PXI-4132 Specifications
- Updated2023-02-18
- 8 minute(s) read
PXI-4132 Specifications
These specifications apply to the PXI-4132.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- Ambient temperature[1] of 23 °C ± 5 ºC
- 30 minutes warm-up time
- niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to On
- niDCPower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 1 power-line cycle (PLC)
- Self-calibration performed within the last 24 hours
Device Capabilities
DC Voltage Ranges (CAT I) | DC Current Source and Sink Ranges |
---|---|
|
|
SMU Specifications
Voltage Programming Accuracy/Resolution
Range | Resolution, Nominal | Accuracy ± (% of Output + Offset) | Peak to Peak Noise, Typical (0.1 Hz to 10 Hz) |
---|---|---|---|
1 Year 23 °C ± 5 °C | |||
±10 V | 50 μV | 0.025% + 3 mV | 70 μV |
±100 V | 500 µV | 0.025% + 10 mV | 300 μV |
Current Programming Accuracy/Resolution[2]
Range | Resolution, Nominal | Accuracy ± (% of Output + Offset) | Peak to Peak Noise, Typical (0.1 Hz to 10 Hz) |
---|---|---|---|
1 Year 23 °C ± 5 °C | |||
10 µA | 500 pA | 0.034% + 2.0 nA | 90 pA |
100 µA | 5 nA | 0.034% + 20 nA | 900 pA |
1 mA | 50 nA | 0.034% + 0.2 µA | 9 nA |
10 mA | 500 nA | 0.034% + 2.0 µA | 90 nA |
100 mA | 5 µA | 0.034% + 20 µA | 900 nA |
Voltage Measurement Accuracy/Resolution
Range | Resolution, Nominal | Accuracy ± (% of Reading + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
±10 V | 10 μV | 0.02% + 2.0 mV |
±100 V | 100 µV | 0.02% + 5.0 mV |
Current Measurement Accuracy/Resolution
Range | Resolution, Nominal | Accuracy ± (% of Reading + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
10 µA | 10 pA | 0.028% + 1.0 nA |
100 µA | 100 pA | 0.028% + 10 nA |
1 mA | 1 nA | 0.028% + 0.1 µA |
10 mA | 10 nA | 0.028% + 1.0 µA |
100 mA | 100 nA | 0.020% + 10 µA |
Additional Specifications
Temperature Coefficient | 15% of accuracy specification per °C | ||||||
Settling time | <300 µs; Settled to 0.1% of final value (1 V step at 50% load of current range), typical | ||||||
Transient response | Recovers to <0.1% of voltage range within 100 µs after a change in load current from 10% to 90% of current range, typical | ||||||
Normal mode noise (source only) | 8 mVp-p into resistive load <1 mV RMS (20 Hz to 20 MHz bandwidth), typical | ||||||
Remote sense | Add 0.5% of HI lead drop to voltage accuracy specification (Maximum lead drop) Up to 1 V drop per lead | ||||||
| |||||||
Guard offset voltage | <4 mV, typical (Current ≤ 10 mA) | ||||||
|
Step Response
Measurement Noise Versus Aperture Time, Typical
Supplemental Specifications
Measurement Speed[4]
ADC Aperture Time | Measure to Host | Source-measure to Host[5] |
---|---|---|
1/64 PLC | 3490 (2900) | 1900 (1700) |
1/8 PLC | 470 (390) | 425 (360) |
1 PLC | 59.9 (49.9) | 59.0 (49.3) |
Maximum source update rate[5] | 4,200 updates/s |
Trigger in to source delay | 500 ns, nominal |
Triggers
| |||||||||||||||||
|
Calibration Interval
Recommended calibration interval | 1 year |
Physical Characteristics
Dimensions | 3U, one-slot, PXI/cPXI module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.), nominal |
Weight | 295 g (10.4 oz), typical |
Front panel connectors | COMBICON, 5.08 mm (8 position), nominal |
Power Requirements
PXI power requirement | 10 W at 5 V 1 W at 3.3 V 2 W at 12 V, typical |
Environment
Maximum altitude | 2,000 m (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 10% to 70%, noncondensing; derate 1.3% per °C above 40 °C (Tested in accordance with IEC 60068-2-56.) |
Storage Environment
Ambient temperature range | -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Shock and Vibration
Operational shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | ||||||
|
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
- 2011/65/EU; Restriction of Hazardous Substances (RoHS)
- 2014/53/EU; Radio Equipment Directive (RED)
- 2014/34/EU; Potentially Explosive Atmospheres (ATEX)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 Minimum programmable current limit/level is 2% of range.
3 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, III, or CAT IV.
4 Does not include load dependent settling time; niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to Off.
5 Source-measure to Host and Maximum Source Update Rate are performed with the source delay set to 200 µs. This is sufficient for the output to settle within 1% of the requested level with a simple resistive load. As you adjust the source delay for your application's requirements, maximum rates vary.
6 Pulse widths and logic levels compliant with PXI Hardware Specification Revision 2.2.
7 Input triggers can be re-exported.