NI-9235 and sbRIO-9236 Specifications
- Updated2024-05-02
- 7 minute(s) read
NI-9235 and sbRIO-9236 Specifications
Connector Types
The NI-9235 is available in two types: push-in spring terminal and spring terminal. The push-in type spring terminal connector is black and orange. The spring terminal connector is black. NI-9235 refers to both types unless the two types are specified. Differences between the two types of spring terminal connectors are noted by the connector color.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted.
Input Characteristics
Number of channels | 8 analog input channels | ||||||
Quarter-bridge completion | 120 Ω, 10 ppm/°C maximum | ||||||
ADC resolution | 24 bits | ||||||
Type of ADC | Delta-Sigma (with analog prefiltering) | ||||||
Sampling mode | Simultaneous | ||||||
| |||||||
| |||||||
| |||||||
Data rates[1]1 The data rate must remain within the appropriate data rate range. (fs) | (fM ÷ 256)/n, n = {2; 4, 5, ..., 63} | ||||||
Full-scale range | ±29.4 mV/V (+62,500 με/-55,500 με) | ||||||
Scaling coefficient | 3.5062 nV/V per LSB | ||||||
Overvoltage protection between any two terminals | ±30 V |
Measurement Conditions | Percent of Reading[2]2 Excusive of lead wire resistance error. (Gain Error) | Percent of Range[3]3 Range equals 29.4 mV/V.,[4]4 Calibrated errors represent offset stability following unstrained measurement. Errors include the effect of completion resistor tolerance and drift. (Offset Error) | ||
---|---|---|---|---|
30 days after cal. (±5 °C) | 1 year after cal. (±5 °C) | |||
Calibrated | Typical (25 °C, ±5 °C) | 0.02% | 0.1% | 0.15% |
Maximum (-40 °C to 70 °C) | 0.07% | 0.17% | 0.4% | |
Uncalibrated[5]5 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. | Typical (25 °C, ±5 °C) | 0.15% | 1.25% | |
Maximum (-40 °C to 70 °C) | 0.53% | 2.14% |
|
Input Signal Frequency (fin) | Gain | Phase | |
---|---|---|---|
Typical | Maximum | Maximum | |
0 to 1 kHz | 0.08% | 0.11% | 0.34°/kHz * fin |
0 to 4 kHz | 0.17% | 0.32% |
| |||||||
Input delay | (39 + 221/1024)/fs + 12 μs | ||||||
| |||||||
| |||||||
Alias-free bandwidth | 0.45 * fs | ||||||
Oversample rate | 64 * fs | ||||||
Rejection at oversample rate[6]6 Rejection by analog prefilter of signal frequencies at oversample rate.(fs = 10 kS/s) | 80 dB @ 640 kHz | ||||||
| |||||||
SFDR (1 kHz, -60 dBFS) | 110 dB | ||||||
THD (1 kHz, -20 dBFS) | -90 dB | ||||||
Crosstalk (fin = 1 kHz) | -100 dB | ||||||
Common-mode voltage, all signals to earth ground | ±60 VDC | ||||||
CMRR (fin = 0 to 60 Hz) | 120 dB | ||||||
MTBF | 566,796 hours at 25 °C; Bellcore Issue 2, Method 1, Case 3, Limited Part Stress Method |
Shunt Calibration Characteristics
Measurement Conditions | Percent of Reading (Gain Error) |
---|---|
Typical (25 °C, ±5 °C) | 0.09% |
Maximum (-40 °C to 70 °C) | 0.22% |
Resistance | 50 kΩ |
Output value | -599.28 μV/V |
Temperature drift | 15 ppm/°C |
Method | Shunt across completion resistor |
Excitation Characteristics
Excitation type | Constant voltage |
Excitation value | 2.0 V ± 1% |
Maximum output current | 80 mA |
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 5,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Safety Voltages
Connect only voltages that are within the following limits:
Between any two terminals | ±30 V maximum | ||||||||||||
|
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Power Requirements
| |||||||
|
Physical Characteristics
| |||||||
Dimensions | Visit ni.com/dimensions and search by module number. |
(Black Connector)
The NI-9235 (black connector) requires a flathead screwdriver with a 2.3 mm × 1.0 mm (0.09 in. × 0.04 in.) blade for signal connection; insert the screwdriver into a spring clamp activation slot to open the corresponding connector terminal, press a wire into the open connector terminal, and then remove the screwdriver from the activation slot to clamp the wire into place.
| |||||||||||
|
Push-In
Style (Black/Orange Connector)
The push-in spring style NI-9235 does not require a tool for signal connection; push the wire into the terminal when using solid wire or stranded wire with a ferrule, or by pressing the push button when using stranded wire without a ferrule.
| |||||||||||
| |||||||||||
|
Calibration
Calibration interval | 1 year |
1 The data rate must remain within the appropriate data rate range.
2 Excusive of lead wire resistance error.
3 Range equals 29.4 mV/V.
4 Calibrated errors represent offset stability following unstrained measurement. Errors include the effect of completion resistor tolerance and drift.
5 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.
6 Rejection by analog prefilter of signal frequencies at oversample rate.