NI-9220 Specifications

NI-9220 Specifications

These specifications apply to the NI-9220.

Revision History

Version Date changed Description
378905B-01 February 2025 Updated accuracy table.
378905A-01 June 2023 Initial release.

Looking For Something Else?

For information not found in the specifications for your product, such as operating instructions, browse Related Information.

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.

Specifications are Typical unless otherwise noted.

Conditions

Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted.

Connector Types

The NI-9220 has more than one connector type: NI-9220 with spring terminal and NI-9220 with DSUB. Unless the connector type is specified, NI-9220 refers to all connector types.

The NI-9220 with spring terminal is available in two types: push-in spring terminal and spring terminal. The push-in type spring terminal connector is black and orange. The spring terminal connector is black. NI-9220 with spring terminal refers to both types unless the two types are specified. Differences between the two types of spring terminal connectors are noted by the connector color.

NI-9220 Spring Terminal Pinout



Table 1. Signal Descriptions
Signal Description
AI+ Positive analog input signal connection
AI- Negative analog input signal connection
COM Common reference connection to isolated ground

NI-9220 Push-In Spring Terminal Pinout



Table 2. Signal Descriptions
Signal Description
AI+ Positive analog input signal connection
AI- Negative analog input signal connection
COM Common reference connection to isolated ground

NI-9220 DSUB Terminal Pinout



Table 3. Signal Descriptions
Signal Description
AI+ Positive analog input signal connection
AI- Negative analog input signal connection
COM Common reference connection to isolated ground

Input Characteristics

Number of channels 16 analog input channels
ADC resolution 16 bits
Type of ADC Successive approximation register (SAR)
Table 4. Input Voltage Ranges
Measurement voltage (AI+ to AI-) Minimum[1]1 The minimum measurement voltage range is the largest voltage the NI-9220 is guaranteed to accurately measure. ±10.4 V
Typical ±10.5 V
Maximum ±10.6 V
Maximum voltage (Signal + Common Mode) Each channel must remain within ±10.4 V of common
Overvoltage protection ±30 V
Conversion time 10 μs minimum
Sample rate 100 kS/s maximum
Table 5. Accuracy
Measurement Conditions Within 1 Year of Calibration Within 10 Years of Calibration
Percent of Reading (Gain Error) Percent of Range[2]2 Range equals ±10.5 V. (Offset Error) Percent of Reading (Gain Error) Percent of Range[3]3 Range equals ±10.5 V. (Offset Error)
Calibrated Maximum (-40 °C to 70 °C) 0.142% ±0.070% 0.226% 0.174%
Typical (23 °C ±5 °C) 0.010% ±0.011% 0.047% 0.063%
Uncalibrated[4]4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. Maximum (-40 °C to 70 °C) 0.350% ±0.360% 0.434% 0.464%
Typical (23 °C ±5 °C) 0.060% ±0.070% 0.097% 0.122%
Note NI recommends a calibration interval of 1 year and only provides calibration services referencing the 1-year specification limits. If you choose to calibrate less often, the 10-year column provides predicted performance over this extended interval. Choose an appropriate interval based on your application requirements. Longer calibration intervals are more likely to result in As-Found calibration failures when the device is sent back for calibration and compared against the 1-year specification limits.
Table 6. Stability Drift
Gain drift 5 ppm/°C
Offset drift 29 μV/°C
Common-mode rejection ratio (CMRR) (fin = 60 Hz) 70 dB
-3 dB bandwidth >100 kHz
Input impedance >1 GΩ
Input noise 0.85 LSBrms
Crosstalk -90 dB
Table 7. Settling Time (to 2 LSBs)
10 V step 19 μs
20 V step 26 μs
No missing codes 15 bits
MTBF 1,522,250 at 25 °C; Bellcore Issue 6, Method 1, Case 3, Limited Part Stress Method

NI-9220 with Spring Terminal (Black Connector)

Connect only voltages that are within the following limits.

Channel-to-COM

±30 V maximum

Isolation

Channel-to-COM

None

Channel-to-earth ground

Continuous

250 V RMS, Measurement Category II

Withstand

3,000 V RMS, verified by a 5 s dielectric withstand test

NI-9220 with Push-In Spring Terminal (Black/Orange Connector)

Connect only voltages that are within the following limits.

Channel-to-channel

None

Channel-to-earth ground

Continuous

250 V RMS, Measurement Category II

Withstand up to 4,000 m

3,000 V RMS, verified by a 5 s dielectric withstand test

NI-9220 with DSUB Safety Voltages

Connect only voltages that are within the following limits.

Channel-to-COM

±30 V maximum

Isolation

Channel-to-COM

None

Channel-to-earth ground

Continuous

60 V DC, Measurement Category I

Withstand up to 2,000 m

1,000 V RMS, verified by a 5 s dielectric withstand test

Measurement Category

Measurement Category I

Caution Do not connect the NI-9220 with DSUB to signals or use for measurements within Measurement Categories II, III, or IV.
Attention Ne pas connecter le NI-9220 with DSUB à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.
Warning Do not connect the NI-9220 with DSUB to signals or use for measurements within Measurement Categories II, III, or IV, or for measurements on MAINs circuits or on circuits derived from Overvoltage Category II, III, or IV which may have transient overvoltages above what the product can withstand. The product must not be connected to circuits that have a maximum voltage above the continuous working voltage, relative to earth or to other channels, or this could damage and defeat the insulation. The product can only withstand transients up to the transient overvoltage rating without breakdown or damage to the insulation. An analysis of the working voltages, loop impedances, temporary overvoltages, and transient overvoltages in the system must be conducted prior to making measurements.
Mise en garde Ne pas connecter le NI-9220 with DSUB à des signaux dans les catégories de mesure II, III ou IV et ne pas l'utiliser pour des mesures dans ces catégories, ou des mesures sur secteur ou sur des circuits dérivés de surtensions de catégorie II, III ou IV pouvant présenter des surtensions transitoires supérieures à ce que le produit peut supporter. Le produit ne doit pas être raccordé à des circuits ayant une tension maximale supérieure à la tension de fonctionnement continu, par rapport à la terre ou à d'autres voies, sous peine d'endommager et de compromettre l'isolation. Le produit peut tomber en panne et son isolation risque d'être endommagée si les tensions transitoires dépassent la surtension transitoire nominale. Une analyse des tensions de fonctionnement, des impédances de boucle, des surtensions temporaires et des surtensions transitoires dans le système doit être effectuée avant de procéder à des mesures.

Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.

Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

Measurement Category II

Caution Do not connect the NI-9220 with spring terminal to signals or use for measurements within Measurement Categories III or IV.
Attention Ne pas connecter le NI-9220 with spring terminal à des signaux dans les catégories de mesure III ou IV et ne pas l'utiliser pour effectuer des mesures dans ces catégories.

Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.

Environmental Characteristics

Temperature

Operating

-40 °C to 70 °C

Storage

-40 °C to 85 °C

Humidity

Operating

10% RH to 90% RH, noncondensing

Storage

5% RH to 95% RH, noncondensing

Ingress protection

IP40

Pollution Degree

2

Maximum altitude

NI-9220 with spring terminal (black connector)

2,000 m

NI-9220 with push-in spring terminal (black/orange connector)

4,000 m

NI-9220 with DSUB

2,000 m

Shock and Vibration
Operating vibration

Random

5 g RMS, 10 Hz to 500 Hz

Sinusoidal

5 g, 10 Hz to 500 Hz

Operating shock

30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations

To meet these shock and vibration specifications, you must panel mount the system.

Power Requirements

Power consumption from chassis (full-scale input, 100 kS/s)

Active mode

1 W maximum

Sleep mode

4 mW maximum

Thermal dissipation (at 70 °C)

Active mode

1.250 W maximum

Sleep mode

510 mW maximum

Physical Characteristics

Weight

NI-9220 with spring terminal (black connector)

143 g (5.0 oz)

NI-9220 with push-in spring terminal (black/orange connector)

148 g (5.2 oz)

NI-9220 with DSUB

147 g (5.2 oz)

NI-9220 with Spring Terminal (Black Connector)

Spring terminal wiring

Gauge

0.08 mm2 to 1.0 mm2 (28 AWG to 18 AWG) copper conductor wire

Wire strip length

7 mm (0.28 in.) of insulation stripped from the end

Temperature rating

90 °C minimum

Wires per spring-terminal

One wire per spring terminal

Connector securement

Securement type

Screw flanges provided

Torque for screw flanges

0.2 N · m (1.80 lb · in.)

NI-9220 with Push-In Spring Terminal (Black/Orange Connector)

Spring terminal wiring

Gauge

0.14 mm2 to 1.5 mm2 (26 AWG to 16 AWG) copper conductor wire

Wire strip length

10 mm (0.394 in.) of insulation stripped from the end

Temperature rating

90 °C minimum

Wires per spring-terminal

One wire per spring-terminal; two wires per spring terminal using a 2-wire ferrule

Ferrules

0.14 mm2 to 1.5 mm2

Connector securement

Securement type

Screw flanges provided

Torque for screw flanges

0.2 N · m (1.80 lb · in.)

Calibration

You can obtain the calibration certificate and information about calibration services for the NI-9220 at ni.com/calibration.
Recommended calibration interval 1 year

1 The minimum measurement voltage range is the largest voltage the NI-9220 is guaranteed to accurately measure.

2 Range equals ±10.5 V.

3 Range equals ±10.5 V.

4 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.