NI-9411 Specifications
- Updated2024-11-05
- 4 minute(s) read
NI-9411 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to COM unless otherwise noted.
NI 9411 Pinout
Signal | Description |
---|---|
COM | Common reference connection to isolated ground |
DIa and DIb | Digital input signal connections |
Supply (+5 Vout) | 5 V power output connection for external devices |
Signal | Description |
---|---|
COM | Common reference connection to isolated ground |
Vsup | Voltage supply connection |
Input Characteristics
Number of channels | 6 digital input channels | ||||||||||||||||||||||||||||
Input type | Differential or single-ended | ||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
Input impedance | 8.4 kΩ | ||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||
Input delay time | 500 ns maximum | ||||||||||||||||||||||||||||
MTBF | 800,319 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method |
Power Requirements
| |||||||
|
Physical Characteristics
Dimensions | Visit ni.com/dimensions and search by module number. | ||||||||||||||
| |||||||||||||||
| |||||||||||||||
Weight | 136 g (4.8 oz) |
Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-COM or Vsup-to-COM | 30 V maximum, Measurement Category I | ||||||||||||
|
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 2,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.