The NI-Digital Pattern Driver 16.0.1 Runtime patch resolves the issues outlined in the table below for the NI-Digital Pattern driver 16.0 and PXIe-6570 Firmware 16.0.0f1. NI recommends this patch for all NI-Digital 16.0 installations and PXIe-6570 Digital Pattern Instruments.
The patch is available through NI Update Service or you can download directly through the Drivers and Updates page linked below. This download also includes an updated PXIe-6570 16.0.0f5 firmware that is required to resolve all issues listed below. The instructions to update to the 16.0.0f5 firmware are listed below.
ID | Description |
---|---|
603581 | A race condition within the PXIe-6570 firmware (version 16.0.0f1) has the potential to cause data corruption issues Note: To fix this issue you must also install the 16.0.0f2 or later firmware. |
637184 | Capture memory acquisition stops to prevent an overwrite when the capture memory is not full and PXIe backplane bandwidth is available. This issue might occur as a result of internal latency, which can cause a failure to transfer data Note: To fix this issue you must also install the 16.0.0f5 firmware. |
637186 | When using source memory, interacting with the device in NI Measurement & Automation Explorer or running the STS Maintenance Software might cause the PXIe-6570 to source incorrect data
|