PXIe-5830 Getting Started Guide
- Updated2023-02-20
- 17 minute(s) read
PXIe-5830 Getting Started Guide
This document explains how to install, configure, and test the PXIe-5830. The PXIe-5830 is an RF vector signal transceiver (VST). You can program the PXIe-5830 instrument using RFmx or NI-RFSA and NI-RFSG driver software.
The PXIe-5830 instrument configuration comprises the following modules:
- PXIe-5820 Vector Signal Transceiver
- PXIe-3621 Vector Signal Up/Down Converter
There is no single instrument labeled "PXIe-5830."
Verifying the System Requirements
To use the PXIe-5830, your system must meet certain requirements. For more information about minimum system requirements, recommended system, and supported application development environments (ADEs), refer to the readme, which is installed or available at ni.com/manuals.
Unpacking the Kit
- Touch the antistatic package to a metal part of the computer chassis.
-
Remove the device from the package and inspect the device for loose components or any other sign of damage.
Notice Never touch the exposed pins of connectors.Note Do not install a device if it appears damaged in any way.
- Unpack any other items and documentation from the kit.
Verifying the Kit Contents
- PXIe-3621 Module
- PXIe-5820 Module
- SMA (m)-to-SMA (m) Cable, 12 in.
- MMPX (m)-to-MMPX (m) Cable, 40 mm (x8)
- MMCX (m)-to-MMCX (m) Cable, 87 mm
- Tweezers (145455B-01)
- Screwdriver, Part Number 772006-01
- Maintain Forced-Air Cooling Note to Users
- PXIe-5830 Getting Started Guide
- PXIe-3621 Safety, Environmental, and Regulatory Information
- PXIe-5820 Safety, Environmental, and Regulatory Information
Other Equipment
There are required items not included in your kit that you need to install or operate the PXIe-5830. Your application may require additional items not included in your kit to install or operate your hardware.
Required Items
- A PXI Express chassis and chassis documentation. To ensure specifications are valid, NI recommends using the PXIe-1095 Chassis or PXIe-1092 Chassis for your hardware. For more information about compatible chassis options, refer to ni.com.
- A PXI Express embedded controller or MXI controller system that meets the system requirements specified in this guide and chassis documentation.
Optional Items
- A 0.9 N · m (8 lb · in.) calibrated torque wrench, 5/16 in. (NI part number 751120-01).
Preparing the Environment
Ensure that the environment you are using the PXIe-5830 in meets the following specifications.
Operating ambient temperature | 0 °C to 45 °C |
Operating relative humidity | 10% to 90%, noncondensing (IEC 60068-2-56) |
Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Choosing and Installing the Software
Software Options
NI provides three software options for programming the PXIe-5830—RFmx, NI-RFSA and NI-RFSG instrument driver software.
Software Option | Description | Use Case |
---|---|---|
RFmx |
Provides a single-handle instrument driver with built-in measurements. You can use instrument driver FPGA extensions to customize the device FPGA using precompiled bitfiles. |
Use RFmx SpecAn to perform spectral measurements. Use RFmx Demod to perform modulation quality measurements, such as EVM. Use RFmx personalities to perform physical layer analysis on supported communications and connectivity signals. |
NI-RFSA and NI-RFSG instrument driver |
The NI-RFSA and NI-RFSG APIs each provide a fixed API for interacting with your hardware. NI-RFSA and NI-RFSG provide standard IVI-based functionality needed for most vector signal analyzer and vector signal generator applications. You can use NI-RFSA and NI-RFSG instrument driver FPGA extensions to customize the device FPGA using precompiled bitfiles. |
Use NI-RFSA to create custom measurements or applications that require I/Q data. Use NI-RFSG to generate continuous wave (CW) or modulated waveforms. Use NI-RFSA and NI-RFSG with the NI Modulation Toolkit to develop software-defined radio transmitters and receivers. Use NI-RFSG with the NI Modulation Toolkit to create and generate modulated signals to test receivers. |
Installing the Software
You must be an Administrator to install NI software on your computer.
- Install an ADE, such as LabVIEW or LabWindows™/CVI™.
-
Download the driver software installer from ni.com/downloads.
NI Package Manager downloads with the driver software to handle the installation. Refer to the NI Package Manager Manual for more information about installing, removing, and upgrading NI software using NI Package Manager.
-
Follow the instructions in the installation prompts.
Note Windows users may see access and security messages during installation. Accept the prompts to complete the installation.
- When the installer completes, select Restart in the dialog box that prompts you to restart, shut down, or restart later.
Installing the PXIe-5830
You must install the software before installing the hardware.
Before you install the hardware, refer to the guidelines in the Maintain Forced-Air Cooling Note to Users included with the module to ensure that the device can cool itself effectively.
To use the included cables, you must install the PXIe-5820 immediately to the left of the PXIe-3621.
-
Ensure the AC power source is connected to the chassis before installing the module.
The AC power cord grounds the chassis and protects it from electrical damage while you install the module.
- Power off the chassis.
- Inspect the slot pins on the chassis backplane for any bends or damage prior to installation. Do not install a module if the backplane is damaged.
- Remove the black plastic covers from all the captive screws on the module front panel.
-
Identify a supported slot in the chassis. The following figure shows the symbols that indicate the slot types.
Figure 1. Chassis Compatibility Symbols
- PXI Express System Controller Slot
- PXI Peripheral Slot
- PXI Express Hybrid Peripheral Slot
- PXI Express System Timing Slot
- PXI Express Peripheral Slot
- Touch any metal part of the chassis to discharge static electricity.
- Ensure that the ejector handle is in the downward (unlatched) position.
-
Place the module edges into the module guides at the top and bottom of the chassis. Slide the module into the slot until it is fully inserted.
Figure 1. Module Installation
- Chassis
- Hardware Module
- Ejector Handle in Downward (Unlatched) Position
- Latch the module in place by pulling up on the ejector handle.
-
Secure the module front panel to the chassis using the front-panel mounting screws.
Note Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis, which can improve the signal quality and electromagnetic performance.
-
Cover all empty slots using either filler panels (standard or EMC) or slot
blockers with filler panels, depending on your application.
Note For more information about installing slot blockers and filler panels, go to ni.com/r/pxiblocker.
- Power on the chassis.
Interconnecting the PXIe-5830 Modules
Take the following considerations into account when interconnecting the PXI Express modules that comprise the PXIe-5830.
- Inspect all RF connectors before attaching mating cables. Clean with dry compressed air, if necessary, to remove any contaminants that may be present.
- Position RF cables with instrument connectors to ensure center pins are straight and aligned before mating.
- Press MMPX connectors until they noticeably snap into place.
- For all threaded RF connectors (SMA and SMA 27 GHz), engage and finger-tighten the connector nuts. The connector nuts should thread easily without binding, but if not, ensure the connector is straight and that center pins have not been damaged.
- Carefully complete tightening all threaded RF connectors to 0.9 N · m(8 lb · in) using an appropriate torque wrench (not included) or torque screwdriver and driver bit. Tighten only until the wrench clicks.
Connecting the PXIe-5820 and PXIe-3621
Complete the following steps for the baseband signal connections and Reference Clock connection, as shown in the following figure.
- MMPX (m)-to-MMPX (m) Cable, 40 mm
- MMCX (m)-to-MMCX (m) Cable, 87 mm
-
Complete the following baseband signal cable connections between the PXIe-5820 and PXIe-3621 front panels using the
included eight 40 mm MMPX (m)-to-MMPX (m) cables
for transmit and receive I/Q signals.
Table 2. Baseband Signal Cable Connections PXIe-5820 Connector PXIe-3621 Connector I/Q OUT I+ I/Q IN I+ I/Q OUT I- I/Q IN I- I/Q OUT Q+ I/Q IN Q+ I/Q OUT Q- I/Q IN Q- I/Q IN I+ I/Q OUT I+ I/Q IN I- I/Q OUT I- I/Q IN Q+ I/Q OUT Q+ I/Q IN Q- I/Q OUT Q- -
Use the 87 mm MMCX (m)-to-MMCX (m)
cable to connect the PXIe-5820 REF IN
front panel connector to the PXIe-3621
REF OUT front panel connector.
This completes the Reference Clock connection.
Direct Connections to the PXIe-5830
The PXIe-5830 is a precision RF instrument that is sensitive to ESD and transients. Ensure you are making proper direct connections to the PXIe-5830 to avoid damaging the hardware.
- Ensure you are properly grounded when manipulating cables or antennas connected to the instrument.
- If you are using noninsulated devices, such as an RF antenna, ensure the devices are maintained in a static-free environment.
- If you are using an active device, such as a preamplifier or switch routed to the PXIe-5830, ensure that the device cannot generate signal transients greater than the RF and DC specifications of the PXIe-5830.
Configuring the PXIe-5830 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX informs other programs about which NI hardware products are in the system and how they are configured. MAX is automatically installed with NI-RFSA, NI-RFSG, and RFmx.
- Launch MAX.
-
In the configuration tree, expand
Devices and Interfaces to see the list of installed NI hardware.
Installed modules appear under the name of their associated chassis.
-
Expand your
Chassis tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note If you do not see your module listed, press <F5> to refresh the list of installed modules. If the module is still not listed, power off the system, ensure the module is correctly installed, and restart. - Record the identifier MAX assigns to the PXIe-3621. Use this identifier when programming the PXIe-5830.
-
Associate the hardware modules that comprise your instrument.
- Select the PXIe-3621 that is identified as not configured in the configuration tree.
- In the Associated Devices section, select the PXIe-5820 from the Baseband list.
-
Click
Save in the MAX toolbar.
Note Module associations may be lost when you move the modules to different chassis slots.
-
Self-test the hardware by selecting the item in the configuration tree and clicking
Self-Test in the MAX toolbar. Repeat this step for all modules in your PXIe-5830 system.
The MAX self-test performs a basic verification of hardware resources.
Self-Calibration
Self-calibration adjusts the PXIe-5830 for variations in the module environment using onboard power detectors.
PXIe-5830 modules are externally calibrated at the factory; however, you should perform a self-calibration in any of the following situations:
- After first installing and interconnecting your PXIe-5830 instrument
- After any module in the chassis is installed, uninstalled, or moved
- After inter-module cabling has been changed, reconnected, or repositioned
- After changing controllers, computers, or reinstalling the driver software
- When the system is in an environment where the ambient temperature varies or the module temperature has drifted more than ±5 °C from the temperature at the last self-calibration
- To periodically adjust for small performance drifts that occur with product aging
Performing a Self-Calibration Using NI-RFSA
- Select to launch LabVIEW.
-
Launch the Example Finder and navigate to the example.
- Select .
- Open the example.
-
Complete the following steps to configure the example:
- Select the device identifier assigned to the PXIe-3621 in MAX in the [Resource Name] drop-down menu.
- Set Clock Source to OnboardClock.
- Set Self Calibration Step Operations to Perform All Self Calibration Steps.
- Run the VI.
Locating the Software and Examples
Software Locations
Software Option | ADE | Location |
---|---|---|
RFmx | LabVIEW |
Available on the LabVIEW Functions palette at . |
LabWindows/CVI |
RFmx functions are available from the LabWindows/CVILibrary menu at and . |
|
Microsoft Visual C/C++ |
Use the header files located in the <NIDir>\Shared\ExternalCompilerSupport\C\include directory and import library files located in one of the following directories:
where <NIDir> is one of the following locations:
|
|
Microsoft .NET |
For the location of .NET class libraries, refer to the installed RFmx readme. |
|
NI-RFSA | LabVIEW |
Available on the LabVIEW Functions palette at . |
LabWindows/CVI |
Available in the <IVIROOTDIR32>\Drivers\niRFSA directory, where <IVIROOTDIR32> is one of the following locations:
|
|
Microsoft Visual C/C++ |
Use the header files located in the
<IVIROOTDIR32>\Include directory and import library files located in the
<IVIROOTDIR32>\Lib directory, where
<IVIROOTDIR32> is one of the following directories:
|
|
Microsoft .NET |
Use the NI-RFSA .NET class library by adding a reference to NationalInstruments.ModularInstruments.NIRfsa.Fx40 or NationalInstruments.ModularInstruments.NIRfsa.Fx45 and any dependent class libraries from within the Solution Explorer in Visual Studio. |
|
NI-RFSG | LabVIEW |
Available on the LabVIEW Functions palette at . |
LabWindows/CVI |
Available in the <IVIROOTDIR32>\Drivers\niRFSG directory, where <IVIROOTDIR32> is one of the following locations:
|
|
Microsoft Visual C/C++ |
Use the header files located in the <IVIROOTDIR32>\Include directory and import library files located in the <IVIROOTDIR32>\Lib directory, where <IVIROOTDIR32> is one of the following directories:
|
|
Microsoft .NET |
Use the NI-RFSG .NET class library by adding a reference to NationalInstruments.ModularInstruments.NIRfsg.Fx40 or NationalInstruments.ModularInstruments.NIRfsg.Fx45 and any dependent class libraries from within the Solution Explorer in Visual Studio. |
Programming Examples Locations
Using the NI Example Finder
If you are using RFmx, NI-RFSA, or NI-RFSG with LabVIEW or LabWindows/CVI, use the NI Example Finder to locate programming examples.
- Launch LabVIEW or LabWindows/CVI.
- Select to open the NI Example Finder.
- Navigate to .
- Open the example that best matches your application requirements.
Using Microsoft Visual C/C++
If you are using RFmx or NI-RFSA with Microsoft Visual C/C++, locate examples in the following directories.
Software Option | Examples Location |
---|---|
RFmx |
<NIDocDir>\RFmx\Demod\Examples <NIDocDir>\RFmx\SpecAn\Examples where <NIDocDir> is the Users\Public\Public Documents\National Instruments directory. |
NI-RFSA |
<NIDocDir>\NI-RFSA\examples where <NIDocDir> is the Users\Public\Public Documents\National Instruments directory. |
Making a First Measurement
You can verify proper installation and configuration of your instrument by making a measurement using the NI-RFSG and RFmx Soft Front Panels (SFPs).
- Connect the PXIe-5830IF IN/OUT 0 and IF IN/OUT 1 ports as appropriate for your application.
- Select NI-RFSG SFP. to launch the
- Select the device identifier assigned to the PXIe-3621 in MAX in the Device drop-down menu.
- Set the RF Frequency to 5 GHz.
- Click the RF On/Off button to generate a continuous wave (CW) tone RF signal at the IF output terminal of your instrument.
- Select to launch the RFmx SFP.
- Select the device identifier assigned to the PXIe-3621 in MAX in the Device drop-down menu.
- Set the Center Frequency to 5 GHz
- Verify that the RFmx SFP is receiving the CW tone.
- Close the RFmx SFP and NI-RFSG SFP.
Troubleshooting
If an issue persists after you complete a troubleshooting procedure, search our KnowledgeBase for additional information our technical support engineers create as they answer common user questions and resolve unexpected issues.
What Should I Do if the PXIe-3621 Does Not Appear in MAX?
- In the MAX configuration tree, expand Devices and Interfaces.
- Expand the Chassis tree to see the list of installed hardware, and press <F5> to refresh the list.
- If the module is still not listed, power off the system, ensure that all hardware is correctly installed, and restart the system.
-
Navigate to the Device Manager.
Operating System Description Windows 10/8.1 Right-click the Start button, and select Device Manager. Windows 7 Select . -
Verify the PXIe-3621 appears in the Device Manager.
-
Under
NI Modular Instruments, confirm that a PXIe-3621 entry appears.
Note If you are using a PC with a device for PXI remote control system, under System Devices, also confirm that no error conditions appear for the PCI-to-PCI Bridge.
- If error conditions appear, reinstall the driver software and the PXIe-3621.
-
Under
NI Modular Instruments, confirm that a PXIe-3621 entry appears.
Why Is the ACCESS LED Off When the Chassis Is On?
The LEDs may not light until the module has been configured in MAX. Before proceeding, verify that the module appears in MAX.
- Disconnect any signals from the PXI Express module front panel.
- Power off the PXI Express chassis.
- Remove the module from the PXI Express chassis and inspect it for damage. Do not reinstall a damaged device.
- Install the module in a different PXI Express chassis slot from which you removed it.
- Power on the PXI Express chassis.
- Restart your computer.
- Verify that the module appears in MAX.
What Should I Do if the PXIe-5830 Fails the Self-Test?
- Restart the system.
- Launch MAX, and perform the self-test again.
- Power off the chassis.
- Reinstall the failed module in a different slot.
- Power on the chassis.
- Perform the self-test again.
What Should I Do if the Instrument Does Not Initialize?
Failure to initialize may indicate a problem with module interconnections or with MAX.
If the niRFSA Initialize VI or niRFSG Initialize VI returns an error and the PXIe-5830 fails to initialize, complete the following steps:
- Power off the system.
- Inspect all PXIe-5830 front panel cables and, if there are no apparent issues, reconnect cables securely, according to the interconnection procedure for your instrument.
- Power on your system and run the MAX configuration and intermediate self-test procedures.
Hardware Connectors and Indicators
PXIe-5820
Connector | Description | Connector Type | |
---|---|---|---|
REF IN | Input connector that allows for the use of an external 10 MHz Reference Clock. | MMPX (f) | |
REF OUT | Output connector that can export a 10 MHz Reference Clock. | MMPX (f) | |
PFI 0 | Programmable-function digital I/O (DIO) connector for use with triggers or events. | MMPX (f) | |
DIO | Multi-signal DIO connector that provides access to FPGA multi-gigabit transceivers (MGTs) and general purpose LVCMOS signals. | Nano-Pitch I/O |
Connector | Description | Connector Type | |
---|---|---|---|
I/Q OUT | I+ | Output connector for I+ signals. | MMPX (f) |
I- | Output connector for I- signals. | MMPX (f) | |
Q+ | Output connector for Q+ signals. | MMPX (f) | |
Q- | Output connector for Q- signals. | MMPX (f) | |
I/Q IN | I+ | Input connector for I+ signals. | MMPX (f) |
I- | Input connector for I- signals. | MMPX (f) | |
Q+ | Input connector for Q+ signals. | MMPX (f) | |
Q- | Input connector for Q- signals. | MMPX (f) |
LED | Indications |
---|---|
ACCESS |
Indicates the basic hardware status of the device. Off—The device is not yet functional or has detected a problem with a PXI Express power rail. Amber—The device is being accessed. Accessed means that you are writing to the device setup registers to control the device, reading from the device to monitor the device status, or transferring data to/from the device. Green—The device is controllable through the software. |
ACTIVE |
Off—The device is idle. Solid green—The device is generating a waveform. Dim amber—The device is waiting for an acquisition Reference Trigger. Solid amber—The device is acquiring a waveform. Solid red—The device has detected an error. The LED remains red until the error condition is removed. Note The indicators are listed in increasing order of priority. For example, if you are generating a waveform using NI-RFSG and waiting on an acquisition Reference Trigger in NI-RFSA, the LED is dim amber.
|
PXIe-3621 Front Panel and LEDs
Connector | Description | Connector Type | |
---|---|---|---|
I/Q IN | I+ | Input connector for I+ signals. | MMPX (f) |
I- | Input connector for I- signals. | MMPX (f) | |
Q+ | Input connector for Q+ signals. | MMPX (f) | |
Q- | Input connector for Q- signals. | MMPX (f) | |
I/Q OUT | I+ | Output connector for I+ signals. | MMPX (f) |
I- | Output connector for I- signals. | MMPX (f) | |
Q+ | Output connector for Q+ signals. | MMPX (f) | |
Q- | Output connector for Q- signals. | MMPX (f) |
Connector | Description | Connector Type | |
---|---|---|---|
LO2 | IN | Input connector for the external local oscillator signal used for the up and down conversion between baseband and IF. | MMPX (f) |
OUT | Output connector for the local oscillator signal used for the up and down conversion between baseband and IF. | MMPX (f) | |
IF IN/OUT | 0 | IF0 bi-directional connector for both input and output. | SMA 27 GHz (f) |
1 | IF1 bi-directional connector for both input and output. | SMA 27 GHz (f) | |
REF | IN | Input connector for the Reference Clock. | MMPX (f) |
OUT | Output connector for the Reference Clock. | MMPX (f) | |
RESERVED | Reserved connectors are not enabled for use on the PXIe-5830 and should remain capped with 50 Ω terminators to ensure specified instrument performance. | — |
LED | Indications |
---|---|
ACCESS |
Indicates the basic hardware status of the module. |
ACTIVE |
|
NI Services
Visit ni.com/support to find support resources including documentation, downloads, and troubleshooting and application development self-help such as tutorials and examples.
Visit ni.com/services to learn about NI service offerings such as calibration options, repair, and replacement.
Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.
NI corporate headquarters is located at 11500 N Mopac Expwy, Austin, TX, 78759-3504, USA.
In This Section
- Verifying the System Requirements
- Unpacking the Kit
- Preparing the Environment
- Choosing and Installing the Software
- Installing the PXIe-5830
- Direct Connections to the PXIe-5830
- Configuring the PXIe-5830 in MAX
- Self-Calibration
- Locating the Software and Examples
- Making a First Measurement
- Troubleshooting
- Hardware Connectors and Indicators
- NI Services
Related Information
- Refer to the RFmx SpecAn Help, the RFmx Demod Help, the NI RF Vector Signal Analyzers Help, or the NI RF Signal Generators Help for more information about using the RFmx, NI-RFSA, and NI-RFSG instrument driver FPGA extensions.
- Installing the Software
- Interconnecting the PXIe-5830 Modules
Take the following considerations into account when interconnecting the PXI Express modules that comprise the PXIe-5830.
- Troubleshooting
If an issue persists after you complete a troubleshooting procedure, search our KnowledgeBase for additional information our technical support engineers create as they answer common user questions and resolve unexpected issues.