PXI-4130 Specifications
- Updated2023-02-18
- 8 minute(s) read
PXI-4130 Specifications
These specifications apply to the PXI-4130 with APS-4100 auxiliary power supply and to the PXI-4130 without auxiliary power supply.
Definitions
Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty.
Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty.
- Typical specifications describe the performance met by a majority of models.
- Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing.
Specifications are Warranted unless otherwise noted.
Conditions
Specifications are valid under the following conditions unless otherwise noted.
- Ambient temperature[1] of 23 °C ± 5 ºC
- 30 minutes warm-up time
- niDCPower Samples to Average property or NIDCPOWER_ATTR_SAMPLES_TO_AVERAGE attribute set to 300 for optimal 50 Hz and 60 Hz rejection
Device Capabilities
Channel | DC Voltage Ranges | Isolation | DC Current Source and Sink Ranges |
---|---|---|---|
SMU Channel (1) |
|
60 VDC, CAT I |
|
Utility Channel (0) |
|
N/A |
|
SMU Channel Specifications (Channel 1)
Voltage Programming Accuracy/Resolution
Range | Resolution | Accuracy ± (% of Output + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
±20 V | 0.33 mV | 0.034% + 1.8 mV |
±6 V | 0.1 mV | 0.034% + 1.5 mV |
Current Programming Accuracy/Resolution[3]
Range | Resolution | Accuracy ± (% of Output + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
200 μA | 10 nA | 0.03% + 0.1 μA |
2 mA | 100 nA | 0.03% + 1 μA |
20 mA | 1 μA | 0.03% + 10 μA |
200 mA | 10 μA | 0.03% + 100 μA |
2 A [4] | 100 μA | 0.12% + 1 mA |
Voltage Measurement Accuracy/Resolution
Range | Resolution | Accuracy ± (% of Reading + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
±20 V | 0.10 mV | 0.03% + 1.5 mV |
±6 V | 0.10 mV | 0.03% + 1.5 mV |
Current Measurement Accuracy/Resolution
Range | Resolution | Accuracy ± (% of Reading + Offset) |
---|---|---|
1 Year 23 °C ± 5 °C | ||
200 μA | 1.0 nA | 0.03% + 0.02 μA |
2 mA | 10 nA | 0.03% + 0.2 μA |
20 mA | 0.1 μA | 0.03% + 2 μA |
200 mA | 1 μA | 0.03% + 40 μA |
2 A [] | 10 μA | 0.12% + 200 μA |
Channel 1 Additional Specifications
Settling time[5] | 500 μs, typical | ||||||
| |||||||
Slew rate | 0.08 V/μs, typical | ||||||
Transient response | Recovers to <0.1% of voltage range within 200 μs after a change in load current from 10% to 90% of current range, typical | ||||||
Normal Mode noise (Source only) | 15 mVp-p into resistive load <5 mV RMS20 Hz to 20 MHz bandwidth, typical | ||||||
Remote sense | Up to 1 V drop per lead using internal power or ≥ 12 V auxiliary power supply; Add 120 μV to accuracy specification per volt of lead drop. | ||||||
| |||||||
|
Temperature coefficient (Tempco) is 10% of accuracy specification per °C.
Utility Channel Specifications (Channel 0)
Programming Accuracy/Resolution[6]
Output Function | Range | Resolution | Accuracy ± (% of output + offset) |
---|---|---|---|
1 Year 23 °C ± 10 °C | |||
Voltage | +6 V | 0.12 mV | 0.05% + 4 mV |
Current | 1 A [7] | 0.02 mA | 0.15% + 4 mA |
Measurement Accuracy/Resolution
Measurement Type | Range | Resolution | Accuracy ± (% of reading + offset) |
---|---|---|---|
1 Year 23 °C ± 10 °C | |||
Voltage | +6 V | 0.06 mV | 0.05% + 4 mV |
Current | 1 A [] | 0.01 mA | 0.15% + 4 mA |
Channel 0 Additional Specifications
Settling time | <1 ms, 10% to 90% of range, measured with full load, typical | ||||||
Output capacitance | 33 µF, typical | ||||||
Transient response | Recovers to <0.1% of voltage range within 50 μs after a change in load current from 50% to 100% of current range, typical | ||||||
Normal Mode noise and ripple (source only, voltage) | <1.5 mV RMS, 20 Hz to 20 MHz bandwidth, typical | ||||||
|
Temperature coefficient (Tempco) is 15% of accuracy specification per °C.
Programming and Measurement Timing[8]
Maximum output update rate | 3000 Updates/s, nominal |
Maximum measurement rate (samples to average = 1) | 3 kS/s, nominal |
Single point update latency | 600 µs, typical |
Protection
|
Calibration Interval
Recommended calibration interval | 1 year |
Accuracy Specification Derating versus Load Current
Maximum Sinking Power versus Ambient Temperature for Channel 1
Physical Characteristics
Dimensions | 3U, one-slot PXI/cPCI module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.), nominal | ||||||
Weight | 312 g (11 oz), typical | ||||||
|
|
Power Requirements
PXI power requirement | 10 W at 5 V 1 W at 3.3 V 6 W at 12 V 2.5 W at -12 V, typical |
Auxiliary power source[9] input requirements | 11 VDC to 15.5 VDC 5 A max |
Environment
Maximum altitude | 2,000 m (at 25 °C ambient temperature) |
Pollution Degree | 2 |
Indoor use only.
Operating Environment
Ambient temperature range | 0 °C to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Storage Environment
Ambient temperature range | -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
Shock and Vibration
Operational shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | ||||||
|
Compliance and Certifications
Safety
|
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
- IEC 61010-1, EN 61010-1
- UL 61010-1, CSA C22.2 No. 61010-1
Electromagnetic Compatibility
- EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
- EN 55011 (CISPR 11): Group 1, Class A emissions
- EN 55022 (CISPR 22): Class A emissions
- EN 55024 (CISPR 24): Immunity
- AS/NZS CISPR 11: Group 1, Class A emissions
- AS/NZS CISPR 22: Class A emissions
- FCC 47 CFR Part 15B: Class A emissions
- ICES-001: Class A emissions
CE Compliance

This product meets the essential requirements of applicable European Directives, as follows:
- 2014/35/EU; Low-Voltage Directive (safety)
- 2014/30/EU; Electromagnetic Compatibility Directive (EMC)
- 2011/65/EU; Restriction of Hazardous Substances (RoHS)
- 2014/53/EU; Radio Equipment Directive (RED)
- 2014/34/EU; Potentially Explosive Atmospheres (ATEX)
Product Certifications and Declarations
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
EU and UK Customers
电子信息产品污染控制管理办法(中国RoHS)
1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake).
2 Current input/output for channel 1 is limited to 2 W or 300 mA when operating under internal power. Continuous sinking power for channel 1 is limited to 10 W, subject to derating above 30 °C. Refer to the figure Maximum Sinking Power versus Ambient Temperature for Channel 1.
3 Minimum programmable current limit/level is 2% of range.
4 For currents ≥ 500 mA, refer to the additional derating information in the figure Accuracy Specification Derating versus Load Current.
5 Settled to 1%, 1 V step, 50% of current range at final value, output capacitance set to low, using auxiliary power supply.
6 Minimum programmable current limit/level is 2% of range.
7 For currents ≥500 mA, refer to the additional derating information in the figure Accuracy Specification Derating versus Load Current.
8 Does not include load dependent settling time.
9 Optional; Channel 1 only.
10 Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, III, or CAT IV.
In This Section
- Definitions
- Conditions
- Device Capabilities
- SMU Channel Specifications (Channel 1)
- Channel 1 Additional Specifications
- Utility Channel Specifications (Channel 0)
- Programming and Measurement Timing[8]
- Protection
- Calibration Interval
- Accuracy Specification Derating versus Load Current
- Maximum Sinking Power versus Ambient Temperature for Channel 1
- Physical Characteristics
- Power Requirements
- Environment
- Shock and Vibration
- Compliance and Certifications