LCR Devices
- Updated2025-10-03
- 7 minute(s) read
Refer to the following sections for information about setting up an LCR instrument in InstrumentStudio.
Calibrating an LCR Device
Self-calibrate an LCR device to ensure accurate measurements.
-
In the LCR device instrument pane, open the instrument header menu (
).
- Click .
- For InstrumentStudio 2025 Q2 and earlier, click Self-calibrate in MAX to proceed to self-calibration in Measurement & Automation Explorer (MAX). Later versions of InstrumentStudio perform self-calibration without MAX.
When to Calibrate an LCR Device
Performing regular LCR device calibration ensures better accuracy in your test measurements.
- You are installing the module in a chassis for the first time.
- You install, uninstall, or move another module in the same chassis as your LCR device.
- The ambient temperature in your test environment has changed significantly.
- The onboard temperature of the LCR device has drifted significantly (±5 °C) since the last self-calibration.
- 24 hours have elapsed since the previous self-calibration.
Performing Compensation on an LCR Device
Typically, you connect an LCR device to a device under test with switches, fixtures, or cables. These devices introduce undesired errors into the measurement. Prevent measurement errors from switches, fixtures, or cables by applying compensation before taking measurements.
- Calibrate the LCR device.
-
Determine the type of cable compensation you want to perform. Click the option
link for detailed instructions on performing the selected compensation.
Option Recommended When Standard cable compensation You use a standard NI cable that is represented with a Cable Length profile in the Instrument Settings dialog (
) and a
simple test setup. That is, the test setup excludes complex circuitry, such as
switches that might affect LCR measurements.Custom cable compensation You use non-NI cabling and a complex test setup. That is, the test setup includes complex circuitry, such as switches that might affect LCR measurements. Note NI recommends applying cable compensation for low-level LCR measurements where the effects of cabling might be significant relative to the quantity you are measuring.Note NI recommends first applying short LCR compensation when using custom non-NI cabling to ensure accurate measurements. -
To determine the type of LCR compensation you want to perform, use the following
criteria. Click the option link for detailed instructions on performing the selected
compensation.
Option Recommended When Open LCR compensation Stray admittance in your test setup influences LCR measurements. Note Such stray admittance is present in most test setups.Short LCR compensation The impedance of your DUT is low (<100 Ω). The residual impedance of cabling and fixtures might be a significant percentage of your DUT impedance. Load LCR compensation The reference load value is similar to your DUT. Your test setup also includes complex circuitry, such as switches that might affect LCR measurements. Note Open and short LCR compensations are prerequisites to performing load LCR compensation.
Applying Standard Cable Compensation
-
Click Instrument Settings (
) in your LCR device instrument panel.
- Select your cable length from the Cable type drop-down.
Performing Custom Cable Compensation
-
Click Custom Compensation (
) in the instrument panel.
- Click Renew.
- Click Compensate.
- Click OK.
Verify that your device is configured for custom cables by opening Instrument
Settings (
) and selecting
"Custom " from the Cable type dropdown.
Verify compensation is applied. Click Custom Compensation
(
) and check that Apply
compensation is enabled.
Performing LCR Load Compensation
- HI CUR
- HI POT
- LO POT
- LO CUR
- Connection between outer conductors
-
Click Load Compensation (
) in the instrument panel to open the compensation
dialog.
- Click Renew.
- Optional: Enter spots to compensate for.
- Click Compensate.
- Click OK.
) and check that Apply compensation is enabled. Performing Open LCR Compensation
- HI CUR
- HI POT
- LO POT
- LO CUR
- Connection between outer conductors
-
Click Open Compensation (
) in the instrument panel to open the compensation
dialog.
- Click Renew.
- Optional: Enter additional frequencies to compensate for.
- Click Compensate.
- Click OK.
) and check that Apply compensation is enabled. Performing Short LCR Compensation
- HI CUR
- HI POT
- LO POT
- LO CUR
- Connection between outer conductors
-
Click Short Compensation (
) in the instrument panel to open the compensation
dialog.
- Click Renew.
- Optional: Enter additional frequencies to compensate for.
-
Click Compensate.
Note You must perform short LCR compensation when using custom non-NI cabling. Complete custom cable compensation before performing short LCR compensation.
- Click OK.
) and check that Apply compensation is enabled. When to Generate New Data for Custom Cable Compensation or LCR Compensation
Performing new custom cable and LCR device compensation ensures better accuracy in test measurements after making changes to your test system setup.
- You perform a new external calibration
- You perform a new self-calibration
- You add complex circuit elements, such as a switch, between the end of the cable and the fixture containing your DUT
- You change the length of the cabling in your test setup
- You change the physical orientation of the cabling in your test setup
- You begin testing a DUT that is different (on the basis of part number) from the DUT for
which you originally generated compensation data; Note For load LCR compensation, choose a new reference load appropriate for the new DUT.