PXIe-4190 Specifications
- Updated2023-08-09
- 29 minute(s) read
PXIe-4190 Specifications
Notes on PXIe-4190 Variants
In this document, the 500 kHz and 2 MHz variants of the PXIe-4190 are referred to inclusively as the PXIe-4190. The information in this document applies to all variants of the PXIe-4190 unless otherwise specified.
- On the device front panel, the PXIe-4190 (2 MHz) shows PXIe-4190 2MHz LCR Meter/SMU. The PXIe-4190 (500 kHz) shows NI PXIe-4190 500kHz LCR Meter/SMU.
- In MAX, the PXIe-4190 (2 MHz) appears as NI PXIe-4190. The PXIe-4190 (500 kHz) appears as NI PXIe-4190 (500 kHz).
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
SMU Specifications
SMU Specifications
Conditions
SMU mode specifications are valid only when the following conditions are met unless otherwise noted.
- Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 5 ºC
- Temperature is within ±5 °C of last self-calibration (Tcal)
- Relative humidity between 10% and 60%, noncondensing
- Chassis with slot cooling capacity ≥58 W
- Calibration interval of 1 year
- 30 minutes warm-up time
- Self-calibration performed within the last 24 hours
- NI-DCPower 23.3 or later installed
- Connections between force and sense leads are required[2]2 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
- niDCPower Aperture Time property set to 2 power-line cycles (PLC)
- niDCPower Cable Length property set when using the lower two current ranges
SMU Instrument Capabilities
PXIe-4190 (2 MHz) | PXIe-4190 (500 kHz) |
---|---|
|
|
PXIe-4190 (2 MHz) | PXIe-4190 (500 kHz) |
---|---|
|
|
PXIe-4190 (2 MHz) | PXIe-4190 (500 kHz) | |
---|---|---|
Sourcing | 4 W | 1 W |
Sinking | 4 W | 1 W |
SMU Voltage
Range | Resolution (Noise Limited) | Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) | Accuracy ± (% of Voltage + Offset) | Tempco† ± (% of Voltage + Offset)/°C | |
---|---|---|---|---|---|
Tambient 23 °C ±5 °C, Tcal* ±5 °C | |||||
% of Voltage | Offset | Tambient 0 °C to 33 °C, Tcal ±5 °C | |||
1 V | 100 nV | 2 µV | 0.009% | 160 µV | 0.0002% + 1 µV |
10 V | 1 µV | 10 µV | 0.008% | 1 mV | |
40 V ‡ | 4 µV | 50 µV | 0.009% | 4.1 mV | |
* Tcal is the internal device temperature recorded by the PXIe-4190 at the completion of the last self-calibration. † Temperature coefficient applies beyond 23 °C ±5 °C ambient within ±5 °C Tcal. ‡ PXIe-4190 (2 MHz) only |
SMU Current
Range | Resolution (Noise Limited) | Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) | Accuracy ± (% of Current + Offset) | Tempco† ± (% of Current + Offset)/°C | |
---|---|---|---|---|---|
Tambient 23 °C ±5 °C, Tcal* ±5 °C | |||||
% of Current | Offset‡,§ | Tambient 0 °C to 33 °C, Tcal ±5 °C | |||
1 nA**,†† | 1 fA | 30 fA | 0.14% | 2 pA | 0.0003% + 20 fA |
1 nA‡‡,†† | 60 fA | ||||
100 nA**,†† | 10 fA | 300 fA | 0.091% | 11 pA | |
100 nA‡‡,†† | 700 fA | ||||
1 µA†† | 100 fA | 2 pA | 0.032% | 140 pA | |
10 µA | 1 pA | 15 pA | 0.026% | 1 nA | |
100 µA | 10 pA | 120 pA | 0.024% | 10 nA | |
1 mA | 100 pA | 1.2 nA | 0.023% | 100 nA | |
10 mA | 1 nA | 12 nA | 0.022% | 1 µA | |
100 mA | 10 nA | 120 nA | 0.028% | 10 µA | |
* Tcal is the internal device temperature recorded by the PXIe-4190 at the completion of the last self-calibration. † Temperature coefficient applies beyond 23 °C ±5 °C ambient within ±5 °C Tcal. ‡ Add 10 pA to current accuracy specifications when using DSUB-DSUB cable accessory (SHDB13W6-DB13W6-LL). § Add 10 pA to current accuracy specifications when operating with Tambient >30 °C. ** Under the following additional conditions: with 10 PLC, and 11-point median filter. †† PXIe-4190 (2 MHz) only ‡‡ Under default specification conditions.
|
SMU Noise
SMU Load Regulation
Voltage | Error included in accuracy specifications |
Current | Error included in accuracy specifications |
SMU Transient Response and Settling
Time
Range | Voltage Mode, ≤5 V Step, Unloaded* | Current Mode, Full-Scale Step† |
---|---|---|
100 mA to 10 µA | <200 μs | <200 μs |
1 µA | <350 μs | <2 ms |
100 nA | <2 ms | <8 ms |
1 nA | <40 ms | <1,100 ms |
Note: Measured as the time to settle to within 0.1% of step amplitude, PXIe-4190 configured for fast transient response, with 1 m cable. * Current limit set to 100% of selected current range for 1 nA and 100 nA ranges, all other ranges set to 50% of selected current range. † Voltage limit set to ≥2 V, resistive load set to 1 V/selected current range. |
Current Range | Recovery Time* | Voltage Dip | Time Constant† |
---|---|---|---|
100 mA | <40 µs | <1.3 V | <10 µs |
10 mA | <40 µs | <1.2 V | <10 µs |
1 mA | <40 µs | <800 mV | <17 µs |
100 µA | <65 µs | <500 mV | <35 µs |
10 µA | <150 µs | <200 mV | <50 µs |
1 µA | <450 µs | <35 mV | <340 µs |
100 nA | — | <8 mV | <3 ms |
1 nA | — | <800 µV | <300 ms |
Note: Load current change from 10% to 90% of range, PXIe-4190 configured for fast transient response, with 1 m cable. * Recovery Time defined as the time to recover within 10 mV after load current change. † Time Constant defined as the time to recover within 63% of Voltage Dip after load current change. |
SMU Remote Sense
Maximum sense lead resistance | 200 Ω |
Maximum lead drop per lead | 1 V |
SMU Guard Output Characteristics
|
SMU Measurement and Update Timing
Sample rate accuracy | Equal to PXIe_CLK100 accuracy, nominal |
Maximum measure rate to host | 600 kS/s, nominal |
Maximum source update rate, sequence mode | 100,000 updates/s (10 μs/update), nominal |
|
LCR Specifications
LCR Specifications
Conditions
LCR mode specifications are valid only when the following conditions are met unless otherwise noted.
- Ambient temperature[5]5 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 10 ºC
- Temperature is within ±5 °C of last self-calibration (Tcal)
- Relative humidity between 10% and 60%, noncondensing
- Chassis with slot cooling capacity ≥58 W
- Calibration interval of 1 year
- 30 minutes warm-up time
- Self-calibration performed within the last 24 hours
- NI-DCPower 23.3 or later installed
- AC Stimulus Automatic Level Control (ALC) is On
- DC Bias Automatic Level Control (ALC) set to On
- Impedance range is within 30% of DUT impedance
- LCR Measurement Time is Long unless otherwise stated
- Source delay set to Automatic
- Open and short LCR compensation has been completed.
- Connections between force and sense leads are required[6]6 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
- Four-terminal pair (4TP) connections to load[7]7 Refer to the PXIe-4190 Getting Started for more information on 4TP connections.
- niDCPower Cable Length property set
LCR Instrument Capabilities
The PXIe-4190 is capable of measuring the following elements using AC stimulus frequencies from 40 Hz to 2 MHz:
- Capacitors—100 fF to 5 mF, with up to 100 aF sensitivity
- Inductors—Greater than 10 nH, with up to 10 pH sensitivity
- Resistors—100 mΩ to 1 GΩ, with up to 10 µΩ sensitivity
Maximum AC voltage | 7.07 V RMS | ||||||
Maximum AC current | 70.7 mA RMS | ||||||
|
Maximum DC bias current range | ±100 mA, including peak AC stimulus |
| |||||||||||
|
Calculating Total LCR Measurement Time per
Setpoint
Total Measurement Time per setpoint = LCR Source Delay + Total LCR Measurement Time
Calculating LCR Source Delay Time
- LCR Source Delay Mode = Automatic
- In Automatic mode, the source delay is 20 cycles of the AC stimulus frequency with a minimum source delay of 1 ms.
- LCR Source Delay Mode = Manual
- LCR Source Delay time is as specified for the Source Delay property.
Note Using a source delay smaller than the default value may not allow the output to sufficiently settle, resulting in measurement inaccuracy. - Setpoint changes that result in a range change add an additional 600 μs of source delay in either mode.
Calculating Total LCR Measurement Time
-
- N—Measurement count
-
- CoercedMeasurementTime—
- The measurement time coerces to a full sinewave cycle boundary regardless of
mode.
Where LCR Measurement Time = Short (1 ms)/Medium (10 ms))/Long (100 ms)/Custom
Note LCR Custom Measurement Time = 0 is a special case that gives 1 cycle for any frequency.
- The measurement time coerces to a full sinewave cycle boundary regardless of
mode.
LCR Measurements
LCR AC Stimulus
Voltage stimulus | ||
Maximum | 7.07 V RMS | |
Minimum | 7.07 mV RMS | |
Resolution | <1 µV RMS | |
Maximum current | 70.7 mA RMS | |
Accuracy (ALC on) | ||
≤10 kHz | ±0.4% | |
>10 kHz | ±4% |
Current stimulus | ||
Maximum | 70.7 mA RMS | |
Minimum | 707 nA RMS | |
Resolution | <100 pA RMS | |
Maximum voltage | 7.07 V RMS | |
Accuracy (ALC on) | ||
≤10 kHz | ±0.5% | |
>10 kHz | ±6% |
LCR DC Bias
| |||||||||
| |||||||||
|
LCR Frequency
Accuracy | Equal to PXIe_CLK100 accuracy, nominal |
Frequency resolution | 1 mHz |
LCR Measurement Accuracy
This topic shows the illustrated LCR measurement accuracy for capacitive DUTs.
The following figure shows capacitor impedance magnitude versus test frequency to help quickly identify the appropriate impedance range for your measurements. Additionally, several important DUT test points across frequency are highlighted, with the corresponding Absolute Measurement Accuracy and AC Stimulus range shown in Table 8.
Complete absolute accuracy specifications are described beginning in Table 10.
Test Point | Capacitor Value | AC Stimulus Frequency | AC Stimulus Level | ZC at AC Stimulus Frequency | Measurement Accuracy | |
---|---|---|---|---|---|---|
Magnitude (Capacitance) | Phase (Dissipation Factor) | |||||
1 | 1 pF | 10 kHz | 708 mV RMS to 7.07 V RMS | 15.9 MΩ | 0.15% (1.5 fF) | 0.08° (0.0014) |
2 | 1 pF | 100 kHz | 708 mV RMS to 7.07 V RMS | 1.59 MΩ | 0.30% (3 fF) | 0.19° (0.0033) |
3 | 1 pF | 2 MHz | 708 mV RMS to 5 V RMS | 79.6 kΩ | 0.60% (6 fF) | 0.26° (0.0045) |
4 | 100 pF | 1 kHz | 708 mV RMS to 7.07 V RMS | 1.59 MΩ | 0.06% (60 fF) | 0.03° (0.0005) |
5 | 1 nF | 2 MHz | 150 mV RMS to 707 mV RMS | 79.6 Ω | 0.50% (50 pF) | 0.18° (0.0031) |
6 | 100 nF | 1 kHz | 150 mV RMS to 707 mV RMS | 1.59 kΩ | 0.05% (50 pF) | 0.02° (0.00035) |
7 | 1 µF | 10 kHz | 150 mV RMS to 707 mV RMS | 15.9 Ω | 0.08% (800 pF) | 0.22° (0.0038) |
8 | 100 µF | 120 Hz | 50 mV RMS to 150 mV RMS | 13.3 Ω | 0.08% (80 nF) | 0.04° (0.0007) |
Capacitor Value | AC Stimulus Voltage | AC Stimulus Frequency | Capacitance Magnitude Accuracy | Phase Accuracy | Df Accuracy |
---|---|---|---|---|---|
13.3 pF < C ≤ 132.6 pF | 1.0 V RMS | 120 Hz | ±0.15% | ±0.08° | ±0.001396 |
132.6 pF < C ≤ 1.3 nF | 1.0 V RMS | 120 Hz | ±0.06% | ±0.03° | ±0.000524 |
1.3 nF < C ≤ 13.3 nF | 1.0 V RMS | 120 Hz | ±0.06% | ±0.02° | ±0.000349 |
13.3 nF < C ≤ 132.6 nF | 1.0 V RMS | 120 Hz | ±0.05% | ±0.02° | ±0.000349 |
132.6 nF < C ≤ 1.3 µF | 1.0 V RMS | 120 Hz | ±0.06% | ±0.02° | ±0.000349 |
1.3 µF < C ≤ 93.8 µF | 1.0 V RMS | 120 Hz | ±0.08% | ±0.03° | ±0.000524 |
93.8 µF < C ≤ 132.6 µF | 0.5 V RMS | 120 Hz | ±0.08% | ±0.03° | ±0.000524 |
1.6 pF < C ≤ 15.9 pF | 1.0 V RMS | 1 kHz | ±0.15% | ±0.08° | ±0.001396 |
15.9 pF < C ≤ 159.2 pF | 1.0 V RMS | 1 kHz | ±0.06% | ±0.03° | ±0.000524 |
159.2 pF < C ≤ 1.6 nF | 1.0 V RMS | 1 kHz | ±0.06% | ±0.02° | ±0.000349 |
1.6 nF < C ≤ 15.9 nF | 1.0 V RMS | 1 kHz | ±0.05% | ±0.02° | ±0.000349 |
15.9 nF < C ≤ 159.2 nF | 1.0 V RMS | 1 kHz | ±0.06% | ±0.02° | ±0.000349 |
159.2 nF < C ≤ 11.3 µF | 1.0 V RMS | 1 kHz | ±0.08% | ±0.03° | ±0.000524 |
11.3 µF < C ≤ 15.9 µF | 0.5 V RMS | 1 kHz | ±0.08% | ±0.03° | ±0.000524 |
624 fF < C ≤ 1.6 pF | 1.0 V RMS | 1 MHz | ±0.30% | ±0.16° | ±0.002793 |
1.6 pF < C ≤ 15.9 pF | 1.0 V RMS | 1 MHz | ±0.30% | ±0.13° | ±0.002269 |
15.9 pF < C ≤ 159.2 pF | 1.0 V RMS | 1 MHz | ±0.20% | ±0.12° | ±0.002094 |
159.2 pF < C ≤ 530.5 pF | 1.0 V RMS | 1 MHz | ±0.20% | ±0.12° | ±0.002094 |
530.5 pF < C ≤ 11.3 nF | 1.0 V RMS | 1 MHz | ±0.20% | ±0.11° | ±0.001920 |
11.3 nF < C ≤ 15.9 nF | 0.5 V RMS | 1 MHz | ±0.20% | ±0.13° | ±0.002269 |
LCR
Magnitude and Phase Accuracy
Impedance Range | AC Stimulus Frequency | ||||||
---|---|---|---|---|---|---|---|
40 Hz to 100 Hz | 100 Hz to 1 kHz | 1 kHz to 10 kHz | 10 kHz to 200 kHz | 200 kHz to 500 kHz | 500 kHz to 1 MHz* | 1 MHz to 2 MHz* | |
100 MΩ to 1 GΩ | 1.00%, typical | 1.00% | — | — | — | — | — |
10 MΩ to 100 MΩ | 0.15%, typical | 0.15% | 0.15% | — | — | — | — |
1 MΩ to 10 MΩ | 0.06%, typical | 0.06% | 0.15% | 0.30% | — | — | — |
100 kΩ to 1 MΩ | 0.05% | 0.06% | 0.08% | 0.30% | 0.30%† | 0.30%‡ | 0.60%§ |
10 kΩ to 100 kΩ | 0.05% | 0.05% | 0.08% | 0.30% | 0.30% | 0.30% | 0.60% |
1 kΩ to 10 kΩ | 0.05% | 0.06% | 0.08% | 0.20% | 0.20% | 0.20% | 0.50% |
300 Ω to 1 kΩ | 0.08% | 0.08% | 0.08% | 0.15% | 0.15% | 0.20% | 0.50% |
10 Ω to 300 Ω | 0.08% | 0.08% | 0.20% | 0.20% | 0.20% | 0.20% | 0.50% |
* PXIe-4190 (2 MHz) only † Up to 640 kΩ impedance range. ‡ Up to 255 kΩ impedance range. § Up to 130 kΩ impedance range. Note: Impedances <10 Ω require a reduced AC stimulus. Refer to the following table for more information. Note: When on boundary, use lower adjacent value. Note: Add the following derating factor to LCR magnitude when AC stimulus level is >5 V RMS and >1 MHz: |
Impedance Range | AC Stimulus Frequency | ||||||
---|---|---|---|---|---|---|---|
40 Hz to 100 Hz | 100 Hz to 1 kHz | 1 kHz to 10 kHz | 10 kHz to 200 kHz | 200 kHz to 500 kHz | 500 kHz to 1 MHz* | 1 MHz to 2 MHz* | |
10 MΩ to 100 MΩ | 0.20%, typical | 0.40% | 1.10% | — | — | — | — |
1 MΩ to 10 MΩ | 0.06%, typical | 0.06% | 0.20% | 0.90% | — | — | — |
100 kΩ to 1 MΩ | 0.05% | 0.06% | 0.08% | 0.90% | 0.60%† | 0.60%‡ | 0.60%§ |
10 kΩ to 100 kΩ | 0.05% | 0.05% | 0.08% | 0.30% | 0.30% | 0.30% | 0.50% |
1 kΩ to 10 kΩ | 0.05% | 0.05% | 0.08% | 0.20% | 0.20% | 0.20% | 0.50% |
300 Ω to 1 kΩ | 0.08% | 0.08% | 0.08% | 0.15% | 0.15% | 0.20% | 0.50% |
10 Ω to 300 Ω | 0.08% | 0.08% | 0.08% | 0.20% | 0.20% | 0.20% | 0.50% |
<10 Ω** | 0.08% + 1 mΩ | 0.08% + 1 mΩ | 0.08% + 1 mΩ | 0.90% + 1mΩ | 0.90% + 1 mΩ | 1.20% + 1 mΩ†† | 2.00% + 2 mΩ†† |
* PXIe-4190 (2 MHz) only † Up to 640 kΩ impedance range. ‡ Up to 255 kΩ impedance range. § Up to 130 kΩ impedance range. ** Typical, offset relative to short compensation. †† Refer to AC Stimulus Current Short Offset Multiplier table for offset multiplier. Note: When on boundary, use lower adjacent value. |
AC Stimulus Current | Short Offset Multiplier |
---|---|
<7.07 mA | 5 |
7.08 mA to 20 mA | 1 |
>20 mA | 2 |
Impedance Range | AC Stimulus Voltage | |
---|---|---|
50 mV RMS to 150 mV RMS | 7.08 mV RMS to 50 mV RMS, typical | |
<10 Ω | 1 | 1 |
10 Ω to 300 Ω | 2 | 5 |
300 Ω to 10 MΩ | 2 | 11 |
10 MΩ to 100 MΩ | 3 | — |
Note: Absolute accuracy is the Absolute Impedance Magnitude Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage table value times the respective multiplier. |
Impedance Range | AC Stimulus Frequency | ||||||
---|---|---|---|---|---|---|---|
40 Hz to 100 Hz | 100 Hz to 1 kHz | 1 kHz to 10 kHz | >10 kHz to 200 kHz | 200 kHz to 500 kHz | 500 kHz to 1 MHz* | 1 MHz to 2 MHz* | |
100 MΩ to 1 GΩ | 0.55 °, typical | 0.55 ° | — | — | — | — | — |
10 MΩ to 100 MΩ | 0.19 °, typical | 0.08 ° | 0.25 ° | — | — | — | — |
1 MΩ to 10 MΩ | 0.02 °, typical | 0.03 ° | 0.21 ° | 0.19 ° | — | — | — |
100 kΩ to 1 MΩ | 0.01 ° | 0.02 ° | 0.19 ° | 0.19 ° | 0.14 °† | 0.16 °‡ | 0.26 °§ |
10 kΩ to 100 kΩ | 0.01 ° | 0.02 ° | 0.10 ° | 0.11 ° | 0.12 ° | 0.13 ° | 0.26 ° |
1 kΩ to 10 kΩ | 0.01 ° | 0.02 ° | 0.09 ° | 0.10 ° | 0.10 ° | 0.12 ° | 0.31 ° |
300 Ω to 1 kΩ | 0.01 ° | 0.03 ° | 0.12 ° | 0.08 ° | 0.13 ° | 0.12 ° | 0.34 ° |
10 Ω to 300 Ω | 0.01 ° | 0.03 ° | 0.13 ° | 0.08 ° | 0.09 ° | 0.11 ° | 0.15 ° |
* PXIe-4190 (2 MHz) only † Up to 640 kΩ impedance range. ‡ Up to 255 kΩ impedance range. § Up to 130 kΩ impedance range. Note: Impedances <10 Ω require a reduced AC stimulus. Refer to the following table for more information. Note: When on boundary, use lower adjacent value. |
Impedance Range | AC Stimulus Frequency | ||||||
---|---|---|---|---|---|---|---|
40 Hz to 100 Hz | 100 Hz to 1 kHz | 1 kHz to 10 kHz | >10 kHz to 200 kHz | 200 kHz to 500 kHz | 500 kHz to 1 MHz* | 1 MHz to 2 MHz* | |
10 MΩ to 100 MΩ | 0.14 °, typical | 0.30 ° | 0.50 ° | — | — | — | — |
1 MΩ to 10 MΩ | 0.03 °, typical | 0.03 ° | 0.14 ° | 0.45 ° | — | — | — |
100 kΩ to 1 MΩ | 0.02 ° | 0.03 ° | 0.14 ° | 0.45 ° | 0.22 °† | 0.22 °‡ | 0.34 °§ |
10 kΩ to 100 kΩ | 0.01 ° | 0.02 ° | 0.07 ° | 0.15 ° | 0.14 ° | 0.14 ° | 0.34 ° |
1 kΩ to 10 kΩ | 0.01 ° | 0.02 ° | 0.07 ° | 0.15 ° | 0.09 ° | 0.11 ° | 0.20 ° |
300 Ω to 1 kΩ | 0.01 ° | 0.02 ° | 0.07 ° | 0.08 ° | 0.09 ° | 0.12 ° | 0.34 ° |
10 Ω to 300 Ω | 0.01 ° | 0.04 ° | 0.22 ° | 0.08 ° | 0.10 ° | 0.13 ° | 0.18 ° |
<10 Ω, typical | 0.01 ° | 0.04 ° | 0.08 ° | 0.03 ° | 0.07 ° | 0.15 ° | 0.20 ° |
* PXIe-4190 (2 MHz) only † Up to 640 kΩ impedance range. ‡ Up to 255 kΩ impedance range. § Up to 130 kΩ impedance range. Note: When on boundary, use lower adjacent value. |
Impedance Range | AC Stimulus Voltage | |
---|---|---|
50 mV RMS to 150 mV RMS | 7.08 mV RMS to 50 mV RMS, typical | |
<10 Ω | 1 | 4 |
10 Ω to 300 Ω | 2 | 20 |
300 Ω to 1 kΩ | 2 | 70 |
1 kΩ to 10 kΩ | 2 | 25 |
>10 kΩ to 100 kΩ | 2 | 25 |
100 kΩ to 1 MΩ | 2 | 10 |
1 MΩ to 10 MΩ | 2 | 8 |
10 MΩ to 100 MΩ | 3 | 8 |
Note: Absolute accuracy is the Absolute Impedance Phase Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage table value times the respective multiplier. |
LCR Noise
Measurement Time | Derating Factor |
---|---|
Medium |
|
Short |
|
Note: Measurement time derating is a function of impedance magnitude and AC stimulus voltage level, and is independent of frequency. Specifications are determined by comparing differences in the standard deviation for different measurement times. |
LCR Accuracy Derating with DC Bias
Above 500 kHz with DC Bias enabled, add the additional error term to the stated magnitude accuracy specification:
Above 500 kHz with DC Bias enabled, add the additional error term to the stated phase accuracy specification:
LCR DC Bias Settling Time for Large Capacitors
Set DC Bias ALC to Off when measuring capacitors over 1 uF to minimize settling time.
Bias | Settling Time |
---|---|
AC Stimulus Voltage / Impedance Range ≤ 7.07 mA | Add 3 ms per µF of DUT capacitance |
AC Stimulus Voltage / Impedance Range > 7.07 mA | Add 600 µs per µF of DUT capacitance |
Note: When applying a bias voltage to capacitors over 1 mF, the bias voltage steps should be no larger than 40 V x 1 mF/C to avoid tripping overcurrent protection. Note: For AC Stimulus Voltage / Impedance Range ≤7.07mA, limit DC bias steps to ≤5 V each, up to 40 V total. |
LCR Cable Accuracy Derating
Cable | Description | NI Part Number | |||
---|---|---|---|---|---|
0.5 m | 1 m | 2 m* | 4 m* | ||
SHDB13W6-4BNCM-LL | DSUB to Male BNC | — | 788280-01 | 788280-02 | 788280-04 |
SHDB13W6-4BNCF-LL | DSUB to Female BNC | 789536-0R5 | 789536-01 | 789536-02 | — |
SHDB13W6-DB13W6-LL | DSUB to DSUB | — | 788279-01 | 788279-02 | 788279-04 |
SHDB13W6-4TriaxM-LL | DSUB to Male Triaxial | — | 788281-01 | 788281-02 | 788281-04 |
* LCR specifications are typical |
≤707 mV RMS | >707 mV RMS | |
---|---|---|
≤10 kHz | ||
>10 kHz |
Determining LCR Measurement Impedance
Range
The impedance range can be calculated and programmed in several ways. The following methods allow you to set the impedance range directly.
Calculating Impedance Range Manually
Use the following formulas to determine the expected impedance based on the load.
The impedance of an ideal capacitor is
The impedance of an ideal inductor is
Setting LCR Impedance Range Source Programmatically
By setting LCR Impedance Range Source to LCR Load Configuration, the range can be determined automatically based on the AC stimulus frequency, and the load settings LCR Load Resistance, LCR Load Inductance, and LCR Load Capacitance.
- LCR Voltage Range
- LCR Current Range
- LCR DC Bias Voltage Range
- LCR DC Bias Current Range
The LCR Voltage Range and LCR Current Range are expressed as RMS values but are equivalent to the corresponding SMU mode ranges when converted to peak value.
Translating LCR Specifications to Other
Impedance Parameters
Accuracy for additional impedance parameters can be derived from the absolute impedance magnitude and phase specifications. For some calculations, the actual DUT impedance must also be known—in these cases, the measured value can be used as an approximation with typically negligible impact on the result.
To determine the specified range of possible Q values, calculate
When DDUT is sufficiently small (<0.1), ΔCSpec ≅ Δ|Z|Spec (%).
For a general solution, determine accuracy using the AC stimulus frequency and reactance specification:
When DDUT is sufficiently small (<0.1), ΔLSpec ≅ Δ|Z|Spec (%).
For a general solution, determine accuracy using the AC stimulus frequency and reactance specification:
For typical non-reactive resistance measurements (DDUT > 10), ΔRSpec ≅ Δ|Z|Spec (%).
To determine accuracy for an arbitrary impedance, first find the maximum and minimum values, RMax and RMin, from four calculations:
For typical L/C measurements (DDUT < 0.1), ΔXSpec ≅ Δ|Z|Spec (%).
To determine accuracy for an arbitrary impedance, first find the maximum and minimum values, RMax and RMin, from four calculations:
The following figure shows the relationship between these parameters when an example vector Z is plotted on the complex impedance plane.
Example 1: Calculating Specifications for Capacitance Measurement
For a capacitor measurement under the stated conditions, complete the following steps to determine and interpret the absolute measurement accuracy.
DUT Actual Capacitance (CDUT) | 10 pF |
DUT Actual Dissipation (DDUT) | 0.005 |
AC Stimulus Frequency (f) | 1 MHz |
AC Stimulus Voltage (Vstim) | 1 V RMS |
DC Bias Voltage (V DC) | 10 V |
Measurement time | Short |
Cable length | 1 m |
- Calculate ideal capacitor impedance as
- Based on the 1 V RMS AC stimulus, the applicable magnitude
and phase specs are found in Table 10 and Table 14, respectively.
- From the calculated impedance, ZDUT, the relevant impedance range is 10 kΩ – 100 kΩ.
- 1 MHz is on the boundary between the 500 kHz to 1 MHz and 1 MHz to 2 MHz frequency ranges, so choose the smaller of the adjacent values.
- The resulting specifications are 0.3% magnitude accuracy and 0.13° phase accuracy.
- Base specifications apply to long measurement time. For short measurement time, apply the
derating factor:
- The derated magnitude specification is then 1.5 * 0.3% = 0.45%
- The derated phase specification is then 1.5 * 0.13° = 0.195°
- Because DC bias is enabled and f >500 kHz, the
additional error terms from LCR Accuracy Derating with DC Bias apply:
- The additional magnitude error is calculated as
- The additional phase error is calculated as
- The additional magnitude error is calculated as
- From the previous steps, the final accuracy specifications under these
measurement conditions:
- Magnitude accuracy, ΔZSpec = 0.45% + 0.03% = 0.48%
- Phase accuracy, ΔθSpec = 0.195° + 0.01° = 0.205°
These specifications can then be used to derive accuracies for other parameters.
- Dissipation factor accuracy, ΔDSpec = ±tan(ΔθSpec) = ±0.0036
- Specified range is DDUT ± ΔDSpec = 0.0013 to 0.0087
- Quality factor specified range is
- Impedance phase has an accuracy of ΔθSpec = 0.205° and can be
expressed as
- Loss angle, δDUT = arctan(DDUT) = arctan(0.005) = 0.286° ± 0.205°
- Impedance phase angle, θDUT = δ – 90° = 0.286° - 90° = -89.714° ± 0.205°
- Resistance accuracy, ΔRSpec, can be calculated by
- Finding the maximum and minimum values, RMax and RMin, from four
calculations:
- Selecting the maximum and minimum values, RMax = 137 Ω, RMin = 22.4 Ω
- Finding the maximum and minimum values, RMax and RMin, from four
calculations:
- Reactance accuracy, ΔXSpec, can be calculated by
- Since DDUT is small, ΔXSpec ≅ Δ|Z|Spec
- Using this simplified approximation, ΔXSpec = 0.48% * 15.915 kΩ = ± 76.4 Ω
- For example, to compare the explicitly calculated specification, first find the maximum
and minimum values, XMax and XMin, from the four
calculations:
- Selecting the maximum and minimum values, XMax = -15.838 kΩ, XMin = -15.991 kΩ
General Specifications
Isolation
Isolation voltage, any pin to earth ground | 40 V DC, Measurement Category I, functional |
Protection
| |||||||||
|
Output channel protection | ||
Output HI to GUARD/Isolated Shield | ||
Overvoltage | Automatic output disable | |
Output LO to all terminals | ||
Overcurrent | Automatic output disable | |
Sense HI/Sense LO to all terminals | ||
Overcurrent | Current limiter protects inputs up to absolute maximum voltage specification | |
Overtemperature | Automatic output disable |
Physical Characteristics
Dimensions | 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm x 13.0 cm x 21.6 cm (0.8 in. x 5.1 in. x 8.5 in.) |
Weight | 481 g (17.1 oz) |
Front panel connectors | Mixed layout DSUB, 13W6 contact arrangement (6 coaxial 50 Ω, 7-signal), female |
Triggers
Input triggers | ||
Types | Start, Source, Sequence Advance, Measure | |
Sources (PXI trigger lines 0 to 7) | ||
Polarity | Active high (not configurable) | |
Minimum pulse width | 100 ns | |
Destinations[8]8 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source. (PXI trigger lines 0 to 7) | ||
Polarity | Active high (not configurable) | |
Minimum pulse width | 200 ns |
Output triggers (events) | ||
Types | Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete | |
Destinations (PXI trigger lines 0 to 7)[9]9 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1. | ||
Polarity | Active high (not configurable) | |
Pulse width | 230 ns |
Calibration Interval
Recommended calibration interval | 1 year |
Power Requirements
+3.3 V | 1.0 A |
+12 V | 2.7 A |
Environmental Characteristics
Pollution Degree | 2 |
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Humidity | |
Operating | 10% RH to 90% RH, noncondensing [11]11 Accuracy specifications are only warranted for operating environments with temperatures below 30 °C and relative humidity levels below 60%. When transitioning the product from a storage or operating environment with relative humidity above 60%, you should allow the product to stabilize in the lower humidity environment for several hours before using it. |
Storage | 5% RH to 95% RH, noncondensing |
|
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
3 10 Hz to 20 MHz bandwidth, PXIe-4190 configured for normal transient response.
4 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate.
5 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
6 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
7 Refer to the PXIe-4190 Getting Started for more information on 4TP connections.
8 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source.
9 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.
10 Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.
11 Accuracy specifications are only warranted for operating environments with temperatures below 30 °C and relative humidity levels below 60%. When transitioning the product from a storage or operating environment with relative humidity above 60%, you should allow the product to stabilize in the lower humidity environment for several hours before using it.
In This Section
- Notes on PXIe-4190 Variants
- Definitions
- SMU Specifications
- SMU Instrument Capabilities
- SMU Voltage
- SMU Current
- SMU Noise
- SMU Load Regulation
- SMU Transient Response and Settling
Time
- SMU Remote Sense
- SMU Guard Output Characteristics
- SMU Measurement and Update Timing
- LCR Specifications
- LCR Instrument Capabilities
- LCR Measurements
- LCR AC Stimulus
- LCR DC Bias
- LCR Frequency
- LCR Measurement Accuracy
- General Specifications