PXI-2531 Specifications
- Updated2023-02-18
- 6 minute(s) read
PXI-2531 Specifications
This document lists specifications for the PXI-2531 . All specifications are subject to change without notice.
PXI-2531
Specifications
Specifications characterize the warranted performance of the instrument under the stated operating conditions. Data in this document are Specifications unless otherwise noted.
Typical Specifications are specifications met by the majority of the instrument under the stated operating conditions and are tested at 23 °C ambient temperature. Typical specifications are not warranted.
All voltages are specified in DC, ACpk, or a combination unless otherwise specified.
Topology
Topologies | 1-wire 4 × 128 matrix 1-wire 8 × 64 matrix 1-wire dual 4 × 64 matrix 1-wire dual 8 × 32 matrix |
Input Characteristics
|
Maximum current (switching or carry, per channel) | 0.5 A |
Maximum switching power (per channel) | 10 W | ||||||||
|
|
RF Performance Characteristics
Bandwidth (-3 dB, 50 Ω termination), 1-wire row/column | ≥20 MHz, typical | ||||||||||
| |||||||||||
|
Dynamic Characteristics
| |||||||
Relay operate time | 0.25 ms |
Release time | 0.25 ms | ||||||||||||||
|
Trigger Characteristics
|
|
Physical Characteristics
Relay type | Reed |
Relay contact material | Rhodium | ||||||
I/O connectors | 2, 150 pos, Samtec ERM8 series | ||||||
| |||||||
Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module, 21.6 × 2.0 × 13.0 cm(8.5 × 0.8 × 5.1 in.) | ||||||
Weight | 454 g (1 lb) |
Environment
Operating temperature | 0 °C to 55 °C |
Storage temperature | -40 °C to 70 °C |
Relative humidity | 5% to 85%, noncondensing |
Pollution Degree | 2 |
Maximum altitude | 2,000 m |
Indoor use only.
Shock and Vibration
Operational Shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | ||||||
|
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for sensitive electrical equipment for measurement, control, and laboratory use:
- EN 61326-2-1 (IEC 61326-2-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.