PXIe-5185 Specifications
- Updated2023-02-19
- 13 minute(s) read
PXIe-5185 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- For 50 Ω input channel, vertical range (Vpk-pk) set to 0.11, 0.2, 0.5, or 1
- For 50 Ω input channel, vertical range (Vpk-pk) set to 0.11, 0.2, 0.5, 1, 2, 5, or 10
- 1 MΩ input channel disconnected for 50 Ω input channel specifications, and 50 Ω input channel disconnected for 1 MΩ input channel specifications
- Sample clock set to 6.25 GS/s or 12.5 GS/s
- Onboard Sample clock locked to PXIe_CLK100 Reference clock
- 0 °C to 50 °C ambient temperature
- PXIe-5185 module part numbers 199363x-0zL (where x is any letter and z is any number) only support 50 Ω input impedance.
- PXIe-5185 module part numbers 152962x-0zL (where x is any letter and z is any number) support both 50 Ω and 1 MΩ input impedance
Warranted specifications are valid under the following conditions unless otherwise noted.
- The PXIe-5185 is warmed up for 25 minutes at ambient temperature
- Self-calibration is completed after warm-up period or when switching from an external Sample and/or Reference clock to the Onboard clock
- Calibration cycle is maintained
- The PXI Express chassis fan speed is set to HIGH, the fan filters are clean if present, and the empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals.
- NI-SCOPE 3.9.6 or later instrument driver is used
- External calibration is performed at 23 °C ± 3 °C
Vertical
Analog Input (Channel 0 and Channel 1)
Number of channels | Two (simultaneously sampled) | ||||||||||
Input type | Reference single-ended | ||||||||||
|
Impedance and Coupling
| |||||||
| |||||||
|
Voltage Levels
Input | Input range (Vpk-pk) | Vertical offset range (V) |
---|---|---|
50 Ω and 1 MΩ inputs | 0.11 to 1 in >0.3 mV steps | ±0.25 |
1 MΩ input only | >1 to 10 in >3 mV steps | ±2.5 |
|
Accuracy
Resolution | 8 bits | ||||||||||||||
| |||||||||||||||
Programmable vertical offset accuracy, warranted[3] | ±1.2% of offset setting | ||||||||||||||
| |||||||||||||||
Programmable vertical offset drift, characteristic[4] | ±0.02% of offset setting per ºC | ||||||||||||||
| |||||||||||||||
AC amplitude drift, characteristic[4] | ±0.014 dB per ºC at 50 kHz | ||||||||||||||
|
Bandwidth and Transient Response
Spectral Characteristics
PXIe-5185 50 Ω Spectral Characteristics
PXIe-5185 1 MΩ Spectral Characteristics
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
|
Noise
Skew
|
Horizontal
Sample Clock
|
Onboard Clock (Internal VCO)
External Sample Clock
Sources | CLK IN (front panel SMA connector) |
Frequency range[20] | 1.6 GHz to 3.125 GHz |
Duty cycle tolerance, typical | 45% to 55% |
Phase-Locked Loop (PLL) Reference Clock
| |||||||
Frequency[21] | 10 MHz or 100 MHz | ||||||
Duty cycle tolerance, characteristic | 45% to 55% |
CLK IN (Sample Clock Input, Front Panel Connector)
Input voltage range, characteristic | Sine wave: 0.45 Vpk-pk to 1.78 Vpk-pk (–3 dBm to 9 dBm) |
Maximum input overload, characteristic | 3 Vrms, |Peaks| ≤ 4.25 V |
Impedance, nominal | 50 Ω |
Coupling | AC |
REF CLK (Reference Clock In, Front Panel Connector)
Input voltage range, characteristic | Sine wave: -2 dBm to 16 dBm |
Maximum input overload, typical | 1.6 Vrms, |Peaks| ≤ 10 V (1 ms peak) |
Impedance, nominal | 50 Ω |
Coupling | AC |
Frequency[22] | 10 MHz or 100 MHz |
Trigger
Supported trigger | Reference (stop) trigger | ||||||||||||
Trigger types | Edge, Digital, Immediate, and Software | ||||||||||||
Trigger sources | CH 0, CH 1, TRIG, PXI_Trig <0..6>, and Software | ||||||||||||
| |||||||||||||
| |||||||||||||
Holdoff | Rearm time to 10.99 s | ||||||||||||
Trigger delay | From 0 to 1,450,000 seconds (15 days) |
Analog Trigger (Edge Trigger Type)
Sources | CH 0, CH 1, or TRIG | ||||||
| |||||||
| |||||||
| |||||||
| |||||||
|
Digital Trigger (Digital Trigger Type)
Sources | PXIe_TRIG <0..6> (backplane connector) |
TRIG (External Trigger, Front Panel Connector)
Connector | SMA |
Impedance, nominal | 50 Ω |
Coupling | DC |
Input voltage range, nominal | ±5 V |
Maximum input overload, characteristic | |Peaks| ≤ 6 V |
TClk Specifications
You can use the National Instruments TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of SMC-based modules in a chassis. Specifications are valid for any number of NI 5185 or NI 5186 modules installed in one PXI Express chassis, with all parameters set to identical values for each SMC-based module. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.
Waveform Specifications
Onboard memory sizes[28] | 32 MB or 1 GB | ||||||
Minimum record length, characteristic | 1 sample | ||||||
Number of pretrigger samples, characteristic[29] | Zero up to full record length | ||||||
Number of posttrigger samples, characteristic[29] | Zero up to full record length | ||||||
| |||||||
Allocated onboard memory per record, characteristic | [(Record length × 1 byte/sample) + 1,500], rounded up to: 4 KB, 8 KB, 16 KB, 32 KB, 64 KB, or an integer multiple of 128 KB |
Memory Sanitization
For information about memory sanitization, refer to the NI PXIe-5185/5186 Letter of Volatility, which is available for download from ni.com/manuals.
Calibration
Power-up calibration | Automatically performed by the device at power-on to calibrate the gain, offset, and phase of the ADCs on the device. Typically takes 5 to 10 minutes to complete. |
Self-calibration | Self-calibration is done on software command. The calibration corrects for gain, offset, triggering, and timing errors for all input ranges, excluding the External Trigger input channel (TRIG). Refer to the NI High-Speed Digitizers Help for information about when to self-calibrate the device. |
External calibration | The external calibration calibrates the onboard references used in self-calibration, the input overload levels, and the external trigger levels. All calibration constants are stored in nonvolatile memory. |
Interval for external calibration | 1 year |
Warm-up time | 25 minutes |
Power
+3.3 VDC | 5.1 A |
+12 VDC | 6.1 A |
+5 Vaux | 12 mA |
Total power | 90 W |
Software
Driver Software
This device is supported in NI-SCOPE 3.9.6 or later. NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5185 . NI-SCOPE provides application programming interfaces for many development environments.
Application Software
NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:
- LabVIEW
- LabWindows™/CVI™
- Measurement Studio
- Microsoft Visual C/C++
- Microsoft Visual Basic
Interactive Soft Front Panel and Configuration
The NI-SCOPE Soft Front Panel version 3.9.6 or later supports interactive control of the PXIe-5185 . The NI-SCOPE Soft Front Panel is included on the NI-SCOPE DVD.
National Instruments Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5185 . MAX is included on the NI-SCOPE DVD.
Physical
Dimensions and Weight
Dimensions | 3U, 3 slot, PXI Express Module, 21.6 × 6.2 × 13.0 cm (8.5 × 2.4 × 5.1 in.) | ||||||
|
Environment
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C |
Relative humidity range | 10% to 90%, noncondensing |
Storage Environment
Ambient temperature range | -40 °C to 71 °C |
Relative humidity range | 5% to 95%, noncondensing |
Shock and Vibration
Operating shock | 30 g peak, half-sine, 11 ms pulse | ||||||
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
1 50 Ω input only.
2 Signals exceeding the maximum input overload may cause damage to the device.
3 Within ±3 °C of self-calibration temperature.
4 Used to calculate errors when temperature changes more than ±3 °C since the last self-calibration.
5 Measured on one channel with test signal applied to other channel. Same range settings used on both channels.
6 Normalized to 50 kHz.
7 1 MΩ input tested using a 50 Ω source and a 50 Ω feed through terminator connected at the input.
8 50% FS input pulse, 23°C ± 10°C.
9 AC coupling available on 1 MΩ only.
10 For ≤100 MHz, -1 dBFS input signal corrected to FS. For >100 MHz, -2 dBFS input signal corrected to FS.
11 For 10 MHz, -1 dBFS input signal corrected to FS. For 300 MHz, -2 dBFS input signal corrected to FS. Includes the 2nd through the 5th harmonics. 18 kHz resolution bandwidth (RBW).
12 For 10 MHz, -1 dBFS input signal corrected to FS. For 300 MHz, -2 dBFS input signal corrected to FS. Includes the 2nd through the 5th harmonics. 18 kHz resolution bandwidth (RBW).
13 50 Ω terminator connected to input. 23°C ± 10°C.
14 50 Ω terminator connected to input. 23°C ± 10°C.
15 Internal Sample clock is locked to the PXIe_CLK100 Reference clock.
16 Divide by n decimation from 6.25 GS/s used for all rates less than maximum sample rate. For more information about Sample clock and decimation, refer to the NI High-Speed Digitizers Help.
17 With one channel enabled, stepped in multiples of 12.5 GS/s. With two channels enabled, stepped in multiples of 6.25 GS/s.
18 Includes the effects of the converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter.
19 Phase-locked to Reference clock. The chassis clock or external Reference clock must be accurate to 25 parts per million (ppm), or (1 × 10-6).
20 Divide by n decimation available where 1 ≤ n ≤ 65535. For more information about Sample clock and decimation, refer to the NI High-Speed Digitizers Help. The effective sample rate can be 1 × Input Frequency or 2 × Input Frequency when acquiring on two channels, or 1 × Input Frequency, 2 × Input Frequency or 4 × Input Frequency when acquiring on one channel; use the Sample Clock Timebase Multiplier property or the NISCOPE_ATTR_SAMP_CLK_TIMEBASE_MULT attribute to specify.
21 The PLL Reference clock frequency must be accurate to ±25 ppm.
22 The PLL Reference clock frequency must be accurate to ±25 ppm.
23 Holdoff set to 0.
24 Within ±5 °C of self-calibration temperature.
25 When same impedance settings used on both input channels. For more information about functionality when using mixed impedances between the input channels, visit ni.com/kb and enter 5W8CFE8P.
26 Caused by clock and analog path delay differences. No manual adjustment performed.
27 Synchronized triggers are synchronized to ±1 Sample clock timebase.
28 Onboard memory is shared between all enabled channels.
29 Single-record and multirecord acquisitions.
30 You can exceed these numbers if you fetch records while acquiring data. For more information, refer to the NI High-Speed Digitizers Help.