Optical sensing can benefit many applications, whether the application needs to span long distances, immunity to EMI and high voltages or the ability to take multiple measurements on a single cable. The PXIe-4844 optical sensor interrogator enables you to take advantage of these benefits as the best in class static interrogator. However, the interrogator is only one piece of an optical sensing system. Another important element in an optical sensing system is the FBG sensors. In order to ensure a properly setup optical sensing system, scaling formulas are required to be input to translate the shift in FBG to a strain or temperature value. This whitepaper will give you the appropriate scaling formulas to be entered into NI-OSI Explorer to ensure you receive the highest quality, accurate and reliable measurements.
Choose from a wide variety of FBG optical sensor, ranging from standard off-the-shelf strain gages to custom downhole oil and gas pressure transducers. You can daisy chain sensors including strain and temperature in any order by using optical connectors or splicing bare fibers with a fusion splicer. You can obtain sensors individually or in a preassembled array to meet your exact requirements.
The method for installing and using FBG optical sensors is very similar to the established methods used for electrical sensors. You can glue, weld, and embed FBG sensors while having fewer cables to manage with no noise, isolation, and/or shielding considerations to address. A variety of cabling exists, from simple low-cost to rugged deep sea cables, meeting the most extreme and rigid requirements.
Below are the scaling formulas that can be directly copied into NI-OSI Explorer once the appropriate variables are entered in (if applicable). Before getting to this step, let’s ensure a few things are completed first.
Each sensor below includes the vendor scaling formula, the NI-OSI scaling formula, and any clarifying variables.
For all strain FBGs, the scaling formula includes temperature compensation. In order to perform temperature compensation a second FBG (temperature) will need to be used. The Temperature FBG will be a reference to the strain FBG in NI-OSI Explorer.
Common terms used in all sensors are below. These terms can be entered in as shown.
Common variables used in sensors are below. These variables need to have a number entered in.
Sensor | Use Case | Scaling Formula |
---|---|---|
os3110 FBG Optical Strain Gage, Spot Weld
| NI-OSI | ε=((x-x0)/(y))*((1*10^6)/(Gage Factor))-((r-r0)*(Gage Constant 1/Gage Factor)+CTE of Test Specimen-Gage Constant 2) |
NI-OSI Ex. | ((x-x0)/(y))*((1*10^6)/(0.89))-((r-r0)*(6.156/0.89)+13-0.70) | |
Vendor | ||
os3120 FBG Optical Strain Gage, Epoxy Mount
| NI-OSI | ε=((x-x0)/(y))*((1*10^6)/(Gage Factor))-((r-r0)*(Gage Constant 1/Gage Factor)+CTE of Test Specimen-Gage Constant 2) |
NI-OSI Ex. | ((x-x0)/(y))*((1*10^6)/(0.89))-((r-r0)*(6.156/0.89)+13-0.70) | |
Vendor | ||
os3150 Rugged FBG Optical Strain Gage | NI-OSI | ε=((x-x0)/(y))*((1*10^6)/(Gage Factor))-((r-r0)*(Gage Constant 1/Gage Factor)+CTE of Test Specimen-Gage Constant 2) |
NI-OSI Ex. | ((x-x0)/(y))*((1*10^6)/(0.811))-((r-r0)*(6.156/0.811+13-0.70) | |
Vendor | ||
os3155 Rugged FBG Optical Strain Gage with Integrated Temperature Compensation | NI-OSI (Strain) | ε=(1*10^6)*((((x-x0)/(y))-((r-r0)/(Nominal Wavelength Temp)))/(Gage Factor))-(((r-r0)/(Nominal Wavelength Temp))/(Temperature Sensitivity))*(CTE of Test Specimen-CTE of Temperature Probe) |
NI-OSI (Temp) | T=(x-x0)+(Temperature of Environment) | |
NI-OSI Ex. (S) | (1*10^6)*((((x-x0)/(y))-((r-r0)/(1582.214)))/(0.807))-(((r-r0)/(1582.214))/(0.00001833))*(13-16) | |
NI-OSI Ex. (T) | (x-x0)+22 | |
Vendor | ||
os3200 Non-metallic FBG Optical Strain Gage
| NI-OSI | ε=((x-x0)/(y))*((1*10^6)/(Gage Factor))-((r-r0)*(Gage Constant 1/Gage Factor)+CTE of Test Specimen-Gage Constant 2) |
NI-OSI Ex. | ((x-x0)/(y))*((1*10^6)/(0.796))-((r-r0)*(6.156/0.796)+13-0.70) | |
Vendor | ||
os3600 Embeddable Long-gage Strain Sensor
| NI-OSI | ε=(((x-x0)/(y))*((1*10^6)/(Gage Factor)) )-(((r-r0)/(Nominal Wavelength Temp))*((1*10^6)/(Gage Constant 1)) )+(CTE of Test Specimen-Gage Constant 2)*(1*10^3 )*((r-r0)/(Temperature Sensitivity)) |
NI-OSI Ex. | (((x-x0)/(y))*((1*10^6)/(0.786)) )-(((r-r0)/(1551.756))*((1*10^6)/(0.796)) )+(13-10.1)*(1*10^3 )*((r-r0)/(23.8)) | |
Vendor | N/A | |
os3610 Surface-mount Long-gage Strain Sensor
| NI-OSI | ε=(((x-x0)/(y))*((1*10^6)/(Gage Factor)))-(((r-r0)/(Nominal Wavelength Temp))*((1*10^6)/(Gage Constant 1)) )+(CTE of Test Specimen-Gage Constant 2)*(1*10^3 )*((r-r0)/(Temperature Sensitivity)) |
NI-OSI Ex. | (((x-x0)/(y))*((1*10^6)/(0.840)))-(((r-r0)/(1521.818))*((1*10^6)/(0.813)) )+(13-11)*(1*10^3 )*((r-r0)/(22.0)) | |
Vendor | N/A |
Sensor | Use Case | Scaling Formula |
---|---|---|
os4100 Temperature Compensation Gage
| NI-OSI | T=(1*10^3 )*((x-x0)/(Temperature Sensitivity) |
NI-OSI Ex. | (1*10^3 )*((x-x0)/(28.9) | |
Vendor | ||
os4210 Temperature Probe
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.7529852856878300)*(((x-x0)+(7.9227) )^3 )+(-17397.533932784500)*(((x-x0)+(7.9227) )^2 )+(26883059.46686950)*((x-x0)+(7.9227) )+(-13846814009.7684) | |
Vendor | ||
os4230 Ruggedized Temperature Probe | NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.7529852856878300)*(((x-x0)+(7.9227) )^3 )+(-17397.533932784500)*(((x-x0)+(7.9227) )^2 )+(26883059.46686950)*((x-x0)+(7.9227) )+(-13846814009.7684) | |
Vendor | ||
os4280 Ruggedized Probe w/ Thermocouple Head
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.7529852856878300)*(((x-x0)+(7.9227) )^3 )+(-17397.533932784500)*(((x-x0)+(7.9227) )^2 )+(26883059.46686950)*((x-x0)+(7.9227) )+(-13846814009.7684) | |
Vendor | ||
os4310 Temperature Sensor
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.4716563220263200)*(((x-x0)+(12.0233) )^3 )+(-16427.242229277400)*(((x-x0)+(12.0233) )^2 )+(25910320.73981810)*((x-x0)+(12.0233) )+(-13622634512.06860) | |
Vendor | ||
os4330 Non-metallic Epoxy Mount Temperature Sensor
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.4716563220263200)*(((x-x0)+(12.0233) )^3 )+(-16427.242229277400)*(((x-x0)+(12.0233) )^2 )+(25910320.73981810)*((x-x0)+(12.0233) )+(-13622634512.06860) | |
Vendor | ||
os4350 Armored Cable, Flange Mount Temperature Sensor
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.4716563220263200)*(((x-x0)+(12.0233) )^3 )+(-16427.242229277400)*(((x-x0)+(12.0233) )^2 )+(25910320.73981810)*((x-x0)+(12.0233) )+(-13622634512.06860) | |
Vendor | ||
os4400 Temperature Sensing Cable (Each FBG will have a scaling formula)
| NI-OSI | T=(Calibration Coefficient 3)*(((x-x0)+(Wavelength Offset) )^3 )+(Calibration Coefficient 2)*(((x-x0)+(Wavelength Offset) )^2 )+(Calibration Coefficient 1)*((x-x0)+(Wavelength Offset) )+(Calibration Coefficient 0) |
NI-OSI Ex. | (3.4716563220263200)*(((x-x0)+(12.0233) )^3 )+(-16427.242229277400)*(((x-x0)+(12.0233) )^2 )+(25910320.73981810)*((x-x0)+(12.0233) )+(-13622634512.06860) | |
Vendor |
Sensor | Use Case | Scaling Formula |
---|---|---|
os5100 Displacement Gage
| NI-OSI | D=(Gage Factor)*((x-x0)-(r-r0) )+Offset |
NI-OSI Ex. | (6.1573)*((x-x0)-(r-r0) )+(-15.72) | |
Vendor | ||
os5500 Long Range Displacement Gage
| NI-OSI | D=(Calibration Coefficient 3)*((x-x0)^3)+(Calibration Coefficient 2)*((x-x0)^2)+((Calibration Coefficient 1)*(x-x0)+(Calibration Coefficient 0) |
NI-OSI Ex. | (-0.00145)*((x-x0)^3)+(0.0260)*((x-x0)^2)+((16.641)*(x-x0)+(-25.25) | |
Vendor | |