PXI-2593 Specifications
- Updated2025-05-22
- 6 minute(s) read
PXI-2593 Specifications
These specifications apply to the PXI-2593.
All specifications are subject to change without notice.
About These Specifications
Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
PXI-2593 Pinout
| Signal | Description |
|---|---|
| CHx | Signal connection |
| COMx | Routing destination for corresponding signal connections |
| TRIG IN | Trigger input connection |
| TRIG OUT | Trigger output connection |
Topology
Topology | 16x1 multiplexer 8x1 terminated multiplexer Dual 4x1 terminated multiplexer Dual 8x1 terminated multiplexer Independent |
Input Characteristics
Maximum switching voltage | 150 V, CAT I (channel-to-channel and channel-to-ground) |
Maximum switching current (per channel) | 0.5 A |
Maximum carry current (per channel) | 1 A |
Simultaneous channels at maximum current | Up to 2 |
Maximum switching power (per channel) | 10 W |
Minimum switch load | 10 μA, 10 mV | ||||||
Maximum RF carry power (per channel) | 10 W up to 500 MHz | ||||||
| |||||||
Path resistance is a combination of relay contact resistance and trace resistance. Contact resistance typically remains low for the life of a relay. At the end of relay life, the contact resistance rises rapidly above 1.0 Ω.
RF Performance Characteristics
Characteristic impedance (Z0) | 50 Ω nominal | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Channel-to-channel skew within each 8-channel bank is less than 100 ps. Only channels from standard topologies are listed in Table 1.
| COM | CH0-CH7 | CH8-CH15 |
|---|---|---|
| 0 | 1.90 to 2.00 | 2.55 to 2.65 |
| 1 | — | 1.90 to 2.00 |
| |||||||
Refer to the following figures for typical insertion loss, typical VSWR, typical isolation, and typical bank-to-bank crosstalk, respectively.
Dynamic Characteristics
| |||||||
| |||||||
Trigger Characteristics
| |||||||||||||
| |||||||||||||
| |||||||||||||
Physical Characteristics
Relay type | Electromechanical, latching | ||||||
Relay contact material | Silver palladium and gold | ||||||
I/O connectors | 18 MCX jacks | ||||||
Trigger connectors | 2 SMB jacks | ||||||
| |||||||
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module, PXI Express compatible 21.6 × 2.0 × 13.0 cm (8.5 × 0.8 × 5.1 in.) | ||||||
Weight | 330 g (12 oz) | ||||||
Environment
| Operating temperature | 0 °C to 50 °C |
| Storage temperature | -20 °C to 70 °C |
| Relative humidity | 5% to 85%, noncondensing |
| Pollution Degree | 2 |
| Maximum altitude | 2,000 m |
Indoor use only.
Shock and Vibration
| Operational Shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) |
| Random vibration, operating | 5 Hz to 500 Hz, 0.3 grms |
| Random vibration, nonoperating | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Compliance and Certifications
Safety Compliance Standards
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.