PXIe-5830 Getting Started Guide

Note Before you begin, install and configure your chassis and controller.

This document explains how to install, configure, and test the PXIe-5830. The PXIe-5830 is an RF vector signal transceiver (VST). You can program the PXIe-5830 instrument using RFmx or NI-RFSA and NI-RFSG driver software.

The PXIe-5830 instrument configuration comprises the following modules:

  • PXIe-5820 Vector Signal Transceiver
  • PXIe-3621 Vector Signal Up/Down Converter

There is no single instrument labeled "PXIe-5830."

Verifying the System Requirements

To use the PXIe-5830, your system must meet certain requirements. For more information about minimum system requirements, recommended system, and supported application development environments (ADEs), refer to the readme, which is installed or available at ni.com/manuals.

Unpacking the Kit

Notice To prevent electrostatic discharge (ESD) from damaging the device, ground yourself using a grounding strap or by holding a grounded object, such as your computer chassis.
  1. Touch the antistatic package to a metal part of the computer chassis.
  2. Remove the device from the package and inspect the device for loose components or any other sign of damage.
    Notice Never touch the exposed pins of connectors.
    Note Do not install a device if it appears damaged in any way.
  3. Unpack any other items and documentation from the kit.
Store the device in the antistatic package when the device is not in use.

Verifying the Kit Contents


1378

  1. PXIe-3621 Module
  2. PXIe-5820 Module
  3. SMA (m)-to-SMA (m) Cable, 12 in.
  4. MMPX (m)-to-MMPX (m) Cable, 40 mm (x8)
  5. MMCX (m)-to-MMCX (m) Cable, 87 mm
  6. Tweezers (145455B-01)
  7. Screwdriver, Part Number 772006-01
  8. Maintain Forced-Air Cooling Note to Users
  9. PXIe-5830 Getting Started Guide
  10. PXIe-3621 Safety, Environmental, and Regulatory Information
  11. PXIe-5820 Safety, Environmental, and Regulatory Information
Notice To prevent damage while disconnecting the right-angle cables, use specified tweezers (145455B-01) to remove them by grasping only the metal end of the connector. Do not grasp or pull on non-metallic portions of the cables.

Other Equipment

There are required items not included in your kit that you need to install or operate the PXIe-5830. Your application may require additional items not included in your kit to install or operate your hardware.

Required Items
  • A PXI Express chassis and chassis documentation. To ensure specifications are valid, NI recommends using the PXIe-1095 Chassis or PXIe-1092 Chassis for your hardware. For more information about compatible chassis options, refer to ni.com.
  • A PXI Express embedded controller or MXI controller system that meets the system requirements specified in this guide and chassis documentation.
Optional Items
  • A 0.9 N · m (8 lb · in.) calibrated torque wrench, 5/16 in. (NI part number 751120-01).

Preparing the Environment

Ensure that the environment you are using the PXIe-5830 in meets the following specifications.

Operating ambient temperature

0 °C to 45 °C

Operating relative humidity

10% to 90%, noncondensing (IEC 60068-2-56)

Maximum altitude

2,000 m (800 mbar) (at 25 °C ambient temperature)

Pollution Degree

2

Indoor use only.

Notice Clean the hardware with a soft, nonmetallic brush or lint free cloth. Make sure that the hardware is completely dry and free from contaminants before returning it to service.
Note Refer to the PXIe-5830 Specifications at ni.com/manuals for complete specifications.

Choosing and Installing the Software

Software Options

NI provides three software options for programming the PXIe-5830RFmx, NI-RFSA and NI-RFSG instrument driver software.

Table 1. PXIe-5830 Software Options
Software Option Description Use Case
RFmx

Provides a single-handle instrument driver with built-in measurements.

You can use instrument driver FPGA extensions to customize the device FPGA using precompiled bitfiles.

Use RFmx SpecAn to perform spectral measurements.

Use RFmx Demod to perform modulation quality measurements, such as EVM.

Use RFmx personalities to perform physical layer analysis on supported communications and connectivity signals.

NI-RFSA and NI-RFSG instrument driver

The NI-RFSA and NI-RFSG APIs each provide a fixed API for interacting with your hardware.

NI-RFSA and NI-RFSG provide standard IVI-based functionality needed for most vector signal analyzer and vector signal generator applications.

You can use NI-RFSA and NI-RFSG instrument driver FPGA extensions to customize the device FPGA using precompiled bitfiles.

Use NI-RFSA to create custom measurements or applications that require I/Q data.

Use NI-RFSG to generate continuous wave (CW) or modulated waveforms.

Use NI-RFSA and NI-RFSG with the NI Modulation Toolkit to develop software-defined radio transmitters and receivers.

Use NI-RFSG with the NI Modulation Toolkit to create and generate modulated signals to test receivers.

Installing the Software

You must be an Administrator to install NI software on your computer.

  1. Install an ADE, such as LabVIEW or LabWindows™/CVI™.
  2. Download the driver software installer from ni.com/downloads.
    NI Package Manager downloads with the driver software to handle the installation. Refer to the NI Package Manager Manual for more information about installing, removing, and upgrading NI software using NI Package Manager.
  3. Follow the instructions in the installation prompts.
    Note Windows users may see access and security messages during installation. Accept the prompts to complete the installation.
  4. When the installer completes, select Restart in the dialog box that prompts you to restart, shut down, or restart later.

Installing the PXIe-5830

Notice To prevent damage to the PXIe-5830 caused by ESD or contamination, handle the module using the edges or the metal bracket.
Notice

You must install the software before installing the hardware.

Before you install the hardware, refer to the guidelines in the Maintain Forced-Air Cooling Note to Users included with the module to ensure that the device can cool itself effectively.

To use the included cables, you must install the PXIe-5820 immediately to the left of the PXIe-3621.

  1. Ensure the AC power source is connected to the chassis before installing the module.
    The AC power cord grounds the chassis and protects it from electrical damage while you install the module.
  2. Power off the chassis.
  3. Inspect the slot pins on the chassis backplane for any bends or damage prior to installation. Do not install a module if the backplane is damaged.
  4. Remove the black plastic covers from all the captive screws on the module front panel.
  5. Identify a supported slot in the chassis. The following figure shows the symbols that indicate the slot types.
    Figure 1. Chassis Compatibility Symbols

    1378

    1. PXI Express System Controller Slot
    2. PXI Peripheral Slot
    3. PXI Express Hybrid Peripheral Slot
    4. PXI Express System Timing Slot
    5. PXI Express Peripheral Slot
  6. Touch any metal part of the chassis to discharge static electricity.
  7. Ensure that the ejector handle is in the downward (unlatched) position.
  8. Place the module edges into the module guides at the top and bottom of the chassis. Slide the module into the slot until it is fully inserted.
    Figure 1. Module Installation

    1378

    1. Chassis
    2. Hardware Module
    3. Ejector Handle in Downward (Unlatched) Position
  9. Latch the module in place by pulling up on the ejector handle.
  10. Secure the module front panel to the chassis using the front-panel mounting screws.
    Note Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis, which can improve the signal quality and electromagnetic performance.
  11. Cover all empty slots using either filler panels (standard or EMC) or slot blockers with filler panels, depending on your application.
    Note For more information about installing slot blockers and filler panels, go to ni.com/r/pxiblocker.
  12. Power on the chassis.

Interconnecting the PXIe-5830 Modules

Take the following considerations into account when interconnecting the PXI Express modules that comprise the PXIe-5830.

  • Inspect all RF connectors before attaching mating cables. Clean with dry compressed air, if necessary, to remove any contaminants that may be present.
  • Position RF cables with instrument connectors to ensure center pins are straight and aligned before mating.
  • Press MMPX connectors until they noticeably snap into place.
  • For all threaded RF connectors (SMA and SMA 27 GHz), engage and finger-tighten the connector nuts. The connector nuts should thread easily without binding, but if not, ensure the connector is straight and that center pins have not been damaged.
  • Carefully complete tightening all threaded RF connectors to 0.9 N · m(8 lb · in) using an appropriate torque wrench (not included) or torque screwdriver and driver bit. Tighten only until the wrench clicks.
Notice Incorrect torque at SMA connections can damage device connectors and degrade signal fidelity.

Connecting the PXIe-5820 and PXIe-3621

Complete the following steps for the baseband signal connections and Reference Clock connection, as shown in the following figure.

Figure 3. PXIe-5830 Equipment Setup

1378

  1. MMPX (m)-to-MMPX (m) Cable, 40 mm
  2. MMCX (m)-to-MMCX (m) Cable, 87 mm
  1. Complete the following baseband signal cable connections between the PXIe-5820 and PXIe-3621 front panels using the included eight 40 mm MMPX (m)-to-MMPX (m) cables for transmit and receive I/Q signals.
    Table 2. Baseband Signal Cable Connections
    PXIe-5820 Connector PXIe-3621 Connector
    I/Q OUT I+ I/Q IN I+
    I/Q OUT I- I/Q IN I-
    I/Q OUT Q+ I/Q IN Q+
    I/Q OUT Q- I/Q IN Q-
    I/Q IN I+ I/Q OUT I+
    I/Q IN I- I/Q OUT I-
    I/Q IN Q+ I/Q OUT Q+
    I/Q IN Q- I/Q OUT Q-
  2. Use the 87 mm MMCX (m)-to-MMCX (m) cable to connect the PXIe-5820 REF IN front panel connector to the PXIe-3621 REF OUT front panel connector.
    This completes the Reference Clock connection.

Direct Connections to the PXIe-5830

The PXIe-5830 is a precision RF instrument that is sensitive to ESD and transients. Ensure you are making proper direct connections to the PXIe-5830 to avoid damaging the hardware.

Notice Apply external signals only while the hardware is powered on. Applying external signals while the hardware is powered off may cause damage.
  • Ensure you are properly grounded when manipulating cables or antennas connected to the instrument.
  • If you are using noninsulated devices, such as an RF antenna, ensure the devices are maintained in a static-free environment.
  • If you are using an active device, such as a preamplifier or switch routed to the PXIe-5830, ensure that the device cannot generate signal transients greater than the RF and DC specifications of the PXIe-5830.

Configuring the PXIe-5830 in MAX

Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX informs other programs about which NI hardware products are in the system and how they are configured. MAX is automatically installed with NI-RFSA, NI-RFSG, and RFmx.

  1. Launch MAX.
  2. In the configuration tree, expand Devices and Interfaces to see the list of installed NI hardware.
    Installed modules appear under the name of their associated chassis.
  3. Expand your Chassis tree item.

    MAX lists all modules installed in the chassis. Your default names may vary.

    Note If you do not see your module listed, press <F5> to refresh the list of installed modules. If the module is still not listed, power off the system, ensure the module is correctly installed, and restart.
  4. Record the identifier MAX assigns to the PXIe-3621. Use this identifier when programming the PXIe-5830.
  5. Associate the hardware modules that comprise your instrument.
    1. Select the PXIe-3621 that is identified as not configured in the configuration tree.
    2. In the Associated Devices section, select the PXIe-5820 from the Baseband list.
    3. Click Save in the MAX toolbar.
      Note Module associations may be lost when you move the modules to different chassis slots.
  6. Self-test the hardware by selecting the item in the configuration tree and clicking Self-Test in the MAX toolbar. Repeat this step for all modules in your PXIe-5830 system.
    The MAX self-test performs a basic verification of hardware resources.

Self-Calibration

Self-calibration adjusts the PXIe-5830 for variations in the module environment using onboard power detectors.

PXIe-5830 modules are externally calibrated at the factory; however, you should perform a self-calibration in any of the following situations:

  • After first installing and interconnecting your PXIe-5830 instrument
  • After any module in the chassis is installed, uninstalled, or moved
  • After inter-module cabling has been changed, reconnected, or repositioned
  • After changing controllers, computers, or reinstalling the driver software
  • When the system is in an environment where the ambient temperature varies or the module temperature has drifted more than ±5 °C from the temperature at the last self-calibration
  • To periodically adjust for small performance drifts that occur with product aging
Note To ensure specifications before you use the instrument or perform self-calibration, power on the PXI Express chassis; set the chassis fan speeds to AUTO or HIGH, depending on the chassis; wait for the operating system to completely load; and then allow the instrument to warm up for 30 minutes.
Note Ensure that any connected external hardware is in an idle state and not transmitting during self-calibration. It is not recommended to perform self-calibration while connected in direct loopback configurations.
Note For the best image rejection and LO leakage performance, it is recommended that you run Self Cal or Self Cal Range for only the image rejection and LO leakage portion if the temperature changes beyond ±1 °C.

Performing a Self-Calibration Using NI-RFSA

NI recommends you perform the self-calibration using the NI-RFSA example when you self-calibrate the module.
  1. Select Start » All Programs » National Instruments » LabVIEW to launch LabVIEW.
  2. Launch the Example Finder and navigate to the example.
    1. Select Hardware Input and Output » Modular Instruments » NI-RFSA » Utilities » RFSA Self Calibration.
    2. Open the example.
  3. Complete the following steps to configure the example:
    1. Select the device identifier assigned to the PXIe-3621 in MAX in the [Resource Name] drop-down menu.
    2. Set Clock Source to OnboardClock.
    3. Set Self Calibration Step Operations to Perform All Self Calibration Steps.
  4. Run the VI.

Locating the Software and Examples

Software Locations

Table 3. Location of PXIe-5830 Software Options
Software Option ADE Location
RFmx LabVIEW

Available on the LabVIEW Functions palette at Measurement I/O » NI-RFmx.

LabWindows/CVI

RFmx functions are available from the LabWindows/CVILibrary menu at Library » RFmx SpecAn Library and Library » RFmx Demod Library.

Microsoft Visual C/C++

Use the header files located in the <NIDir>\Shared\ExternalCompilerSupport\C\include directory and import library files located in one of the following directories:

  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-<NIDir>\Shared\ExternalCompilerSupport\C\include
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit) :
    • 32-bit installation-<NIDir>\Shared\ExternalCompilerSupport\C\lib32\msvc
    • 64-bit installation-<NIDir>\Shared\ExternalCompilerSupport\C\lib64\msvc

where <NIDir> is one of the following locations:

  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-Program Files\National Instruments
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit)-Program Files (x86)\National Instruments
Microsoft .NET

For the location of .NET class libraries, refer to the installed RFmx readme.

NI-RFSA LabVIEW

Available on the LabVIEW Functions palette at Measurement I/O » NI-RFSA.

LabWindows/CVI

Available in the <IVIROOTDIR32>\Drivers\niRFSA directory, where <IVIROOTDIR32> is one of the following locations:

  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-Program Files\IVI Foundation\IVI
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit)-Program Files (x86)\IVI Foundation\IVI
Microsoft Visual C/C++
Use the header files located in the <IVIROOTDIR32>\Include directory and import library files located in the <IVIROOTDIR32>\Lib directory, where <IVIROOTDIR32> is one of the following directories:
  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-Program Files\IVI Foundation\IVI
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit)-Program Files (x86)\IVI Foundation\IVI
Microsoft .NET

Use the NI-RFSA .NET class library by adding a reference to NationalInstruments.ModularInstruments.NIRfsa.Fx40 or NationalInstruments.ModularInstruments.NIRfsa.Fx45 and any dependent class libraries from within the Solution Explorer in Visual Studio.

NI-RFSG LabVIEW

Available on the LabVIEW Functions palette at Measurement I/O » NI-RFSG.

LabWindows/CVI

Available in the <IVIROOTDIR32>\Drivers\niRFSG directory, where <IVIROOTDIR32> is one of the following locations:

  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-Program Files\IVI Foundation\IVI
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit)-Program Files (x86)\IVI Foundation\IVI
Microsoft Visual C/C++

Use the header files located in the <IVIROOTDIR32>\Include directory and import library files located in the <IVIROOTDIR32>\Lib directory, where <IVIROOTDIR32> is one of the following directories:

  • Windows 10 (32-bit)/8.1 (32-bit)/7 (32-bit)-Program Files\IVI Foundation\IVI
  • Windows 10 (64-bit)/8.1 (64-bit)/7 (64-bit)-Program Files (x86)\IVI Foundation\IVI
Microsoft .NET

Use the NI-RFSG .NET class library by adding a reference to NationalInstruments.ModularInstruments​​​.NIRfsg.Fx40 or NationalInstruments.ModularInstruments​​.NIRfsg.Fx45 and any dependent class libraries from within the Solution Explorer in Visual Studio.

Programming Examples Locations

Using the NI Example Finder

If you are using RFmx, NI-RFSA, or NI-RFSG with LabVIEW or LabWindows/CVI, use the NI Example Finder to locate programming examples.

  1. Launch LabVIEW or LabWindows/CVI.
  2. Select Help » Find Examples to open the NI Example Finder.
  3. Navigate to Hardware Input and Output » Modular Instruments.
  4. Open the example that best matches your application requirements.

Using Microsoft Visual C/C++

If you are using RFmx or NI-RFSA with Microsoft Visual C/C++, locate examples in the following directories.

Table 4. Location of Microsoft Visual C/C++ Programming Examples
Software Option Examples Location
RFmx

<NIDocDir>\RFmx\Demod\Examples

<NIDocDir>\RFmx\SpecAn\Examples

where <NIDocDir> is the Users\Public\Public Documents\National Instruments directory.

NI-RFSA

<NIDocDir>\NI-RFSA\examples

where <NIDocDir> is the Users\Public\Public Documents\National Instruments directory.

Making a First Measurement

You can verify proper installation and configuration of your instrument by making a measurement using the NI-RFSG and RFmx Soft Front Panels (SFPs).

  1. Connect the PXIe-5830IF IN/OUT 0 and IF IN/OUT 1 ports as appropriate for your application.
  2. Select Start » All Programs » National Instruments » NI-RFSG » NI-RFSG Soft Front Panel to launch the NI-RFSG SFP.
  3. Select the device identifier assigned to the PXIe-3621 in MAX in the Device drop-down menu.
  4. Set the RF Frequency to 5 GHz.
  5. Click the RF On/Off button to generate a continuous wave (CW) tone RF signal at the IF output terminal of your instrument.
  6. Select Start » All Programs » National Instruments » RFmx SpecAn » RFmx Soft Front Panel to launch the RFmx SFP.
  7. Select the device identifier assigned to the PXIe-3621 in MAX in the Device drop-down menu.
  8. Set the Center Frequency to 5 GHz
  9. Verify that the RFmx SFP is receiving the CW tone.
  10. Close the RFmx SFP and NI-RFSG SFP.

Troubleshooting

If an issue persists after you complete a troubleshooting procedure, search our KnowledgeBase for additional information our technical support engineers create as they answer common user questions and resolve unexpected issues.

What Should I Do if the PXIe-3621 Does Not Appear in MAX?

  1. In the MAX configuration tree, expand Devices and Interfaces.
  2. Expand the Chassis tree to see the list of installed hardware, and press <F5> to refresh the list.
  3. If the module is still not listed, power off the system, ensure that all hardware is correctly installed, and restart the system.
  4. Navigate to the Device Manager.
    Operating SystemDescription
    Windows 10/8.1 Right-click the Start button, and select Device Manager.
    Windows 7 Select Start » Control Panel » Device Manager.
  5. Verify the PXIe-3621 appears in the Device Manager.
    1. Under NI Modular Instruments, confirm that a PXIe-3621 entry appears.
      Note If you are using a PC with a device for PXI remote control system, under System Devices, also confirm that no error conditions appear for the PCI-to-PCI Bridge.
    2. If error conditions appear, reinstall the driver software and the PXIe-3621.

Why Is the ACCESS LED Off When the Chassis Is On?

The LEDs may not light until the module has been configured in MAX. Before proceeding, verify that the module appears in MAX.

If the module appears in MAX but the ACCESS LED fails to light after you power on the chassis, a problem may exist with the chassis power rails, the hardware module, or the LED.
Notice Apply external signals only while the hardware is powered on. Applying external signals while the hardware is powered off may cause damage.
  1. Disconnect any signals from the PXI Express module front panel.
  2. Power off the PXI Express chassis.
  3. Remove the module from the PXI Express chassis and inspect it for damage. Do not reinstall a damaged device.
  4. Install the module in a different PXI Express chassis slot from which you removed it.
  5. Power on the PXI Express chassis.
  6. Restart your computer.
  7. Verify that the module appears in MAX.

What Should I Do if the PXIe-5830 Fails the Self-Test?

  1. Restart the system.
  2. Launch MAX, and perform the self-test again.
  3. Power off the chassis.
  4. Reinstall the failed module in a different slot.
  5. Power on the chassis.
  6. Perform the self-test again.

What Should I Do if the Instrument Does Not Initialize?

Failure to initialize may indicate a problem with module interconnections or with MAX.

If the niRFSA Initialize VI or niRFSG Initialize VI returns an error and the PXIe-5830 fails to initialize, complete the following steps:

  1. Power off the system.
  2. Inspect all PXIe-5830 front panel cables and, if there are no apparent issues, reconnect cables securely, according to the interconnection procedure for your instrument.
  3. Power on your system and run the MAX configuration and intermediate self-test procedures.

Hardware Connectors and Indicators

PXIe-5820

Figure 4. PXIe-5820 Front Panel and LEDs

1378

Table 5. General Connector Descriptions
Connector Description Connector Type
REF IN Input connector that allows for the use of an external 10 MHz Reference Clock. MMPX (f)
REF OUT Output connector that can export a 10 MHz Reference Clock. MMPX (f)
PFI 0 Programmable-function digital I/O (DIO) connector for use with triggers or events. MMPX (f)
DIO Multi-signal DIO connector that provides access to FPGA multi-gigabit transceivers (MGTs) and general purpose LVCMOS signals. Nano-Pitch I/O
Table 6. I/Q Connector Descriptions
Connector Description Connector Type
I/Q OUT I+ Output connector for I+ signals. MMPX (f)
I- Output connector for I- signals. MMPX (f)
Q+ Output connector for Q+ signals. MMPX (f)
Q- Output connector for Q- signals. MMPX (f)
I/Q IN I+ Input connector for I+ signals. MMPX (f)
I- Input connector for I- signals. MMPX (f)
Q+ Input connector for Q+ signals. MMPX (f)
Q- Input connector for Q- signals. MMPX (f)
Table 7. LED Indicators
LED Indications
ACCESS

Indicates the basic hardware status of the device.

Off—The device is not yet functional or has detected a problem with a PXI Express power rail.

Amber—The device is being accessed. Accessed means that you are writing to the device setup registers to control the device, reading from the device to monitor the device status, or transferring data to/from the device.

Green—The device is controllable through the software.

ACTIVE

Off—The device is idle.

Solid green—The device is generating a waveform.

Dim amber—The device is waiting for an acquisition Reference Trigger.

Solid amber—The device is acquiring a waveform.

Solid red—The device has detected an error. The LED remains red until the error condition is removed.

Note The indicators are listed in increasing order of priority. For example, if you are generating a waveform using NI-RFSG and waiting on an acquisition Reference Trigger in NI-RFSA, the LED is dim amber.
Figure 5. PXIe-5820 DIO Connector Pinout

1378

PXIe-3621 Front Panel and LEDs


1378

Table 8. I/Q Connector Descriptions
Connector Description Connector Type
I/Q IN I+ Input connector for I+ signals. MMPX (f)
I- Input connector for I- signals. MMPX (f)
Q+ Input connector for Q+ signals. MMPX (f)
Q- Input connector for Q- signals. MMPX (f)
I/Q OUT I+ Output connector for I+ signals. MMPX (f)
I- Output connector for I- signals. MMPX (f)
Q+ Output connector for Q+ signals. MMPX (f)
Q- Output connector for Q- signals. MMPX (f)
Table 9. General Connector Descriptions
Connector Description Connector Type
LO2 IN Input connector for the external local oscillator signal used for the up and down conversion between baseband and IF. MMPX (f)
OUT Output connector for the local oscillator signal used for the up and down conversion between baseband and IF. MMPX (f)
IF IN/OUT 0 IF0 bi-directional connector for both input and output. SMA 27 GHz (f)
1 IF1 bi-directional connector for both input and output. SMA 27 GHz (f)
REF IN Input connector for the Reference Clock. MMPX (f)
OUT Output connector for the Reference Clock. MMPX (f)
RESERVED Reserved connectors are not enabled for use on the PXIe-5830 and should remain capped with 50 Ω terminators to ensure specified instrument performance.
Table 10. LED Indicators
LED Indications
ACCESS

Indicates the basic hardware status of the module.

  • OFF—The driver software has not yet loaded, or the module exceeded approved operating temperature and thermal shutdown occurred or has detected a problem with a power rail.
  • AMBER—The module is being accessed. Accessed means that the device setup registers are being written to in order to control the device.
  • GREEN—The module is ready to be programmed by software.
  • ACTIVE
  • OFF—The module is idle.
  • RED—The module detected an error state.
  • NI Services

    Visit ni.com/support to find support resources including documentation, downloads, and troubleshooting and application development self-help such as tutorials and examples.

    Visit ni.com/services to learn about NI service offerings such as calibration options, repair, and replacement.

    Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.

    NI corporate headquarters is located at 11500 N Mopac Expwy, Austin, TX, 78759-3504, USA.