PXIe-2525 Specifications
- Updated2025-05-22
- 6 minute(s) read
PXIe-2525 Specifications
PXIe-2525 Specifications
These specifications apply to the PXIe-2525 .
All specifications are subject to change without notice.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid at 23 °C unless otherwise noted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
PXIe-2525 Pinout
| Signal | Description |
|---|---|
| CHx+ | Positive signal connection |
| CHx- | Negative signal connection |
| COMx+ | Routing destination for corresponding positive signal connections |
| COMx- | Routing destination for corresponding negative signal connections |
Topology
Topology | 2-wire 16 Bank 4 × 1 2-wire Octal 8 × 1 2-wire Quad 16 × 1 2-wire Dual 32 × 1 2-wire 64 × 1 |
Input Characteristics
| Maximum switching voltage, channel-to-channel | 150 V |
|---|---|
| Maximum switching voltage, channel-to-ground | 150 V, CAT O |
Measurement Categories CAT I and CAT O (Other) are equivalent and are for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. This category is for measurement of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special hardware, limited-energy parts of hardware, circuits powered by regulated low-voltage sources, and electronics.
| Maximum switching power | 60 W, 62.5 VA |
|---|
| CH-COM DC isolation resistance | >1 GΩ, typical at 25 °C |
|---|---|
| Maximum switching current | 2 A (per channel) |
| Maximum carry current | 2 A (per channel) |
| Minimum switch load | 20 mV/1 mA |
| DC path resistance, initial | <0.8 Ω, warranted |
|---|---|
| DC path resistance, end-of-life | ≥1.8 Ω |
DC path resistance is a combination of relay contact resistance and trace resistance. Measure path resistance by combining the resistance of the high and low signal paths from one row to one column. Contact resistance typically remains low for the life of a relay. At the end of relay life, the contact resistance rises rapidly at ≥1.8 Ω.
| Thermal EMF | <10 µV |
|---|
RF Performance Characteristics
Bandwidth (-3 dB, typical at 23 ºC) 50 Ω termination | ≤10 MHz | ||||||||
| |||||||||
| |||||||||
Dynamic Characteristics
| Relay operate time |
|
|---|---|
| Expected relay life |
|
Physical Characteristics
| Relay type | Electromechanical, non-latching |
|---|---|
| Relay contact material | Palladium-ruthenium, gold covered |
| Front panel connector | 160 DIN 41612, 160 positions, male |
| PXI Express power requirement |
|
| Dimensions (L × W × H) |
|
| Weight | 230 g (8.1 oz) |
Environment
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
|---|---|
| Pollution Degree | 2 |
Indoor use only.
Operating Environment
| Ambient temperature range, operating | 0 °C to 55 °C |
|---|---|
| Relative humidity range, operating | 10% to 90%, noncondensing |
Storage Environment
| Ambient temperature, storage | -40 °C to 71 °C |
|---|---|
| Relative humidity range, storage | 5% to 95%, noncondensing |
Shock and Vibration
| Operating shock | 30 g peak, half-sine, 11 ms pulse |
|---|---|
| Random vibration |
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.