PXIe-5114 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured specifications describe the measured performance of a representative model.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • All filter settings
  • All impedance selections
  • Sample clock set to 250 MS/s

Warranted specifications are valid under the following conditions unless otherwise noted.

  • Temperature range of 0 °C to 55 °C
  • The PXIe-5114 is warmed up for 15 minutes at ambient temperature
  • Self-calibration is completed after warm-up period
  • Calibration cycle is maintained
  • The PXI Express chassis fan speed is set to HIGH, the foam fan filters are removed if present, and the empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users available at ni.com/manuals.
  • External calibration is performed at 23 °C ± 3 °C

Vertical

Analog Input

Number of channels

Two (simultaneously sampled)

Input type

Referenced single-ended

Connectors

BNC

Impedance and Coupling

Input impedance

50 Ω

50 Ω ± 1.5%

1 MΩ

1 MΩ ± 1% in parallel with a nominal capacitance of 26 pF

Input coupling

AC[1], DC, GND

Voltage Levels

Table 1. Full-Scale (FS) Input Range and Programmable Vertical Offset Range
Input Range (Vpk-pk) Vertical Offset Range (V)
50 Ω Input 1 MΩ Input
0.04 V ±0.8 ±0.8
0.1 V
0.2 V
0.4 V
1 V ±6.5 ±8.0
2 V ±6.0
4 V ±5.0
10 V ±2.0 ±30
20 V ±25
40 V ±15
Maximum input overload

50 Ω

7 V RMS with |Peaks| ≤10 V

1 MΩ

|Peaks| ≤35 V

Accuracy

Resolution

8 bits

DC accuracy[2]

±[(1.5% × |Reading - Vertical Offset| + 2% of Vertical Offset + 0.3% of FS + 200 µV), warranted

DC drift

±(0.03% × Reading + 0.06% of FS + 40 µV) per °C, nominal

Crosstalk[3]

At 10 MHz input frequency

≤-60 dB

At 100 MHz input frequency

≤-45 dB

Bandwidth and Transient Response

Bandwidth (-3 dB)

0.04 Vpk-pk input range

100 MHz, warranted

All other input ranges

125 MHz, warranted

Rise/fall time

0.04 Vpk-pk input range

3.5 ns

All other input ranges

2.8 ns

Bandwidth-limiting filter

20 MHz noise filter

AC coupling[4] cutoff (-3 dB)

12 Hz

Passband flatness[5]

±1 dB up to 50 MHz, warranted

Figure 1. PXIe-5114 Frequency Response, Measured

1378

Spectral Characteristics

Spurious-free dynamic range (SFDR)[6]

58 dBc

Total harmonic distortion (THD)[6]

-58 dBc

Effective number of bits (ENOB), calculated[6]

0.04 Vpk-pk input range

6.2

All other input ranges

7.2

Signal to noise and distortion (SINAD)[6]

0.04 Vpk-pk input range

38 dB

All other input ranges

44 dB

Noise

Table 2. RMS Noise[7]
Input Range (Vpk-pk) Full Bandwidth 20 MHz Filter Enabled
0.04 V 0.45% FS 0.28% FS
All other input ranges 0.28% FS 0.28% FS

Horizontal

Sample Clock

Sources

Internal[8]

Onboard clock (internal VCXO)

External

CLK IN (front panel SMB connector)

Onboard Clock (Internal VCXO)

Sample rate range

Real-time sampling (single-shot)[9]

3.815 kS/s to 250 MS/s

Random interleaved sampling (RIS)

250 MS/s to 5 GS/s in 250 MS/s increments

Timebase frequency[10]

250 MHz

Timebase accuracy

Not phase-locked to Reference clock

±25 ppm, warranted

Phase-locked to Reference clock

Equal to the Reference clock accuracy

Sample clock delay range

±1 Sample clock period

Sample clock delay resolution

≤20 ps

External Sample Clock

Source

CLK IN (front panel SMB connector)

Frequency range[11]

50 MHz to 250 MHz

Duty cycle tolerance

45% to 55%

Phase-Locked Loop (PLL) Reference Clock

Sources

PXI_CLK10 (backplane connector)

CLK IN (front panel SMB connector)

Frequency range[12]

5 MHz to 20 MHz in 1 MHz increments

Default: 10 MHz

Duty cycle tolerance

45% to 55%

Exported Reference clock destinations

PFI <0..1> (front panel 9-pinmini-circular DIN connector)

PXI_Trig <0..7>

External Sample Clock and External Reference Clock In

Source

CLK IN (front panel SMB connector)

Input voltage range

Sine wave

0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)

Square wave

0.2 Vpk-pk to 2.8 Vpk-pk

Maximum input overload

7 V RMS with |Peaks| ≤10 V

Impedance

50 Ω

Coupling

AC

Trigger

Reference (Stop) Trigger

Trigger types[13]

Edge

Window

Hysteresis

Video

Digital

Immediate

Software

Trigger sources

CH 0

CH 1

TRIG

PXI_Trig <0..6>

PFI <0..1>

RTSI <0..6>

Software

Time resolution

Onboard clock, time-to-digital conversion circuit (TDC) on

40 ps

Onboard clock, TDC off

4 ns

External clock, TDC off

External clock period

Minimum rearm time[14]

TDC on

10 µs

TDC off

2 µs

Holdoff

From rearm time up to [(235 - 1) × Sample Clock Period]

Trigger delay

From 0 up to [(235 - 1) - Posttrigger Samples] × (1 / Sample Rate), in seconds

Analog Trigger

Trigger types

Edge

Window

Hysteresis

Sources

CH 0 (front panel BNC connector)

CH 1 (front panel BNC connector)

TRIG (front panel BNC connector)

Trigger level resolution

8 bits (1 in 256)

Trigger level range

CH 0, CH 1

100% FS

TRIG (external trigger)

±5 V

Edge trigger sensitivity

CH 0, CH 1

5% FS up to 100 MHz, warranted

TRIG (external trigger)

0.5 Vpk-pk up to 125 MHz, warranted

Level accuracy

CH 0, CH 1

±5% FS up to 100 MHz

TRIG (external trigger)

±0.5 V up to 10 MHz

Jitter

≤65 ps RMS

Trigger filters

Low frequency (LF) reject

50 kHz

High frequency (HF) reject

50 kHz

Digital Trigger

Trigger type

Digital

Sources

PXI_Trig <0..6> (backplane connector)

PFI <0..1> (front panel DIN connector)

Video Trigger

Trigger type

Video

Sources

CH 0 (front panel BNC connector)

CH 1 (front panel BNC connector)

TRIG (front panel BNC connector)

Video trigger types

Specific Line

Any Line

Specific Field

Standards[15]

SDTV

M-NTSC

B/G-PAL

SECAM

M-PAL

EDTV

480i/59.94 fps

480i/60 fps

480p/59.94 Fps

480p/60 Fps

576i/50 fps

576p/50 Fps

HDTV[16]

720p/50 Fps

720p/59.94 Fps

720p/60 Fps

1080i/50 fps

1080i/59.94 fps

1080i/60 fps

1080p/24 Fps

External Trigger

Connector

TRIG (front panel BNC connector)

Impedance

1 MΩ in parallel with 22 pF

Coupling

AC, DC

AC-coupling cutoff (-3 dB)

12 Hz

Input voltage range

±5 V

Maximum input overload

|Peaks| ≤42 V

Programmable Function Interface (PFI 0 and PFI 1)

Connector

AUX I/O (front panel 9-pinmini-circular DIN connector)

Direction

Bidirectional

As an Input (Trigger)

Destinations

Start trigger (acquisition arm)

Reference (Stop) trigger

Arm reference trigger

Advance trigger

Input impedance

150 kΩ, nominal

VIH

2.0 V

VIL

0.8 V

Maximum input overload

-0.5 V, 5.5 V

Maximum frequency

25 MHz

As an Output (Event)

Sources

Start trigger (acquisition arm)

Reference (Stop) trigger

End of Record

Done (end of acquisition)

Probe Compensation[17]

Output impedance

50 Ω

Logic type

3.3 V CMOS

Maximum drive current

±24 mA

Maximum frequency

25 MHz

Waveform

Onboard memory sizes[18]

8 MB per channel (standard)

64 MB per channel (option)

256 MB per channel (option)

Minimum record length

1 sample

Number of pretrigger samples[19]

Zero up to full record length

Number of posttrigger samples[19]

Zero up to full record length

Maximum number of records in onboard memory[20]

8 MB

32,768

64 MB

100,000

256 MB

100,000

Allocated onboard memory per record

(Record Length × 1 byte/sample) + 240 bytes, rounded up to next multiple of 128 bytes (minimum 256 bytes)

Calibration

External Calibration

External calibration calibrates the onboard references used in self-calibration and the external trigger levels. All calibration constants are stored in nonvolatile memory.

Calibration Specifications

Interval for external calibration

2 years

Warm-up time[21]

15 minutes

Software

Driver Software

Driver support for this device was first available in NI-SCOPE14.1.

NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5114. NI-SCOPE provides application programming interfaces for many development environments.

Application Software

NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:

  • LabVIEW
  • LabWindows™/CVI™
  • Measurement Studio
  • Microsoft Visual C/C++
  • .NET (C# and VB.NET)

Interactive Soft Front Panel and Configuration

When you install NI-SCOPE on a 64-bit system, you can monitor, control, and record measurements from the PXIe-5114 using InstrumentStudio.

InstrumentStudio is a software-based front panel application that allows you to perform interactive measurements on several different device types in a single program.

Note InstrumentStudio is supported only on 64-bit systems. If you are using a 32-bit system, use the NI-SCOPE–specific soft front panel instead of InstrumentStudio.

Interactive control of the PXIe-5114 was first available via InstrumentStudio in NI-SCOPE18.1 and via the NI-SCOPE SFP in NI-SCOPE14.1. InstrumentStudio and the NI-SCOPE SFP are included on the NI-SCOPE media.

NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5114. MAX is included on the driver media.

TClk Specifications

You can use the NI TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of supported devices, in one or more chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.

Intermodule SMC Synchronization Using NI-TClk for Identical Modules

Specifications are valid for modules installed in one NI PXI-1042 chassis. These specifications do not apply to PCI modules. Specifications are valid under the following conditions:

  • All parameters are set to identical values for each SMC-based module.
  • Sample clock set to 250 MS/s.
  • All filters are disabled.
Note Although you can use NI-TClk to synchronize non-identical modules, these specifications apply only to synchronizing identical modules.

Skew [22]

500 ps

Average skew after manual adjustment [23]

<20 ps

Sample clock adjustment resolution

<20 ps

Power

Current draw

+3.3 VDC

1.1 A

+12 VDC

1.5 A

Total power

21.63 W

Physical

Dimensions and Weight

Dimensions

21.4 cm × 2.0 cm × 13.0 cm

(8.4 in × 0.8 in × 5.1 in)

Weight

501 g (17.7 oz)

Figure 1. PXIe-5114

1378

Front Panel

Table 3. PXIe-5114 Front Panel Connectors
Label Connector Type Description
CH 0 BNC female Analog input connection; digitizes data and triggers acquisitions.
CH 1 Analog input connection; digitizes data and triggers acquisitions.
TRIG External analog trigger; signals on the TRIG connector cannot be digitized.
CLK IN SMB jack Imports an external Reference clock or Sample clock to the oscilloscope.
AUX I/O 9-pin mini-circular DIN PFI 0 and PFI 1 lines for digital trigger input/output and probe compensation.
Table 4. PXIe-5114 Front Panel Indicators[24]
Label Description
ACCESS The ACCESS LED indicates the status of the PCI bus and the interface from the device to the controller.
ACTIVE The ACTIVE LED indicates the status of the onboard acquisition hardware of the device.

Environment

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient temperature)

Pollution Degree

2

Indoor use only.

Operating Environment

Ambient temperature range

0 °C to 55 °C

Relative humidity range

10% to 90%, noncondensing

Storage Environment

Ambient temperature range

-40 °C to 71 °C

Relative humidity range

5% to 95%, noncondensing

Shock and Vibration

Operating shock

30 g peak, half-sine, 11 ms pulse

Random vibration

Operating

5 Hz to 500 Hz, 0.3 grms

Nonoperating

5 Hz to 500 Hz, 2.4 grms

Compliance and Certifications

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • EN 55022 (CISPR 22): Class A emissions
  • EN 55024 (CISPR 24): Immunity
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • AS/NZS CISPR 22: Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the Product Certifications and Declarations section.

CE Compliance 1378

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)
  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
  • 2011/65/EU; Restriction of Hazardous Substances (RoHS)

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • 1378 Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 1378 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 AC coupling available on 1 MΩ input only.

    2 Within ±5 °C of self-calibration temperature.

    3 CH 0 to/from CH 1, and External Trigger to CH 0 or CH 1

    4 AC coupling available on 1 MΩ input only.

    5 Referenced to 50 kHz, with bandwidth-limiting filter disabled.

    6 10 MHz, 10 MHz - 1 dBFS input signal. Includes the second through the fifth harmonics. Measured from DC to 125 MHz, with the 20 MHz bandwidth-limiting filter disabled. ENOB values are corrected to full scale.

    7 Verified using a 50 Ω terminator connected to input.

    8 Internal Sample clock is locked to the Reference clock or derived from the onboard VCXO.

    9 Divide by n sampling used for all rates less than 250 MS/s. For more information about the Sample clock and decimation, refer to the NI High-Speed Digitizers Help.

    10 When not using external Sample clock.

    11 Divide by n decimation available where 1 ≤ n ≤ 65,535. For more information about the Sample clock and decimation, refer to the NI High-Speed Digitizers Help.

    12 The PLL Reference clock frequency must be accurate to ±50 ppm.

    13 Refer to the following sources and the NI High-Speed Digitizers Help for more information about which sources are available for each trigger type.

    14 Holdoff set to 0. Onboard Sample clock at maximum rate.

    15 fps = fields per second. Fps = frames per second.

    16 This trigger type does not support negative trigger polarity.

    17 1 kHz, 50% duty cycle square wave, PFI 1 only.

    18 Onboard memory is shared between enabled channels.

    19 Single-record and multirecord acquisitions.

    20 You can exceed these numbers if you fetch records while acquiring data. For more information, refer to the NI High-Speed Digitizers Help.

    21 Warm-up time begins after the NI-SCOPE driver is loaded.

    22 Caused by clock and analog path delay differences. No manual adjustment performed.

    23 For more information about manual adjustment, refer to the Synchronization Repeatability Optimization topic in the NI-TClk Synchronization Help.

    24 For more information, refer to the NI High-Speed Digitizers Help.