CAN Physical Layers
Transceiver | NXP PCA82C251 |
Max baud rate | 1 Mb/s |
Min baud rate | 10 Kb/s |
CAN_H, CAN_L bus lines | –36 to +36 VDC |
Parameter | Min | Max |
Output0, Output1 | ||
VOH (IOH = –8 mA) | 3.8 V | — |
VOL (IOL = 8 mA) | — | 0.44 V |
NERR | ||
VIH | 2.0 V | — |
VIL | — | 0.8 V |
Physical Dimensions
Dimensions (PCI devices) | 10.67 x 16.76 cm (4.2 x 6.6 in.) |
Dimensions (PXI devices) | 10.00 x 16.00 cm (3.9 x 6.3 in.) |
I/O connector | 9-pin male D-SUB for each port |
Power Requirements
+5 VDC (±5%) | 640 mA typical |
+3.3 VDC (±5%) | 940 mA typical |
Shock (PXI Devices)
Operating | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC-60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) |
Random Vibration (PXI Devices)
Operating | 5 to 500 Hz, 0.3 grms |
Non-operating | 5 to 500 Hz, 2.4 grms (Tested in accordance with IEC-60068-2-64. Non-operating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Safety
Withstand | 500 Vrms, verified by a 5 s dielectric withstand test |
Continuous | 60 VDC, Measurement Category I |
Withstand | 500 Vrms, verified by a 5 s dielectric withstand test |
Continuous | 60 VDC, Measurement Category I |
Safety Standards
This product meets the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA 61010-1
For UL and other safety certifications, refer to the product label or the Online Product Certification section.
Environmental
Operating temperature | 0 to 55 °C |
Storage temperature | –20 to 70 °C (Tested in accordance with IEC-60068-2-1 and IEC-60068-2-2.) |
Operating humidity | 10 to 90% RH, noncondensing |
Storage humidity | 5 to 95% RH, noncondensing (Tested in accordance with IEC-60068-2-56.) |
Maximum altitude | 2000 m |
Pollution Degree (IEC 60664) | 2 |
Indoor use only.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
- EN 61326 (IEC 61326): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- AS/NZS CISPR 11: Group 1, Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance 
This product meets the essential requirements of applicable European Directives as follows:
- 2006/95/EC; Low-Voltage Directive (safety)
- 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.