PXIe-4137 Specifications

These specifications apply to the PXIe-4137.

Note In this document, the PXIe-4137 (40W) and PXIe-4137 (20W) are referred to inclusively as the PXIe-4137. The information in this document applies to all versions of the PXIe-4137 unless otherwise specified. To determine which version of the module you have, locate the device name in one of the following places:
  • In MAX—The PXIe-4137 (40W) shows NI PXIe-4137 (40W), and the PXIe-4137 (20W) shows as NI PXIe-4137.
  • Device front panel—The PXIe-4137 (40W) shows PXIe-4137 40W System SMU, and the PXIe-4137 (20W) shows NI PXIe-4137 Precision System SMU on the front panel.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured specifications describe the measured performance of a representative model.

Specifications are Warranted unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature[1]1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). of 23 °C ± 5 ºC
  • Chassis with slot cooling capacity ≥38 W[2]2 For increased capability, NI recommends installing the PXIe-4137 (40W) in a chassis with slot cooling capacity ≥58 W.
    • For chassis with slot cooling capacity = 38 W, fan speed set to HIGH
  • Calibration interval of 1 year
  • 30 minutes warm-up time
  • Self-calibration performed within the last 24 hours
  • NI-DCPower Aperture Time is set to 2 power-line cycles (PLC)

Cleaning Statement

Notice Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service.

Device Capabilities

The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4137.

Table 1. Current Source and Sink Ranges
DC voltage ranges DC current source and sink ranges

600 mV

6 V

20 V

200 V[3]3 Voltage levels and limits >|40 VDC| require the safety interlock input to be closed.

1 μA

10 μA

100 μA

1 mA

10 mA

100 mA

1 A

3 A [4]4 Current is limited to 1 A DC. Higher levels are pulsing only.

Figure 1. Quadrant Diagram for PXIe-4137 (40W)

40W Quadrant Diagram

For additional information related to the Pulse Voltage or Pulse Current settings of the Output Function, for the PXIe-4137 (40W), including pulse on time and duty cycle limits for a particular operating point, refer to Pulsed Operation. For supplementary examples, refer to Examples of Determining Extended Range Pulse Parameters and Optimizing Slew Rate using NI SourceAdapt.

Figure 1. Quadrant Diagram for PXIe-4137 (20W)

20W Quadrant Diagram

DC sourcing power and sinking power are limited to the values in the following table, regardless of output voltage. [5]5 Power limit defined by voltage measured between HI and LO terminals.

Table 2. DC Sourcing & Sinking Power
Model Variant Chassis Type DC Sourcing Power DC Sinking Power
PXIe-4137 (40W) ≥58 W Slot Cooling Capacity 40 W 40 W
<58 W Slot Cooling Capacity 20 W 12 W
PXIe-4137 (20W) ≥58 W Slot Cooling Capacity 20 W 12 W
<58 W Slot Cooling Capacity 20 W 12 W
Caution Limit DC power sinking to 12 W where applicable as indicated in the above table. For <58 W cooling slots,
  • Additional derating applies to sinking power when operating at an ambient temperature of >45 °C.
  • If the PXI Express chassis has multiple fan speed settings, set the fans to the highest setting.

Voltage

Table 3. Voltage Programming and Measurement Accuracy/Resolution
Range Resolution (noise limited) Noise (0.1 Hz to 10 Hz, peak to peak), Typical Accuracy (23 °C ±5 °C) ± (% of voltage + offset)[6]6 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions. Tempco ± (% of voltage + offset)/°C, 0 °C to 55 °C
Tcal ±5 °C [7] 7 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration. Tcal ±1 °C [7]
600 mV 100 nV 2 μV 0.020% + 50 μV 0.017% + 30 μV 0.0005% + 1 μV
6 V 1 μV 6 μV 0.020% + 320 μV 0.017% + 90 μV
20 V 10 μV 20 μV 0.022% + 1 mV 0.017% + 400 μV
200 V 100 μV 200 μV 0.025% + 10 mV 0.020% + 2.5 mV

Current

Table 4. Current Programming and Measurement Accuracy/Resolution
Range Resolution (noise limited) Noise (0.1 Hz to 10 Hz, peak to peak), Typical Accuracy (23 °C ± 5 °C) ± (% of current + offset) Tempco ± (% of current + offset)/°C, 0 °C to 55 °C
Tcal ± 5 °C [8] 8 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration. Tcal ± 1 °C [8]
1 μA 100 fA 4 pA 0.03% + 100 pA 0.022% + 40 pA 0.0006% + 4 pA
10 μA 1 pA 30 pA 0.03% + 700 pA 0.022% + 300 pA 0.0006% + 22 pA
100 μA 10 pA 200 pA 0.03% + 6 nA 0.022% + 2 nA 0.0006% + 200 pA
1 mA 100 pA 2 nA 0.03% + 60 nA 0.022% + 20 nA 0.0006% + 2 nA
10 mA 1 nA 20 nA 0.03% + 600 nA 0.022% + 200 nA 0.0006% + 20 nA
100 mA 10 nA 200 nA 0.03% + 6 μA 0.022% + 2 μA 0.0006% + 200 nA
1 A 100 nA 2 μA 0.04% + 60 μA 0.035% + 20 μA 0.0006% + 2 μA
3 A [9]9 3 A range above 1 A is for pulsing only. 1 μA 20 μA 0.08% + 900 μA 0.075% + 600 μA 0.0018% + 20 μA

Noise

Wideband source noise

<20 mV peak-to-peak in 20 V range, device configured for normal transient response, 10 Hz to 20 MHz, typical

The following figures illustrate measurement noise as a function of measurement aperture for the PXIe-4137.
Figure 1. Voltage Measurement Noise vs. Measurement Aperture, Nominal


Note When the aperture time is set to 2 power-line cycles (PLCs), measurement noise differs slightly depending on whether the Power Line Frequency is set to 50 Hz or 60 Hz.
Figure 1. Current Measurement Noise vs. Measurement Aperture, Nominal


Note When the aperture time is set to 2 power-line cycles (PLCs), measurement noise differs slightly depending on whether the Power Line Frequency is set to 50 Hz or 60 Hz.

Sinking Power vs. Ambient Temperature Derating

The following figure illustrates sinking power derating as a function of ambient temperature.

This applies to the PXIe-4137 (20W) when used with any chassis and only applies to the PXIe-4137 (40W) when used with a chassis with slot cooling capacity <58 W.
Figure 1. Sinking Power vs. Ambient Temperature Derating


Note When using the PXIe-4137 (40W) with a chassis with slot cooling capacity ≥58 W, ambient temperature derating does not apply.

Output Resistance Programming Accuracy

Table 5. Output Resistance Programming Accuracy Characteristics
Current Level/Limit Range Programmable Resistance Range, Voltage Mode Programmable Resistance Range, Current Mode Accuracy ± (% of resistance setting), Tcal ± 5 °C [10]10 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.
1 μA 0 to ±5 MΩ ±5 MΩ to ±infinity 0.03%
10 μA 0 to ±500 kΩ ±500 kΩ to ±infinity
100 μA 0 to ±50 kΩ ±50 kΩ to ±infinity
1 mA 0 to ±5 kΩ ±5 kΩ to ±infinity
10 mA 0 to ±500 Ω ±500 Ω to ±infinity
100 mA 0 to ±50 Ω ±50 Ω to ±infinity
1 A 0 to ±5 Ω ±5 Ω to ±infinity
3 A [11]11 3 A range above 1 A is for pulsing only. 0 to ±500 mΩ ±500 mΩ to ±infinity

Overvoltage Protection

Accuracy[12]12 Overvoltage protection accuracy is valid with an ambient temperature of 23 °C ± 5 °C and with Tcal ±5 °C. Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration. (% of OVP limit + offset)

0.1% + 200 mV, typical

Temperature coefficient (% of OVP limit + offset)/°C

0.01% + 3 mV/°C, typical

Measurement location

Local sense

Maximum OVP limit value

210 V

Minimum OVP limit value

2 V

Pulsed Operation

Dynamic load, minimum pulse cycle time[13]13 For example, given a continuous pulsing load, if the largest dynamic step in current that the load sources/sinks is from 0.5 A to 1.0 A, then the maximum SMU current step is 0.5 A. Thus, the minimum dynamic load pulse cycle time is 50 μs. Minimum dynamic load pulse cycle time is independent of output voltage.[14]14 Measurable unit of μs/A is used because the minimum pulse cycle time is independent of output voltage

100 μs/A

The following figure visually explains the terms used in the extended range pulsing sections.

Figure 1. Definition of Pulsing Terminology


Extended Range Pulsing for PXIe-4137 (40W)

Note Extended range pulses fall outside DC range limits for either current or power. In-range pulses fall within DC range limits and are not subject to extended range pulsing limitations. Extended range pulsing is enabled by setting the Output Function to Pulse Voltage or Pulse Current.

The following figures illustrate the maximum pulse on time and duty cycle for the PXIe-4137 (40W) in a ≥58 W cooling slot, for a desired pulse voltage and pulse current given zero bias voltage and current. The shaded areas allow for a quick approximation of output limitations and limiting parameters. Actual limits are described by equations in Table 6.

Figure 1. Pulse On-time vs Pulse Current and Pulse Voltage


Note Equations to solve for maximum pulse on time, tonMax, are shown in Table 6. Additionally, Equation 8 solves for pulse on time, ton, in terms of maximum pulse energy in Example 1: Determining Extended Range Pulse On Time and Duty Cycle Parameters for the (40W).
Figure 1. Duty Cycle vs Pulse Current and Pulse Voltage


Note Equations to solve for maximum duty cycle, DMax, are shown in Table 6. Additionally, Equation 9 solves for pulse off time, toff, in terms of maximum pulse energy in Example 1: Determining Extended Range Pulse On Time and Duty Cycle Parameters for the (40W).
Bias level limits

Maximum voltage, Vbias

200 V

Maximum current, Ibias

1 A

Table 6. PXIe-4137 (40W) Pulse Level Limits
Specification Value Equation
Maximum voltage, VpulseMax 160 V
Maximum current, IpulseMax 3 A
Maximum on time, tonMax[15]15 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB. If Ipulse > 1 A and ≥58 W Slot Cooling Capacity Chassis Calculate using the equation or refer to Figure 7 to estimate the value.
tonMax=100 ms*2 A|Ipulse|1A
, where tonMax is167 s

(Equation 1)

If Ipulse > 1 A and <58 W Slot Cooling Capacity Chassis Calculate using the equation.
tonMax=10 ms*2 A|Ipulse|1A
, where tonMax is167 s

(Equation 2)

If Ipulse1 A tonMax = 167 s
Maximum pulse energy, EpulseMax[16]16 Refer to Figure 7 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis. 0.4 J

Epulse=|Vpulse*Ipulse*ton|
, where Epulse< EpulseMax

(Equation 3)

Maximum duty cycle, DMax[17]17 Refer to Figure 8 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis. If D≥100%, consider switching Output Function from Pulse mode to DC mode. If ≥58 W Slot Cooling Capacity Chassis Calculate using the equation or refer to Figure 8 to estimate the value.
DMax=(1.18 A)2|Ibias|2|Ipulse|2|Ibias|2*100%

(Equation 4)

If <58 W Slot Cooling Capacity Chassis Calculate using the equation.
DMax=(1 A)2|Ibias|2|Ipulse|2|Ibias|2*100%

(Equation 5)

Minimum pulse cycle time, tcycleMin 5 ms

tcycle=ton+toff
, where tcycle>tcycleMin

(Equation 6)

Maximum cycle average power, PCAMax[18]18 Refer to Figure 8 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis. ≥58 W Slot Cooling Capacity Chassis 20 W

PCA=|Vpulse*Ipulse*ton|+|Vbias*Ibias*toff|ton+toff
, where PCA<PCAMax

(Equation 7)

<58 W Slot Cooling Capacity Chassis 10 W
Note Software will not allow settings that violate these limiting equations and will generate an error.

Extended Range Pulsing for PXIe-4137 (20W)

Note Extended range pulses fall outside DC range limits for either current or power. In-range pulses fall within DC range limits and are not subject to extended range pulsing limitations. Extended range pulsing is enabled by configuring the Output Function to Pulse Voltage or Pulse Current.
Bias level limits

Maximum voltage

200 V

Maximum current

1 A

Pulse level limits

Maximum voltage

160 V

Maximum current

3 A

Maximum on time[19]19 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB.

1 ms

Minimum pulse cycle time

5 ms

Energy

0.2 J

Maximum cycle average power

10 W

Maximum duty cycle

5%

Transient Response and Settling Time

Transient response

<70 μs to recover within 0.1% of voltage range after a load current change from 10% to 90% of range, device configured for fast transient response, typical

Maximum slew rate[20]20 Optimize transient response, overshoot, and slew rate with NI SourceAdapt by adjusting the Transient Response. ,[21]21 To improve the slew rate, see Examples of Determining Extended Range Pulse Parameters and Optimizing Slew Rate using NI SourceAdapt.

0.5A/μs

Settling time[22]22 Measured as the time to settle to within 0.1% of step amplitude, device configured for fast transient response.

Voltage mode, 180 V step, unloaded[23]23 Current limit set to ≥60 μA and ≥60% of the selected current limit range.

<500 μs, typical

Voltage mode, 5 V step or smaller, unloaded[24]24 Current limit set to ≥20 μA and ≥20% of selected current limit range.

<70 μs, typical

Current mode, full-scale step, 3 A to 100 μA ranges25 Voltage limit set to ≥2 V, resistive load set to 1 V/selected current range.[25]

<50 μs, typical

Current mode, full-scale step, 10 μA range[25]

<150 μs, typical

Current mode, full-scale step, 1 μA range[25]

<300 μs, typical

The following figures illustrate the effect of the transient response setting on the step response of the PXIe-4137 for different loads.

Figure 1. 1 mA Range, No Load Step Response, Nominal


Figure 1. 1 mA Range, 100 nF Load Step Response, Nominal


Load Regulation

Voltage

Device configured for local sense

200 mV per A of output load change (measured between output channel terminals), typical

Device configured for remote sense

100 μV per A of output load change (measured between sense terminals), typical

Current, device configured for local or remote sense

Load regulation effect included in current accuracy specifications, typical

Expected Relay Life

Output Connected

≥100 k cycles

Note To avoid excessive relay wear, do not set Output Connected to TRUE when a non-zero voltage is connected to the output.

Measurement and Update Timing Characteristics

Available sample rates[26]26 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate.

(1.8 MS/s)/N where N = 1, 2, 3, … 224, nominal

Sample rate accuracy

Equal to PXIe_CLK100 accuracy, nominal

Maximum measure rate to host

1.8 MS/s per channel, continuous, nominal

Maximum source update rate[27]27 As the source delay is adjusted or if advanced sequencing is used, maximum source rates vary. Timed output mode is enabled in Sequence Mode by setting Sequence Step Delta Time Enabled to True. Additional timing limitations apply when operating in pulse mode (Output Function is set to Pulse Voltage or Pulse Current).

Sequence mode

100,000 updates/s (10 μs/update), nominal

Timed output mode

80,000 updates/s (12.5 μs/update), nominal

Input trigger to

Source event delay

10 μs, nominal

Source event jitter

1 μs, nominal

Measure event jitter

1 μs, nominal

Pulse mode timing and accuracy[28]28 Pulse mode is enabled when the Output Function is set to Pulse Voltage or Pulse Current. This mode enables access to extended range pulsing capabilities. For PXIe-4137 (20W), shorter minimum on times for in-range pulses can be achieved using Sequence mode or Timed Output mode with the Output Function set to Voltage or Current.
Minimum pulse on time[29]29 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB.

PXIe-4137 (40W)[30]30 Optimize transient response, overshoot, and slew rate with NI SourceAdapt by adjusting the Transient Response.

10 μs, nominal

PXIe-4137 (20W)

50 μs, nominal

Minimum pulse off time [31]31 Pulses fall inside DC limits. Pulse off time is measured from the start of the trailing edge to the start of a subsequent leading edge.

50 μs, nominal

Pulse on time or off time programming resolution

100 ns, nominal

Pulse on time or off time programming accuracy

±5 μs, nominal

Pulse on time or off time jitter

1 μs, nominal

Remote Sense

Voltage accuracy

Add 3 ppm of voltage range per volt of HI lead drop plus 1 μV per volt of lead drop per ohm of corresponding sense lead resistance to voltage accuracy specifications

Maximum sense lead resistance

100 Ω

Maximum lead drop per lead

3 V, maximum 202 V between HI and LO terminals

Note Exceeding the maximum lead drop per lead value may cause the driver to report a sense lead error.

Safety Interlock

The safety interlock feature is designed to prevent users from coming in contact with hazardous voltage generated by the SMU in systems that implement protective barriers with controlled user access points.

Caution Hazardous voltages of up to the maximum voltage of the PXIe-4137 may appear at the output terminals if the safety interlock terminal is closed. Open the safety interlock terminal when the output connections are accessible. With the safety interlock terminal open, the output voltage level/limit is limited to ±40 V DC, and protection will be triggered if the voltage measured between the device HI and LO terminals exceeds ±(42 V peak ±0.4 V).
Attention Des tensions dangereuses allant jusqu'à la tension maximale du PXIe-4137 peuvent apparaître aux terminaux de sortie si le terminal de verrouillage de sécurité est fermé. Ouvrez le terminal de verrouillage de sécurité lorsque les connexions de sortie sont accessibles. Lorsque le terminal de verrouillage de sécurité est ouvert, le niveau ou la limite de tension de sortie est limité à ± 40 V CC, et la protection se déclenchera si la tension mesurée entre les terminaux HI et LO de l'appareil dépasse ± (42 Vpic ± 0,4 V).
Caution Do not apply voltage to the safety interlock connector inputs. The interlock connector is designed to accept passive, normally open contact closure connections only.
Attention N'appliquez pas de tension aux entrées du connecteur de verrouillage de sécurité. Le connecteur de verrouillage est conçu pour accepter uniquement des connexions à fermeture de contact passives, normalement ouvertes.
Safety interlock terminal open

Output

<±42.4 V peak

Setpoint

<±40 V DC

Safety interlock terminal closed

Output

Maximum voltage of the device

Setpoint

Maximum selected voltage range

Examples of Calculating Accuracy Specifications

Note Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device.

Example 1: Calculating 5 °C Accuracy

Calculate the accuracy of 900 nA output in the 1 µA range under the following conditions:

Ambient temperature 28 °C
Internal device temperature within Tcal ±5 °C[32]32 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.
Self-calibration within the last 24 hours

Solution: Because the device internal temperature is within Tcal ±5 °C and the ambient temperature is within 23 °C ±5 °C, the appropriate accuracy specification is the following value:

0.03% + 100 pA

Calculate the accuracy using the following formula:

Accuracy=900nA*0.03%+100pA
  =270pA+100pA

=370pA

Therefore, the actual output is within 370 pA of 900 nA.

Example 2: Calculating Remote Sense Accuracy

Calculate the remote sense accuracy of 500 mV output in the 600 mV range. Assume the same conditions as in Example 1, with the following differences:

HI path lead drop 3 V
HI sense lead resistance 2 Ω
LO path lead drop 2.5 V
LO sense lead resistance 1.5 Ω

Solution: Because the device internal temperature is within Tcal ±5 °C and the ambient temperature is within 23 °C ±5 °C, the appropriate accuracy specification is the following value:

0.02% + 50 μV

Because the device is using remote sense, use the following remote sense accuracy specification:

Add 3 ppm of voltage range per volt of HI lead drop plus 1 μV per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications.

Calculate the remote sense accuracy using the following formula:

Accuracy=(500mV*0.02%+50μV)+600mV*3ppm1Vofleaddrop*3V+1μVV*Ω*3V*2Ω+1μVV*Ω*2.5V*1.5Ω
  =100μV+50μV+1.8μV*3+6μV+3.75μV
=165.15μV

Therefore, the actual output is within 165.15 µV of 500 mV.

Example 3: Calculating Accuracy with Temperature Coefficient

Calculate the accuracy of 900 nA output in the 1 µA range. Assume the same conditions as in Example 1, with the following differences:

Ambient temperature 15 °C

Solution: Because the device internal temperature is within Tcal ±5 °C, the appropriate accuracy specification is the following value:

0.03% + 100 pA

Because the ambient temperature falls outside of 23 °C ±5 °C, use the following temperature coefficient per °C outside the 23 °C ±5 °C range:

0.0006% + 4 pA

Calculate the accuracy using the following formula:

TemperatureVariation=(23°C5°C)15°C=3°C

Accuracy=(900nA*0.03%+100pA)+900nA*0.0006%+4pA1°C*3°C

  =370pA+28.2pA

=398.2pA

Therefore, the actual output is within 398.2 pA of 900 nA.

Examples of Determining Extended Range Pulse Parameters and Optimizing Slew Rate using NI SourceAdapt

Note Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device.

Example 1: Determining Extended Range Pulse On Time and Duty Cycle Parameters for the PXIe-4137 (40W)

Determine the extended range pulsing parameters, assuming the following operating point.

Output function Pulse Current
Pulse voltage limit, Vpulse 80 V
Pulse current level, Ipulse 3 A
Bias voltage limit, Vbias 0.1 V
Bias current level, Ibias 0 A
Pulse on time, ton 1.5 ms
Chassis' slot cooling capacity ≥58 W

Solution

Begin by calculating the pulse power using the following equation.

Pulse power=Vpulse*Ipulse
=80 V* 3 A
=240 W

For PXIe-4137 (40W), refer to the following figures to identify next steps. First, verify the the region of operation using Figure 1, which shows 240 W is in the extended range pulsing region.

Next, refer to GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290.html#GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290__FIG_LNR_2ZZ_5MB, which shows the maximum pulse on time, ton, is limited by the maximum pulse energy, EpulseMax. Use the pulse energy equation (Equation 3) from GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290.html#GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290__TABLE_CJS_41C_1NB to calculate the maximum pulse on time, tonMax(Equation 8).

tonMax=|EpulseMaxVpulse*Ipulse|    (Eq.8)
=|0.4 J80 V* 3 A|
=1.67 ms

Next, refer to GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290.html#GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290__FIG_EQB_BZZ_5MB, which shows the maximum duty cycle, D, is limited by the cycle average power, PCA.If the required pulse on time is 1.5 ms and the module is installed in a chassis with slot cooling capacity ≥58 W, use the cycle average power equation (Equation7) from GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290.html#GUID-CABA0054-BF9F-4E2F-99B3-67B9894D8290__TABLE_CJS_41C_1NB to calculate the minimum pulse off time, toffMin(Equation 9).

toffMin=|PCA*tonVpulse*Ipulse*tonPCAVbias*Ibias|    (Eq.9)
=|20 W* 1.5 ms80 V* 3 A* 1.5 ms20 W0.1 V* 0 A|
=16.5 ms

Finally, verify that the pulse cycle time, tcycle, is greater than or equal to the minimum pulse cycle time, tcycleMin (5 ms). To calculate the pulse cycle time, use the following equation:

tcycle=ton+toff    (Eq. 6)
=1.5 ms+16.5 ms
=18 ms

In this case, the pulse cycle time meets the minimum pulse cycle time specification.

Therefore, a 80 V, 3 A pulse with an on time of 1.5 ms and a pulse off time of 16.5 ms is supported, since it fulfills the following criteria:

  • Greater than the minimum pulse on time of 10 μs
  • Equal to the minimum pulse off time of 16.5 ms to meet maximum cycle average power
  • Greater than the minimum pulse cycle time of 5 ms

Example 2: Determining Extended Range Pulse On Time and Duty Cycle Parameters for the PXIe-4137 (20W)

Determine the extended range pulsing parameters, assuming the following operating point.

Output function Pulse Current
Pulse voltage limit, Vpulse 80 V
Pulse current level, Ipulse 3 A
Bias voltage limit, Vbias 0.1 V
Bias current level, Ibias 0 A
Pulse on time, ton 1.5 ms
Chassis' slot cooling capacity ≥58 W

Solution

Begin by calculating the pulse power using the following equation.

Pulse power=Vpulse*Ipulse
=80 V* 3 A
=240 W

Since the pulse power of 240 W is within the 480 W region of GUID-F6A69CB1-8128-4008-937F-4F11666F190F.html#GUID-F6A69CB1-8128-4008-937F-4F11666F190F__FIG_20WQUADDIAGRAM, the maximum configurable on time is 400 μs and maximum duty cycle is 2%.

For example, if the required pulse on time is 100 μs, and the required pulse cycle time is 10 ms, calculate the pulse off time and verify the duty cycle using the following equations.

toff=tcycleton
=10 ms100μs
=9.9 ms

Duty cycle=tontcycle* 100%
=1%

Therefore, a pulse with an on time of 100 μs and 1% duty cycle would be supported, since it fulfills the following criteria:

  • Greater than the minimum pulse on time of 50 μs
  • Less than the maximum pulse on time of 400 μs and duty cycle of 2%
  • Greater than the minimum pulse cycle time of 5 ms

Example 3: Using NI SourceAdapt to Increase the Slew Rate of the Pulse

Determine the appropriate operating parameters and custom transient response settings, assuming the following example parameters.

Output function Pulse Current
Pulse voltage limit, Vpulse 160 V
Pulse current level, Ipulse 3 A
Bias voltage limit, Vbias 0.1 V
Bias current level, Ibias 0 A
Transient response Fast
Load, cable impedance 22.3 Ω, 1.8 μH
Pulse on time, ton 10 μs
Pulse off time, toff 4.99 ms

The SMU Transient Response can be configured to three predefined settings, Slow, Normal, and Fast. If these settings do not provide the desired pulse response, a fourth setting, Custom, enables NI SourceAdapt[33]33 Visit ni.com for more information about NI SourceAdapt Next-Generation SMU Technology. technology which provides the ability to customize the SMU response to any load, and achieve an ideal response with minimum rise times and no overshoots or oscillations.

Figure 1. 10 μs Pulse Output with Load, Fast Transient Response


Solution

SourceAdapt allows users to set the desired gain bandwidth, compensation frequency, and pole-zero ratio through custom transient response to obtain the desired pulse waveform. To use SourceAdapt, first set the Transient Response to Custom.

To achieve the resulting waveform in the following figure, use the parameters in the following table.

Figure 1. 10 μs Pulse Output with Load, Custom Transient Response


Transient response Custom
Current: Gain bandwidth 900 kHz
Current: Compensation frequency 200 kHz
Current: Pole-zero ratio 2

Gain bandwidth is directly proportional to the step response slew rate. The higher the gain bandwidth, the higher the slew rate. It is worth noting that increasing the gain bandwidth will likely increase ringing. However, this can likely be removed by appropriately setting the compensation frequency and the pole-zero ratio.

Figure 1. Example of Ringing Frequency


Compensation frequency and pole-zero ratio are used to determine the frequencies of the SMU control loop pole and zero, which can be used to optimize the system transient response by increasing phase margin and reducing ringing. To reduce the overshoot, it is recommended to set the compensation frequency close to the overshoot ringing frequency, see Fc in the figure above, and set the pole-zero ratio to be greater than 1.

For reference, the pole frequency and zero frequency are derived by the following equations.

Pole frequency=Compensation frequency*Pole-zero ratio

Zero frequency=Compensation frequencyPole-zero ratio

These settings can be accessed through the Transient Response set to Custom: Voltage or Current.

Trigger Characteristics

Input triggers

Types

Start, Source, Sequence Advance, Measure, Pulse

Sources (PXI trigger lines <0...7>) [34]34 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.

Polarity

Configurable

Minimum pulse width

100 ns, nominal

Destinations[35]35 Input triggers can be re-exported. (PXI trigger lines <0...7>)

Polarity

Active high (not configurable)

Pulse width

>200 ns, typical

Output triggers (events)

Types

Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete, Pulse Complete, Ready for Pulse

Destinations (PXI trigger lines <0...7>)

Polarity

Configurable

Pulse width

Configurable between 250 ns and 1.6 μs, nominal

Protection

Output channel protection

Overcurrent or overvoltage

Automatic shutdown, output disconnect relay opens

Sink overload protection

Automatic shutdown, output disconnect relay opens

Overtemperature

Automatic shutdown, output disconnect relay opens

Safety interlock

Disable high voltage output, output disconnect relay opens

Safety Voltage and Current

Notice The protection provided by the PXIe-4137 can be impaired if it is used in a manner not described in the user documentation.
Warning Take precautions to avoid electrical shock when operating this product at hazardous voltages.
Caution Isolation voltage ratings apply to the voltage measured between any channel pin and the chassis ground. When operating channels in series or floating on top of external voltage references, ensure that no terminal exceeds this rating.
Attention Les tensions nominales d'isolation s'appliquent à la tension mesurée entre n'importe quelle broche de voie et la masse du châssis. Lors de l'utilisation de voies en série ou flottantes en plus des références de tension externes, assurez-vous qu'aucun terminal ne dépasse cette valeur nominale.

DC voltage

±200 V

Channel-to-earth ground isolation

Continuous

250 V DC, CAT I

Withstand

1,000 V RMS, verified by a 5 s withstand

Caution Do not connect the PXIe-4137 to signals or use for measurements within Measurement Categories II, III, or IV.
Attention Ne connectez pas le PXIe-4137 à des signaux et ne l'utilisez pas pour effectuer des mesures dans les catégories de mesure II, III ou IV.

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

DC current range

±1 A; ±3 A, pulse only

Guard Output Characteristics

Cable guard

Output impedance

3 kΩ, nominal

Offset voltage

1 mV, typical

Calibration Interval

Recommended calibration interval

1 year

Power Requirement

PXIe-4137 (40W)

3.0 A from the 3.3 V rail and 6.0 A from the 12 V rail

PXIe-4137 (20W)

2.5 A from the 3.3 V rail and 2.7 A from the 12 V rail

Physical

Dimensions

3U, one-slot, PXI Express/CompactPCI Express module

2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.)

Weight

PXIe-4137 (20W)

419 g (14.8 oz)

PXIe-4137 (40W)

428 g (15.1 oz)

Front panel connectors

5.08 mm (8 position) combicon, 1 × 4.08 mm(3 position) combicon

Environmental Characteristics

Temperature

Operating

0 °C to 55 °C

Storage

-40 °C to 71 °C

Humidity

Operating

10% to 90%, noncondensing

Storage

5% to 95%, noncondensing

Pollution Degree

2

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient temperature)

Shock and Vibration

Operating vibration

5 Hz to 500 Hz, 0.3 g RMS

Non-operating vibration

5 Hz to 500 Hz, 2.4 g RMS

Operating shock

30 g, half-sine, 11 ms pulse

1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).

2 For increased capability, NI recommends installing the PXIe-4137 (40W) in a chassis with slot cooling capacity ≥58 W.

3 Voltage levels and limits >|40 VDC| require the safety interlock input to be closed.

4 Current is limited to 1 A DC. Higher levels are pulsing only.

5 Power limit defined by voltage measured between HI and LO terminals.

6 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions.

7 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.

8 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.

9 3 A range above 1 A is for pulsing only.

10 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.

11 3 A range above 1 A is for pulsing only.

12 Overvoltage protection accuracy is valid with an ambient temperature of 23 °C ± 5 °C and with Tcal ±5 °C. Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.

13 For example, given a continuous pulsing load, if the largest dynamic step in current that the load sources/sinks is from 0.5 A to 1.0 A, then the maximum SMU current step is 0.5 A. Thus, the minimum dynamic load pulse cycle time is 50 μs. Minimum dynamic load pulse cycle time is independent of output voltage.[14]14 Measurable unit of μs/A is used because the minimum pulse cycle time is independent of output voltage

14 Measurable unit of μs/A is used because the minimum pulse cycle time is independent of output voltage

15 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB.

16 Refer to Figure 7 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis.

17 Refer to Figure 8 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis. If D≥100%, consider switching Output Function from Pulse mode to DC mode.

18 Refer to Figure 8 to estimate the value and determine the limiting equation for a PXIe-4137 (40W) in a ≥58 W Slot Cooling Capacity Chassis.

19 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB.

20 Optimize transient response, overshoot, and slew rate with NI SourceAdapt by adjusting the Transient Response.

21 To improve the slew rate, see Examples of Determining Extended Range Pulse Parameters and Optimizing Slew Rate using NI SourceAdapt.

22 Measured as the time to settle to within 0.1% of step amplitude, device configured for fast transient response.

23 Current limit set to ≥60 μA and ≥60% of the selected current limit range.

24 Current limit set to ≥20 μA and ≥20% of selected current limit range.

25 Voltage limit set to ≥2 V, resistive load set to 1 V/selected current range.

26 When sourcing while measuring, both the Source Delay and Aperture Time affect the sampling rate. When taking a measure record, only the Aperture Time affects the sampling rate.

27 As the source delay is adjusted or if advanced sequencing is used, maximum source rates vary. Timed output mode is enabled in Sequence Mode by setting Sequence Step Delta Time Enabled to True. Additional timing limitations apply when operating in pulse mode (Output Function is set to Pulse Voltage or Pulse Current).

28 Pulse mode is enabled when the Output Function is set to Pulse Voltage or Pulse Current. This mode enables access to extended range pulsing capabilities. For PXIe-4137 (20W), shorter minimum on times for in-range pulses can be achieved using Sequence mode or Timed Output mode with the Output Function set to Voltage or Current.

29 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. See GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C.html#GUID-2E9B16BD-1A40-4F06-8896-84F9B325AF7C__FIG_JFH_D4Z_DNB.

30 Optimize transient response, overshoot, and slew rate with NI SourceAdapt by adjusting the Transient Response.

31 Pulses fall inside DC limits. Pulse off time is measured from the start of the trailing edge to the start of a subsequent leading edge.

32 Tcal is the internal device temperature recorded by the PXIe-4137 at the completion of the last self-calibration.

33 Visit ni.com for more information about NI SourceAdapt Next-Generation SMU Technology.

34 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1.

35 Input triggers can be re-exported.