This document contains the InstrumentStudio known issues that were discovered before and since the release of InstrumentStudio 2023 Q2. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community; see Additional Resources.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1154387 |
Digital Trigger Sources Do Not Update After Renaming a Device in MAXIf you rename an Oscilloscope device in MAX, the new name will not automatically appear in the list of digital trigger sources in the InstrumentStudio Oscilloscope panel. After renaming one or more devices, errors occur when attempting to use those digital trigger sources.
Workaround: Select a different trigger type in the InstrumentStudio Oscilloscope panel and then select Digital triggering again. This will force the list of trigger sources to refresh.
|
Reported Version: InstrumentStudio 2019 Resolved Version: N/A Added: Oct 22, 2021 |
|
1063539, 992871 |
Pins Using Multiplexed Connections Are Not Shown Under the Correct Site in InstrumentStudioInstrumentStudio does not currently support multiplexed connections in pin maps. If the site filter is set to a specific site, pins under that site will not appear in the panel if they are mapped in the pin map using multiplexed connections. The channels using multiplexed connections will appear under Unmapped channels if the Site filter is set to All sites, but will not display the pin names from the pin map. Workaround: There is currently no known workaround for this issue.
|
Reported Version: InstrumentStudio 2019 Resolved Version: N/A Added: Oct 22, 2021 |
|
1067929 |
SMU Waveform Mode May Not Graph Data for All Selected Channels With High Capture TimesWhen multiple channels are enabled in SMU Waveform mode and selected for display on the graph, InstrumentStudio may fail to acquire waveform data for all of the selected channels when you increase the capture time, depending on the channel count and capture time used.
Workaround: Reduce the capture time or limit the number of channels you are acquiring data from.
|
Reported Version: InstrumentStudio 2019 Resolved Version: N/A Added: Oct 22, 2021 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.