In this document, the NI-9215 with screw terminal, the NI-9215 with spring terminal, and the NI-9215 with BNC are referred to inclusively as the NI-9215. The information in this document applies to all versions of the NI-9215 unless otherwise specified.
NI-9215 Block Diagram
- Input signals on each channel are buffered, conditioned, and then sampled by an ADC.
- Each AI channel provides an independent track-and-hold amplifier, enabling you to sample all channels simultaneously.
NI-9215 Pinout
Signal | Description |
---|---|
AI+ | Positive analog input signal connection |
AI- | Negative analog input signal connection |
COM | Common reference connection to isolated ground |
NC | No connection |
NI-9215
Grounded Differential Connections
Floating Differential Connections
- Connect the negative lead to COM through a 1 MΩ resistor to keep the signal source within the common-mode voltage range.
- The NI-9215 with screw terminal and the NI-9215 with spring terminal does not read data accurately if the signal source is outside of the common-mode voltage range.
The NI-9215 with BNC has internal circuitry that maintains the common mode voltage range.
Single-Ended Connections
Connect the ground signal to COM to keep the signal source within the common-mode voltage range.
The NI-9215 with BNC has internal circuitry that maintains the common mode voltage range.
Connection Guidelines
- Make sure that devices you connect to the NI-9215 are compatible with the module specifications.
- You must use 2-wire ferrules to create a secure connection when connecting more than one wire to a single terminal on the NI-9215 with screw terminal or NI-9215 with spring terminal.
High-Vibration Application Connections
If your application is subject to high vibration, NI recommends that you follow these guidelines to protect connections to the
NI-9215 with screw
terminal or the
NI-9215 with spring
terminal:
- Use ferrules to terminate wires to the detachable connector.
- Use the NI-9927 backshell kit with the NI-9215 with screw terminal or the NI-9981 backshell kit with the NI-9215 with spring terminal.