This document contains the Calibration Executive known issues that were discovered before and since the release of Calibration Executive 6.0. Known issues are performance issues or technical bugs that NI has acknowledged exist within this version of the product.
Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that you may encounter and provide workarounds when possible. Other technical issues that you may encounter could occur through normal product use or system compatibility issues. You may find more information on these issues in NI’s Product Documentation, Knowledgebase, or Community.
Bug Number |
Legacy ID |
Description |
Details |
---|---|---|---|
208070 | 677002 |
Procedure Throws Error or Terminates When a PXIe-4080/4081/4082 DMM Is UsedPXIe-408x digital multimeters are not compatible with several Calibration Executive procedures for which a PXI-407x digital multimeter is recommended as an instrument standard. Workaround: Do not substitute a PXIe-4080/4081/4082 in a procedure in which a PXI-4070/4071/4072 DMM is recommended. |
Reported Version: Calibration Executive 4.2 Resolved Version: N/A Added: Aug 14, 2020 |
1090478 |
NIPM Does Not Enforce Dependencies in Calibration Executive FeedNI Package Manager allows the Calibration Executive feed to be modified to remove core components that are required for certain procedures. Dependencies should not be modified or deselected during installation. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1087818 |
Calibration Executive 6.0 and Later Cannot Be Installed on 32-Bit OSCalibration Executive 6.0 removes 32-bit operating system and application support. If you try to install Calibration Executive 6.0 or later on a 32-bit OS, NI Package Manager attempts to install the package, and then displays an error: Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1057234 |
In Read-Only Mode, Uncertainty Database Error Handling Is ImpairedWhen the uncertainty database is read only (for example, the operator logs in without administrator privileges), an initial attempt to manage uncertainties causes Error -2147217911 when launching the Manage Uncertainty UI, as Calibration Executive tries to update the legacy database.
Workaround: To resolve these errors, you should remove the read-only attribute of the uncertainty database, and then re-open the Manage Uncertainty UI. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Oct 8, 2020 |
|
1094770 |
Error Occurs Generating a Calibration Report on a Non-English OSWindows OS for languages other than English (Spanish, for example) may use a date format with extended characters. In this case, PDF, Word, and Excel reports cannot be generated using default report options; Calibration Executive displays an error: Workaround: Configure report options to remove Calibration Date from the report file name. For details about customizing report names, refer to "Configuring Report Options" in Calibration Executive Help. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1091448 |
Installing Version 6.0 on Japanese OS with Suggested Components Causes ErrorWhen you install Calibration Executive 6.0 on a Japanese operating system and accept the pre-selected additional items, an error occurs at the NI-HSDIO-Documentation package. Workaround: When you install Calibration Executive on a Japanese OS, deselect all NI-HSDIO components from the additional items list in NI Package Manager. In this case, however, the Japanese OS system cannot be used to calibrate high-speed digital I/O hardware. |
Reported Version: N/A Resolved Version: N/A Added: Oct 8, 2020 |
|
1088175 |
Unable to Silently Install Calibration ExecutiveWorkaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1084777 |
Error Occurs If Path to Custom Report Stylesheet or CustomReports Folder ChangesIf the file path for a custom report stylesheet or the CustomReports folder changes (for example, if a user's Documents folder is cleaned up), Calibration Executive displays an error when the report generator is launched: Workaround: Do one of the following:
|
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Aug 14, 2020 |
|
1066924 |
Fatal Error Occurs at AdjustmentNI 4461, NI 6704, and NI 6713 calibration procedures fail to create adjustment sessions and return a fatal error. When this error occurs, you must terminate the procedure and either use Windows Device Manager to disable and then re-enable the PCI or PXI device or shut down the system completely and power it back on again. Workaround: Use a previous version of Calibration Executive to execute the Verify & Adjust run mode on PCI/PXI-4461, -6704, and -6713 devices until a fix becomes available for version 6.0. |
Reported Version: Calibration Executive 6.0 Resolved Version: N/A Added: Oct 8, 2020 |
|
1114526 |
Error -300401 Occurs at "Verify ALC Limits Table," Calibration Results are EmptyCalibration Executive throws an error during the "Verify ALC Limits Table" step for the NI 565x, PXIe-5663, and PXIe-5673 procedures. Before the step is initialized, the procedure terminates and cannot be continued. This issue was introduced in Calibration Executive 6.0.
Workaround: Use a previous version of Calibration Executive to execute NI 565x, PXIe-5663, and PXIe-5673 procedures until a fix becomes available for version 6.0. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.0.1 Added: Aug 27, 2020 |
|
1138884 |
TestStepName Field Is Logged Incorrectly for Framework 2.0 ProceduresThe calibration reports database is logging the TestStepName field as "Adjust Only" for all test points. Workaround: There is currently no known workaround for this issue. |
Reported Version: Calibration Executive 6.0 Resolved Version: Calibration Executive 6.0.1 Added: Oct 8, 2020 |
Issues found in this section will not be listed in future known issues documents for this product.
There are currently no issues to list.