NI-9209 Specifications
- Updated2024-11-19
- 6 minute(s) read
NI-9209 Specifications
Connector Types
The NI-9209 has more than one connector type: NI-9209 with spring terminal and NI-9209 with DSUB. Unless the connector type is specified, NI-9209 refers to all connector types.
The NI-9209 with spring terminal is available in two types: push-in spring terminal and spring terminal. The push-in type spring terminal connector is black and orange. The spring terminal connector is black. NI-9209 with spring terminal refers to both types unless the two types are specified. Differences between the two types of spring terminal connectors are noted by the connector color.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to COM unless otherwise noted.
NI-9209 Pinout
Signal | Description |
---|---|
AI | Analog input signal connection |
COM | Common reference connection to isolated ground |
NC | No connection |
NI-9209
Signals
You can connect single-ended or differential signals to the NI-9209; use a differential measurement configuration to attain more accurate measurements and less noise. Specific signal pairs are valid for differential connections.
The following table shows the signal pairs that are valid for differential connection configurations with the NI-9209.
Channel | AI+ | AI- |
---|---|---|
0 | AI0 | AI8 |
1 | AI1 | AI9 |
2 | AI2 | AI10 |
3 | AI3 | AI11 |
4 | AI4 | AI12 |
5 | AI5 | AI13 |
6 | AI6 | AI14 |
7 | AI7 | AI15 |
16 | AI16 | AI24 |
17 | AI17 | AI25 |
18 | AI18 | AI26 |
19 | AI19 | AI27 |
20 | AI20 | AI28 |
21 | AI21 | AI29 |
22 | AI22 | AI30 |
23 | AI23 | AI31 |
Input Characteristics
Number of channels | 16 differential/32 single-ended channels |
ADC resolution | 24 bits |
Type of ADC | Delta-Sigma |
Sampling mode | Scanned |
Calibrated Measurement Conditions | Percent of Reading (Gain Error) | Percent of Range[1]1 Range equals 10.4 V(Offset Error) |
---|---|---|
Maximum (-40 °C to 70 °C) | ±0.46% | ±0.011% |
Typical (25 °C ±5 °C) | ±0.06% | ±0.003% |
| |||||||
Maximum working voltage for analog inputs (signal voltage + common mode voltage) | Each channel must remain within ±10.2 V of common | ||||||
| |||||||
Overvoltage protection, channel-to-COM | ±30 V maximum on one channel at a time | ||||||
Input impedance | >1 GΩ | ||||||
| |||||||
| |||||||
| |||||||
CMRR (fin= 0 Hz to 60 Hz) | 68 dB | ||||||
| |||||||
|
Power Requirements
| |||||||
|
Physical Characteristics
| |||||||
Dimensions | Visit ni.com/dimensions and search by module number. |
with Spring Terminal
| |||||||||||||
|
NI-9209 with
Spring Terminal Safety Voltages
Connect only voltages that are within the following limits:
|
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
NI-9209 with
DSUB Safety Voltages
Connect only voltages that are within the following limits:
|
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 5,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Calibration
Calibration interval | 2 years |