) |
|---|---|---|
| 20 MHz to 2.0 GHz | ±0.55 | ±0.6 |
| >2.0 GHz to 3.0 GHz | ±0.55 | ±0.6 |
| >3.0 GHz to 7.0 GHz | ±0.55 | ±0.6 |
Low-Frequency Bypass Path
Maximum insertion loss | <1.8 dB, nominal |
Noise Figure
| Frequency | 23 °C ± 5 °C (dB) | 0 °C to 55 °C (dB) |
|---|---|---|
| 20 MHz to 150 MHz | 13 | 13 |
| 12, typical | 12, typical | |
| >150 MHz to 2.5 GHz | 14.5 | 15 |
| 13, typical | 14.5, typical | |
| >2.5 GHz to 3.2 GHz | 15 | 16 |
| 14, typical | 15, typical | |
| >3.2 GHz to 5.5 GHz | 18 | 18.5 |
| 16, typical | 17, typical | |
| >5.5 GHz to 7.0 GHz | 18.5 | 19.1 |
| 16, typical | 17, typical |
Settling Time[9]
| |||||||||
Gain switching | 200 μs, nominal | ||||||||
| |||||||||
Linearity
Third-Order Intercept Points
| Filter Path | Frequency (MHz) | Third-Order Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Preselector path | 20 to 150 | -2.0 | -2.0 |
| -1, typical | -1, typical | ||
| >150 to 2,500 | -7.0 | -7.5 | |
| -5.0, typical | -5.0, typical | ||
| >2,500 to 3,200 | -5.5 | -6 | |
| -5.0, typical | -5.0, typical | ||
| >3,200 to 5,500 | -5.0 | -6.5 | |
| -1.0, typical | -2.0, typical | ||
| >5,500 to 7,000 | -4.5 | -5.5 | |
| 0, typical | -1.0, typical | ||
| Filter Path | Frequency (MHz) | Third-Order Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Preselector path | 20 to 150 | +20 | +20 |
| +22, typical | +22, typical | ||
| >150 to 2,500 | +18 | +17 | |
| +21, typical | +20, typical | ||
| >2,500 to 3,200 | +17 | +16 | |
| +21, typical | +20, typical | ||
| >3,200 to 5,500 | +14 | +14 | |
| +20, typical | +18, typical | ||
| >5,500 to 7,000 | +15 | +14 | |
| +20, typical | +18, typical | ||
| External filter[12] | 20 to 150 | +23, nominal | +22.5, nominal |
| >150 to 2,500 | +20, nominal | +19.5, nominal | |
| >2,500 to 3,000 | +21, nominal | +20.5, nominal | |
| Filter Path | Frequency (MHz) | Third-Order Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Preselector path | 20 to 150 | +10 | +9 |
| +17, typical | +15, typical | ||
| >150 to 2,500 | 0 | -1 | |
| +4, typical | +3, typical | ||
| >2,500 to 7,000 | +9 | +8 | |
| +17, typical | +17, typical | ||
| Filter Path | Frequency (MHz) | Third-Order Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Preselector path | 20 to 150 | +38 | +38 |
| +42, typical | +40, typical | ||
| >150 to 2,500 | +30 | +28 | |
| +35, typical | +32, typical | ||
| >2,500 to 7,000 | +29 | +29 | |
| +34, typical | +33, typical | ||
Second Harmonic Intercept Points[15]
| Filter Path | Source Frequency (MHz) | Second Harmonic Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Internal | 20 to 250 | +61 | +60 |
| >250 to 2,000 | +48 | +47 | |
| >2,000 to 3,500 | +64 | +63 | |
| Filter Path | Source Frequency (MHz) | Second Harmonic Intercept Point (dBm) | |
|---|---|---|---|
| 23 °C ± 5 °C | 0 °C to 55 °C | ||
| Internal | 20 to 250 | +69 | +68 |
| >250 to 3,500 | +74 | +73 | |
| External | 20 to 1,000 | +37, nominal | +35, nominal |
| >1,000 to 1,500 | +40, nominal | +36, nominal | |
Input Power at 1 dB Gain Compression
| Frequency (MHz) | 23 °C ± 5 °C (dBm) | 0 °C to 55 °C (dBm) | ||
|---|---|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | Preamplifier Enabled | Preamplifier Disabled | |
| 20 to 150 | -16 | +6 | -19 | +4 |
| >150 to 2,500 | -21 | +5 | -26 | +3 |
| >2,500 to 3,200 | -16 | +5 | -20 | +1.5 |
| >3,200 to 5,500 | -12 | +4 | -16 | +1.5 |
| >5,500 to 7,000 | -11 | +5 | -15 | +1.5 |
Spurious Responses
Non-Input-Related (Residual) Spurs[17]
>20 MHz to 150 MHz | -60 dBm, typical |
>150 MHz to 2,500 MHz | -81 dBm, typical |
>2,500 MHz to 7,000 MHz | -95 dBm, typical |
RF Input Connector Emissions Level
| Frequency (MHz) | 23 °C ± 5 °C (dBm) | |
|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | |
| 20 to 3,000 | < -92 | < -91 |
| 3,000 to 7,000 | < -95 | < -94 |
Reverse Isolation for Frequencies 4.2 GHz to 8.2 GHz
| Signal Path | 23 °C ± 5 °C (dBm) | |
|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | |
| Preselector filters 1 to 4 | >39 | >39 |
| >46, typical | >46, typical | |
| Preselector filters 5 to 8 | >38 | >38 |
| >45, typical | >45, typical | |
| Preselector filters 9 to 12 | >35 | >35 |
| >41, typical | >41, typical | |
| Preselector filters 13 to 16 | >35 | >35 |
| >43, typical | >41, typical | |
| Notch filter | >35 | >35 |
| >37, typical | >37, typical | |
| External filter | — | >33, typical |
| Low-frequency bypass path | — | >33, typical |
Reverse Isolation
| Signal Path | 23 °C ± 5 °C (dB) | |
|---|---|---|
| Preamplifier Enabled | Preamplifier Disabled | |
| Preselector filters 1 to 4 | >49 | >35 |
| Preselector filters 5 to 8 | >50 | >36 |
| Preselector filters 9 to 12 | >57 | >36 |
| Preselector filters 13 to 16 | >43 | >41 |
Calibration Signal
Frequency range | 34.5 MHz to 7.0 GHz, nominal |
| Start Frequency (MHz) | Stop Frequency (MHz) | Frequency Resolution (kHz) |
|---|---|---|
| 34.5 | 68.75 | 1.5625 |
| >68.75 | 137.5 | 3.125 |
| >137.5 | 275 | 6.250 |
| >275 | 550 | 12.50 |
| >550 | 1,100 | 25.00 |
| >1,100 | 2,200 | 50.00 |
| >2,200 | 4,200 | 100.0 |
| >4,200 | 4,400 | 100.0 |
| >4,400 | 7,000 | 200.0 |
Calibration Signal Amplitude Accuracy
| Frequency (MHz) | 23 °C ± 5 °C (dB) | 0 °C to 55 °C (dB) |
|---|---|---|
| 34.5 MHz to 3.2 GHz | ±0.9 | ±1.1 |
| >3.2 GHz to 5.5 GHz | ±1.2 | ±1.2 |
| >5.5 GHz to 7 GHz | ±1.1 | ±1.8 |
Physical Characteristics
Front Panel Connectors
RF Input (RF IN)
Connector | SMA female | ||||||||||||||
Reference impedance | 50 Ω | ||||||||||||||
| |||||||||||||||
| |||||||||||||||
RF Output (RF OUT)
Connector | SMA female | ||||||||||||||
Reference impedance | 50 Ω | ||||||||||||||
| |||||||||||||||
| |||||||||||||||
External Filter Input/Output (EXT FILTER IN/EXT FILTER OUT)
Connector | SMA female | ||||||
Reference impedance | 50 Ω | ||||||
| |||||||
VSWR | <2.0 : 1 | ||||||
Power
Power Limits
| Signal Path | RF IN Maximum Continuous Safe RF Power (dBm) | RF OUT Saturated Power (dBm) | RF OUT Maximum Continuous Safe RF Power (dBm) |
|---|---|---|---|
| Preselector bypass | +30 | — | +30 |
| Preselector with preamplifier disabled | +20 | +15 | +20 |
| Preselector with preamplifier enabled | +20 | +15 | +20 |
| External filter | +20 | +15 | +20 |
Power Requirements
3.3 V | 1.30 A, nominal |
12 V | 0.85 A, nominal |
Physical Dimensions
Size | 3U, Single-slot, PXI Express module 21.6 cm × 2.0 cm × 13.0 cm(8.5 in. × 0.8 in. × 5.1 in.) |
Weight | 465 g (16.4 oz) |
Environment
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL PRF-28800F Class 3 low temperature limit and MIL PRF-28800F Class 2 high temperature limit.) |
Relative humidity range | 10% to 90%, noncondensing |
Storage Environment
Ambient temperature range | - 41 °C to +71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL PRF-28800F Class 3 limits.) |
Relative humidity range | 5% to 95%, noncondensing |
Shock and Vibration
Operating shock | 30 g peak, half-sine, 11 ms pulse | ||||||
| |||||||
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
1 The preselector filter band selection is based on the center frequency tuning range. The lowest frequency preselector band is selected at the band-crossing frequencies.
2 Passband Frequency Range is the calibrated range of the Preselector Filter Band.
3 Notch filter bands are a cascade of a preselector filter and a notch filter. Stopband rejection specifications for the preselector filter band apply in the notch filter bands.
4 Gain is the maximum PXIe-5693 gain setting.
5 Control the PXIe-5693 module gain by setting the reference level and the IF output power level. The module gain is defined by the following equation: Module Gain = IF Output Power Level - Reference Level.
6 Gain range is the difference between the maximum and minimum PXIe-5693 module gain settings.
7 The downconverter gain value is read-only when using NI-RFSA. Gain accuracy is the difference between the value reported by NI-RFSA and the measured gain.
8 Noise figure verification uses the maximum PXIe-5693 module gain with the preamplifier enabled.
9 Settled to within <0.1 dB of final amplitude.
10 The values in this table were obtained from two -30 dBm tones spaced 700 kHz apart and the preselector set to the maximum PXIe-5693 module gain with the preamplifier enabled.
11 Two -10 dBm tones spaced 700 kHz apart and the preselector set to the maximum PXIe-5693 module gain with the preamplifier disabled.
12 The external filter path uses an external 3-inch RF coaxial cable connected between the EXT FILTER IN and the EXT FILTER OUT connectors on the PXIe-5693.
13 Two -25 dBm tones placed outside the preselector passband such that the third-order intermodulation product falls in-band. Preselector set to maximum module gain and preamplifier enabled.
14 Two -5 dBm tones placed outside the preselector passband such that the third-order intermodulation product falls in-band. Preselector set to maximum module gain and preamplifier disabled.
15 Single tone, power at RF IN connector at -5 dBm (preamplifier disabled) or -25 dBm (preamplifier enabled). For each preselector filter band, the tone frequency is set to the maximum center frequency tune range frequency divided by 2.
16 The gain compression measurement uses a two-tone desensitization method with an input referred power level and maximum module gain. This method places two tones within the preselector filter center frequency range. The lower amplitude tone power is set to -30 dBm with the preamplifier disabled and -50 dBm with the preamplifier enabled. The amplitude variation for the lower amplitude tone level cannot be great than 1 dB, as the higher amplitude tone power is increased from low power to the 1 dB gain compression specified input power.
17 23 °C ± 5 °C. Applies only to the RF OUT connector with the RF IN connector terminated. Measurements use the maximum gain setting and calibration signal disabled.
18 Calibration signal disabled.
19 Center frequency set to device center frequency.
20 Calibration tone power is referenced to the RF IN connector and is reported by the NI-RFSACal Tone Power Referred to RF IN property. The calibration tone power is measured using an RF substitution technique that compares the RF power of a calibrated source as measured at the RF OUT connector with the calibration tone signal at the RF OUT connector.