PXI-2532B Specifications

PXI-2532B Specifications

These specifications apply to the PXI-2532B.

All specifications are subject to change without notice.

Notice The protection provided by the PXI-2532B can be impaired if it is used in a manner not described in this document.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Warranted unless otherwise noted.

Conditions

Specifications are valid at 23 °C unless otherwise noted.

All voltages are specified in DC, ACpk, or a combination unless otherwise specified.

PXI-2532B Pinout

Figure 1. PXI-2532B Pinout


Topology

Table 1. Topologies
Topologies

1-wire 4 x 128 matrix

1-wire 8 x 64 matrix

1-wire 16 x 32 matrix

1-wire dual 4 x 64 matrix

1-wire dual 8 x 32 matrix

1-wire dual 16 x 16 matrix

1-wire quad 4 x 32 matrix

1-wire sixteen 2 x 16 matrix

2-wire 4 x 64 matrix

2-wire 8 x 32 matrix

2-wire 16 x 16 matrix

2-wire dual 4 x 32 matrix

Input

Caution

This module is rated for Measurement Category I and intended to carry signal voltages no greater than 100 V. This module can withstand up to 500 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III, or IV. Do not connect to MAINs supply circuits (for example, wall outlets) of 115 VAC or 230 VAC.

Caution When hazardous voltages (>42.4 Vpk/ 60 VDC) are present on any relay terminal, safety low-voltage (<42.4 Vpk/60 VDC) cannot be connected to any other relay terminal.
Table 2. Maximum Switching Voltage, Current, and Power
Maximum switching voltage, channel-to-channel 100 V
Maximum switching voltage, channel-to-ground 100 V, CAT I
Maximum current 0.5 A (switching or carry, per channel)
Maximum switching power 10 W (per channel)
Note The maximum switching power is limited by the maximum switching current and the maximum voltage, and must not exceed 10 W.
Table 3. DC Path Resistance
Initial <1 Ω, warranted
End-of-life ≥2 Ω
Open channel >1 × 109
Note DC path resistance typically remains low for the life of the relay. At the end of relay life, the path resistance rapidly rises above 2 Ω. Load ratings apply to relays used within the specification before the end of relay life.
Table 4. Thermal EMF
1-wire <50 µV
2-wire <20 µV
Table 5. Bandwidth (-3 dB, 50 Ω termination)
1-wire row/column ≥30 MHz
2-wire row/column ≥25 MHz
Table 6. Channel-to-channel crosstalk (50 Ω termination) Channel-to-channel
10 kHz <-85 dB
100 kHz <-65 dB
1 MHz <-45 dB
Table 7. Open channel isolation (50 Ω termination) Open channel
10 kHz >84 dB
100 kHz >64 dB
1 MHz >44 dB

Dynamic

Table 8. Dynamic
Simultaneous drive limit 64 relays
Relay operate time 0.25 ms
Release time 0.25 ms
Relay life (no load), mechanical 1 × 109 cycles, typical
Note Relay operate time may be longer; certain applications may require additional time for proper settling.
Note Optional series protection resistance, available for the terminal blocks, increases the expected relay life at higher voltages. This series protection resistance shields the reed relays from the effects of cable and load capacitance. For more information, refer to the NI-SWITCH User Manual.
Note Reed relays are highly susceptible to damage caused by switching capacitive and inductive loads. Capacitive loads can cause high inrush currents while inductive loads can cause high flyback voltages. The addition of appropriate resistive protection can greatly improve contact lifetime. For more information about adding protection circuitry to a capacitive load, refer to the NI-SWITCH User Manual. For information about inductive loads, visit ni.com/r/relayflyback.
Note The relays used in the PXI-2532B are field replaceable.
Table 9. Electrical (Resistive, <10 pF Load)
10 V, 100 mA 1 × 107 cycles, typical
20 V, 500 mA 5 × 106 cycles, typical
100 V, 10 mA 5 × 105 cycles, typical

Trigger

Table 10. Input Trigger
Sources PXI trigger lines <0...7>
Minimum pulse width 150 ns
Note The PXI-2532B can recognize trigger pulse widths less than 150 ns if you disable digital filtering. Refer to the NI-SWITCH User Manual for information about disabling digital filtering.
Table 11. Output Trigger
Destinations PXI trigger lines <0...7>
Pulse width Software-selectable: 1 µs to 62 µs

Physical

Table 12. Physical
Relay type Reed
Relay contact material Rhodium
I/O connectors 2, 150 pos, Samtec ERM8-EM series. Terminal block connectivity is with standard 0.050 inch pitch headers.
Dimensions (L × W × H) 3U, one slot, PXI/cPCI module, PXI Express compatible 18.5 cm × 2.0 cm × 13.0 cm (7.3 in. × 0.8 in. × 5.1 in.)
Weight 454 g (1 lb)
Note NI advises against installing reed relay modules directly adjacent to an embedded controller with a magnetic hard drive because of the sensitivity of reed relays and the possibility of interference.
Table 13. PXI Power Requirement
5 V 10 W
3.3 V 2 W

Environment

Table 14. Environment
Operating temperature 0 °C to 55 °C
Storage temperature -20 °C to 70 °C
Relative humidity 5% to 85%, noncondensing
Pollution Degree 2
Maximum altitude 2,000 m

Indoor use only.

Shock and Vibration

Table 15. Shock and Vibration
Operational Shock 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.)
Random vibration, operating 5 Hz to 500 Hz, 0.3 grms
Random vibration, nonoperating 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.)

Compliance and Certifications

Safety Compliance Standards

This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:

  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA C22.2 No. 61010-1
Note For safety certifications, refer to the product label or the Product Certifications and Declarations section.

Electromagnetic Compatibility

This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • EN 55022 (CISPR 22): Class A emissions
  • EN 55024 (CISPR 24): Immunity
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • AS/NZS CISPR 22: Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations.
Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
Note For EMC declarations, certifications, and additional information, refer to Product Certifications and Declarations.

Product Certifications and Declarations

Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.

Environmental Management

NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.