PXIe-4190 Specifications

Notes on PXIe-4190 Variants

In this document, the 500 kHz and 2 MHz variants of the PXIe-4190 are referred to inclusively as the PXIe-4190. The information in this document applies to all variants of the PXIe-4190 unless otherwise specified.

To determine which version of the PXIe-4190 you have, locate the device name in one of the following places:
  • On the device front panel, the PXIe-4190 (2 MHz) shows PXIe-4190 2MHz LCR Meter/SMU. The PXIe-4190 (500 kHz) shows NI PXIe-4190 500kHz LCR Meter/SMU.
  • In MAX, the PXIe-4190 (2 MHz) appears as NI PXIe-4190. The PXIe-4190 (500 kHz) appears as NI PXIe-4190 (500 kHz).

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Warranted unless otherwise noted.

SMU Specifications

SMU Specifications Conditions

SMU mode specifications are valid only when the following conditions are met unless otherwise noted.

  • Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 5 ºC
  • Temperature is within ±5 °C of last self-calibration (Tcal)
  • Relative humidity between 10% and 60%, noncondensing
  • Chassis with slot cooling capacity ≥58 W
  • Calibration interval of 1 year
  • 30 minutes warm-up time
  • Self-calibration performed within the last 24 hours
  • NI-DCPower 23.3 or later installed
  • Connections between force and sense leads are required[2]2 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
  • niDCPower Aperture Time property set to 2 power-line cycles (PLC)
  • niDCPower Cable Length property set when using the lower two current ranges
Note To avoid excessive relay wear, avoid setting Output Connected to TRUE with a non-zero voltage connected to the output.

SMU Instrument Capabilities

Table 1. DC Voltage Ranges
PXIe-4190 (2 MHz) PXIe-4190 (500 kHz)
  • 1 V
  • 10 V
  • 40 V
  • 1 V
  • 10 V
Table 2. DC Current Ranges
PXIe-4190 (2 MHz) PXIe-4190 (500 kHz)
  • 1 nA
  • 100 nA
  • 1 µA
  • 10 µA
  • 100 µA
  • 1 mA
  • 10 mA
  • 100 mA
  • 10 µA
  • 100 µA
  • 1 mA
  • 10 mA
  • 100 mA
Table 3. Available DC output power
PXIe-4190 (2 MHz) PXIe-4190 (500 kHz)
Sourcing 4 W 1 W
Sinking 4 W 1 W
Figure 1. PXIe-4190 (2 MHz) Quadrant Diagram

1378

Figure 2. PXIe-4190 (500 kHz) Quadrant Diagram

1378

SMU Voltage

Table 4. Voltage Programming and Measurement Accuracy/Resolution
Range Resolution (Noise Limited) Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) Accuracy ± (% of Voltage + Offset) Tempco ± (% of Voltage + Offset)/°C
Tambient 23 °C ±5 °C, Tcal* ±5 °C
% of Voltage Offset Tambient 0 °C to 33 °C, Tcal ±5 °C
1 V 100 nV 2 µV 0.009% 160 µV 0.0002% + 1 µV
10 V 1 µV 10 µV 0.008% 1 mV
40 V 4 µV 50 µV 0.009% 4.1 mV

* Tcal is the internal device temperature recorded by the PXIe-4190 at the completion of the last self-calibration.

Temperature coefficient applies beyond 23 °C ±5 °C ambient within ±5 °C Tcal.

PXIe-4190 (2 MHz) only

SMU Current

Table 5. Current Programming and Measurement Accuracy/Resolution
Range Resolution (Noise Limited) Noise (0.1 Hz to 10 Hz, peak-to-peak, typical) Accuracy ± (% of Current + Offset) Tempco ± (% of Current + Offset)/°C
Tambient 23 °C ±5 °C, Tcal* ±5 °C
% of Current Offset‡,§ Tambient 0 °C to 33 °C, Tcal ±5 °C
1 nA**,†† 1 fA 30 fA 0.14% 2 pA 0.0003% + 20 fA
1 nA‡‡,†† 60 fA
100 nA**,†† 10 fA 300 fA 0.091% 11 pA
100 nA‡‡,†† 700 fA
1 µA†† 100 fA 2 pA 0.032% 140 pA
10 µA 1 pA 15 pA 0.026% 1 nA
100 µA 10 pA 120 pA 0.024% 10 nA
1 mA 100 pA 1.2 nA 0.023% 100 nA
10 mA 1 nA 12 nA 0.022% 1 µA
100 mA 10 nA 120 nA 0.028% 10 µA

* Tcal is the internal device temperature recorded by the PXIe-4190 at the completion of the last self-calibration.

Temperature coefficient applies beyond 23 °C ±5 °C ambient within ±5 °C Tcal.

Add 10 pA to current accuracy specifications when using DSUB-DSUB cable accessory (SHDB13W6-DB13W6-LL).

§ Add 10 pA to current accuracy specifications when operating with Tambient >30 °C.

** Under the following additional conditions: with 10 PLC, and 11-point median filter.

†† PXIe-4190 (2 MHz) only

‡‡ Under default specification conditions.

SMU Noise

Wideband source noise

<20 mV peak-to-peak, typical[3]3 10 Hz to 20 MHz bandwidth, PXIe-4190 configured for normal transient response.

Figure 3. Voltage RMS Noise versus Aperture Time, Nominal

1378

Figure 4. Current RMS Noise versus Aperture Time, Nominal

1378

Note Use an aperture time of at least 1 PLC to minimize powerline noise pickup in the 1 nA range.

SMU Load Regulation

Voltage

Error included in accuracy specifications

Current

Error included in accuracy specifications

SMU Transient Response and Settling Time

Table 6. Settling Time, Typical
Range Voltage Mode, ≤5 V Step, Unloaded* Current Mode, Full-Scale Step
100 mA to 10 µA <200 μs <200 μs
1 µA <350 μs <2 ms
100 nA <2 ms <8 ms
1 nA <40 ms <1,100 ms

Note: Measured as the time to settle to within 0.1% of step amplitude, PXIe-4190 configured for fast transient response, with 1 m cable.

* Current limit set to 100% of selected current range for 1 nA and 100 nA ranges, all other ranges set to 50% of selected current range.

Voltage limit set to ≥2 V, resistive load set to 1 V/selected current range.

Table 7. Transient Response, Typical
Current Range Recovery Time* Voltage Dip Time Constant
100 mA <40 µs <1.3 V <10 µs
10 mA <40 µs <1.2 V <10 µs
1 mA <40 µs <800 mV <17 µs
100 µA <65 µs <500 mV <35 µs
10 µA <150 µs <200 mV <50 µs
1 µA <450 µs <35 mV <340 µs
100 nA <8 mV <3 ms
1 nA <800 µV <300 ms

Note: Load current change from 10% to 90% of range, PXIe-4190 configured for fast transient response, with 1 m cable.

* Recovery Time defined as the time to recover within 10 mV after load current change.

Time Constant defined as the time to recover within 63% of Voltage Dip after load current change.

SMU Remote Sense

Maximum sense lead resistance

200 Ω

Maximum lead drop per lead

1 V

SMU Guard Output Characteristics

Cable guard

Output impedance

<100 mΩ, nominal

Offset voltage

1 mV, typical

SMU Measurement and Update Timing

Available sample rates[4]4 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate.

(600 kS/s)/N, nominal

where

  • N = 1, 2, 3, … 224
  • S is samples

Sample rate accuracy

Equal to PXIe_CLK100 accuracy, nominal

Maximum measure rate to host

600 kS/s, nominal

Maximum source update rate, sequence mode

100,000 updates/s (10 μs/update), nominal

Input trigger to

Source event delay

10 µs nominal

Source event jitter

2 µs peak-to-peak, nominal

Measure event jitter

2 µs peak-to-peak, nominal

LCR Specifications

LCR Specifications Conditions

LCR mode specifications are valid only when the following conditions are met unless otherwise noted.

  • Ambient temperature[5]5 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 10 ºC
  • Temperature is within ±5 °C of last self-calibration (Tcal)
  • Relative humidity between 10% and 60%, noncondensing
  • Chassis with slot cooling capacity ≥58 W
  • Calibration interval of 1 year
  • 30 minutes warm-up time
  • Self-calibration performed within the last 24 hours
  • NI-DCPower 23.3 or later installed
  • AC Stimulus Automatic Level Control (ALC) is On
  • DC Bias Automatic Level Control (ALC) set to On
  • Impedance range is within 30% of DUT impedance
  • LCR Measurement Time is Long unless otherwise stated
  • Source delay set to Automatic
  • Open and short LCR compensation has been completed.
  • Connections between force and sense leads are required[6]6 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.
  • Four-terminal pair (4TP) connections to load[7]7 Refer to the PXIe-4190 Getting Started for more information on 4TP connections.
  • niDCPower Cable Length property set
Note To avoid excessive relay wear, avoid setting Output connected to TRUE with a non-zero voltage connected to the output.

LCR Instrument Capabilities

The PXIe-4190 is capable of measuring the following elements using AC stimulus frequencies from 40 Hz to 2 MHz:

  • Capacitors—100 fF to 5 mF, with up to 100 aF sensitivity
  • Inductors—Greater than 10 nH, with up to 10 pH sensitivity
  • Resistors—100 mΩ to 1 GΩ, with up to 10 µΩ sensitivity

Maximum AC voltage

7.07 V RMS

Maximum AC current

70.7 mA RMS

Maximum DC bias voltage range

PXIe-4190 (2 MHz)

±40 V, including peak AC stimulus

PXIe-4190 (500 kHz)

±10 V, including peak AC stimulus

Maximum DC bias current range ±100 mA, including peak AC stimulus
AC stimulus frequency range

PXIe-4190 (2 MHz)

40 Hz to 2 MHz

PXIe-4190 (500 kHz)

40 Hz to 500 kHz

Measurement time settings

Short

1 ms

Medium

10 ms

Long

100 ms

Custom

0 to 0.99999 s

Note Measurement times round up to the nearest positive integer number of cycles of the AC stimulus frequency.

Calculating Total LCR Measurement Time per Setpoint

Total Measurement Time per setpoint = LCR Source Delay + Total LCR Measurement Time

Calculating LCR Source Delay Time

  • LCR Source Delay Mode = Automatic
    • In Automatic mode, the source delay is 20 cycles of the AC stimulus frequency with a minimum source delay of 1 ms.
    • LCR Source Delay=Maximum(20×1f,1 ms)
  • LCR Source Delay Mode = Manual
    • LCR Source Delay time is as specified for the Source Delay property.
    Note Using a source delay smaller than the default value may not allow the output to sufficiently settle, resulting in measurement inaccuracy.
  • Setpoint changes that result in a range change add an additional 600 μs of source delay in either mode.

Calculating Total LCR Measurement Time

  • Total LCR Measurement Time=N×(CoercedMeasurementTime+10 μs)
    • N—Measurement count
    • CoercedMeasurementTime
      • The measurement time coerces to a full sinewave cycle boundary regardless of mode.
        x=Ceiling function
      • CoercedMeasurementTime=LCR Measurement Time×ff

        Where LCR Measurement Time = Short (1 ms)/Medium (10 ms))/Long (100 ms)/Custom

        Note LCR Custom Measurement Time = 0 is a special case that gives 1 cycle for any frequency.

LCR Measurements

  • Z—Impedance
  • Y—Admittance
  • Ls—Inductance using series-equivalent circuit model
  • Cs—Capacitance using series-equivalent circuit model
  • Rs—Resistance using series-equivalent circuit model
  • Lp—Inductance using parallel-equivalent circuit model
  • Cp—Capacitance using parallel-equivalent circuit model
  • Rp—Resistance using parallel-equivalent circuit model
  • D—Dissipation factor
  • Q—Quality factor
  • V DC—DC voltage measurement
  • I DC—DC current measurement
  • AC voltage—AC voltage magnitude and phase angle
  • AC current—AC current magnitude and phase angle
  • LCR AC Stimulus

    Voltage stimulus
    Maximum 7.07 V RMS
    Minimum 7.07 mV RMS
    Resolution <1 µV RMS
    Maximum current 70.7 mA RMS
    Accuracy (ALC on)
    ≤10 kHz ±0.4%
    >10 kHz ±4%
    Current stimulus
    Maximum 70.7 mA RMS
    Minimum 707 nA RMS
    Resolution <100 pA RMS
    Maximum voltage 7.07 V RMS
    Accuracy (ALC on)
    ≤10 kHz ±0.5%
    >10 kHz ±6%

    LCR DC Bias

    Voltage DC bias - PXIe-4190 (2 MHz)

    Maximum

    ±40 V, including peak AC stimulus

    Resolution

    <10 µV

    Accuracy

    0.02% + 5 mV

    Voltage DC bias - PXIe-4190 (500 kHz)

    Maximum

    ±10 V, including peak AC stimulus

    Resolution

    <10 µV

    Accuracy

    0.02% + 5 mV

    Current DC bias

    Maximum

    ±100 mA, including peak AC stimulus

    Resolution

    <10 nA

    Accuracy

    0.04% + 10 µA

    LCR Frequency

    Accuracy

    Equal to PXIe_CLK100 accuracy, nominal

    Frequency resolution

    1 mHz

    LCR Measurement Accuracy

    This topic shows the illustrated LCR measurement accuracy for capacitive DUTs.

    The following figure shows capacitor impedance magnitude versus test frequency to help quickly identify the appropriate impedance range for your measurements. Additionally, several important DUT test points across frequency are highlighted, with the corresponding Absolute Measurement Accuracy and AC Stimulus range shown in Table 8.

    Complete absolute accuracy specifications are described beginning in Table 10.

    Figure 5. Capacitor Impedance versus Frequency, with Representative Test Points Identified

    1378

    Table 8. Specifications for Representative DUT Test Points
    Test Point Capacitor Value AC Stimulus Frequency AC Stimulus Level ZC at AC Stimulus Frequency Measurement Accuracy
    Magnitude (Capacitance) Phase (Dissipation Factor)
    1 1 pF 10 kHz 708 mV RMS to 7.07 V RMS 15.9 MΩ 0.15% (1.5 fF) 0.08° (0.0014)
    2 1 pF 100 kHz 708 mV RMS to 7.07 V RMS 1.59 MΩ 0.30% (3 fF) 0.19° (0.0033)
    3 1 pF 2 MHz 708 mV RMS to 5 V RMS 79.6 kΩ 0.60% (6 fF) 0.26° (0.0045)
    4 100 pF 1 kHz 708 mV RMS to 7.07 V RMS 1.59 MΩ 0.06% (60 fF) 0.03° (0.0005)
    5 1 nF 2 MHz 150 mV RMS to 707 mV RMS 79.6 Ω 0.50% (50 pF) 0.18° (0.0031)
    6 100 nF 1 kHz 150 mV RMS to 707 mV RMS 1.59 kΩ 0.05% (50 pF) 0.02° (0.00035)
    7 1 µF 10 kHz 150 mV RMS to 707 mV RMS 15.9 Ω 0.08% (800 pF) 0.22° (0.0038)
    8 100 µF 120 Hz 50 mV RMS to 150 mV RMS 13.3 Ω 0.08% (80 nF) 0.04° (0.0007)
    Note Equations to solve for capacitor impedance, inductor impedance, and dissipation factor are shown in Example 1.
    Table 9. Calculated Accuracy for Capacitive DUTs (Cp, Cs) for Common MLCC AC Stimulus Frequencies from Absolute Impedance Magnitude Accuracy and Absolute Impedance Phase Accuracy Tables
    Capacitor Value AC Stimulus Voltage AC Stimulus Frequency Capacitance Magnitude Accuracy Phase Accuracy Df Accuracy
    13.3 pF < C ≤ 132.6 pF 1.0 V RMS 120 Hz ±0.15% ±0.08° ±0.001396
    132.6 pF < C ≤ 1.3 nF 1.0 V RMS 120 Hz ±0.06% ±0.03° ±0.000524
    1.3 nF < C ≤ 13.3 nF 1.0 V RMS 120 Hz ±0.06% ±0.02° ±0.000349
    13.3 nF < C ≤ 132.6 nF 1.0 V RMS 120 Hz ±0.05% ±0.02° ±0.000349
    132.6 nF < C ≤ 1.3 µF 1.0 V RMS 120 Hz ±0.06% ±0.02° ±0.000349
    1.3 µF < C ≤ 93.8 µF 1.0 V RMS 120 Hz ±0.08% ±0.03° ±0.000524
    93.8 µF < C ≤ 132.6 µF 0.5 V RMS 120 Hz ±0.08% ±0.03° ±0.000524
    1.6 pF < C ≤ 15.9 pF 1.0 V RMS 1 kHz ±0.15% ±0.08° ±0.001396
    15.9 pF < C ≤ 159.2 pF 1.0 V RMS 1 kHz ±0.06% ±0.03° ±0.000524
    159.2 pF < C ≤ 1.6 nF 1.0 V RMS 1 kHz ±0.06% ±0.02° ±0.000349
    1.6 nF < C ≤ 15.9 nF 1.0 V RMS 1 kHz ±0.05% ±0.02° ±0.000349
    15.9 nF < C ≤ 159.2 nF 1.0 V RMS 1 kHz ±0.06% ±0.02° ±0.000349
    159.2 nF < C ≤ 11.3 µF 1.0 V RMS 1 kHz ±0.08% ±0.03° ±0.000524
    11.3 µF < C ≤ 15.9 µF 0.5 V RMS 1 kHz ±0.08% ±0.03° ±0.000524
    624 fF < C ≤ 1.6 pF 1.0 V RMS 1 MHz ±0.30% ±0.16° ±0.002793
    1.6 pF < C ≤ 15.9 pF 1.0 V RMS 1 MHz ±0.30% ±0.13° ±0.002269
    15.9 pF < C ≤ 159.2 pF 1.0 V RMS 1 MHz ±0.20% ±0.12° ±0.002094
    159.2 pF < C ≤ 530.5 pF 1.0 V RMS 1 MHz ±0.20% ±0.12° ±0.002094
    530.5 pF < C ≤ 11.3 nF 1.0 V RMS 1 MHz ±0.20% ±0.11° ±0.001920
    11.3 nF < C ≤ 15.9 nF 0.5 V RMS 1 MHz ±0.20% ±0.13° ±0.002269

    LCR Magnitude and Phase Accuracy

    Table 10. Absolute Impedance Magnitude Accuracy, 708 mV RMS to 7.07 V RMS AC Stimulus Voltage
    Impedance Range AC Stimulus Frequency
    40 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz* 1 MHz to 2 MHz*
    100 MΩ to 1 GΩ 1.00%, typical 1.00%
    10 MΩ to 100 MΩ 0.15%, typical 0.15% 0.15%
    1 MΩ to 10 MΩ 0.06%, typical 0.06% 0.15% 0.30%
    100 kΩ to 1 MΩ 0.05% 0.06% 0.08% 0.30% 0.30% 0.30% 0.60%§
    10 kΩ to 100 kΩ 0.05% 0.05% 0.08% 0.30% 0.30% 0.30% 0.60%
    1 kΩ to 10 kΩ 0.05% 0.06% 0.08% 0.20% 0.20% 0.20% 0.50%
    300 Ω to 1 kΩ 0.08% 0.08% 0.08% 0.15% 0.15% 0.20% 0.50%
    10 Ω to 300 Ω 0.08% 0.08% 0.20% 0.20% 0.20% 0.20% 0.50%

    * PXIe-4190 (2 MHz) only

    Up to 640 kΩ impedance range.

    Up to 255 kΩ impedance range.

    § Up to 130 kΩ impedance range.

    Note: Impedances <10 Ω require a reduced AC stimulus. Refer to the following table for more information.

    Note: When on boundary, use lower adjacent value.

    Note: Add the following derating factor to LCR magnitude when AC stimulus level is >5 V RMS and >1 MHz:

    Additional magnitude error(%)=(f1 MHz)2×(Vstim5 V)2×0.025%

    Table 11. Absolute Impedance Magnitude Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage
    Impedance Range AC Stimulus Frequency
    40 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz* 1 MHz to 2 MHz*
    10 MΩ to 100 MΩ 0.20%, typical 0.40% 1.10%
    1 MΩ to 10 MΩ 0.06%, typical 0.06% 0.20% 0.90%
    100 kΩ to 1 MΩ 0.05% 0.06% 0.08% 0.90% 0.60% 0.60% 0.60%§
    10 kΩ to 100 kΩ 0.05% 0.05% 0.08% 0.30% 0.30% 0.30% 0.50%
    1 kΩ to 10 kΩ 0.05% 0.05% 0.08% 0.20% 0.20% 0.20% 0.50%
    300 Ω to 1 kΩ 0.08% 0.08% 0.08% 0.15% 0.15% 0.20% 0.50%
    10 Ω to 300 Ω 0.08% 0.08% 0.08% 0.20% 0.20% 0.20% 0.50%
    <10 Ω** 0.08% + 1 mΩ 0.08% + 1 mΩ 0.08% + 1 mΩ 0.90% + 1mΩ 0.90% + 1 mΩ 1.20% + 1 mΩ†† 2.00% + 2 mΩ††

    * PXIe-4190 (2 MHz) only

    Up to 640 kΩ impedance range.

    Up to 255 kΩ impedance range.

    § Up to 130 kΩ impedance range.

    ** Typical, offset relative to short compensation.

    †† Refer to AC Stimulus Current Short Offset Multiplier table for offset multiplier.

    Note: When on boundary, use lower adjacent value.

    Table 12. AC Stimulus Current Short Offset Multiplier
    AC Stimulus Current Short Offset Multiplier
    <7.07 mA 5
    7.08 mA to 20 mA 1
    >20 mA 2

    Table 13. Absolute Impedance Magnitude Accuracy Multiplier for AC Stimuli Below 150 mV RMS
    Impedance Range AC Stimulus Voltage
    50 mV RMS to 150 mV RMS 7.08 mV RMS to 50 mV RMS, typical
    <10 Ω 1 1
    10 Ω to 300 Ω 2 5
    300 Ω to 10 MΩ 2 11
    10 MΩ to 100 MΩ 3

    Note: Absolute accuracy is the Absolute Impedance Magnitude Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage table value times the respective multiplier.

    Table 14. Absolute Impedance Phase Accuracy, 708 mV RMS to 7.07 V RMS AC Stimulus Voltage
    Impedance Range AC Stimulus Frequency
    40 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz to 10 kHz >10 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz* 1 MHz to 2 MHz*
    100 MΩ to 1 GΩ 0.55 °, typical 0.55 °
    10 MΩ to 100 MΩ 0.19 °, typical 0.08 ° 0.25 °
    1 MΩ to 10 MΩ 0.02 °, typical 0.03 ° 0.21 ° 0.19 °
    100 kΩ to 1 MΩ 0.01 ° 0.02 ° 0.19 ° 0.19 ° 0.14 ° 0.16 ° 0.26 °§
    10 kΩ to 100 kΩ 0.01 ° 0.02 ° 0.10 ° 0.11 ° 0.12 ° 0.13 ° 0.26 °
    1 kΩ to 10 kΩ 0.01 ° 0.02 ° 0.09 ° 0.10 ° 0.10 ° 0.12 ° 0.31 °
    300 Ω to 1 kΩ 0.01 ° 0.03 ° 0.12 ° 0.08 ° 0.13 ° 0.12 ° 0.34 °
    10 Ω to 300 Ω 0.01 ° 0.03 ° 0.13 ° 0.08 ° 0.09 ° 0.11 ° 0.15 °

    * PXIe-4190 (2 MHz) only

    Up to 640 kΩ impedance range.

    Up to 255 kΩ impedance range.

    § Up to 130 kΩ impedance range.

    Note: Impedances <10 Ω require a reduced AC stimulus. Refer to the following table for more information.

    Note: When on boundary, use lower adjacent value.

    Table 15. Absolute Impedance Phase Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage
    Impedance Range AC Stimulus Frequency
    40 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz to 10 kHz >10 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz* 1 MHz to 2 MHz*
    10 MΩ to 100 MΩ 0.14 °, typical 0.30 ° 0.50 °
    1 MΩ to 10 MΩ 0.03 °, typical 0.03 ° 0.14 ° 0.45 °
    100 kΩ to 1 MΩ 0.02 ° 0.03 ° 0.14 ° 0.45 ° 0.22 ° 0.22 ° 0.34 °§
    10 kΩ to 100 kΩ 0.01 ° 0.02 ° 0.07 ° 0.15 ° 0.14 ° 0.14 ° 0.34 °
    1 kΩ to 10 kΩ 0.01 ° 0.02 ° 0.07 ° 0.15 ° 0.09 ° 0.11 ° 0.20 °
    300 Ω to 1 kΩ 0.01 ° 0.02 ° 0.07 ° 0.08 ° 0.09 ° 0.12 ° 0.34 °
    10 Ω to 300 Ω 0.01 ° 0.04 ° 0.22 ° 0.08 ° 0.10 ° 0.13 ° 0.18 °
    <10 Ω, typical 0.01 ° 0.04 ° 0.08 ° 0.03 ° 0.07 ° 0.15 ° 0.20 °

    * PXIe-4190 (2 MHz) only

    Up to 640 kΩ impedance range.

    Up to 255 kΩ impedance range.

    § Up to 130 kΩ impedance range.

    Note: When on boundary, use lower adjacent value.

    Table 16. Absolute Impedance Phase Accuracy Multiplier for AC Stimuli Below 150 mV RMS
    Impedance Range AC Stimulus Voltage
    50 mV RMS to 150 mV RMS 7.08 mV RMS to 50 mV RMS, typical
    <10 Ω 1 4
    10 Ω to 300 Ω 2 20
    300 Ω to 1 kΩ 2 70
    1 kΩ to 10 kΩ 2 25
    >10 kΩ to 100 kΩ 2 25
    100 kΩ to 1 MΩ 2 10
    1 MΩ to 10 MΩ 2 8
    10 MΩ to 100 MΩ 3 8

    Note: Absolute accuracy is the Absolute Impedance Phase Accuracy, 150 mV RMS to 707 mV RMS AC Stimulus Voltage table value times the respective multiplier.

    LCR Noise

    Figure 6. Impedance Magnitude Measurement Noise versus Measurement Time, 1 kΩ, Nominal

    1378

    Figure 7. Impedance Phase Measurement Noise versus Measurement Time, 1 kΩ, Nominal

    1378

    Table 17. Specification Derating for Short and Medium Measurement Time, Typical

    Use the following table to multiply the respective magnitude and phase accuracy specification value by the derating factor for the applicable measurement time.

    Measurement Time Derating Factor
    Medium
    Maximum(1, log(|Z|VacStimulus×106))
    Short
    Maximum(1.5, log(|Z|VacStimulus×5×104))
    Note: Measurement time derating is a function of impedance magnitude and AC stimulus voltage level, and is independent of frequency. Specifications are determined by comparing differences in the standard deviation for different measurement times.
    Figure 8. Capacitance versus Frequency, 1 pF, 1 V RMS AC Stimulus, Long Measurement Time, Measured Data for 20 Modules

    1378

    Figure 9. Dissipation Factor versus Frequency, 1 pF, 1 V RMS AC Stimulus, Long Measurement Time, Measured Data for 20 Modules

    1378

    Figure 10. Capacitance versus Frequency, 100 pF, 1 V RMS AC Stimulus, Long Measurement Time, Measured Data for 20 Modules

    1378

    Figure 11. Dissipation Factor versus Frequency, 100 pF, 1 V RMS AC Stimulus, Long Measurement Time, Measured Data for 20 Modules

    1378

    LCR Accuracy Derating with DC Bias

    Above 500 kHz with DC Bias enabled, add the additional error term to the stated magnitude accuracy specification:

    Additional magnitude error(%)=f500 kHz×|VDCBias|×0.0015%, typical

    Above 500 kHz with DC Bias enabled, add the additional error term to the stated phase accuracy specification:

    Additional phase error(°)=f500 kHz×|VDCBias|×0.0005°, typical

    LCR DC Bias Settling Time for Large Capacitors

    Set DC Bias ALC to Off when measuring capacitors over 1 uF to minimize settling time.

    Figure 12. DC Bias Settling Time (40 V Step Settled to 10 mV)

    1378

    Table 18. DC Bias Settling Time Required in Addition to LCR Source Delay for Large Capacitors
    Bias Settling Time
    AC Stimulus Voltage / Impedance Range ≤ 7.07 mA Add 3 ms per µF of DUT capacitance
    AC Stimulus Voltage / Impedance Range > 7.07 mA Add 600 µs per µF of DUT capacitance

    Note: When applying a bias voltage to capacitors over 1 mF, the bias voltage steps should be no larger than 40 V x 1 mF/C to avoid tripping overcurrent protection.

    Note: For AC Stimulus Voltage / Impedance Range ≤7.07mA, limit DC bias steps to ≤5 V each, up to 40 V total.

    LCR Cable Accuracy Derating

    Table 19. Cable Accuracy Derating
    Cable Description NI Part Number
    0.5 m 1 m 2 m* 4 m*
    SHDB13W6-4BNCM-LL DSUB to Male BNC 788280-01 788280-02 788280-04
    SHDB13W6-4BNCF-LL DSUB to Female BNC 789536-0R5 789536-01 789536-02
    SHDB13W6-DB13W6-LL DSUB to DSUB 788279-01 788279-02 788279-04
    SHDB13W6-4TriaxM-LL DSUB to Male Triaxial 788281-01 788281-02 788281-04
    * LCR specifications are typical
    For cable lengths >1 m, LCR measurement magnitude specifications are typical with the following additional derating. Add the following term to the absolute accuracy, where
  • L—is the cable length in meters
  • Additional magnitude error (%)=f×L8×106
    For cable lengths >1 m, LCR measurement phase specifications are typical with the following additional derating. Refer to the following table and add the term that corresponds to your measurement frequency and AC stimulus amplitude to the absolute accuracy, where
  • L—is the cable length in meters
  • |Z|—is impedance magnitude
  • Table 20. Additional Phase Error (°)
    ≤707 mV RMS >707 mV RMS
    ≤10 kHz
    L×|Z|5×107
    L×|Z|5×108
    >10 kHz
    L×|Z|1×108
    L×|Z|1×109

    Determining LCR Measurement Impedance Range

    The impedance range can be calculated and programmed in several ways. The following methods allow you to set the impedance range directly.

    Calculating Impedance Range Manually

    Use the following formulas to determine the expected impedance based on the load.

    The impedance of an ideal capacitor is

    |ZC|=12πfC=0.159fC
    where

  • ZC—Capacitor impedance (Ω)
  • f—AC stimulus frequency (Hz)
  • C—Nominal capacitance value (F)
  • The impedance of an ideal inductor is

    |ZL|=2πfL=6.283×f×L

  • ZL—Inductor impedance (Ω)
  • f—AC stimulus frequency (Hz)
  • L—Nominal inductance value (H)
  • Setting LCR Impedance Range Source Programmatically

    By setting LCR Impedance Range Source to LCR Load Configuration, the range can be determined automatically based on the AC stimulus frequency, and the load settings LCR Load Resistance, LCR Load Inductance, and LCR Load Capacitance.

    Note The PXIe-4190 LCR impedance ranges do not directly correspond to the underlying hardware ranges. When a measurement is configured, NI-DCPower will determine the best hardware range based on the requested impedance range, frequency, AC stimulus level, and bias settings. To determine the active hardware ranges for the configured measurement—or to set them manually—NI-DCPower provides these settings:
    • LCR Voltage Range
    • LCR Current Range
    • LCR DC Bias Voltage Range
    • LCR DC Bias Current Range

    The LCR Voltage Range and LCR Current Range are expressed as RMS values but are equivalent to the corresponding SMU mode ranges when converted to peak value.

    Translating LCR Specifications to Other Impedance Parameters

    Accuracy for additional impedance parameters can be derived from the absolute impedance magnitude and phase specifications. For some calculations, the actual DUT impedance must also be known—in these cases, the measured value can be used as an approximation with typically negligible impact on the result.

  • xSpec—Specified accuracy for a parameter x (for example, Δ|Z|Spec is the magnitude accuracy specification)
  • xDUT—The actual value of parameter x for a DUT
  • |Z|—Impedance magnitude. Δ|Z|Spec corresponds to the numbers listed in the magnitude accuracy tables (in percent).
  • θ—Impedance phase angle. ΔθSpec corresponds to the numbers listed in the phase accuracy tables (in degrees).
  • δ—Phase angle between the impedance vector and the reactive axis.
    ΔδSpec=ΔθSpec(°)
  • D—Dissipation factor.
    D=tan(δ)=R|X|

    ΔDSpec=±tan(ΔθSpec)
    (when DDUT < 0.1).

  • Q—Quality factor,
    Q=1D

    To determine the specified range of possible Q values, calculate

    1DDUT±ΔDSpec
    (when DDUT < 0.1).

  • C—Capacitance, series (CS) or parallel (CP) equivalent model.

    When DDUT is sufficiently small (<0.1), ΔCSpecΔ|Z|Spec (%).

    For a general solution, determine accuracy using the AC stimulus frequency and reactance specification:

    ΔCSpec=±12πf×ΔXSpec(F)

  • L—Inductance, series (LS) or parallel (LP) equivalent model.
  • When DDUT is sufficiently small (<0.1), ΔLSpecΔ|Z|Spec (%).

    For a general solution, determine accuracy using the AC stimulus frequency and reactance specification:

    ΔLSpec=±(2πf×ΔXSpec)(H)

  • R—Resistance, the real component of complex impedance.

    For typical non-reactive resistance measurements (DDUT > 10), ΔRSpecΔ|Z|Spec (%).

    To determine accuracy for an arbitrary impedance, first find the maximum and minimum values, RMax and RMin, from four calculations:

    (|Z|DUT±Δ|Z|Spec)×cos(θDUT±ΔθSpec),
    then
    ΔRSpec=±RMaxRMin2(Ω)

  • X—Reactance, the imaginary component of complex impedance.

    For typical L/C measurements (DDUT < 0.1), ΔXSpecΔ|Z|Spec (%).

    To determine accuracy for an arbitrary impedance, first find the maximum and minimum values, RMax and RMin, from four calculations:

    (|Z|DUT±Δ|Z|Spec)×sin(θDUT±ΔθSpec),
    then
    ΔXSpec=±XMaxXMin2(Ω)

  • The following figure shows the relationship between these parameters when an example vector Z is plotted on the complex impedance plane.

    Figure 13. Impedance Specification Representation on a Complex Impedance Plane

    1378

    Note When computing tan(PhaseInDegrees) using tan(Radians), note that
    Degrees×π180=Radians

    Example 1: Calculating Specifications for Capacitance Measurement

    For a capacitor measurement under the stated conditions, complete the following steps to determine and interpret the absolute measurement accuracy.

    DUT Actual Capacitance (CDUT)

    10 pF

    DUT Actual Dissipation (DDUT)

    0.005

    AC Stimulus Frequency (f)

    1 MHz

    AC Stimulus Voltage (Vstim)

    1 V RMS

    DC Bias Voltage (V DC)

    10 V

    Measurement time

    Short

    Cable length

    1 m

    1. Calculate ideal capacitor impedance as
      ZDUT=12πfC=12×π×1 MHz×10 pF=15.915kΩ
    2. Based on the 1 V RMS AC stimulus, the applicable magnitude and phase specs are found in Table 10 and Table 14, respectively.
      • From the calculated impedance, ZDUT, the relevant impedance range is 10 kΩ100 kΩ.
      • 1 MHz is on the boundary between the 500 kHz to 1 MHz and 1 MHz to 2 MHz frequency ranges, so choose the smaller of the adjacent values.
      • The resulting specifications are 0.3% magnitude accuracy and 0.13° phase accuracy.
    3. Base specifications apply to long measurement time. For short measurement time, apply the derating factor:
      Max(1.5, log(|Z|Vstim×5×104))=Max(1.5, log(15.915 kΩ1 V×5×104))=Max(1.5, 0.5)=1.5
      • The derated magnitude specification is then 1.5 * 0.3% = 0.45%
      • The derated phase specification is then 1.5 * 0.13° = 0.195°
    4. Because DC bias is enabled and f >500 kHz, the additional error terms from LCR Accuracy Derating with DC Bias apply:
      • The additional magnitude error is calculated as
        f×VDC×0.0015%500 kHz=1 MHz×10 V×0.0015%500 kHz=0.03%
      • The additional phase error is calculated as
        f×VDC×0.005°500 kHz=1 MHz×10 V×0.0005°500 kHz=0.01°
    5. From the previous steps, the final accuracy specifications under these measurement conditions:
      • Magnitude accuracy, ΔZSpec = 0.45% + 0.03% = 0.48%
      • Phase accuracy, ΔθSpec = 0.195° + 0.01° = 0.205°

    These specifications can then be used to derive accuracies for other parameters.

    • Dissipation factor accuracy, ΔDSpec = ±tan(ΔθSpec) = ±0.0036
      • Specified range is DDUT ± ΔDSpec = 0.0013 to 0.0087
    • Quality factor specified range is
      1DDUT±ΔDSpec=115 to 769
    • Impedance phase has an accuracy of ΔθSpec = 0.205° and can be expressed as
      • Loss angle, δDUT = arctan(DDUT) = arctan(0.005) = 0.286° ± 0.205°
      • Impedance phase angle, θDUT = δ – 90° = 0.286° - 90° = -89.714° ± 0.205°
    • Resistance accuracy, ΔRSpec, can be calculated by
      • Finding the maximum and minimum values, RMax and RMin, from four calculations:
        (ZDUT±ΔZSpec)×cos(θDUT±ΔθSpec)
        =(15.915 kΩ±0.48%)×cos(-89.714°±0.205°)
        =[15.991 kΩ×cos(-89.919°),15.991 kΩ×cos(-89.509°),15.839 kΩ×cos(-89.919°),15.839 kΩ×cos(-89.509°)]
        =[22.6Ω,137Ω,22.4Ω,135.7Ω]
      • Selecting the maximum and minimum values, RMax = 137 Ω, RMin = 22.4 Ω
      • ΔRSpec=RMaxRMin2=137Ω22.4Ω2=±57.3Ω
    • Reactance accuracy, ΔXSpec, can be calculated by
      • Since DDUT is small, ΔXSpec ≅ Δ|Z|Spec
      • Using this simplified approximation, ΔXSpec = 0.48% * 15.915 kΩ = ± 76.4 Ω
      • For example, to compare the explicitly calculated specification, first find the maximum and minimum values, XMax and XMin, from the four calculations:
        (ZDUT±ΔZSpec)×sin(θDUT±ΔθSpec)
        =(15.915 kΩ±0.48%)×sin(-89.714°±0.205°)
        =[15.991 kΩ×sin(-89.919°),15.991 kΩ×sin(-89.509°),15.839 kΩ×sin(-89.919°),15.839 kΩ×sin(-89.509°)]
        =[-15.991 kΩ,-15.990 kΩ,-15.839 kΩ,-15.838 kΩ]
      • Selecting the maximum and minimum values, XMax = -15.838 kΩ, XMin = -15.991 kΩ
      • ΔXSpec=XMaxXMin2=-15.838 kΩ-15.991 kΩ2=±76.5Ω

    General Specifications

    Isolation

    Isolation voltage, any pin to earth ground

    40 V DC, Measurement Category I, functional

    Note Pins are functionally isolated from chassis ground to prevent ground loops, but do not meet IEC 61010-1 for safety isolation.
    Note The PXIe-4190 contains an internal switch controlled by the niDCPower Isolation State property that can connect the GUARD terminal to chassis ground and prevent the module output from floating. Isolation ratings only apply when this property/attribute is set to Isolated.

    Protection

    Absolute maximum voltage

    Output HI/Output LO/Sense HI/Sense LO to Output HI/Output LO/Sense HI/Sense LO

    ±42 V

    Output HI/Sense HI to GUARD/Isolated Shield

    ± 6 V

    GUARD/Isolated Shield to Chassis GND

    ±42 V

    Absolute maximum current

    All terminals

    ±300 mA

    Output channel protection
    Output HI to GUARD/Isolated Shield
    Overvoltage Automatic output disable
    Output LO to all terminals
    Overcurrent Automatic output disable
    Sense HI/Sense LO to all terminals
    Overcurrent Current limiter protects inputs up to absolute maximum voltage specification
    Overtemperature Automatic output disable

    Physical Characteristics

    Dimensions

    3U, one-slot, PXI Express/CompactPCI Express module

    2.0 cm x 13.0 cm x 21.6 cm (0.8 in. x 5.1 in. x 8.5 in.)

    Weight

    481 g (17.1 oz)

    Front panel connectors

    Mixed layout DSUB, 13W6 contact arrangement (6 coaxial 50 Ω, 7-signal), female

    Triggers

    Input triggers
    Types Start, Source, Sequence Advance, Measure
    Sources (PXI trigger lines 0 to 7)
    Polarity Active high (not configurable)
    Minimum pulse width 100 ns
    Destinations[8]8 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source. (PXI trigger lines 0 to 7)
    Polarity Active high (not configurable)
    Minimum pulse width 200 ns
    Output triggers (events)
    Types Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete
    Destinations (PXI trigger lines 0 to 7)[9]9 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.
    Polarity Active high (not configurable)
    Pulse width 230 ns

    Calibration Interval

    Recommended calibration interval

    1 year

    Power Requirements

    +3.3 V

    1.0 A

    +12 V

    2.7 A

    Environmental Characteristics

    Temperature

    Operating

    0 °C to 55 °C [10]10 Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.

    Storage

    -40 °C to 71 °C

    Pollution Degree 2
    Maximum altitude 2,000 m (800 mbar) (at 25 °C ambient temperature)
    Humidity
    Operating 10% RH to 90% RH, noncondensing [11]11 Accuracy specifications are only warranted for operating environments with temperatures below 30 °C and relative humidity levels below 60%. When transitioning the product from a storage or operating environment with relative humidity above 60%, you should allow the product to stabilize in the lower humidity environment for several hours before using it.
    Storage 5% RH to 95% RH, noncondensing
    Shock and Vibration

    Operating vibration

    5 Hz to 500 Hz, 0.3 g RMS

    Non-operating vibration

    5 Hz to 500 Hz, 2.4 g RMS

    Operating shock

    30 g, half-sine, 11 ms pulse

    1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).

    2 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.

    3 10 Hz to 20 MHz bandwidth, PXIe-4190 configured for normal transient response.

    4 When source-measuring, both the NI-DCPower Source Delay and Aperture Time properties affect the sampling rate. When taking a measure record, only the Aperture Time property affects the sampling rate.

    5 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).

    6 For the PXIe-4190 revision D and earlier—niDCPower Output Enabled or niDCPower Output Connected properties must be set to FALSE when making connections between force and sense leads. Disconnecting the sense leads while both these properties are set to TRUE may result in output protection errors or long settling tails due to the feedback path for the control loop being open. If the PXIe-4190 is run open loop due to accidental sense lead disconnection, allow a minimum of 1 minute after establishing proper lead connections before making measurements.

    7 Refer to the PXIe-4190 Getting Started for more information on 4TP connections.

    8 Input triggers can come from any source (PXI trigger or software trigger) and be exported to any PXI trigger line. This allows for easier multi-board synchronization regardless of the trigger source.

    9 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.

    10 Not all chassis can achieve this ambient temperature range. Refer to PXI chassis specifications to determine the ambient temperature ranges your chassis can achieve.

    11 Accuracy specifications are only warranted for operating environments with temperatures below 30 °C and relative humidity levels below 60%. When transitioning the product from a storage or operating environment with relative humidity above 60%, you should allow the product to stabilize in the lower humidity environment for several hours before using it.