NI-568x 17.0 Known Issues

Overview

This document contains the NI-568x 17.0 known issues that were discovered since the release of NI-568x 17.0. Not every issue known to NI appears on this list; it is intended to show the most severe and common issues that can be encountered.

Each issue appears as a row in the table and includes the following fields:

  • Issue ID - The number in at the top of each of the cells in the first column. When you report an issue to NI, you may be given this ID, you can also find IDs posted by NI on the discussion forums or in Knowledge Base articles.  "N/A" indicates that there is no ID assigned to the issue.
  • Issue Title (in italics) - Describes the issue in one sentence or less.
  • Problem Description - A few sentences which describe the problem in further detail. The brief description given does not necessarily describe the problem in full detail, and it is expected that you may want more information on an issue. If you would like more information on an issue, contact NI and reference the ID number given in the document.
  • Workaround - Possible ways to work around the problem. The workarounds that appear in the document are not always tested by NI and are not guaranteed to resolve the issue. If a workaround refers you to the NI KnowledgeBase, visit www.ni.com/kb/ and enter the KnowledgeBase number in the search field to locate the specific document.
  • Reported Version - The earliest version of NI-568x in which the issue was reported. If you discover the issue appears in an earlier version of NI-568x than is reported in this field, report the discrepancy to NI to have the field updated.
  • Resolved Version - Version in which the issue was resolved or was no longer applicable. "N/A" indicates that the issue has not been resolved.
  • Date Added - The date the issue was added to the document (not the reported date).

Known Issues

666625Installing the NI-568x 17.0 driver with default LabVIEW settings prevents calling NI-568x VIs when using LabVIEW 2016 (64-bit)
413281The NI-568x Soft Front Panel (SFP) crashes when the computer wakes from standby mode
552187Rapidly disconnecting and reconnecting NI 568x devices may cause Measurement & Automation Explorer (MAX) to timeout.
542379Manually setting COM ports is not supported
480524Windows 8.0 to Windows 8.1 upgrade issue
544015The NI 5680/5681 may not appear in MAX without a refresh on some USB ports
545868The NI 5681 may fail to recognize an external trigger following a command to initiate the device
522841The NI 5681 may become unresponsive if the ni568x Fetch VI is aborted during execution
536935Calling the ni568x Initiate VI while zeroing the NI 5680 delays error reporting
536460Calling advanced functions on the NI 5680/5681 may return an unexpected error
545649The NI 5683/5684 may not receive all external triggers
412319Initializing the NI 568x devices in Windows 8 with high memory addressing enabled results in an error
412196Disconnecting an NI 5681 device while a measurement is in progress causes unexpected behavior
411345The NI-568x SFP cannot connect with the power sensor after the device is disconnected and reconnected.
393077The NI-568x RF power meter instrument driver returns an error during ni568x Fetch after calling niVNA Initialize
258340Closing a session to an NI 568x device takes longer to complete if the device is still waiting on a trigger
IDKnown Issue
666625Installing the NI-568x 17.0 driver with default LabVIEW settings prevents calling NI-568x VIs when using LabVIEW 2016 (64-bit).
When installing the NI-568x 17.0 driver and selecting the default LabVIEW (64-bit) settings, you cannot access the NI-568x VIs when using LabVIEW 2016 (64-bit).

Workaround: If you have not yet installed the NI-568x 17.0 driver, during installation unselect the latest LabVIEW version and select the version of LabVIEW you wish to use.

If you have already installed the NI-568x 17.0 driver, complete the following steps:
  1. Run the installer again, and select "LabVIEW (latest version)."
  2. In the drop down menu, select "Remove this feature."
  3. Complete the installation, and restart your computer.
  4. Select Start>>Control Panel>>Programs and Features>>National Instruments>>NI-568x 17.0.
  5. Select Repair.

If you have NI Package Manager installed, run the setup.exe from the original installation media or download, unselect the latest LabVIEW version, and select the version of LabVIEW you wish to use.
Reported Version: 17.0  Resolved Version: 18.0  Added: 3/8/2018
413281The NI-568x Soft Front Panel (SFP) crashes when the computer wakes from standby mode.
If you put the computer in standby mode while the NI-568x SFP is open, the SFP crashes when the computer wakes.

Workaround: Close the SFP and any other applications that have an open session to the power sensor before going into standby mode.

Reported Version: 1.2  Resolved Version: N/A  Added: 8/31/2015
552187Rapidly disconnecting and reconnecting NI 568x devices may cause Measurement & Automation Explorer (MAX) to timeout.
MAX requires time to process NI 5683/5684 connections and disconnections. Connecting multiple devices in quick succession or reconnecting the same device shortly after disconnecting may cause MAX to return an error or timeout.

Workaround: Restart MAX or restart the host computer.

Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
542379Manually setting COM ports is not supported.
Windows Device Manager allows changing COM ports manually. Manually changing COM ports may produce undefined behavior, especially when trying to initialize the NI 5680/5681.

Workaround: Use the COM port number automatically assigned by Windows when the device is connected. Use NI Measurement & Automation Explorer (MAX) to rename a device if desired.

Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
480524
Windows 8.0 to Windows 8.1 upgrade issue.
Upgrading the version of Windows on a system with NI-568x installed may result in the loss of device names, the loss of device configuration, and/or devices appearing disconnected. For more information about how to resolve this issue, visit ni.com/info and enter the Info Code excrxm.

Workaround: N/A

Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
544015The NI 5680/5681  may not appear in MAX without a refresh on some USB ports.
The NI 5680/5681 may appear in MAX with a red 'X' after the device is initially plugged in. This may happen on some USB ports on some PCs.

Workaround: In the MAX configuration tree, select Devices and Interfaces to see the list of devices, and press <F5> to refresh the list.
Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
545868The NI 5681 may fail to recognize an external trigger following a command to initiate the power sensor.
On fast host computers it is possible to complete an initiate call before the NI 5681 is ready for an external trigger. This may occur when using the program flow to initiate an external trigger following a command to initiate the power sensor.

Workaround: After calling initiate on the power sensor, call a 1 ms delay before initiating the external trigger.

Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
522841The NI 5681 may become unresponsive if the ni568x Fetch VI is aborted during execution.
Pressing the LabVIEW abort button or resetting the host computer while the NI 5681 is executing the ni568x Fetch VI may make the device unresponsive.

Workaround: Power cycle the NI 5681 by disconnecting and reconnecting the device or by restarting the host computer.

Reported Version:15.0  Resolved Version: N/A  Added: 8/31/2015
544015Calling the ni568x Initiate VI while zeroing the NI 5680 delays error reporting.
Calling the ni568x Initiate VI while zeroing the NI 5680 is not supported. Error -1074119641 occurs when  you make the next call to the driver. For instance, calling the ni568x Zero and then ni568x Initiate does not return an error. However, if you were to call ni568x Is Zero Complete next, Error -1074119641 occurs.

Workaround: When calling ni568x Zero with the NI 5680, you must confirm that the zeroing operation is complete with ni568x Is Zero Complete before calling ni568x Initiate.
Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
536460Calling advanced functions on the NI 5680/5681 may return an unexpected error.
When calling a function that is not supported by the NI 5680/5681, the error returned may state that the attribute is not supported, instead of stating that the function is not supported.

Workaround: N/A

Reported Version: 15.0  Resolved Version: N/A  Added: 8/31/2015
545649The NI 5683/5684 may not receive all external triggers.
The NI 5683/5684 does not receive approximately 0.1% of external triggers sent to the device.

Workaround: Use repeating external triggers when possible to increase trigger detection. If the NI 5683/5684 misses a trigger, it will most likely receive the following trigger. A firmware fix for this issue is in work.

Reported Version:15.0  Resolved Version: N/A  Added: 8/31/2015
412319Initializing the NI 568x devices in Windows 8 with high memory addressing enabled results in an error.
Initializing the NI 568x devices in Windows 8 with high memory addressing enabled results in the following error:
ni568x Initialize With Options.vi<ERR>
ni568x Initialize With Options.vi<ERR>
Driver Status: (Hex 0xBFFA0060) COM4

Workaround: Disable high memory addressing in Windows 8 systems.

Reported Version:1.2  Resolved Version: N/A  Added: 8/31/2015
412196Disconnecting the NI 5681 while a measurement is in progress causes unexpected behavior.
When you disconnect the NI 5681 while it is executing a measurement operation, such as averaging, subsequent calls to the Is Measurement Complete VI/function returns measurement status unknown without an error. If Is Measurement Complete is called within a loop, the code could hang because it is waiting for measurement complete or an error to be returned. In LabVIEW, if you abort the VI after encountering this issue, initialization will take a long time to run and returns error -1073807345, even after you reconnect the device.

Workaround: Abort the executing code and reconnect the device to a different USB port.

Reported Version:1.2  Resolved Version: N/A  Added: 8/31/2015
411345The NI-568x SFP cannot connect with the power sensor after the device is disconnected and reconnected.
When you disconnect and then reconnect the power meter with the NI-568x SFP before the SFP can release the handle (before "No Sensor" is displayed), the device enters a state in which it cannot be connected either through the SFP or any other development environment. When you attempt to open a session to the device, the system returns an error of BFFF0011 (-1073807343).

Workaround: The two workarounds for this issue are as follows: 1) If the device was running at the time of disconnection, wait until "No Sensor" is shown in the SFP when you unplug the device from the system. 2) Unplug and then reconnect the device to the system while the SFP is running. This causes the SFP to reconnect properly to the device and to continue with normal operation.

Reported Version: 1.2  Resolved Version: N/A  Added: 8/31/2015
393077When using the NI-568x instrument driver, the ni568x Fetch VI or ni568x_Fetch function returns an error after calling the niVNA Initialize VI or niVNA_init function if the RF power meter is initialized before the NI VNA device.
Words.

Workaround: Call the niVNA Initialize VI or niVNA_init function before calling the ni568x Initialize VI or ni568x_init function.

Reported Version: Unknown  Resolved Version: N/A  Added: 8/31/2015
258340Closing a session to an NI 568x device takes longer to complete if the device is still waiting on a trigger.
If you attempt to close a session to the NI 568x device while the device is waiting to receive a trigger, the call to close can take up to 2 minutes to execute

Workaround: Send the trigger before closing the session.

Reported Version: 1.1  Resolved Version: N/A  Added: 8/31/2015

Contacting NI

Contact NI regarding this document or issues in the document. If you contact NI in regards to a specific issue, reference the ID number given in the document. The ID number contains the current issue ID number as well as the legacy ID number (use the current ID number when contacting NI). You can contact us through any of the normal support channels including phone, email, or the discussion forums. Visit the NI Website to contact us. Also contact us if you find a workaround for an issue that is not listed in the document.

Glossary of Terms

 

  • Bug ID - When an issue is reported to NI, you may be given this ID or find it on ni.com.  You may also find IDs posted by NI on the discussion forums or in KnowledgeBase articles.
  • Legacy ID – An older issue ID that refers to the same issue.  You may instead find this issue ID in older known issues documents.
  • Description - A few sentences which describe the problem. The brief description given does not necessarily describe the problem in full detail.
  • Workaround - Possible ways to work around the problem.
  • Reported Version - The earliest version in which the issue was reported.
  • Resolved Version - Version in which the issue was resolved or was no longer applicable. "N/A" indicates that the issue has not been resolved.
  • Date Added - The date the issue was added to the document (not the reported date).