Measuring Current
- Updated2025-10-10
- 2 minute(s) read
This section describes how to measure current on NI DMMs.
Measuring Current Within the Normal Range
To configure current measurements within the normal range on the DMM, use one of the following two methods:
Method 1
- Select the current measurement function.
- Take a reading from the DMM immediately.
Method 2
- Configure the DMM for a software trigger.
- Initiate the DMM for a current measurement to cause the input relay to switch.
- Send a software trigger to the DMM to take a measurement.
Using Current Shunt Modules
When taking current measurements outside of the specified range on the DMM, you must use a current shunt module. NI offers the NI CSM-10A and the NI CSM-200mA current shunt modules. The NI CSM-10A has a 0.01 Ω sense resistor, and the NI CSM-200mA has a 1.0 Ω sense resistor.
Each current shunt module (CSM) operates by passing the input current through a precise resistor. The following figure shows the internal circuitry of the CSM:
- Voltage Across Resistor (V OUT)
- Precision Resistor
- Fuse
- Current Source (I IN)
To configure current measurements using the NI CSM-200mA or the NI CSM-10A, complete the following steps:
- Measure the voltage drop across the resistor of the CSM.
- Use this value to calculate the current using Ohm's Law, as
follows:
I IN = ( V OUT)/ R
where
I is the input current
V is the voltage across the precision resistor
R is the resistance of the precision resistor
For example, assume that you are using the CSM-200mA, which has a 1 Ω precision resistor, and the measured voltage is 50 mV. Apply these values to Ohm's Law to determine the current, as follows:
50 mA = 50 mV / 1 Ω