The following items are known issues in the LabVIEW 2018 ELVIS III Toolkit.
716129 | — | IV Analyzer Read does not produce accurate diode characteristic curves. |
726398 | — | The Logic Configure Pulse Output VI introduces an output delay on the pattern generator under certain conditions. |
ID | Known Issue | |||||
---|---|---|---|---|---|---|
716129 Return | IV Analyzer Read does not produce accurate diode characteristic curves. IV 2Wire Read.vi and IV 3Wire Read.vi do not produce valid characteristic curves for diodes. Solution: Update to LabVIEW 2018 ELVIS III Toolkit SP2 or later.
| |||||
726398 Return | The Logic Configure Pulse Output VI introduces an output delay on the pattern generator under certain conditions. There is an output delay on the pattern generator when generating a constant logic high signal using Logic Configure Pulse Output.vi with the Low Time (s) input set to 0. The maximum delay of about 6.5 seconds is expected when the High Time (s) input is equal to 1. Workaround: Set the High Time (s) input equal to a smaller number to reduce the pattern generator output delay.
|