NI-9244 Specifications
- Updated2024-11-11
- 5 minute(s) read
NI-9244 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted.
NI-9244 Pinout
Signal | Description |
---|---|
AI | Analog input signal connection referenced to the Neutral signal |
Neutral | Referenced, single-ended analog input connection |
Input Characteristics
Scaling coefficient | 118,911 nV/LSB | ||||||
Number of channels | 4 analog input channels | ||||||
ADC resolution | 24 bits | ||||||
Type of ADC | Delta-Sigma (with analog prefiltering) | ||||||
Sampling mode | Simultaneous | ||||||
| |||||||
| |||||||
| |||||||
Data rates (fs)[1]1 The data rate must remain within the appropriate data rate range. |
| ||||||
| |||||||
Overvoltage withstand | 800 Vrms continuous, 1000 Vrms for 1 s | ||||||
Surge withstand | 8 kV (1.2 µs/50 µs) | ||||||
Input coupling | DC | ||||||
Input impedance, AIx-to-Ground and Neutral-to-Ground | 2 MΩ |
Measurement Conditions | Percent of Reading (Gain Error) | Percent of Range (Offset Error)[2]2 Range equals 565.7 V (400 Vrms) | |
---|---|---|---|
Calibrated | Maximum, (-40 °C to 70 °C) | 0.17% | 0.14% |
Typical, (23 °C±5 °C) | 0.043% | 0.012% |
Nonlinearity (at 25 °C) | 24 ppm | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Phase mismatch (channel-to-channel) | 0.202°/ kHz maximum | ||||||||||||||||||
Phase mismatch (module-to-module, maximum) | 0.202°/kHz + 360° * fin/ fM | ||||||||||||||||||
| |||||||||||||||||||
Input delay | 40 | ||||||||||||||||||
Passband Frequency | 0.453 * fs | ||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
| |||||||||||||||||||
Alias-free bandwidth | 0.453 * fs | ||||||||||||||||||
Anti-alias rejection (fs = 50 kS/s) | 50 dB | ||||||||||||||||||
-3 dB bandwidth (fs = 50 kS/s) | 0.49 * fs | ||||||||||||||||||
| |||||||||||||||||||
CMRR (fin = 60 Hz) | -78 dB | ||||||||||||||||||
SFDR (1 kHz, -60 dBFS) | -120 dB | ||||||||||||||||||
Total Harmonic Distortion (THD), up to 1 kHz | -98 dB |
Power Requirements
| |||||||
|
Physical Characteristics
Screw-terminal wiring
Gauge | 0.2 mm 2 to 3.0 mm2 (24 AWG to 12 AWG) copper conductor wire | ||||||
Wire strip length | 7 mm (0.28 in.) of insulation stripped from the end | ||||||
Temperature rating | 90 °C minimum | ||||||
Torque for screw terminals | 0.5 N · m to 0.6 N · m (4.4 lb · in. to 5.3 lb · in.) | ||||||
Wires per screw terminal | One wire per screw terminal | ||||||
Ferrules | 0.25 mm2 to 2.5 mm2 | ||||||
Weight | 150 g (5.3 oz) | ||||||
|
Safety Voltages
Connect only voltages that are within the following limits:
Maximum working voltage | 400 V RMS L-N, 800 V RMS L-L | ||||||||
| |||||||||
Overvoltage withstand | 800 V RMS continuous, 1,000 V RMS for 1 s | ||||||||
| |||||||||
|
Measurement Categories
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Measurement Category III
Measurement Category III is for measurements performed in the building installation at the distribution level. This category refers to measurements on hard-wired hardware such as hardware in fixed installations, distribution boards, and circuit breakers. Other examples are wiring, including cables, bus bars, junction boxes, switches, socket outlets in the fixed installation, and stationary motors with permanent connections to fixed installations.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 5,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Calibration
Calibration interval | 1 year |