Power On and Reset Conditions

The PXI-5661 hardware is in the following state after powering on or restarting the system and allowing the PC operating system and NI-RFSA to fully load. These conditions are also true after a device reset that you perform directly from NI Measurement & Automation Explorer (MAX).
  • Thermal shutdown monitoring is activated.
  • PXI-5600 RF signal downconverter module RF enclosure is powered on.
  • PXI-5600 module front panel POWER LED is green.
  • PXI-5600 module front panel STATUS LED is off.
  • PXI-5142 oscilloscope module front panel ACCESS LED is green.
  • PXI-5142 module front panel ACTIVE LED is off.
  • PXI-5600 onboard 10 MHz Reference Clock is selected as the vector signal analyzer Reference Clock.
  • Warm-up begins (if applicable).