mmWave OTA Validation Test Software 20.5 Known Issues

Overview

This document contains the mmWave OTA Validation Test Software known issues that were discovered before and since the release of mmWave OTA Validation Test Software 20.5. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.

 

Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community.

Bug Number

Legacy ID

Description

Details

1097076

Error -379415 occurs when initializing an OTA session

On the Generator Calibration Settings section of the RF System Calibration Assistant Configurator window, if there is more than one Powers (dBm) selected for a single Frequencies (Hz), the following error occurs:

"-379415 Invalid table input. The input frequency array contains a duplicate value."

Workaround:

Run the RF System Calibration Assistant using only one power level and one reference level.

Reported Version:

mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5

Resolved Version:

N/A

Added:

Sep 2, 2020

1097098

LoadLibrary failed with error 87

If you connect to a system using a remote desktop and open the mmWave OTA Validation Configurator or the mmWave OTA Validation Visualizer, the following error occurs:

"LoadLibrary failed with error 87: The parameter is incorrect."

The error occurs because there is a Windows problem when running OpenGL hardware-accelerated programs through remote desktop (RDP) sessions.

Workaround:

  • Disable the graphic card driver if the option to disable is available.
  • For ATI E6460 (PXIe-8880)/E4690 (PXIe-8135) and Windows 10, complete either of the following options:
    • Uninstall the ATI/AMD drivers until only the Microsoft compatibility drivers are in use, as this driver does not have hardware acceleration support.
    • Rename the driver file that is responsible for the OpenGL support:
      • Navigate to C:\windows\system32
      • Rename atig6pxx.dll to atig6pxx.bak

Reported Version:

mmWave OTA Validation Test Software 1.0 | mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5

Resolved Version:

N/A

Added:

Sep 2, 2020

1112059

An initialization error occurs in TestStand if you run the OTA sequence directly from TestStand

If you open the mmWave OTA Validation Configurator and run the example sequence file directly in the TestStand Sequence Editor instead of starting the sequence from the mmWave OTA Validation Configurator, TestStand returns the following error:

“Cannot initialize RF session.”

Workaround:

  1. Run the TestStand sequence by clicking the Start button on the Execution tab in the mmWave OTA Validation Configurator.
  2. Close the mmWave OTA Validation Configurator.

Reported Version:

mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5

Resolved Version:

N/A

Added:

Sep 2, 2020

1274014

Cannot Run the OTA Test when the mmWave OTA Validation Positioner Control Panel is Launched

If you run the OTA test when the mmWave OTA Validation Positioner Control Panel is launched, error 56 occurs and the positioner is disconnected.

Possible reason(s): The network operation exceeded the user-specified or system time limit.

Workaround:

  1. Open Internet Explorer.
  2. Type 192.168.2.10 in the address bar and press <enter>.
  3. Click Restart on the upper right corner of the pop-up window.
  4. Wait until the restart progress complete.

Reported Version:

mmWave OTA Validation Test Software 20.5

Resolved Version:

N/A

Added:

Mar 12, 2021

1248593

The mmWave OTA Validation Test Software does not Support the German Operating System due to Driver Error

When you run the OTA examples or use the RF System Calibration Assistant, error -1074097903 occurs because the NI-RFSA/NI-RFSG drivers (19.2.1 or later) required for using the mmWave OTA Validation Test Software do not support the German operating system.

Possible reason(s): Data read from non-volatile memory on the device is invalid.

Workaround:

Change the operating system language to English or contact NI services for technical support.

Reported Version:

mmWave OTA Validation Test Software 20.5

Resolved Version:

N/A

Added:

Mar 12, 2021

Final Time Issue Listed

Issues found in this section will not be listed in future known issues documents for this product.

There are currently no issues to list.

Additional Resources

Explore Support Content and Product Documentation 

 

 

Ask the NI Community

 

 

Request Support from an Engineer

 

A valid service agreement may be required, and support options vary by country

 

Glossary of Terms

 

  • Bug ID - When an issue is reported to NI, you may be given this ID or find it on ni.com.  You may also find IDs posted by NI on the discussion forums or in KnowledgeBase articles.
  • Legacy ID – An older issue ID that refers to the same issue.  You may instead find this issue ID in older known issues documents.
  • Description - A few sentences which describe the problem. The brief description given does not necessarily describe the problem in full detail.
  • Workaround - Possible ways to work around the problem.
  • Reported Version - The earliest version in which the issue was reported.
  • Resolved Version - Version in which the issue was resolved or was no longer applicable. "N/A" indicates that the issue has not been resolved.
  • Date Added - The date the issue was added to the document (not the reported date).