NI-9263 Specifications
- Updated2023-04-03
- 4 minute(s) read
NI-9263 Specifications
NI-9263
Nomenclature
In this article, the NI-9263 with screw terminal and NI-9263 with spring terminal are referred to inclusively as the NI-9263.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to COM unless otherwise noted.
Safety Voltages
Connect only voltages that are within the following limits:
Channel-to-channel | None | ||||||
|
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
| |||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Power Requirements
| |||||||
|
Physical Characteristics
Dimensions | Visit ni.com/dimensions and search by module number. | ||||||||||||||
| |||||||||||||||
| |||||||||||||||
| |||||||||||||||
|
Output Characteristics
Measurement Conditions | Percent of Reading (Gain Error) | Percent of Range[3] (Offset Error) | |
---|---|---|---|
Calibrated | Maximum (-40 °C to 70 °C) | 0.35% | 0.75% |
Typical (25 °C, ±5 °C) | 0.03% | 0.1% | |
Uncalibrated[4] | Maximum (-40 °C to 70 °C) | 2.2% | 1.7% |
Typical (25 °C, ±5 °C) | 0.3% | 0.25% |
| |||||||
|
Number of Channels | Update Time for All Other Chassis | Update Time for NI cRIO-9151 R Series Expansion Chassis |
---|---|---|
1 | 3 μs min | 3.5 μs min |
2 | 5 μs min | 6.5 μs min |
3 | 7.5 μs min | 9 μs min |
4 | 9.5 μs min | 12 μs min |
| |||||||||
Slew rate | 4 V/μs | ||||||||
Crosstalk | 76 dB | ||||||||
| |||||||||
Capacitive drive | 1,500 pF minimum | ||||||||
Monotonicity | 16 bits | ||||||||
DNL | ±1 LSB maximum | ||||||||
INL (endpoint) | ±12 LSB maximum | ||||||||
MTBF | 1,732,619 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method |
Calibration
Calibration interval | 1 year |
1 When the module powers on, a glitch occurs for 20 μs peaking at -1.5 V.
2 The power-down voltage peaks at 1.8 V before exponentially discharging to 0 V in 100 μs. You can add a 10 kΩ load to reduce the peak voltage.
3 Range equals ±10.7 V
4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.