NI-9215 and sbRIO-9215 Specifications
- Updated2024-05-02
- 6 minute(s) read
NI-9215 and sbRIO-9215 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted.
Connector Types
The NI-9215 has more than one connector type: NI-9215 with screw terminal, NI-9215 with spring terminal, and NI-9215 with BNC. Unless the connector type is specified, NI-9215 refers to all connector types.
Input Characteristics
Number of channels | 4 analog input channels | ||||||||||||||||||||||
ADC resolution | 16 bits | ||||||||||||||||||||||
Type of ADC | Successive approximation register (SAR) | ||||||||||||||||||||||
Input range | ±10.0 V | ||||||||||||||||||||||
| |||||||||||||||||||||||
Overvoltage protection | ±30 V | ||||||||||||||||||||||
|
Measurement Conditions | Percent of Reading (Gain Error) | Percent of Range[2]2 Range equals ±10.4 V. (Offset Error) | |
---|---|---|---|
Calibrated | Maximum (-40 °C to 70 °C) | 0.2% | 0.082% |
Typical (23 °C ±5 °C) | 0.02% | 0.014% | |
Uncalibrated[3]3 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. | Maximum (-40 °C to 70 °C) | 1.05% | 0.82% |
Typical (23 °C ±5 °C) | 0.6% | 0.38% |
| |||||||||||||||||||||||||||
CMRR (fin = 60 Hz) | 73 dB min | ||||||||||||||||||||||||||
Input bandwidth (-3 dB) | 420 kHz minimum | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
Input bias current | 10 nA | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
Crosstalk | -80 dB | ||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
No missing codes | 15 bits guaranteed | ||||||||||||||||||||||||||
DNL | -1.9 to 2 LSB | ||||||||||||||||||||||||||
INL | ±6 LSB maximum | ||||||||||||||||||||||||||
MTBF | 1,167,174 hours at 25 °C; Bellcore Issue 6, Method 1, Case 3, Limited Part Stress Method |
Power Requirements
| |||||||
|
Physical Characteristics
| |||||||||||||||
| |||||||||||||||
| |||||||||||||||
|
NI-9215 with Screw Terminal and NI-9215 with Spring Terminal Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-COM | ±30 V maximum | ||||||||||||
|
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
NI-9215 with BNC Safety Voltages
Connect only voltages that are within the following limits.
AI+-to-AI- | ±30 V maximum | ||||||||||||
|
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 2,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Calibration
Calibration interval | 1 year |
1 The minimum measurement voltage range is the largest voltage the NI-9215 is guaranteed to accurately measure.
2 Range equals ±10.4 V.
3 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data.