NI-9228 Specifications
- Updated2024-11-11
- 7 minute(s) read
NI-9228 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted.
NI-9228
Pinout
Signal | Description |
---|---|
AI+ | Positive analog input signal connection |
AI- | Negative analog input signal connection |
Safety Voltages
Connect only voltages that are within the following limits:
| |||||||||||||||||||||
|
Measurement Categories
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 5,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Power Requirements
| |||||||
|
Physical Characteristics
If you need to clean the module, wipe it with a dry towel.
Dimensions | Visit ni.com/dimensions and search by module number. | ||||||||||||||
| |||||||||||||||
| |||||||||||||||
Weight | 152 g (5.36 oz) |
Input Characteristics
Number of channels | 8 analog input channels |
ADC resolution | 24 bits |
Type of ADC | Delta-Sigma |
Sampling mode | Simultaneous |
Timing Mode | Conversion Time (ms) | Sample Rate (S/s) |
---|---|---|
High resolution | 500 | 2 |
Medium resolution | 83.3 | 12 |
Medium speed | 10 | 100 |
High speed | 1 | 1000 |
| |||||||
Overvoltage protection, AI+ to AI- | 250 Vrms | ||||||
Input coupling | DC |
Measurement Conditions | Percent of Reading (Gain Error) | Percent of Range[1]1 Range equals 63.8 V (Offset Error) | |
---|---|---|---|
Calibrated | Typical (25 °C, ±5 °C) | ±0.11% | ±0.02% |
Maximum (-40 °C to 70 °C) | ±0.40% | ±0.06% |
| |||||||||||
| |||||||||||
Post-calibration gain match (fin = 100 Hz) | ±76 mdB maximum | ||||||||||
Phase mismatch, channel-to-channel (fin = 100 Hz) | ±0.002 °/Hz maximum | ||||||||||
|
| |||||||||||
|
| |||||||
Crosstalk (fin = 1 kHz) | -120 dB | ||||||
Alias hole rejection ( (fin = 614.4 kHz) | 75 dB |
Calibration
Calibration interval | 2 years |
1 Range equals 63.8 V