ECUTS-16000 Specifications
- Updated2024-12-11
- 12 minute(s) read
ECUTS-16000 Specifications
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Typical-95 specifications describe the performance met by 95% (≈2σ) of models with a 95% confidence.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
- Measured specifications describe the measured performance of a representative model.
Specifications are Warranted unless otherwise noted.
Conditions
These specifications apply to all ECUTS-16000 configuration options unless otherwise noted. Specifications are valid for the system and all included instruments under the following conditions unless otherwise noted.
- ECU Test System environmental characteristics are met
- Instrument-level conditions are met
Calibration Conditions
The performance of an externally calibrated instrument is defined in the instrument specifications. Additionally, specifications for externally calibrated instruments are only valid if the conditions defined in the instrument specifications are met.
ECUTS-16001 Mass Interconnect
Pinouts
The tables in this document show the pin definitions for each slot type on the mass interconnect.
J1 Pinout
J2 Pinout
J3-J8 Pinouts
J9, J10, and J16 Pinout
J12 and J14 Pinout
J18-J21 Pinouts
J22 Pinouts
J23 Pinout
J24 Pinout
Physical Characteristics
| |||||||||||||||||||||
| |||||||||||||||||||||
| |||||||||||||||||||||
| |||||||||||||||||||||
|
The circuit breaker on the power entry panel PEP-116 is rated for a mechanical impact energy level of IK06(1 J), when tested with a direct vertical impact per IEC 61010-1, 3rd Ed., Table 15 and Clause 8.2. The circuit breaker should be guarded against impacts exceeding 1 J.
Protective Earthing
High levels of leakage current may be present on the ECUTS-16000. Connect the ECUTS-16000 protective earth terminal before connecting to AC power.
|
Power Requirements
Input voltage range | 200 V AC to 240 V AC, 50/60 Hz, single phase, 16 A, maximum |
Input power receptacle | IEC 60320 C20 |
The ECUTS-16000 shipment contains a power cord that corresponds to your country of operation.
Maximum Thermal Load
Minimum rating for internal equipment | 50 °C | ||||||||
|
Compressed Air
Air type | Clean, dry air |
Connector type | 6 mm outer diameter |
Maximum pressure | 100 PSI (690 kPa) |
DC Output
PDU | Nominal Voltage | Rated Output Current | Overcurrent Protection | Location and Use | Interlock |
---|---|---|---|---|---|
DC1 | 12 V | 4 A | 20 A | Connector J22, pin 1 referenced to pin 2, fixture power | Connector J1, short, dry contact between pins B43 and B44 |
DC2 | 24 V | 2 A | 10 A | Connector J22, pin 34 referenced to pin 35, fixture power |
Connector J1, short, dry contact between connectors C43 and C44 |
DC3[3]3 Reserved for NI internal use only. | 24 V | 5 A | 10 A | Internal, fans and instrumentation | — |
DC4 | 48 V | 2 A | 5 A | Internal, network switch | — |
| |||||||
|
I/O
| |||||||||||||||||||||||||||
| |||||||||||||||||||||||||||
Mass interconnect model | Virginia Panel Corporation 9025/9025TR |
Environmental Characteristics
| |||||||
| |||||||
Pollution degree | 2 | ||||||
Maximum altitude | 2,000 m | ||||||
|
Hardware Components
The following components are used within the ECUTS-16000.
Model | Description | |
---|---|---|
PXI Chassis | PXIe-1084 | 18-Slot PXI Chassis |
PXIe-1095 | 18-Slot High Performance PXI Chassis | |
PXI Controller | PXIe-8861 | 2.8 GHz Quad-Core Intel Xeon PXI Controller |
PXIe-8881 | 3 GHz 18-Core Intel Xeon PXI Controller | |
RMX-10011 | Low Power 24U Rack with Low Speed Fan Panel | |
RMX-10050 | Single Phase PDU with DC Output |
Model | Description | Overriding Ratings in ECUTS[4]4 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS. | |
---|---|---|---|
Maximum Working Voltage | Maximum Transient Voltage | ||
PXI-2520 | 80-Channel, 2 A, SPST Relay Module | 100 V, CAT I | 500 V |
PXI-2527 | 32-Channel, 2-Wire Multiplexer | 100 V, CAT I | 500 V |
PXI-2564 | 16-Channel, 5 A, SPST Relay Module | 60 V | 60 V |
PXI-2567 [5]5 The maximum total drive current per module in the ECUTS-16000 is 10 A. | 64-Channel Relay Driver Module | — | — |
PXI-8517 | 2-Port FlexRay Interface Module | — | — |
PXIe-2532B | 512-Crosspoint Matrix Switch Module | 100 V, CAT I | 500 V |
PXIe-2737 | 4 x 64, 2-Wire Matrix Switch Module | 100 V, CAT I | 500 V |
PXIe-2738 | 8 x 32, 2-Wire Matrix Switch Module | 100 V, CAT I | 500 V |
PXIe-2739 | 16 x 16, 2-Wire Matrix Switch Module | 100 V, CAT I | 500 V |
PXIe-4081 [6]6 Add ±15 μV to the DC accuracy specifications in the PXIe-4081 Specifications on ni.com/docs when used in the ECUTS. Do not add this value to the performance limits specified in a calibration procedure or calibration certificate for the PXIe-4081. | 7½-Digit Multimeter with 1.8 MS/s Isolated Digitizer | 100 V, CAT I (slots J3-J8) 60 V (slots J18-J21) | 500 V (slots J3-J8) 60 V (slots J18-J21) |
PXIe-4082 | 6½-Digit Multimeter with 1.8 MS/sIsolated Digitizer, L and C Measurement Support | 100 V, CAT I (slots J3-J8) 60 V (slots J18-J21) | 500 V (slots J3-J8) 60 V (slots J18-J21) |
PXIe-4112 | 2-Channel, 60 V, 1 A Programmable Power Supply | 60 V | 60 V |
PXIe-4139 (40 W) | ±60 V, ±3 A DC, ±10 A Pulsed, 40 W DC, 100 fA precision source measure unit | — | — |
PXIe-4139 (20 W) | ±60 V, ±3 A DC, ±10 A Pulsed, 20 W DC, 100 fA precision source measure unit | — | — |
PXIe-4309 | 2 MS/s, 28-Bit Flexible Resolution Analog Input | — | — |
PXIe-4310 with TB-4310 (600V) | 8-Channel, 400 kS/s, 16-Bit Isolated Analog Input Module | 60 V | 60 V |
PXIe-4322 | 8-Channel, 250 kS/s, 16-Bit Isolated Analog Output Module | Isolation and DC Overvoltage: 100 V, CAT I (slots J3-J8)60 V (slots J18-J21) | Isolation: 500 V (slots J3-J8)60 V (slots J18-J21) |
PXIe-5105 | 8-Channel, 60 MHz Bandwidth, 12-Bit Oscilloscope | — | — |
PXIe-5110 | 2-Channel, 100 MHz Bandwidth, 8-Bit Oscilloscope | 100 V, CAT I | 250 V |
PXIe-5163 | 2-Channel, 200 MHz Bandwidth, 14-Bit Oscilloscope | 100 V, CAT I | 250 V |
PXIe-5413 | 2-Channel, 20 MHz Bandwidth, 16-Bit Waveform Generator | — | — |
PXIe-5433 | 2-Channel, 80 MHz Bandwidth, 16-Bit Waveform Generator | — | — |
PXIe-6366 | 8 AI (16-Bit, 2 MS/s/ch), 2 AO, 24 DIO Multifunction I/O Module | — | — |
PXIe-6386 | 8 AI (16-Bit, 14 MS/s/ch), 2 AO, 24 DIO Multifunction I/O Module | — | — |
PXI-6515 | 64-Channel, ±30 VDC, 32 Sink/Source Inputs, 32 Sink Outputs, Bank-Isolated Digital I/O Module | — | — |
PXIe-6738 | 16-bit, 32-Channel, 1 MS/s PXI Analog Output Module | — | — |
PXIe-8430/8 | 8-Port, RS232 Serial Interface Module | — | — |
PXIe-8510 with TRC-8543 and/or TRC-8546 Transceiver Cable | 6-Port Vehicle Multiprotocol Interface Module with CAN or LIN Transceivers | — | — |
PXIe-8523 | PXIe, 4-Port, 100/1000BASE-T1[7]7 The PXIe-8523 is limited to 100 Mbit when operating the channels in parallel or up to 1000 Mbit on a single channel. PXI Automotive Ethernet Interface Module | — | — |
Model | Description | Overriding Ratings in ECUTS[8]8 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS. | |
---|---|---|---|
Maximum Working Voltage | Maximum Transient Voltage | ||
SLSC-12001 | 12-Slot Chassis for SLSC | — | — |
SLSC-12251 [9]9 Limit maximum current to 5 A per channel for SLSC-12251 modules routed to the J3-J8 QuadraPaddle slots. | 16-Channel, 8 A Fault Insertion Module | 100 V, CAT I | 500 V |
SLSC-12252 [10]10 The maximum allowable current in an SLSC-12252 channel is 30 A. | 8-Channel, 30 A Fault Insertion Module | 100 V, CAT I | 500 V |
PXIe-6375 | PXIe, 208 AI (16-Bit, 3.8 MS/s), 2 AO, 24 DIO, PXI Multifunction I/O Module | — | — |
Model | Description |
---|---|
RMX-4101 | 20 V, 10 A, Programmable Power Supply Device |
36 V, 6 A, Programmable Power Supply Device | |
60 V, 3.5 A, Programmable Power Supply Device | |
RMX-4102 | 20 V, 20 A, Programmable Power Supply Device |
36 V, 12 A, Programmable Power Supply Device | |
60 V, 7 A, Programmable Power Supply Device | |
RMX-4002 | 4-Slot Electronic Load Device Mainframe |
RMX-4003 | 2-Channel, 100 W, Electronic Load Device |
RMX-4005 [11]11 The single channel on an RMX-4005 module is routed to two sets of pins on the mass interconnect. Each load terminal on the RMX-4005 is rated to 40 A maximum. | 1-Channel, 350 W, Electronic Load Device |
N6700C | 4-Slot 400 W Power System Mainframe |
N6752A/761 | 50 V, 10 A, 100 W Supply, with Output Disconnect |
N6753A/761 | 20 V, 50 A, 300 W Supply, with Output Disconnect |
N6754A/761 | 60 V, 20 A, 300 W Supply, with Output Disconnect |
N6762A/761 | Precision 50 V, 3 A, 100 W Supply, with Output Disconnect |
N6763A/761 | Precision 20 V, 50 A, 300 W Supply, with Output Disconnect |
N6764A/761 | Precision 60 V, 20 A, 300 W Supply, with Output Disconnect |
N6791A | 60 V, 20 A, 100 W Electronic Load |
N6792A | 60 V, 40 A, 200 W Electronic Load |
Part Number | Description | Maximum Working Voltage | Maximum Transient Voltage |
---|---|---|---|
132503-03 | ECUTS B3-3 PXIe-4322 Cable | 100 V pk | 500 V pk |
132504-01 | ECUTS B4-1 Switch Cable | 100 V pk | 500 V pk |
132504-02 | ECUTS B4-2 Switch, DMM, and PXIe-4322 Cable | 100 V pk | 500 V pk |
132504-04 | ECUTS B4-4 SLSC-12251, DMM, and PXIe-4322 Cable | 100 V pk | 500 V pk |
132504-05 | ECUTS B4-5 PXIe-2527 and DMM Cable | 100 V pk | 500 V pk |
132504-06 | ECUTS, B4-6, VPC Quadrapaddle Signal to DMM and 2x 78 Pin Female DSUB | 100 V pk | 500 V pk |
136249-01 | ECUTS B4-7 PXIe-2532B and DMM Interface Assembly | 100 V pk | 500 V pk |
132505-0x | ECUTS B5-x SLSC-12251 Cable | 100 V pk | 500 V pk |
132506-01 | ECUTS B6 SLSC-12252 Cable | 100 Vpk | 500 V pk |
132508-05 | ECUTS B8-5 Multifunction Cable | 100 V pk | 500 V pk |
132508-06 | ECUTS B8-6 Multifunction Cable | 100 V pk | 500 V pk |
132508-07 | ECUTS B8-7 Multifunction Cable | 100 V pk | 500 V pk |
132508-08 | ECUTS B8-8 Multifunction Cable | 100 V pk | 500 V pk |
132509-0x | ECUTS B9-x RMX-400x Cable | 80 V pk | 80 V pk |
1 The weight of the system depends on which options are purchased and installed.
2 Maximum weight limit when a fixture is attached to the ECUTS-16000 without the 9025TR platform.
3 Reserved for NI internal use only.
4 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.
5 The maximum total drive current per module in the ECUTS-16000 is 10 A.
6 Add ±15 μV to the DC accuracy specifications in the PXIe-4081 Specifications on ni.com/docs when used in the ECUTS. Do not add this value to the performance limits specified in a calibration procedure or calibration certificate for the PXIe-4081.
7 The PXIe-8523 is limited to 100 Mbit when operating the channels in parallel or up to 1000 Mbit on a single channel.
8 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.
9 Limit maximum current to 5 A per channel for SLSC-12251 modules routed to the J3-J8 QuadraPaddle slots.
10 The maximum allowable current in an SLSC-12252 channel is 30 A.
11 The single channel on an RMX-4005 module is routed to two sets of pins on the mass interconnect. Each load terminal on the RMX-4005 is rated to 40 A maximum.