This document contains the mmWave OTA Validation Test Software known issues that were discovered before and since the release of mmWave OTA Validation Test Software 20.5. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
1097076 |
Error -379415 occurs when initializing an OTA sessionOn the Generator Calibration Settings section of the RF System Calibration Assistant Configurator window, if there is more than one Powers (dBm) selected for a single Frequencies (Hz), the following error occurs: "-379415 Invalid table input. The input frequency array contains a duplicate value." Workaround: Run the RF System Calibration Assistant using only one power level and one reference level.
|
Reported Version: mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5 Resolved Version: N/A Added: Sep 2, 2020 |
|
1097098 |
LoadLibrary failed with error 87If you connect to a system using a remote desktop and open the mmWave OTA Validation Configurator or the mmWave OTA Validation Visualizer, the following error occurs: "LoadLibrary failed with error 87: The parameter is incorrect." The error occurs because there is a Windows problem when running OpenGL hardware-accelerated programs through remote desktop (RDP) sessions. Workaround:
|
Reported Version: mmWave OTA Validation Test Software 1.0 | mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5 Resolved Version: N/A Added: Sep 2, 2020 |
|
1112059 |
An initialization error occurs in TestStand if you run the OTA sequence directly from TestStandIf you open the mmWave OTA Validation Configurator and run the example sequence file directly in the TestStand Sequence Editor instead of starting the sequence from the mmWave OTA Validation Configurator, TestStand returns the following error: “Cannot initialize RF session.” Workaround:
|
Reported Version: mmWave OTA Validation Test Software 20.0 | mmWave OTA Validation Test Software 20.5 Resolved Version: N/A Added: Sep 2, 2020 |
|
1274014 |
Cannot Run the OTA Test when the mmWave OTA Validation Positioner Control Panel is LaunchedIf you run the OTA test when the mmWave OTA Validation Positioner Control Panel is launched, error 56 occurs and the positioner is disconnected. Possible reason(s): The network operation exceeded the user-specified or system time limit.
Workaround:
|
Reported Version: mmWave OTA Validation Test Software 20.5 Resolved Version: N/A Added: Mar 12, 2021 |
|
1248593 |
The mmWave OTA Validation Test Software does not Support the German Operating System due to Driver ErrorWhen you run the OTA examples or use the RF System Calibration Assistant, error -1074097903 occurs because the NI-RFSA/NI-RFSG drivers (19.2.1 or later) required for using the mmWave OTA Validation Test Software do not support the German operating system. Possible reason(s): Data read from non-volatile memory on the device is invalid.
Workaround: Change the operating system language to English or contact NI services for technical support.
|
Reported Version: mmWave OTA Validation Test Software 20.5 Resolved Version: N/A Added: Mar 12, 2021 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.