ECUTS-16000 Specifications

Definitions

Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration.

Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.

  • Typical specifications describe the performance met by a majority of models.
  • Typical-95 specifications describe the performance met by 95% (≈2σ) of models with a 95% confidence.
  • Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
  • Measured specifications describe the measured performance of a representative model.

Specifications are Warranted unless otherwise noted.

Conditions

Refer to ni.com/docs for detailed specifications on the specific instruments used within the ECUTS-16000.
Note Other product and company names listed are trademarks or trade names of their respective companies.

These specifications apply to all ECUTS-16000 configuration options unless otherwise noted. Specifications are valid for the system and all included instruments under the following conditions unless otherwise noted.

  • ECU Test System environmental characteristics are met
  • Instrument-level conditions are met

Calibration Conditions

The performance of an externally calibrated instrument is defined in the instrument specifications. Additionally, specifications for externally calibrated instruments are only valid if the conditions defined in the instrument specifications are met.

ECUTS-16001 Mass Interconnect Pinouts

The tables in this document show the pin definitions for each slot type on the mass interconnect.

J1 Pinout



J2 Pinout



J3-J8 Pinouts

Figure 1. J3-J8 Pinout when Populated with a PXIe-2520


Figure 2. J3-J8 Pinout when Populated with a PXIe-2527


Figure 3. J3-J8 Pinout when Populated with a PXIe-2527 (1-Wire Topology)


Figure 4. J3-J8 Pinout when Populated with a PXIe-2527 (2-Wire Topology)


Figure 5. J3-J8 Pinout when Populated with a PXIe-2527 (4-Wire Topology)


Figure 6. J3-J8 Pinout when Populated with a PXIe-2737


Figure 7. J3-J8 Pinout when Populated with a PXIe-2738


Figure 8. J3-J8 Pinout when Populated with a PXIe-2739


Figure 9. J3-J8 Pinout when Populated with a SLSC-12251


J9, J10, and J16 Pinout



J12 and J14 Pinout



J18-J21 Pinouts

Figure 10. J18-J21 Default Pinout


Figure 11. J18-J21 Pinout when Populated with a PXIe-6378


Figure 12. J18-J21 Pinout when Populated with a PXIe-4309


J22 Pinouts

Figure 13. J22 Pinout when Populated with PXIe-5105 Oscilloscopes


Figure 14. J22 Pinout when Populated with PXIe-5110 Oscilloscopes


Figure 15. J22 Pinout when Populated with PXIe-5163 Oscilloscopes


J23 Pinout

Figure 16. J23 Pinout


Figure 17. J23 Pinout CAN Port Pair


Figure 18. J23 Pinout LIN Port Pair


J24 Pinout



Physical Characteristics

Dimensions

Width, all configurations

584 mm (23.0 in.)

Depth

Without 9025TR platform

853 mm (33.6 in.)

With 9025TR platform

1,650 mm (65.0 in.)

Height

With monitor mount, before monitor arm assembly

1,440 mm (56.5 in.)

Maximum, with monitor and keyboard tray

2,080 mm (81.8 in.)

Total system weight[1]1 The weight of the system depends on which options are purchased and installed.

Minimum

185 kg (407 lb)

Maximum

270 kg (595 lb)

Maximum fixture weight

9025TR platform

18 kg at 510 mm (40 lb at 20 in.)

Fixture[2]2 Maximum weight limit when a fixture is attached to the ECUTS-16000 without the 9025TR platform.

18 kg at 510 mm (40 lb at 20 in.)

Keyboard tray

0.9 kg (2 lb)

Human-Machine interface weight

Keyboard tray

0.9 kg (2 lb)

Monitor

3.5 kg (8 lb)

Rack paint

Color

RAL-7035 gray

Paint type

ESD dissipative paint as defined in IEC 61340-5-1

The circuit breaker on the power entry panel PEP-116 is rated for a mechanical impact energy level of IK06(1 J), when tested with a direct vertical impact per IEC 61010-1, 3rd Ed., Table 15 and Clause 8.2. The circuit breaker should be guarded against impacts exceeding 1 J.

Protective Earthing

High levels of leakage current may be present on the ECUTS-16000. Connect the ECUTS-16000 protective earth terminal before connecting to AC power.

Note The facility installation shall provide a means for connection to protective earth, and qualified personnel shall install a protective earthing conductor from the ECUTS-16000 protective earthing terminal to the protective earth wire in the facility.
Protective earth terminal wiring

Grounding wire

2.1 mm2 (14 AWG)

Ring lug

Size

M8

Length

20 mm (0.8 in.)

Minimum protective earth terminal torque

1.29 N · m (11.5 lb · in.)

Power Requirements

Note The AC power cords used in the ECUTS-16000 are specially designed for the ECUTS-16000. Do not use these power cords with other electrical appliances. Contact your assigned NI Hardware Services Project Manager for system-specific power cords and associated part numbers.
Note All equipment inside the rack must be powered, directly or indirectly, through the power entry panel PEP-116. Other sources of external power must not be used.

Input voltage range

200 V AC to 240 V AC, 50/60 Hz, single phase, 16 A, maximum

Input power receptacle

IEC 60320 C20

The ECUTS-16000 shipment contains a power cord that corresponds to your country of operation.

Maximum Thermal Load

Minimum rating for internal equipment

50 °C

Maximum allowable internal dissipation

40 °C ambient

1,350 W

28 °C ambient

3,400 W

23 °C ambient

3,790 W

Compressed Air

Air type

Clean, dry air

Connector type

6 mm outer diameter

Maximum pressure

100 PSI (690 kPa)

DC Output

Note Refer to the ECUTS-16000 User Manual for more information about connector and pin locations.
Table 1. DC Output
PDU Nominal Voltage Rated Output Current Overcurrent Protection Location and Use Interlock
DC1 12 V 4 A 20 A Connector J22, pin 1 referenced to pin 2, fixture power Connector J1, short, dry contact between pins B43 and B44
DC2 24 V 2 A 10 A Connector J22, pin 34 referenced to pin 35, fixture power

Connector J1, short, dry contact between connectors C43 and C44

DC3[3]3 Reserved for NI internal use only. 24 V 5 A 10 A Internal, fans and instrumentation
DC4 48 V 2 A 5 A Internal, network switch
RMX-410x (optional)

Interlock

Connector J1, short, dry contact between pins A43 and A44

Maximum total output power

1,500 W

N67xx (optional)

Interlock

Connector J1, short, dry contact between pins D43 and D44

Maximum total output power

1,500 W

I/O

Network ports
External

Type

I219

Standard

IEEE 802.3 Ethernet, 10BASE-T, 100BASE-TX, 1000BASE-T

Speed

10 Mbps, 100 Mbps, 1,000 Mbps

Internal

Speed

10 Mbps, 100 Mbps

Subnet

169.254.160.xxx

Number of ports

15 switched ports (two for the mass interconnect, up to 13 for instrumentation)

USB ports
Rear

Number of ports

2

Type

USB 3.0

Top

Number of ports

1

Type

USB 2.0 cable (to monitor hub)

Internal (optional)

Number of ports

7

Type

USB 2.0

Mass interconnect model

Virginia Panel Corporation 9025/9025TR

Environmental Characteristics

Note This product is intended for use in indoor applications only.
Temperature

Operating

5 °C to 40 °C

Storage

0 °C to 60 °C

Humidity

Operating

20% to 80%, noncondensing

Storage

10% to 80%, noncondensing

Pollution degree

2

Maximum altitude

2,000 m

Ventilation clearances

Top

760 mm (30 in.)

Front and back

300 mm (12 in.)

Hardware Components

The following components are used within the ECUTS-16000.

Note Detailed specifications for hardware components are available online at ni.com/docs.
Table 2. Required Hardware Components
Model Description
PXI Chassis PXIe-1084 18-Slot PXI Chassis
PXIe-1095 18-Slot High Performance PXI Chassis
PXI Controller PXIe-8861 2.8 GHz Quad-Core Intel Xeon PXI Controller
PXIe-8881 3 GHz 18-Core Intel Xeon PXI Controller
RMX-10011 Low Power 24U Rack with Low Speed Fan Panel
RMX-10050 Single Phase PDU with DC Output

Table 3. PXI Instruments
Model Description Overriding Ratings in ECUTS[4]4 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.
Maximum Working Voltage Maximum Transient Voltage
PXI-2520 80-Channel, 2 A, SPST Relay Module 100 V, CAT I 500 V
PXI-2527 32-Channel, 2-Wire Multiplexer 100 V, CAT I 500 V
PXI-2564 16-Channel, 5 A, SPST Relay Module 60 V 60 V
PXI-2567 [5]5 The maximum total drive current per module in the ECUTS-16000 is 10 A. 64-Channel Relay Driver Module
PXI-8517 2-Port FlexRay Interface Module
PXIe-2532B 512-Crosspoint Matrix Switch Module 100 V, CAT I 500 V
PXIe-2737 4 x 64, 2-Wire Matrix Switch Module 100 V, CAT I 500 V
PXIe-2738 8 x 32, 2-Wire Matrix Switch Module 100 V, CAT I 500 V
PXIe-2739 16 x 16, 2-Wire Matrix Switch Module 100 V, CAT I 500 V
PXIe-4081 [6]6 Add ±15 μV to the DC accuracy specifications in the PXIe-4081 Specifications on ni.com/docs when used in the ECUTS. Do not add this value to the performance limits specified in a calibration procedure or calibration certificate for the PXIe-4081. 7½-Digit Multimeter with 1.8 MS/s Isolated Digitizer 100 V, CAT I (slots J3-J8)

60 V (slots J18-J21)
500 V (slots J3-J8)

60 V (slots J18-J21)
PXIe-4082 6½-Digit Multimeter with 1.8 MS/sIsolated Digitizer, L and C Measurement Support 100 V, CAT I (slots J3-J8)

60 V (slots J18-J21)
500 V (slots J3-J8)

60 V (slots J18-J21)
PXIe-4112 2-Channel, 60 V, 1 A Programmable Power Supply 60 V 60 V
PXIe-4139 (40 W) ±60 V, ±3 A DC, ±10 A Pulsed, 40 W DC, 100 fA precision source measure unit
PXIe-4139 (20 W) ±60 V, ±3 A DC, ±10 A Pulsed, 20 W DC, 100 fA precision source measure unit
PXIe-4309 2 MS/s, 28-Bit Flexible Resolution Analog Input
PXIe-4310 with TB-4310 (600V) 8-Channel, 400 kS/s, 16-Bit Isolated Analog Input Module 60 V 60 V
PXIe-4322 8-Channel, 250 kS/s, 16-Bit Isolated Analog Output Module Isolation and DC Overvoltage:

100 V, CAT I (slots J3-J8)

60 V (slots J18-J21)
Isolation:

500 V (slots J3-J8)

60 V (slots J18-J21)
PXIe-5105 8-Channel, 60 MHz Bandwidth, 12-Bit Oscilloscope
PXIe-5110 2-Channel, 100 MHz Bandwidth, 8-Bit Oscilloscope 100 V, CAT I 250 V
PXIe-5163 2-Channel, 200 MHz Bandwidth, 14-Bit Oscilloscope 100 V, CAT I 250 V
PXIe-5413 2-Channel, 20 MHz Bandwidth, 16-Bit Waveform Generator
PXIe-5433 2-Channel, 80 MHz Bandwidth, 16-Bit Waveform Generator
PXIe-6366 8 AI (16-Bit, 2 MS/s/ch), 2 AO, 24 DIO Multifunction I/O Module
PXIe-6386 8 AI (16-Bit, 14 MS/s/ch), 2 AO, 24 DIO Multifunction I/O Module
PXI-6515 64-Channel, ±30 VDC, 32 Sink/Source Inputs, 32 Sink Outputs, Bank-Isolated Digital I/O Module
PXIe-6738 16-bit, 32-Channel, 1 MS/s PXI Analog Output Module
PXIe-8430/8 8-Port, RS232 Serial Interface Module
PXIe-8510 with TRC-8543 and/or TRC-8546 Transceiver Cable 6-Port Vehicle Multiprotocol Interface Module with CAN or LIN Transceivers
PXIe-8523 PXIe, 4-Port, 100/1000BASE-T1[7]7 The PXIe-8523 is limited to 100 Mbit when operating the channels in parallel or up to 1000 Mbit on a single channel. PXI Automotive Ethernet Interface Module

Table 4. Load Switching
Model Description Overriding Ratings in ECUTS[8]8 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.
Maximum Working Voltage Maximum Transient Voltage
SLSC-12001 12-Slot Chassis for SLSC
SLSC-12251 [9]9 Limit maximum current to 5 A per channel for SLSC-12251 modules routed to the J3-J8 QuadraPaddle slots. 16-Channel, 8 A Fault Insertion Module 100 V, CAT I 500 V
SLSC-12252 [10]10 The maximum allowable current in an SLSC-12252 channel is 30 A. 8-Channel, 30 A Fault Insertion Module 100 V, CAT I 500 V
PXIe-6375 PXIe, 208 AI (16-Bit, 3.8 MS/s), 2 AO, 24 DIO, PXI Multifunction I/O Module

Table 5. Power Supplies and Electronic Loads
Model Description
RMX-4101 20 V, 10 A, Programmable Power Supply Device
36 V, 6 A, Programmable Power Supply Device
60 V, 3.5 A, Programmable Power Supply Device
RMX-4102 20 V, 20 A, Programmable Power Supply Device
36 V, 12 A, Programmable Power Supply Device
60 V, 7 A, Programmable Power Supply Device
RMX-4002 4-Slot Electronic Load Device Mainframe
RMX-4003 2-Channel, 100 W, Electronic Load Device
RMX-4005 [11]11 The single channel on an RMX-4005 module is routed to two sets of pins on the mass interconnect. Each load terminal on the RMX-4005 is rated to 40 A maximum. 1-Channel, 350 W, Electronic Load Device
N6700C 4-Slot 400 W Power System Mainframe
N6752A/761 50 V, 10 A, 100 W Supply, with Output Disconnect
N6753A/761 20 V, 50 A, 300 W Supply, with Output Disconnect
N6754A/761 60 V, 20 A, 300 W Supply, with Output Disconnect
N6762A/761 Precision 50 V, 3 A, 100 W Supply, with Output Disconnect
N6763A/761 Precision 20 V, 50 A, 300 W Supply, with Output Disconnect
N6764A/761 Precision 60 V, 20 A, 300 W Supply, with Output Disconnect
N6791A 60 V, 20 A, 100 W Electronic Load
N6792A 60 V, 40 A, 200 W Electronic Load

Table 6. High-Voltage Cables and Assemblies
Part Number Description Maximum Working Voltage Maximum Transient Voltage
132503-03 ECUTS B3-3 PXIe-4322 Cable 100 V pk 500 V pk
132504-01 ECUTS B4-1 Switch Cable 100 V pk 500 V pk
132504-02 ECUTS B4-2 Switch, DMM, and PXIe-4322 Cable 100 V pk 500 V pk
132504-04 ECUTS B4-4 SLSC-12251, DMM, and PXIe-4322 Cable 100 V pk 500 V pk
132504-05 ECUTS B4-5 PXIe-2527 and DMM Cable 100 V pk 500 V pk
132504-06 ECUTS, B4-6, VPC Quadrapaddle Signal to DMM and 2x 78 Pin Female DSUB 100 V pk 500 V pk
136249-01 ECUTS B4-7 PXIe-2532B and DMM Interface Assembly 100 V pk 500 V pk
132505-0x ECUTS B5-x SLSC-12251 Cable 100 V pk 500 V pk
132506-01 ECUTS B6 SLSC-12252 Cable 100 Vpk 500 V pk
132508-05 ECUTS B8-5 Multifunction Cable 100 V pk 500 V pk
132508-06 ECUTS B8-6 Multifunction Cable 100 V pk 500 V pk
132508-07 ECUTS B8-7 Multifunction Cable 100 V pk 500 V pk
132508-08 ECUTS B8-8 Multifunction Cable 100 V pk 500 V pk
132509-0x ECUTS B9-x RMX-400x Cable 80 V pk 80 V pk

1 The weight of the system depends on which options are purchased and installed.

2 Maximum weight limit when a fixture is attached to the ECUTS-16000 without the 9025TR platform.

3 Reserved for NI internal use only.

4 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.

5 The maximum total drive current per module in the ECUTS-16000 is 10 A.

6 Add ±15 μV to the DC accuracy specifications in the PXIe-4081 Specifications on ni.com/docs when used in the ECUTS. Do not add this value to the performance limits specified in a calibration procedure or calibration certificate for the PXIe-4081.

7 The PXIe-8523 is limited to 100 Mbit when operating the channels in parallel or up to 1000 Mbit on a single channel.

8 In some cases, integrating a device in an ECUTS imposes a new specification. This table lists device ratings that change in ECUTS.

9 Limit maximum current to 5 A per channel for SLSC-12251 modules routed to the J3-J8 QuadraPaddle slots.

10 The maximum allowable current in an SLSC-12252 channel is 30 A.

11 The single channel on an RMX-4005 module is routed to two sets of pins on the mass interconnect. Each load terminal on the RMX-4005 is rated to 40 A maximum.