Definitions and Conditions
Specifications are subject to change without notice. For the most recent PXIe-5622 specifications, visit ni.com/manuals. Unless otherwise noted, the following conditions were used for each specification:
- Direct path filter setting enabled
- Sample clock set to internal 150 MS/s, unlocked
- 1 V vertical range
Specifications describe the warranted, traceable performance of the device over an ambient temperature range of 0 °C to 55 °C and include guardband for measurement uncertainty, unless otherwise noted. Specifications are valid under the following conditions unless otherwise noted:
- The PXIe-5622 module is warmed up for 15 minutes at ambient temperature.
- Calibration cycle is maintained.
- NI-SCOPE self-calibration performed after device temperature is stable.
- The PXI Express chassis fan speed is set to HIGH, the foam fan filters are removed if present, and the empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users available at ni.com/manuals.
- External calibration is performed at 23 °C ± 3 °C.
Typical Specifications are unwarranted values that describe the expected performance of the device over ambient temperature ranges of 23 °C ± 5 °C with a 90% confidence level.
Characteristics (or supplemental information) describe basic functions and attributes of the device established by design.
Data in this document are Specifications unless otherwise noted.
To access PXIe-5622 documentation, including the PXIe-5622 Getting Started Guide, go to .
Vertical
Analog Input (IF IN)
Number of channels | One (IF IN) |
Input impedance | 50 Ω, characteristic |
Input return loss | <-15 dB, 5 MHz to 300 MHz, typical |
Input coupling | AC, GND |
Full scale (FS) input voltage range (Vpk-pk)[1] | 0.7 V (+1 dBm), 1 V (+4 dBm), 1.4 V (+7 dBm) |
Maximum voltage input overload (Vpk-pk) | 6.3 V (+20 dBm) |
Accuracy
Resolution | 16-bit | ||||||
| |||||||
| |||||||
|
Absolute Amplitude Accuracy Examples at 40 °C in the Bandpass Path
Amplitude accuracy specification: 0.5 + 0.01 × (40 - 23) = ± 0.67 dB
Amplitude accuracy, typical: 0.3 + 0.01 × (40 - 23) = ± 0.47 dB
Bandwidth and Frequency Response
| |||||||
Dither signal, frequency range[4] | 100 kHz to 12 MHz, typical |
Bandpass Path | Direct Path | |
---|---|---|
Passband amplitude flatness, valid for 1 V range | <+0.35, -0.6 dB (equalized)[5]
187.5 MHz ±25 MHz |
<±0.35 dB (equalized) 53 MHz ± 19 MHz |
< ±0.6 dB (equalized) 10 MHz to 250 MHz (referenced to 100 MHz) |
||
Passband amplitude flatness, valid for all ranges, typical |
<+0.25, -0.4 dB (equalized) <+0.7, -3.5 dB (unequalized) 187.5 MHz ±25 MHz |
< ±0.25 dB (equalized) < ±0.6 dB (unequalized) 53 MHz ±19 MHz |
<±0.5 dB (equalized) <±1.8 dB (unequalized) 10 to 250 MHz (referenced to 100 MHz) |
Bandwidth | Bandpass Path Phase | Direct Path Phase |
---|---|---|
10 MHz | ±0.5° | ±0.5° |
20 MHz | ±1° | ±1° |
40 MHz | ±1.75° | n/a |
50 MHz | ±2.5° |
Spectral Characteristics[6]
| |||||||
| |||||||
|
Noise
| |||||||
|
Bandwidth | Bandpass path (187.5 MHz) | Direct path (53 MHz) | |
---|---|---|---|
SSB phase noise | 100 Hz | <-80 dBc/Hz | <-90 dBc/Hz |
1 kHz | <-117 dBc/Hz | <-128 dBc/Hz | |
10 kHz and above | <-134 dBc/Hz | <-141 dBc/Hz | |
SSB phase noise, typical | 100 Hz | <-83 dBc/Hz | <-94 dBc/Hz |
1 kHz | <-120 dBc/Hz | <-132 dBc/Hz | |
10 kHz and above | <-140 dBc/Hz | <-144 dBc/Hz |
Range | Value | |
---|---|---|
Average noise density[7] | 0.7 V/+1 dBm | <-146 dBm/Hz |
1 V/+4 dBm | <-143 dBm/Hz | |
1.4 V/+7 dBm | <-140 dBm/Hz | |
Average noise density, typical[8] | 0.7 V/+1 dBm | <-149 dBm/Hz |
1 V/+4 dBm | <-146 dBm/Hz | |
1.4 V/+7 dBm | <-143 dBm/Hz |
Horizontal
Sample Clock
|
Onboard Clock (Internal VCXO)
Sample rate[9] | 150 MS/s with decimation by N | ||||||||||
Accuracy | ±5.0 × 10-6, typical | ||||||||||
Accuracy over temperature | ±12 × 10-6, typical | ||||||||||
|
Phase-Locked Loop (PLL) External Reference Clock
Reference Clock sources (used to phase lock onboard VCXO) | CLK IN (front panel connector), PXIe 100 MHz (PXIe backplane) |
Sample Clock delay range (delay relative to Reference Clock when VCXO is locked) | ±1 Sample Clock period |
Sample Clock delay resolution (delay relative to Reference Clock when VCXO is locked) | ≤4 ps |
Reference Clock frequency range | 1 MHz to 100 MHz, in 1 MHz increments |
Reference Clock frequency accuracy[11] | ±25 × 10-6 |
Reference Clock duty cycle tolerance | 45% to 55%, typical |
Reference Clock export ports | CLK OUT (front panel connector) |
External Sample Clock
Frequency range | 20 MHz to 150 MHz |
Duty cycle tolerance | 45% to 55%, typical |
Export ports | CLK OUT (front panel connector) |
CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector)
Input impedance | 50 Ω, typical | ||||||
Coupling | AC | ||||||
| |||||||
Maximum input overload (Vpk-pk) | 6.3 V (+20 dBm) |
CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector)
Output impedance | 50 Ω, typical | ||||||
Coupling | AC | ||||||
|
PFI 1 (Programmable Function Interface)
PFI 1 (programmable function interface) direction | Bi-directional |
Trigger
Trigger types | Digital | ||||||||||||||
| |||||||||||||||
|
TClk Specifications
You can use the NI TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of supported devices, in one or more chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.
PXIe-5622 TClk Specifications
- Specifications measured in an NI PXIe-1062Q chassis.
- All parameters set to identical values for each PXIe-5622.
- Sample Clock set to 150 MS/s and all filters are disabled.
|
Waveform Specifications
| |||||||
| |||||||
Minimum record length | 1 sample | ||||||
Number of pretrigger samples, single-record mode and multiple-record mode | Zero up to full record length | ||||||
Number of posttrigger samples, single-record mode and multiple-record mode | Zero up to full record length | ||||||
Maximum number of records in onboard memory[15] | 100,000 |
Onboard Signal Processing (OSP)
The following OSP operations are available:
- Send one IF signal to CH 0 and perform quadrature downconversion on the signal (complex data is returned).
- Send a signal to CH 0 and perform alias-protected decimation (real data is returned).
- Send a signal to CH 0 and perform real downconversion on the signal (real data is returned).
Number of digital downconverters (DDCs) | One | ||||||||||||
Data processing modes | Real (I path only); Complex (IQ) | ||||||||||||
| |||||||||||||
| |||||||||||||
|
Complex Flat Bandwidth Example
Complex bandwidth is 60 MHz with a complex sample rate of 75 MS/s.
Using a decimation rate of 1 (sample rate of 150 MS/s with internal clock) bypasses the filters in the OSP block.
OSP Digital Gain and Offset
OSP Numerically-Controlled Oscillator (NCO)
| |||||||
| |||||||
I and Q phase resolution | 0.0055° |
OSP Digital Performance
Maximum NCO spur | < -100 dBFS |
Decimating filter passband ripple, passband is from 0 to (0.4 × IQ Rate) | < 0.1 dB |
Decimating filter Out-of-Band suppression, stopband suppression from (0.6 × IQ Rate ) | > 80 dB |
OSP IF Demodulation Typical Performance: Modulation Error Ratio (MER)
Bandpass path carrier frequency: 187.5 MHz (signal source: NI PXIe-5673) | Direct path carrier frequency: 20 MHz (signal source: NI PXI-5441) | |||
---|---|---|---|---|
Internal Reference Clocks (source and receiver unlocked to any external reference) | PXI chassis Reference Clocks (source and receiver locked to PXIe 100 MHz or PXI 10 MHz chassis backplane clock) | Internal Reference Clocks (source and receiver unlocked to any external reference) | PXI chassis Reference Clocks (source and receiver locked to PXIe 100 MHz or PXI 10 MHz chassis backplane clock) | |
GSM physical layer, typical.[23] | 50 dB | 59 dB | 48 dB | 62 dB[24] |
W-CDMA physical layer, typical.[25] | 47 dB | 50 dB | 39 dB | 58 dB |
DVB physical layer, typical.[26] | 46 dB | 48 dB | 40 dB | 56 dB |
20 MSymbols/s QAM, typical.[27] | 43 dB | 44 dB | 37 dB | 49 dB |
26 MSymbols/s QAM, typical.[28] | 39 dB | 37 dB | 36 dB | 40 dB |
34 MSymbols/s QAM, typical.[29] | 38 dB | 37 dB | 38 dB | 37 dB |
Calibration
Self-calibration | Calibrates absolute amplitude accuracy. |
External calibration | Calibrates absolute and relative (flatness) amplitude accuracy, VCXO accuracy. |
External calibration interval | 1 year |
Warm-up time | 15 minutes |
Software
Driver Software
Driver support for this device was first available in NI-SCOPE 3.5.
NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5622. NI-SCOPE provides application programming interfaces for many development environments.
Application Software
NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:
- LabVIEW
- LabWindows™/CVI™
- Measurement Studio
- Microsoft Visual C/C++
- .NET (C# and VB.NET)
Interactive Soft Front Panel and Configuration
The NI-SCOPE Soft Front Panel (SFP) allows interactive control of the PXIe-5622.
Interactive control of the PXIe-5622 was first available in NI-SCOPE SFP version 3.5. The NI-SCOPE SFP is included on the NI-SCOPE media.
NI Measurement Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5622. MAX is included on the NI-SCOPE media.
Power
|
Physical Characteristics
Dimensions | 21.6 cm × 2.0 cm × 13.0 cm (8.5 in. × 0.8 in. × 5.1 in.) 3U, one slot, PXI/cPCI Module, PXI Express compatible |
Weight | 400 g (14.1 oz) |
Environment
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) |
Relative humidity range | 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Storage Environment
Ambient temperature range | -40 °C to 71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) |
Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Shock and Vibration
Operating shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) | ||||||
|
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
Waste Electrical and Electronic Equipment (WEEE)
电子信息产品污染控制管理办法(中国RoHS)
1 Dither enabled. Can overrange up to 3 dB with Dither disabled.
2 Valid over 23 °C ± 5 °C. Maximum drift of ± 2 °C from last self-calibration.
3 Valid over 23 °C ± 5 °C. Maximum drift of ± 2 °C from last self-calibration.
4 Dither is disabled by default in NI-SCOPE. To enable dithering, refer to the NI High-Speed Digitizers Help.
5 Equalization requires using the Digital Filter Design Toolkit to compute equalization filter coefficients. This software is not included with the NI-SCOPE driver.
6 +3 dBm total power at 1 V range, Dither ON
7 Verified using a 50 Ω terminator connected to input; valid for all filter paths.
8 Verified using a 50 Ω terminator connected to input; valid for all filter paths.
9 Refer to the Onboard signal processing (OSP) section for possible N values (with and without fractional resampling). Non-OSP decimation does not protect the acquired data from undersampling aliasing. Non-OSP decimation and OSP decimation are mutually exclusive.
10 Internal VCXO, unlocked.
11 Refer to your chassis documentation to ensure it meets this requirement.
12 For information about manual adjustment, refer to the Synchronization Repeatability Optimization topic in the NI-TClk Synchronization Help; for additional help with the adjustment process, contact NI technical support at ni.com/support.
13 Assumes 2-byte samples. In Complex data processing mode (only available when using onboard signal processing), each sample is 4 bytes, so this number is halved.
14 Record length refers to the number of samples, or data points, the NI-SCOPE device acquires for each channel in a single acquisition.
15 It is possible to exceed this number if you fetch records while acquiring data. For more information, refer to the NI High-Speed Digitizers Help .
16 Non-OSP decimation does not protect against high-frequency aliasing. Non-OSP decimation and OSP decimation are mutually exclusive.
17 For sample rates less than 9.155 kS/s, use an external sample clock or perform additional software decimation.
18 Fractional resampling not available.
19 Using a decimation rate of 1 (sample rate of 150 MS/s with internal clock) bypasses the filters in the OSP block.
20 Gain <1 attenuates user data
21 (-0.5 × Vertical Range) ≤ Output ≤ (+0.5 × Vertical Range)
22 Undersampling can be used for carrier frequencies >75 MHz.
23 MSK modulation, 270.833 kSymbols/s, 1024 symbols, gaussian, BT = 0.3.
24 In this case, the direct path carrier frequency is 35 MHz using the NI PXIe-5450 as the source.
25 QPSK modulation, 3.84 MSymbols/s, 1024 symbols, root raised cosine, alpha = 0.22.
26 32 QAM modulation, 6.92 MSymbols/s, 1024 symbols, root raised cosine, alpha = 0.15.
27 64 QAM modulation, 20 MSymbols/s, 1024 symbols, root raised cosine, alpha = 0.15.
28 64 QAM modulation, 26.09 MSymbols/s, 1024 symbols, root raised cosine, alpha = 0.15.
29 64 QAM modulation, 34.78 MSymbols/s, 1024 symbols, root raised cosine, alpha = 0.15.