PXI-4132 Specifications

These specifications apply to the PXI-4132.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Warranted unless otherwise noted.

Conditions

Specifications are valid under the following conditions unless otherwise noted.

  • Ambient temperature[1] of 23 °C ± 5 ºC
  • 30 minutes warm-up time
  • niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to On
  • niDCPower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 1 power-line cycle (PLC)
  • Self-calibration performed within the last 24 hours

Device Capabilities

The following table and figure illustrate the voltage and current source and sink ranges of the PXI-4132.
Table 1. PXI-4132 Current Source and Sink Ranges
DC Voltage Ranges (CAT I) DC Current Source and Sink Ranges
  • ±10 V
  • ±100 V
  • 10 μA
  • 100 μA
  • 1 mA
  • 10 mA
  • 100 mA
Caution Shock hazards exist when voltage levels are greater than 30 VRMS, 42.4 V peak, or 60 VDC. Use extreme caution when a shock hazard is present. Always ensure the test system is de-energized before connecting or disconnecting the backshell assembly or cables from the PXI-4132.
Figure 1. PXI-4132 Quadrant Diagram, Characteristic

1378

SMU Specifications

Voltage Programming Accuracy/Resolution

Range Resolution, Nominal Accuracy ± (% of Output + Offset) Peak to Peak Noise, Typical (0.1 Hz to 10 Hz)
1 Year 23 °C ± 5 °C
±10 V 50 μV 0.025% + 3 mV 70 μV
±100 V 500 µV 0.025% + 10 mV 300 μV

Current Programming Accuracy/Resolution[2]

Range Resolution, Nominal Accuracy ± (% of Output + Offset) Peak to Peak Noise, Typical (0.1 Hz to 10 Hz)
1 Year 23 °C ± 5 °C
10 µA 500 pA 0.034% + 2.0 nA 90 pA
100 µA 5 nA 0.034% + 20 nA 900 pA
1 mA 50 nA 0.034% + 0.2 µA 9 nA
10 mA 500 nA 0.034% + 2.0 µA 90 nA
100 mA 5 µA 0.034% + 20 µA 900 nA

Voltage Measurement Accuracy/Resolution

Range Resolution, Nominal Accuracy ± (% of Reading + Offset)
1 Year 23 °C ± 5 °C
±10 V 10 μV 0.02% + 2.0 mV
±100 V 100 µV 0.02% + 5.0 mV

Current Measurement Accuracy/Resolution

Range Resolution, Nominal Accuracy ± (% of Reading + Offset)
1 Year 23 °C ± 5 °C
10 µA 10 pA 0.028% + 1.0 nA
100 µA 100 pA 0.028% + 10 nA
1 mA 1 nA 0.028% + 0.1 µA
10 mA 10 nA 0.028% + 1.0 µA
100 mA 100 nA 0.020% + 10 µA

Additional Specifications

Temperature Coefficient

15% of accuracy specification per °C

Settling time

<300 µs; Settled to 0.1% of final value (1 V step at 50% load of current range), typical

Transient response

Recovers to <0.1% of voltage range within 100 µs after a change in load current from 10% to 90% of current range, typical

Normal mode noise (source only)

8 mVp-p into resistive load <1 mV RMS (20 Hz to 20 MHz bandwidth), typical

Remote sense

Add 0.5% of HI lead drop to voltage accuracy specification

(Maximum lead drop) Up to 1 V drop per lead

Load Regulation

Voltage

0.5 mV per mA of output load using local sense

Current

0.01% of range per volt of output change

Guard offset voltage

<4 mV, typical (Current ≤ 10 mA)

Isolation voltage (continuous)

Channel-to-earth ground

150 VDC, CAT I[3], verified by dielectric withstand test, 5 s, characteristic

Caution Do not connect to MAINs. Do not connect to signals or use for measurements within CAT II, III, or IV.

Step Response

The following figures illustrate the step response of the PXI-4132 for different loads.
Figure 2. 100 mA Range Step Response, Typical

1378

Figure 3. 10 mA Range Step Response, Typical

1378

Figure 4. 1 mA, 100 µA, and 10 µA Range Step Response, Typical

1378

Measurement Noise Versus Aperture Time, Typical


1378

Supplemental Specifications

Measurement Speed[4]

Table 2. Maximum Operation Rates Per Second for 60 Hz (50 Hz)
ADC Aperture Time Measure to Host Source-measure to Host[5]
1/64 PLC 3490 (2900) 1900 (1700)
1/8 PLC 470 (390) 425 (360)
1 PLC 59.9 (49.9) 59.0 (49.3)

Maximum source update rate[5]

4,200 updates/s

Trigger in to source delay

500 ns, nominal

Triggers

Input triggers

Types

Start, Source, Sequence Advance, Measure

Sources

PXI Trigger lines <0...7>[6]

Polarity, sources

Configurable

Minimum pulse width, sources

100 ns

Destinations[7]

PXI Trigger Lines <0...7>[6]

Polarity, destinations

Active high (unconfigurable)

Pulse width, destinations

150 ns, nominal

Output Triggers (Events)

Types

Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete

Destinations

PXI Trigger Lines <0...7>[6]

Polarity, destinations

Configurable

Pulse width, destinations

Configurable between 150 ns and 1.6 µs

Calibration Interval

Recommended calibration interval

1 year

Physical Characteristics

Dimensions

3U, one-slot, PXI/cPXI module

2.0 cm × 13.0 cm × 21.6 cm

(0.8 in. × 5.1 in. × 8.5 in.), nominal

Weight

295 g

(10.4 oz), typical

Front panel connectors

COMBICON, 5.08 mm (8 position), nominal

Note Front panel connectors can accept wire gauges from 12 AWG to 28 AWG.

Power Requirements

PXI power requirement

10 W at 5 V

1 W at 3.3 V

2 W at 12 V, typical

Environment

Maximum altitude

2,000 m (at 25 °C ambient temperature)

Pollution Degree

2

Indoor use only.

Operating Environment

Ambient temperature range

0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

Relative humidity range

10% to 70%, noncondensing; derate 1.3% per °C above 40 °C (Tested in accordance with IEC 60068-2-56.)

Storage Environment

Ambient temperature range

-40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.)

Relative humidity range

5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.)

Shock and Vibration

Operational shock

30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)

Random vibration

Operating

5 Hz to 500 Hz, 0.3 grms (Tested in accordance with IEC 60068-2-64.)

Nonoperating

5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

Compliance and Certifications

Caution You can impair the protection provided by the PXI-4132 if you use it in a manner not described in this document.

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • EN 55022 (CISPR 22): Class A emissions
  • EN 55024 (CISPR 24): Immunity
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • AS/NZS CISPR 22: Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to the Product Certifications and Declarations section.

CE Compliance 1378

This product meets the essential requirements of applicable European Directives, as follows:

  • 2014/35/EU; Low-Voltage Directive (safety)
  • 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
  • 2011/65/EU; Restriction of Hazardous Substances (RoHS)
  • 2014/53/EU; Radio Equipment Directive (RED)
  • 2014/34/EU; Potentially Explosive Atmospheres (ATEX)

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.

For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

EU and UK Customers

  • 1378 Waste Electrical and Electronic Equipment (WEEE)—At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee.
  • 电子信息产品污染控制管理办法(中国RoHS)

  • 1378 中国RoHSNI符合中国电子信息产品中限制使用某些有害物质指令(RoHS)。关于NI中国RoHS合规性信息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.)
  • 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).

    2 Minimum programmable current limit/level is 2% of range.

    3 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, III, or CAT IV.

    4 Does not include load dependent settling time; niDCPower Auto Zero property or NIDCPOWER_ATTR_AUTO_ZERO attribute set to Off.

    5 Source-measure to Host and Maximum Source Update Rate are performed with the source delay set to 200 µs. This is sufficient for the output to settle within 1% of the requested level with a simple resistive load. As you adjust the source delay for your application's requirements, maximum rates vary.

    6 Pulse widths and logic levels compliant with PXI Hardware Specification Revision 2.2.

    7 Input triggers can be re-exported.