PXIe-5171 Specifications

PXIe-5171 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design, or verified during production and calibration.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured (meas) specifications describe the measured performance of a representative model.

Specifications are Nominal unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • All vertical ranges
  • Sample rate set to 250 MS/s
  • Onboard sample clock locked to onboard reference clock
  • PXIe-5171 module warmed up for 15 minutes at ambient temperature.[1]1 Warm-up begins after the chassis is powered, the device is recognized by the host, and the ADC clock is configured using either instrument design libraries or the NI-SCOPE device driver.
  • PXI Express chassis fan speed set to HIGH, foam fan filters removed if present, and empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users available at ni.com/docs.
  • Calibration IP used properly when using LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes (instrument design libraries) to create FPGA bitfiles. Refer to the NI Reconfigurable Oscilloscopes Help for more information about the calibration API.

Warranted specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature range of 0 °C to 45 °C
  • External calibration cycle maintained
  • External calibration performed at 23 °C ± 3 °C

Typical specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature ranges of 0 °C to 45 °C

Nominal and Measured specifications are valid under the following conditions unless otherwise noted.

  • Room temperature, approximately 23 °C

PXIe-5171 Front Panel

This section describes the front panel and connectors of the PXIe-5171.

Front Panel Connectors

Label Connector Type Function
CH 0—CH 7 SMA connector Analog input terminal
AUX I/O MHDMR connector Sample Clock or Reference Clock input, Reference Clock output, bidirectional digital PFI, and 3.3 V power output
Note This connector is referred to as AUX 0 in this document.
Note The AUX 0 connector accepts a standard, third-party HDMI™ type C cable, but the AUX 0 port is not an HDMI interface and the specified performance of the AUX 0 connector is not guaranteed if a third-party HDMI cable is used. Use NI cable type SHH19-MH19-AUX for all AUX 0 connections. Do not connect the AUX 0 port on the PXIe-5171 to the HDMI port of another device. NI is not liable for any damage resulting from such signal connections.

PXIe-5171 Pinout

Use the pinout to connect to terminals on the PXIe-5171.

Figure 1. PXIe-5171 AUX 0 Connector Pinout


Table 1. AUX 0 Connector Pin Assignments
Pin Signal Signal Description
1 GND Ground reference for signals
2 CLK IN Used to import an external Reference Clock or Sample Clock
3 GND Ground reference for signals
4 GND Ground reference for signals
5 CLK OUT Used to export the Reference Clock
6 GND Ground reference for signals
7 GND Ground reference for signals
8 AUX 0/PFI 0 Bidirectional PFI line
9 AUX 0/PFI 1 Bidirectional PFI line
10 GND Ground reference for signals
11 AUX 0/PFI 2 Bidirectional PFI line
12 AUX 0/PFI 3 Bidirectional PFI line
13 GND Ground reference for signals
14 AUX 0/PFI 4 Bidirectional PFI line
15 AUX 0/PFI 5 Bidirectional PFI line
16 AUX 0/PFI 6 Bidirectional PFI line
17 AUX 0/PFI 7 Bidirectional PFI line
18 +3.3 V +3.3 V power (200 mA maximum)
19 GND Ground reference for signals

PXIe-5171 SCB-19 Pinout

You can use the SCB-19 connector block to connect digital signals to the AUX 0 connector on the PXIe-5171 front panel. Refer to the following figure and table for information about the SCB-19 signals when connected to the AUX 0 front panel connector.



Table 2. SCB-19 Signal Descriptions
Pin Signal Signal Description
1 PFI 0 Bidirectional PFI line
2 PFI 1 Bidirectional PFI line
3 PFI 2 Bidirectional PFI line
4 PFI 3 Bidirectional PFI line
5 NC No connection
6 CLK IN Used to import an external reference clock or sample clock
7 NC No connection
8 CLK OUT Used to export the reference clock
9 PFI 4 Bidirectional PFI line
10 PFI 5 Bidirectional PFI line
11 PFI 6 Bidirectional PFI line
12 PFI 7 Bidirectional PFI line
13 +3.3 V +3.3 V power (200 mA maximum)
14 to 26 GND Ground reference for signals

PXIe-5171 AUX 0 Breakout Cable to 6 BNCs Pinout

You can use the AUX 0 Breakout Cable to 6 BNCs to connect digital signals to the AUX 0 connector on the PXIe-5171 front panel. Refer to the following figure and table for information about the AUX 0 Breakout Cable to 6 BNCs signals when connected to the AUX 0 front panel connector.



Table 3. AUX 0 Breakout Cable to 6 BNCs Signal Descriptions
Signal Connector Type Description
CLK IN BNC female Used to import an external reference clock
CLK OUT Used to export the reference clock
PFI 0 Bidirectional PFI line
PFI 1 Bidirectional PFI line
PFI 2 Bidirectional PFI line
PFI 3 Bidirectional PFI line

Vertical

Analog Input

Number of channels

8 (simultaneously sampled)

Input type

Referenced single-ended

Connectors

SMA

Impedance and Coupling

Input impedance

50 Ω ± 1.5%, typical

Input coupling

AC, DC

Figure 1. Voltage Standing Wave Ratio (VSWR)


Figure 1. Input Return Loss


Voltage Levels

Full-scale (FS) input range (Vpk-pk)

0.2 V

0.4 V

1 V

2 V

5 V

Maximum input overload[2]2 Signals exceeding the maximum input overload may cause damage to the device.

|Peaks| ≤ 5 V

Accuracy

Notice Electromagnetic interference can adversely affect the measurement accuracy of this product. The coaxial channel inputs of this device (CH 0 to CH 7) are not protected for electromagnetic interference. As a result, this device may experience reduced measurement accuracy or other temporary performance degradation when connected cables are routed in an environment with radiated or conducted radio frequency electromagnetic interference. To limit radiated emissions and to ensure that this device functions within specifications in its operational electromagnetic environment, take precautions when designing, selecting, and installing measurement probes and cables.

Resolution

14 bits

Table 4. DC Accuracy[3]
Input Range Accuracy Drift
Typical[4] Warranted[5] Nominal[6]
Vpk-pk ±(% of |Reading|+ % of FS+ mV) ±(% of |Reading|+ % of FS+ mV) ±(% of |Reading|+ % of FS+ mV) per °C
0.2 V ±(0.45 + 0.6 + 0.2)3 Verification of these specifications requires the DC Adjustment Device Temperature (°C) value. If you are using version 14.0 of the software, visit ni.com/info and enter the Info Code exxpmp for information on how to read this value. Otherwise, use NI-SCOPE to read the value.4 When the reading from the Device Temperature sensor is within ±10 °C of the DC Adjustment Device Temperature (°C) value. ±(0.90 + 0.65 + 0.7) 5 When the reading from the Device Temperature sensor is within ±38 °C of the DC Adjustment Device Temperature (°C) value. This increased temperature span encompasses the majority of temperature differences between the last external calibration environment and the operating environment. ±(0.015 + 0.002 + 0.004) 6 Used to calculate additional temperature error when the difference between the Device Temperature sensor and the DC Adjustment Device Temperature (°C) value is greater than ±10 °C (for typical specifications) or ±38 °C (for warranted specifications).
0.4 V ±(0.45 + 0.24 + 0.2) ±(0.80 + 0.25 + 0.7) ±(0.012 + 0.002 + 0.004)
1 V ±(0.45 + 0.2 + 0.2) ±(0.80 + 0.25 + 0.7) ±(0.010 + 0.002 + 0.004)
2 V ±(0.40 + 0.2 + 0.2) ±(0.60 + 0.25 + 0.7) ±(0.005 + 0.002 + 0.004)
5 V ±(0.40 + 0.2 + 0.2) ±(0.55 + 0.25 + 0.7) ±(0.005 + 0.002 + 0.004)

DC accuracy sampling drift, full bandwidth (±% of |Reading| per MHz from 250 MHz)[7]7 Used to calculate additional DC accuracy error when using an external sample clock of frequency <250 MHz. To calculate the additional error, solve the following for the analog path of interest:

250MHzfrequency1,000,000×DCaccuracysamplingdrift

±0.03, nominal

AC amplitude accuracy[3]

Accuracy

±0.095 dB at 50 kHz, typical[4]

±0.15 dB at 50 kHz, warranted[5]

Drift[6]

±0.0013 dB per °C

Conversion error rate[8]8 A conversion error is defined as deviation greater than 0.6% of full scale.

250 MS/sec

<1 × 10-10

200 MS/sec

<1 × 10-15

150 MS/sec

<1 × 10-20

Figure 1. Channel-to-Channel Crosstalk[9]9 Measured on one channel with test signal applied to another channel, with the same range setting on both channels.


Bandwidth and Transient Response

Bandwidth-limiting filter

100 MHz anti-alias filter

Bandwidth (-3 dB)GUID-18EAD082-4C17-40EC-9D69-1193909CD1BE.html#GUID-18EAD082-4C17-40EC-9D69-1193909CD1BE__NORMALIZEDFN

Anti-alias filter

100 MHz

Full bandwidth

0.2 Vpk-pk input range

260 MHz

All other input ranges

270 MHz

Table 5. Passband Amplitude Flatness, Warranted[10]10 Normalized to 50 kHz.
Input Frequency Anti-Alias Filter Enabled Full Bandwidth
<50 MHz -0.5 dB to 0.5 dB -0.5 dB to 0.5 dB
≥50 MHz to <90 MHz -1.0 dB to 0.5 dB -0.75 dB to 0.5 dB
≥90 MHz to <100 MHz -0.75 dB to 0.5 dB
≥100 MHz to <150 MHz -1 dB to 0.5 dB

AC-coupling cutoff (-3 dB)[11]11 With AC coupling enabled, the input impedance is 260 kΩ to ground. Verified using a 50 Ω source.

120 kHz





Spectral Characteristics

Table 6. Spurious-Free Dynamic Range (SFDR)[12]12 -1 dBFS input signal corrected to FS. 358 Hz resolution bandwidth (RBW).
Input Range (Vpk-pk) Input Frequency Full Bandwidth
0.2 V to 2 V <10 MHz -80.0 dBc
≥10 MHz to <30 MHz -76.0 dBc
5 V <10 MHz -77.0 dBc
≥10 MHz to <30 MHz -73.0 dBc
Table 7. Total Harmonic Distortion (THD)[13]13 Includes the second through the fifth harmonics. -1 dBFS input signal.
Input Frequency Full Bandwidth
<10 MHz -77.0
≥10 MHz to <30 MHz -73.0
Table 8. Effective Number of Bits (ENOB)[12]
Input Range (Vpk-pk) Input Frequency Full Bandwidth
0.2 V <30 MHz 10.8
All other input ranges <30 MHz 11.0


Noise
RMS noiseGUID-204DFAF7-77C3-4329-B5A1-BDEA7EF6BB2A.html#GUID-204DFAF7-77C3-4329-B5A1-BDEA7EF6BB2A__50OHMNOTE

Anti-alias filter enabled

0.017% of FS, typical

Full bandwidth

0.2 Vpk-pk input range

0.037% of FS, typical

0.4 Vpk-pk input range

0.025% of FS, typical

All other input ranges

0.024% of FS, typical

Table 9. Average Noise Density (dBm/Hz), Typical[14]14 Verified using a 50 Ω terminator connected to input.
Input Range (Vpk-pk) Anti-Alias Filter Enabled (dBm/Hz) Full Bandwidth (dBm/Hz)
0.2 V -159.2 dBm/Hz -153.6 dBm/Hz
0.4 V -153.7 dBm/Hz -150.4 dBm/Hz
1 V -145.7 dBm/Hz -142.4 dBm/Hz
2 V -139.7 dBm/Hz -136.4 dBm/Hz
5 V -131.7 dBm/Hz -128.4 dBm/Hz
Table 10. Average Noise Density (dBFS/Hz), Typical[14]
Input Range (Vpk-pk) Anti-Alias Filter Enabled (dBFS/Hz) Full Bandwidth (dBFS/Hz)
0.2 V 149.2 dBFS/Hz 143.6 dBFS/Hz
All other input ranges 149.7 dBFS/Hz 146.4 dBFS/Hz
Table 11. Average Noise Density (nV/√Hz), Typical[14]
Input Range (Vpk-pk) Anti-Alias Filter Enabled (nV/√Hz) Full Bandwidth (nV/√Hz)
0.2 V 3.5 nV/√Hz 6.6 nV/√Hz
0.4 V 6.5 nV/√Hz 9.6 nV/√Hz
1 V 16.4 nV/√Hz 23.9 nV/√Hz
2 V 32.7 nV/√Hz 47.9 nV/√Hz
5 V 81.8 nV/√Hz 119.7 nV/√Hz

Horizontal

Sample Clock

Sources

Internal

Onboard clock (internal VCXO)

External

AUX I/O CLK IN (front panel MHDMR connector)

PXIe_DStarA (backplane connector)

Sample rate range, real-time[15]15 Divide by n decimation from 250 MS/s. For more information about the sample clock and decimation, refer to the NI Reconfigurable Oscilloscopes Help at ni.com/manuals.

3.815 kS/s to 250 MS/s

Timebase frequency

Internal

250 MHz

External

150 MHz – 250 MHz[16]16 Variable external sample clock support was added in NI-SCOPE 18.7.

Timebase accuracy

Phase-locked to onboard clock

±25.0 ppm, warranted

Phase-locked to external clock

Equal to the external clock accuracy

Duty cycle tolerance

45% to 55%

Phase-Locked Loop (PLL) Reference Clock

Sources

Internal

Onboard clock (internal VCXO)

PXI_Clk10 (backplane connector)

External (10 MHz)

AUX I/O CLK IN (front panel MHDMR connector)

Duty cycle tolerance

45% to 55%

External Sample Clock

Source

AUX I/O CLK IN (front panel MHDMR connector)

Impedance

50 Ω

Coupling

AC

Input voltage range

As a 250 MHz sine wave

1 dBm through 18 dBm

As a fast slew rate input (square wave, Vpk-pk)

0.4 V to 5 V

Maximum input overload

As a 250 MHz sine wave

20 dBm

As a fast slew rate input (square wave, Vpk-pk)

6 V

External Reference Clock In

Source

AUX I/O CLK IN (front panel MHDMR connector)

Impedance

50 Ω

Coupling

AC

Frequency[17]17 The PLL reference clock frequency must be accurate to ±25 ppm.

10 MHz

Input voltage range

As a 250 MHz sine wave

1 dBm through 18 dBm

As a fast slew rate input (square wave, Vpk-pk)

6 V

Duty cycle tolerance

45% to 55%

Reference Clock Out

Source

PXI_Clk10 (backplane connector)

Destination

AUX I/O CLK OUT (front panel MHDMR connector)

Output impedance

50 Ω

Logic type

3.3 V LVCMOS

Maximum current drive

±8 mA

PXIe_DStarA

Source

System timing slot

Destinations

Onboard clock (internal VCXO)

FPGA

PXI_Clk100

Source

PXI backplane

Destination

FPGA

Trigger

Note The following characteristic behaviors are valid when using the device with the NI-SCOPE API. When using instrument design libraries, these characteristics may not be valid.

Supported trigger

Reference (Stop) Trigger

Trigger types

Edge

Window

Hysteresis

Digital

Immediate

Software

Trigger sources

CH 0 to CH 7

PFI <0..7>

PXI_Trig <0..6>

Software

Time resolution
Analog triggers

With interpolation

Sample Clock period / 1024

Without interpolation

Sample Clock period

Digital triggers

2x Sample Clock period

Approximate trigger delay difference between analog edge trigger and digital trigger source[18]18 This value is approximate because changes to the digital trigger routing or the analog signal path affect propagation delay. You can compensate for the delay difference by adjusting the NI-SCOPE trigger delay value. Add an additional 80 ns trigger delay when passing a trigger between PXIe-5171 modules. With the same hardware and software configuration, the trigger delay difference is consistent within the timing resolution across modules of the same model. For more information about the trigger delay difference, refer to Characterizing Setup to Account for Delay on Digital Trigger.

630 ns, nominal

Minimum dead time

With interpolation

240 x Sample Clock period

Without interpolation

130 x Sample Clock period

Holdoff

From dead time to [(264 - 1) × Sample Clock timebase period]

Trigger delay

From 0 to [(251 - 1) × Sample Clock timebase period]

Analog trigger accuracy with input frequencies less than 90 MHz

0.5% of full scale

Analog trigger jitter with input frequencies less than 90 MHz

15 psrms

Minimum threshold duration[19]19 Data must exceed each corresponding trigger threshold for at least the minimum duration to ensure analog triggering.

Sample Clock period

Note Trigger interpolation is used when the Enable TDCNI-SCOPE attribute is set to TRUE. Otherwise, trigger interpolation is not used.

For more information about triggers, refer to Triggering in NI-SCOPE.

Programmable Function Interface (PFI 0..7, AUX I/O Front Panel Connector)

Connector

AUX I/O

Direction

Bidirectional per channel

Direction control latency

25 ns

As an Input (Trigger)

Destination

FPGA diagram

Start Trigger (Acquisition Arm)

Reference (Stop) Trigger

Arm Reference Trigger

Advance Trigger

Input impedance

10 kΩ

VIH

2 V

VIL

0.8 V

Maximum input overload

0 V to 3.3 V, 5 V tolerant

Minimum pulse width

10 ns

As an Output (Event)

Sources

FPGA diagram

Ready for Start

Start Trigger (Acquisition Arm)

Ready for Reference

Reference (Stop) Trigger

End of Record

Ready for Advance

Advance Trigger

Done (End of Acquisition)

Output impedance

50 Ω

Logic type

3.3 V CMOS

Maximum current drive

12 mA

Minimum pulse width

10 ns

Power Output (+3.3 V)

Connector

AUX I/O/+3.3 V front panel connector

Voltage output

3.3 V ± 10%

Maximum current drive

200 mA

Output impedance

<1 Ω

Waveform Specifications

Onboard memory size[20]20 Onboard memory is shared among all enabled channels.

1.5 GB

Minimum record length

1 sample

Number of pretrigger samples

Zero up to (record length - 1)

Number of posttrigger samples

Zero up to record length

Channels Max Records per Channel Record Length
1 1 805306192
1 10 80530432
1 1000 805120
1 100,000 7840
1 1M 592
2 1 402653096
2 10 40265216
2 1000 402560
2 100,000 3920
2 1M 296
4 1 201326548
4 10 20132608
4 1000 201280
4 100,000 1960
4 1M 148
8 1 100663274
8 10 10066304
8 1000 100640
8 100,000 980
8 1M 74

Memory Sanitization

For information about memory sanitization, refer to the letter of volatility for your device, which is available at ni.com/manuals.

FPGA

FPGA support

Xilinx Kintex-7 XC7K410T FPGA

Xilinx Kintex-7 XC7K410T FPGA Resources

Slice registers

508,400

Slice look-up tables (LUT)

254,200

DSPs

1,540

18 Kb block RAMs

1,590

Note Note that some of these resources are consumed by the logic necessary to operate the device and integrate with software, and are thus out of the control of users.

External Calibration

External calibration corrects for gain, offset, and timing errors at all input ranges.

All calibration constants are stored in nonvolatile memory.

Self-Calibration

Self-calibration is done on software command. The calibration corrects for intermodule synchronization errors.

For information about when to self-calibrate the device, refer to the documentation of your device at ni.com.

Calibration Specifications

Interval for external calibration

2 years

Warm-up time[21]21 Warm-up begins after the chassis is powered, the device is recognized by the host, and the device is configured using the instrument design libraries or NI-SCOPE. Running an included sample project or running self-calibration using NI-MAX will configure the device and start warm-up.

15 minutes

Software

Driver Software

This device was first supported in LabVIEW Instrument Design Libraries for Reconfigurable Oscilloscopes 14.0 and NI-SCOPE 15.1. NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the device. NI-SCOPE provides application programming interfaces for many development environments.

For information about the available software options, refer to the documentation of your device at ni.com.

Application Software

NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:

  • LabVIEW
  • LabWindows™/CVI™
  • Measurement Studio
  • Microsoft Visual C/C++
  • .NET (C# and VB.NET)

Interactive Soft Front Panel and Configuration

When you install NI-SCOPE on a 64-bit system, you can monitor, control, and record measurements from the PXIe-5171 using InstrumentStudio.

InstrumentStudio is a software-based front panel application that allows you to perform interactive measurements on several different device types in a single program.

Note InstrumentStudio is supported only on 64-bit systems. If you are using a 32-bit system, use the NI-SCOPE–specific soft front panel instead of InstrumentStudio.

Interactive control of the PXIe-5171 was first available via InstrumentStudio in NI-SCOPE18.1 and via the NI-SCOPE SFP in NI-SCOPE15.1. InstrumentStudio and the NI-SCOPE SFP are included on the NI-SCOPE media.

NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5171. MAX is included on the driver media.

Synchronization

Channel-to-channel skew

Anti-alias filter enabled

<120 ps, nominal[22]22 For input frequencies less than 75 MHz.

Full bandwidth

<120 ps, nominal

Synchronization with the NI-TClk API

NI-TClk is an API that enables system synchronization of supported PXI modules in one or more PXI chassis, which you can use with the PXIe-5171 and NI-SCOPE. NI-TClk installs with NI-SCOPE.

NI-TClk uses a shared Reference Clock and triggers to align the Sample Clocks of PXI modules and synchronize the distribution and reception of triggers. These signals are routed through the PXI chassis backplane without external cable connections between PXI modules in the same chassis.

Module-to-module skew, between PXIe-5171 modules using NI-TClk [23]23 Although you can use NI-TClk to synchronize non-identical modules, these specifications apply only to synchronizing identical modules. Specifications are valid under the following conditions:
  • All modules installed in the same PXI Express chasses.
  • NI-TClk used to align the sample clocks of each module.
  • All parameters set to identical values for each module.
  • Self-calibration completed.
  • Ambient temperature within ±1 °C of self-calibration.
For other configurations, including multi-chassis systems, contact NI Technical Support at ni.com/support.
NI-TClk synchronization without manual adjustment 24 Manual adjustment is the process of minimizing synchronization jitter and skew by adjusting Trigger Clock (TClk) signals using the instrument driver. [24]

Skew, Peak-to-Peak 25 Caused by clock and analog delay differences. Tested with a PXIe-1082 chassis with maximum slot to slot skew of 100 ps. Valid within ±1 °C of self-calibration. [25]

300 ps

NI-TClk synchronization with manual adjustment [24]

Skew after manual adjustment

≤10 ps

Sample Clock delay/adjustment resolution

3.5 ps

Power

Note Power consumed depends on the FPGA image and driver software used. This specifications represents the maximum power for the NI-SCOPE use case or the typical value when using the Instrument Design Libraries (IDL).
Table 12. PXIe-5171 Power Consumption
Instrument Design Libraries NI-SCOPE
+3.3 VDC 6.4 W 6.3 W
+12 VDC 16.2W 17.2W
Total power 22.6 W 23.5 W

Total maximum power allowed

38.25 W

Dimensions and Weight

Dimensions

18.5 cm × 2.0 cm × 13.0 cm (7.3 in. × 0.8 in. × 5.1 in.)

3U, 1 slot, PXI Express Gen 2 x8 Module

Environment

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient temperature)

Pollution Degree

2

Indoor use only.

Operating Environment

Ambient temperature range

0 °C to 40 °C

Relative humidity range

10% to 90%, noncondensing

Storage Environment

Ambient temperature range

-40 °C to 71 °C

Relative humidity range

5% to 95%, noncondensing

Shock and Vibration

Operating shock

30 g peak, half-sine, 11 ms pulse

Random vibration

Operating

5 Hz to 500 Hz, 0.3 g RMS

Nonoperating

5 Hz to 500 Hz, 2.4 g RMS

Compliance and Certifications

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • EN 55022 (CISPR 22): Class A emissions
  • EN 55024 (CISPR 24): Immunity
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • AS/NZS CISPR 22: Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the Product Certifications and Declarations section.

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers
  • Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)
  • 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 Warm-up begins after the chassis is powered, the device is recognized by the host, and the ADC clock is configured using either instrument design libraries or the NI-SCOPE device driver.

    2 Signals exceeding the maximum input overload may cause damage to the device.

    3 Verification of these specifications requires the DC Adjustment Device Temperature (°C) value. If you are using version 14.0 of the software, visit ni.com/info and enter the Info Code exxpmp for information on how to read this value. Otherwise, use NI-SCOPE to read the value.

    4 When the reading from the Device Temperature sensor is within ±10 °C of the DC Adjustment Device Temperature (°C) value.

    5 When the reading from the Device Temperature sensor is within ±38 °C of the DC Adjustment Device Temperature (°C) value. This increased temperature span encompasses the majority of temperature differences between the last external calibration environment and the operating environment.

    6 Used to calculate additional temperature error when the difference between the Device Temperature sensor and the DC Adjustment Device Temperature (°C) value is greater than ±10 °C (for typical specifications) or ±38 °C (for warranted specifications).

    7 Used to calculate additional DC accuracy error when using an external sample clock of frequency <250 MHz. To calculate the additional error, solve the following for the analog path of interest:

    250MHzfrequency1,000,000×DCaccuracysamplingdrift

    8 A conversion error is defined as deviation greater than 0.6% of full scale.

    9 Measured on one channel with test signal applied to another channel, with the same range setting on both channels.

    10 Normalized to 50 kHz.

    11 With AC coupling enabled, the input impedance is 260 kΩ to ground. Verified using a 50 Ω source.

    12 -1 dBFS input signal corrected to FS. 358 Hz resolution bandwidth (RBW).

    13 Includes the second through the fifth harmonics. -1 dBFS input signal.

    14 Verified using a 50 Ω terminator connected to input.

    15 Divide by n decimation from 250 MS/s. For more information about the sample clock and decimation, refer to the NI Reconfigurable Oscilloscopes Help at ni.com/manuals.

    16 Variable external sample clock support was added in NI-SCOPE 18.7.

    17 The PLL reference clock frequency must be accurate to ±25 ppm.

    18 This value is approximate because changes to the digital trigger routing or the analog signal path affect propagation delay. You can compensate for the delay difference by adjusting the NI-SCOPE trigger delay value. Add an additional 80 ns trigger delay when passing a trigger between PXIe-5171 modules. With the same hardware and software configuration, the trigger delay difference is consistent within the timing resolution across modules of the same model. For more information about the trigger delay difference, refer to Characterizing Setup to Account for Delay on Digital Trigger.

    19 Data must exceed each corresponding trigger threshold for at least the minimum duration to ensure analog triggering.

    20 Onboard memory is shared among all enabled channels.

    21 Warm-up begins after the chassis is powered, the device is recognized by the host, and the device is configured using the instrument design libraries or NI-SCOPE. Running an included sample project or running self-calibration using NI-MAX will configure the device and start warm-up.

    22 For input frequencies less than 75 MHz.

    23 Although you can use NI-TClk to synchronize non-identical modules, these specifications apply only to synchronizing identical modules. Specifications are valid under the following conditions:

    • All modules installed in the same PXI Express chasses.
    • NI-TClk used to align the sample clocks of each module.
    • All parameters set to identical values for each module.
    • Self-calibration completed.
    • Ambient temperature within ±1 °C of self-calibration.
    For other configurations, including multi-chassis systems, contact NI Technical Support at ni.com/support.

    24 Manual adjustment is the process of minimizing synchronization jitter and skew by adjusting Trigger Clock (TClk) signals using the instrument driver.

    25 Caused by clock and analog delay differences. Tested with a PXIe-1082 chassis with maximum slot to slot skew of 100 ps. Valid within ±1 °C of self-calibration.