NI-9207 Specifications
- Updated2024-11-19
- 6 minute(s) read
NI-9207 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Typical unless otherwise noted.
Conditions
Specifications are valid for the range -40 °C to 70 °C unless otherwise noted. All voltages are relative to COM unless otherwise noted.
Connector Types
The NI-9207 has more than one connector type: NI-9207 with spring terminal and NI-9207 with DSUB. Unless the connector type is specified, NI-9207 refers to all connector types.
NI-9207
Pinout
Signal | Description |
---|---|
AI+ | Positive analog input voltage connection |
AI- | Negative analog input signal connection |
AI | Analog input current connection |
Vsup | Voltage supply connection |
COM | Common reference connection to isolated ground |
Input Characteristics
Number of channels | 16 analog input channels: 8 voltage and 8 current | ||||||||||||||||||
ADC resolution | 24 bits | ||||||||||||||||||
Type of ADC | Delta-Sigma | ||||||||||||||||||
Sampling mode | Scanned | ||||||||||||||||||
| |||||||||||||||||||
Maximum working voltage for analog inputs (signal voltage + common mode voltage), voltage channels only | Each channel must remain within ±10.2 V of common | ||||||||||||||||||
| |||||||||||||||||||
Overvoltage protection, channel-to-COM, all channels | ±30 V maximum on one channel at a time | ||||||||||||||||||
| |||||||||||||||||||
|
Calibrated Measurement Conditions | Channels | Percent of Reading (Gain Error) | Percent of Range[1]1 Range equals 10.4 V for voltage channels and 22.0 mA for current channels.(Offset Error) |
---|---|---|---|
Maximum (-40 °C to 70 °C) | Voltage channels | ±0.52% | ±0.04% |
Current channels | ±0.87% | ±0.05% |
| |||||||||||||||||||
| |||||||||||||||||||
CMRR (fin= 0 Hz to 60 Hz), voltage channels only | 86 dB | ||||||||||||||||||
| |||||||||||||||||||
|
NI-9207 with
Spring Terminal Safety Voltages
Connect only voltages that are within the following limits:
|
NI-9207 with
DSUB Safety Voltages
Connect only voltages that are within the following limits:
|
Measurement Category
Measurement Category I
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics.
Measurement Category II
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Environmental Characteristics
| |||||||||||||
| |||||||||||||
Ingress protection | IP40 | ||||||||||||
Pollution Degree | 2 | ||||||||||||
Maximum altitude | 5,000 m | ||||||||||||
|
To meet these shock and vibration specifications, you must panel mount the system.
Power Requirements
| |||||||
|
Physical Characteristics
| |||||||||||||
Dimensions | Visit ni.com/dimensions and search by module number. | ||||||||||||
| |||||||||||||
|
Calibration
Calibration interval | 2 years |